Source Medtronic Diabetes
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The purpose of the study is to assess the performance of a new subcutaneous glucose sensor over a seven-day sensor life when used with currently marketed Medtronic Diabetes devices. In addition performance will be calculated for use of the new sensor with proposed new devices using new calibration algorithms.


Description:

The first generation Medtronic MiniMed Subcutaneous Glucose Sensor (Sof-Sensor) was originally approved by the FDA for commercialization as part of the Continuous Glucose Monitoring System (CGMS) on June 15, 1999 (PMA 980022). The Sensor is composed of a microelectrode with a thin coating of glucose oxidase beneath several layers of biocompatible membrane. This same sensor was used as part of subsequent continuous glucose monitoring (CGM) systems, such as the Guardian REAL-Time and Paradigm REAL-Time sensor augmented insulin pump. The current configuration of sensor has undergone in vitro and in vivo testing. A new second-generation glucose sensor (herein referred to as the Comfort Sensor) has been developed. The first-generation glucose sensor was approved with reported sensor accuracy (MAD) of 18%; it was labeled for a maximum use duration of 72 hours, using only the abdomen as an insertion site. The new sensor is shorter and has a smaller diameter, with a smaller gauge introducer needle. The new sensor inserter is designed to be used with the new sensor. The objectives of this study are to 1) Assess performance of the Comfort Sensor when used over a period of seven days with currently available devices, and 2) Assess performance of the Comfort Sensor when used over a period of seven days using new calibration algorithms (post-processed with algorithm for future devices). Accuracy data will be calculated based on comparing calibrated glucose sensor values to a "gold standard" (YSI plasma glucose values) in adult subjects during in-clinic testing, and on comparing calibrated glucose sensor values to glucose meter values during in-home testing. The devices that will be used for gathering sensor data during this study will be: 1) the Comfort Sensor, 2) the MiniLink (Transmitter), 3) the Guardian REAL-Time Display Device and 4) the CGMS iPro.


Recruitment information / eligibility

Status Completed
Enrollment 100
Est. completion date November 2010
Est. primary completion date November 2010
Accepts healthy volunteers No
Gender All
Age group 18 Years to 75 Years
Eligibility Inclusion Criteria:

1. Males and females aged 18 through 75 inclusive

2. Diagnosed with Type 1 or type 2 Diabetes Mellitus, using insulin to treat their diabetes (Each site's and overall study enrollment of subjects with Type 2 diabetes will target 20% of the total number)

3. Willingness to perform required study and data collection procedures and to adhere to operating requirements of the Guardian REAL-TIME and CGMS iPro Systems

4. Willingness to perform at least 4 capillary blood glucose tests per day while wearing the Guardian REAL-TIME and CGMS iPro Systems

5. Willingness to participate in a 10 hour frequent blood sampling session during the study

6. Subject agrees to comply with the study protocol requirements

7. Informed Consent, HIPAA Authorization, and California Experimental Subject Bill of Rights (if applicable) signed by the subject

8. The Subject is willing to wear both the Guardian REAL-Time and CGMS iPro Systems for 14 days (~340 hours)

Exclusion Criteria:

1. The Subject has a history of tape allergies that have not been resolved

2. The Subject has any skin abnormality (e.g. psoriasis, rash, staphylococcus infection) that has not been resolved and would inhibit them from wearing the sensors

3. Any additional condition(s) that in the Investigator's opinion would warrant exclusion from the study or prevent the Subject from completing the study

4. Subject is currently participating in an investigational study (drug or device)

Study Design


Intervention

Device:
Sensor wear
All subjects to wear sensors

Locations

Country Name City State
United States AMCR Institute, Inc. Escondido California
United States Rainier Clinical Research Center, Inc. Renton Washington
United States Sansum Diabetes Research Institute Santa Barbara California

Sponsors (1)

Lead Sponsor Collaborator
Medtronic Diabetes

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Glucose Sensor Accuracy When Compared to Laboratory Standard (YSI): Proportion of Glucose Sensor Readings That Met Accuracy Criteria The primary accuracy parameter (primary effectiveness endpoint) was the comparative readings of paired sensor and YSI glucose readings, measured on days 1 through 6. Accuracy is defined as within 20% agreement between YSI and paired sensor (within 20 mg/dL if YSI <80 mg/dL). Accuracy ranges from 0 - 100, with higher number suggests better accuracy. Days one through six of sensor use
Secondary Device Related Moderate or Device Related Severe Adverse Events Device related moderate adverse event: low level of inconvenience or concern to the subject and may interfere with daily activities but is usually improved by simple therapeutic remedy.
Device related severe adverse event: interrupts a subject's daily activity and typically requires intervening treatment.
Note: device related determination is made by the site that there is a reasonable possibility that the adverse event may have been caused by the device.
days one through six of sensor wear
See also
  Status Clinical Trial Phase
Completed NCT02771093 - An Exploratory Study of the Effects of Trelagliptin and Alogliptin on Glucose Variability in Patients With Type 2 Diabetes Mellitus Phase 4
Completed NCT02545842 - Assessment Study of Three Different Fasting Plasma Glucose Targets in Chinese Patients With Type 2 Diabetes Mellitus (BEYOND III/FPG GOAL) Phase 4
Recruiting NCT03436212 - Real-Life Home Glucose Monitoring Over 14 Days in T2D Patients With Intensified Therapy Using Insulin Pump. N/A
Completed NCT03244800 - A Study to Investigate Different Doses of 0382 in Overweight and Obese Subjects With Type 2 Diabetes Mellitus. Phase 2
Completed NCT03960424 - Diabetes Management Program for Hispanic/Latino N/A
Withdrawn NCT02769091 - A Study in Adult Patients With Nonalcoholic Steatohepatitis Who Also Have Type 2 Diabetes Phase 2
Recruiting NCT06065540 - A Research Study to See How Well CagriSema Compared to Semaglutide, Cagrilintide and Placebo Lowers Blood Sugar and Body Weight in People With Type 2 Diabetes Treated With Metformin With or Without an SGLT2 Inhibitor Phase 3
Recruiting NCT05008276 - Puberty, Diabetes, and the Kidneys, When Eustress Becomes Distress (PANTHER Study)
Completed NCT04091373 - A Study Investigating the Pharmacokinetics of a Single Dose Administration of Cotadutide Phase 1
Completed NCT03296800 - Study to Evaluate Effects of Probenecid, Rifampin and Verapamil on Bexagliflozin in Healthy Subjects Phase 1
Recruiting NCT06212778 - Relationship Between Nutritional Status, Hand Grip Strength, and Fatigue in Hospitalized Older Adults With Type 2 Diabetes Mellitus.
Completed NCT05979519 - Fresh Carts for Mom's to Improve Food Security and Glucose Management N/A
Recruiting NCT05579314 - XW014 in Healthy Subjects and Patients With Type 2 Diabetes Mellitus (T2DM) Phase 1
Completed NCT03859934 - Metabolic Effects of Melatonin Treatment Phase 1
Terminated NCT03684642 - Efficacy and Safety of Efpeglenatide Versus Dulaglutide in Patients With Type 2 Diabetes Mellitus Inadequately Controlled With Metformin Phase 3
Completed NCT03248401 - Effect of Cilostazol on Carotid Atherosclerosis Estimated by 3D Ultrasound in Patients With Type 2 Diabetes Phase 4
Completed NCT03644134 - A Personalized Intervention to Manage Physiological Stress and Improve Sleep Patterns N/A
Completed NCT05295160 - Fasting-Associated Immune-metabolic Remission of Diabetes N/A
Completed NCT02836873 - Safety and Efficacy of Bexagliflozin in Type 2 Diabetes Mellitus Patients With Moderate Renal Impairment Phase 3
Completed NCT02226003 - Efficacy and Safety of Ertugliflozin (MK-8835/PF-04971729) With Sitagliptin in the Treatment of Participants With Type 2 Diabetes Mellitus (T2DM) With Inadequate Glycemic Control on Diet and Exercise (MK-8835-017) Phase 3