Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT02360020
Other study ID # NCTCM03
Secondary ID
Status Completed
Phase Phase 4
First received February 2, 2015
Last updated May 9, 2016
Start date August 2014
Est. completion date May 2016

Study information

Verified date May 2016
Source Clinica Mediterranea
Contact n/a
Is FDA regulated No
Health authority Italy: Ministry of Health
Study type Interventional

Clinical Trial Summary

Stent delivery failure occurs in 4% of all percutaneous coronary interventions (PCI) and >90% of these failures are due to vessel tortuosity and/or calcification. The XLIMUS eluting coronary stent (CARDIONOVUM GmbH, Bonn, Germany) is a new type of endovascular prostheses characterised by better mechanical properties than traditional DES. This is a prospective, non-randomized, single-center pilot study, aiming to evaluate the performance of the XLIMUS DES in severely complex coronary lesions in real-world clinical practice.


Description:

All consecutive patients who will undergo elective PCI in native coronary arteries at the Clinica Mediterranea (Naples, Italy) will be considered for eligility. Study participants wiil require to have symptomatic ischemic heart disease attributable to critical (that is, >70% visual estimate) stenotic lesions of native coronary arteries. Inclusion criteria in this pilot study are 1) chronic total occlusion (CTO), 2) severe target vessel calcification, and 3) severe target vessel tortuosity. CTO is defined as the presence of TIMI 0 flow within the occluded segment and angiographic or clinical evidence or high likelihood of an occlusion duration of ≥3 months. Calcification is defined severe when larger than 3x vessel diameter, and comprising the vessel wall totally in two perpendicular views. Tortuosity is defined severe when it satisfies the following criteria: one or more bends of 90° or more, or three or more bends of 45-90° proximal to the diseased segment.

200 patients will be enrolled into the study. Stents will be implanted according to current clinical practice. Techniques attempted for facilitating stent delivery in such a complex lesions are: maximize guide catheter support, optimize predilatation of the stenosis, use of a stiffer guidewire. Specific tricks include: a) buddy-wire; anchoring balloon; GuideLiner catheter. In case of severe calcification, rotational atherectomy will be electively performed with the Rotablator® system (Boston Scientific Corporation, Natick, MA, U.S.A.). Following stent implantation, postdilatation will be performed in all instances with a non-compliant balloon. All patients will receive aspirin 325 mg and clopidogrel (75 mg daily) before stent deployment, with a loading dose (600 mg of clopidogrel) given to patients not pretreated. All patients will receive 70 IU/Kg intra-arterial bolus of unfractionated heparin in order to achieve and activated clotting time >250 seconds. Glycoprotein IIb/IIIa inhibitors will be administered according to operator preference. Estimated glomerular filtration rate (eGFR) will be calculated by applying the Levey Modification of Diet in Renal Disease (MDRD) formula. Chronic kidney disease was defined as a eGFR <60 ml/min/1.73 m2 .

XLIMUS eluting-stent is made of cobalt chromium L 605 and the stent is available in a 6-, 8-, or 10-cell structure design (closed cell architecture). The struts thickness is 73µm. The 6-cell design is for stenting of coronary artery diameter of 2.25mm-2.50mm, 8-cell structure is used for stenting of 2.75-3.50 mm artery diameters, and the 10-cell is for larger artery diameter lesions (up to 5mm). The XLIMUS has an innovative hydrophilic-coated shaft and an extra-low tip profile (crossing profile = 0.90 mm) to access the most tortuous lesions. The highly biocompatible polylactid acid (PLLA) drug containing release matrix degrades smoothly and provides an optimal release kinetic profile. Within 30 days, about 70% of the anti-proliferative drug is distributed into the surrounding arterial tissue of the stent struts, ensuring a highly effective inhibition of smooth muscle cell migration and proliferation. Pharmacokinetic study result confirm sustained anti-proliferative drug efficacy up to 120 days.

The primary objective of the study is the assessment of the clinical performance of the XLILMUS DES, using the following criteria 1) device success, defined as the ability to insert the stent into the target lesion and the attainment of <20% residual stenosis (by visual estimate), 2) lesion success, defined as attainment of <20% residual stenosis of the target lesion using any percutaneous method, and 3) procedural success, defined as lesion success without any in-hospital and MACE.


Recruitment information / eligibility

Status Completed
Enrollment 200
Est. completion date May 2016
Est. primary completion date December 2015
Accepts healthy volunteers No
Gender Both
Age group 18 Years to 90 Years
Eligibility Inclusion Criteria:

- chronic total occlusion (CTO)

- severe calcification

- severe tortuosity

Exclusion Criteria:

• coronary artery lesions non satisfying the inclusion criteria

Study Design

Endpoint Classification: Safety/Efficacy Study, Intervention Model: Single Group Assignment, Masking: Open Label, Primary Purpose: Treatment


Related Conditions & MeSH terms


Intervention

Device:
XLimus
Techniques attempted for facilitating stent delivery in such a complex lesions are: maximize guide catheter support, optimize predilatation of the stenosis, use of a stiffer guidewire. Specific tricks include: a) buddy-wire; anchoring balloon; GuideLiner catheter. In case of severe calcification, rotational atherectomy was electively performed with the Rotablator® system (Boston Scientific Corporation, Natick, MA, U.S.A.). Following stent implantation, postdilatation is performed in all instances with a non-compliant balloon

Locations

Country Name City State
Italy Clinica Mediterranea Naples

Sponsors (1)

Lead Sponsor Collaborator
Clinica Mediterranea

Country where clinical trial is conducted

Italy, 

References & Publications (13)

Ashikaga T, Sakurai K, Satoh Y. Tools & Techniques: stent delivery in distal lesions. EuroIntervention. 2010 Nov;6(5):660-1. doi: 10.4244/EIJV6I5A109. — View Citation

Briguori C, Visconti G, Focaccio A, Donahue M. Performance of the XLIMUS sirolimus-eluting coronary stent in very complex lesions. Minerva Cardioangiol. 2014 Feb;62(1):1-8. — View Citation

Cantor WJ, Lazzam C, Cohen EA, Bowman KA, Dolman S, Mackie K, Natarajan MK, Strauss BH. Failed coronary stent deployment. Am Heart J. 1998 Dec;136(6):1088-95. — View Citation

Feldman T. Tricks for overcoming difficult stent delivery. Catheter Cardiovasc Interv. 1999 Nov;48(3):285-6. — View Citation

Fernandes V, Kaluza GL, Godlewski B, Li G, Raizner AE. Novel technique for stent delivery in tortuous coronary arteries: report of three cases. Catheter Cardiovasc Interv. 2002 Apr;55(4):485-90. — View Citation

Gyöngyösi M, Yang P, Khorsand A, Glogar D. Longitudinal straightening effect of stents is an additional predictor for major adverse cardiac events. Austrian Wiktor Stent Study Group and European Paragon Stent Investigators. J Am Coll Cardiol. 2000 May;35(6):1580-9. — View Citation

Kumar S, Gorog DA, Secco GG, Di Mario C, Kukreja N. The GuideLiner "child" catheter for percutaneous coronary intervention - early clinical experience. J Invasive Cardiol. 2010 Oct;22(10):495-8. — View Citation

Lohavanichbutr K, Webb JG, Carere RG, Solankhi N, Jarochowski M, D'yachkova Y, Dodek A. Mechanisms, management, and outcome of failure of delivery of coronary stents. Am J Cardiol. 1999 Mar 1;83(5):779-81, A9. — View Citation

Morice MC, Colombo A, Meier B, Serruys P, Tamburino C, Guagliumi G, Sousa E, Stoll HP; REALITY Trial Investigators. Sirolimus- vs paclitaxel-eluting stents in de novo coronary artery lesions: the REALITY trial: a randomized controlled trial. JAMA. 2006 Feb 22;295(8):895-904. — View Citation

Mortier P, De Beule M, Segers P, Verdonck P, Verhegghe B. Virtual bench testing of new generation coronary stents. EuroIntervention. 2011 Jul;7(3):369-76. doi: 10.4244/EIJV7I3A62. — View Citation

Nikolsky E, Gruberg L, Pechersky S, Kapeliovich M, Grenadier E, Amikam S, Boulos M, Suleiman M, Markiewicz W, Beyar R. Stent deployment failure: reasons, implications, and short- and long-term outcomes. Catheter Cardiovasc Interv. 2003 Jul;59(3):324-8. — View Citation

Rieu R, Barragan P, Garitey V, Roquebert PO, Fuseri J, Commeau P, Sainsous J. Assessment of the trackability, flexibility, and conformability of coronary stents: a comparative analysis. Catheter Cardiovasc Interv. 2003 Aug;59(4):496-503. — View Citation

Schmidt W, Lanzer P, Behrens P, Topoleski LD, Schmitz KP. A comparison of the mechanical performance characteristics of seven drug-eluting stent systems. Catheter Cardiovasc Interv. 2009 Feb 15;73(3):350-60. doi: 10.1002/ccd.21832. — View Citation

* Note: There are 13 references in allClick here to view all references

Outcome

Type Measure Description Time frame Safety issue
Primary Stent Performance The primary objective of the study was the assessment of the clinical performance of the XLIMUS DES, using the following criteria 1) device success, defined as the ability to insert the stent into the target lesion and the attainment of <20% residual stenosis (by visual estimate), 2) lesion success, defined as attainment of <20% residual stenosis of the target lesion using any percutaneous method, and 3) procedural success, defined as lesion success without any in-hospital Man 30-day MACE up to 1 month Yes
Secondary Inhospital, 30-day and 1-year MACE Major adverse cardiac events included death of any cause, nonfatal myocardial infarction, and repeated revascularization by PCI or surgery occurring within 30-day and 1-year. Myocardial infarction was defined as the presence of pathological and new Q waves on an ECG or as an increase in creatine kinase-myocardial band level to >3 times the upper limit of normal (ULN). Periprocedural myocardial infarction was defined as an increase of troponin I =5 times ULN. Target lesion revascularization was defined as a clinically-driven repeat percutaneous coronary angioplasty or coronary artery bypass surgery. Stent thrombosis was defined according to the Academic Research Consortium definitions 1 year Yes
See also
  Status Clinical Trial Phase
Terminated NCT03959072 - Cardiac Cath Lab Staff Radiation Exposure
Completed NCT03475888 - Incidence of Ventricular Arrhythmias in Patients With Chronic Total Occlusion Recanalization N/A
Not yet recruiting NCT04917432 - Early and Midterm Outcomes of Intravascular Ultrasound (IVUS) Versus Non-IVUS Guidance in Complex Coronary Chronic Total Occlusion (CTO) Revascularization. N/A
Recruiting NCT04944615 - To Evaluate Whether IVUS-guided Drug-eluting Stent (DES) Implantation Leads to Better Clinical Outcomes Compared to Conventional Angiography in the Treatment of Chronic Complete Occlusion (CTO) Disease. N/A
Completed NCT01861860 - OPtimized Stenting Using Intravascular Ultrasound(IVUS) in Long lEsion: Rationale for Simplified criteriA N/A
Recruiting NCT05142215 - A Placebo-controlled Trial of Chronic Total Occlusion Percutaneous Coronary Intervention for the Relief of Stable Angina N/A
Active, not recruiting NCT05464147 - DYNAMX Bioadaptor ImplanTation for the trEatment of Complex Coronary Lesions N/A
Recruiting NCT04145167 - Observational Registry on Clinical Outcome After Diagnosis of Chronic Total Occlusions
Not yet recruiting NCT04965207 - EvaLuAtion On the cHaracteristics of the True/False lUmen and Its Prognostic Value for cOronary CTO Patients Just Before Stent Implantation
Active, not recruiting NCT05197361 - Microvascular Coronary Resistance and Absolute Coronary FLOW in Patients With Percutaneous Intervention of a Chronic Total Occlusion
Recruiting NCT05813704 - Coronary Crossing System in Patients With Coronary Chronic Total Occlusions N/A
Recruiting NCT03563417 - ISCHEMIA-CTO Trial - Revascularisation or Optimal Medical Therapy of CTO N/A
Completed NCT05377866 - Holo CTO Proctoring Study
Completed NCT01978860 - A Prospective, First in Man Study to Evaluate the Safety and Performance of the NovaCross™ Micro-catheter N/A
Withdrawn NCT02784418 - The SHINE-CTO Trial N/A
Completed NCT03988166 - Chronic Total Occlusion Percutaneous Coronary Intervention Study N/A
Completed NCT02477579 - A Prospective, Multi-Center, Pivotal Study to Evaluate the Safety and Effectiveness of the NovaCross™ Micro-catheter in Facilitating Crossing Chronic Total Occlusion (CTO) Coronary Lesions N/A
Completed NCT03209843 - Post-stenting Assessment of Reendothelialization With OFDI After CTO Procedure (PERFECTO) N/A
Completed NCT03947398 - The BLIMP Balloon in Coronary Interventions N/A
Active, not recruiting NCT04060615 - Properties of Myocardial Microcirculation in Patients With Different Pathomorphological Substrates, Before and After Recanalization of Coronary Artery CTO N/A