Cataract Clinical Trial
Official title:
Spherical Aberration and Contrast Sensitivity in Eyes Implanted With Aspheric and Spherical Intraocular Lenses: Clinical Comparative Study
Purpose: To determine whether implantation of an intraocular lens (IOL) with a modified
anterior aspheric surface results in reduced spherical aberration and improved contrast
sensitivity after cataract surgery.
Design: Prospective, comparative, interventional case series.
Methods: In an intraindividual randomized prospective study of 25 patients with bilateral
cataract, an IOL with a modified anterior surface (Tecnis Z9001, AMO- Group 1) was compared
with biconvex lens with spherical surfaces (ClariFlex®, AMO- Group 2). Ocular aberrations
for a 5.0 mm pupil and 6.0 mm pupil were measured with Hartmann-Shack aberrometer. Quality
of vision was measured by visual acuity and contrast sensitivity under mesopic and photopic
conditions. All patients were followed for 3 months.
With improvement in the quality of life and the resulting expansion of the elderly
population, the prevalence of cataract cases will continue to rise. The population is on
average older, and the environment has become more demanding for older people, who therefore
need to regain excellent quality of vision. Advances in both IOL and phacoemulsification
technology have enable cataract surgery to evolve from a procedure concerned with the safe
removal of the cataract to a procedure refined to achieve the best possible postoperative
refractive result.
Snellen visual acuity insufficiently describes the quality of the eye's optics before and
after cataract surgery. 1-3 Multiple scientific studies have demonstrated that contrast
sensitivity represents a robust indicator of functional vision. 4-7 The contrast sensitivity
function, measured under varying conditions of luminance and glare, establishes the limits
of visual perception across the spectrum of spatial frequencies.
Studies have demonstrated a nearly linear decline in image quality with age, suggesting a
significant increase in the optical aberrations in the eye over time. 8-9 Advances in
wavefront technology have opened a new door to the measurement of ocular aberrations. 9-10
Also, this technology allowed the development of a new type of IOL designed to compensate
for the positive spherical aberration of the cornea, which is one of the most important
aberrations contributing to visual deterioration of the pseudophakic eye. 11-14 Based on
these findings, the approach to compensate for increasing spherical aberration in older eyes
is to develop an IOL that introduces negative spherical aberration into the system. Such an
IOL, the Tecnis Z9001® IOL (Model Z9001, Advanced Medical Optics, Santa Ana, CA), has been
developed by AMO. This is a polysiloxane foldable 3-piece IOL, with anterior aspheric
surface (negative spherical aberration) designed to decrease the total amount of ocular
spherical aberration after cataract surgery.7
The purpose was to determine whether implantation of an intraocular lens (IOL) with a
modified anterior aspheric surface results in reduced spherical aberration and improved
contrast sensitivity after cataract surgery.
Patients and Methods
This double-blind prospective randomized study included patients with age-related cataract,
no indication of existing ocular pathology, unsatisfactory correction with glasses and less
than 2.50 diopters (D) of topography cylinder. Patients were offered the opportunity to be
part of a clinical trial in which they would be allocated to have cataract surgery with
implantation of aspherical or spherical IOL. Patients were randomized to receive a Tecnis
Z9001 aspheric IOL in one eye and a ClariFlex spheric IOL in the fellow eye. To protect
patient safety, after the first eye surgery, they were asked whether they wished to have the
same lens type implanted in the second eye. Written informed consent was obtained from all
patients before surgery and the study was approved by the local ethics committee. Exclusion
criteria were previous ocular surgery, central endothelial cell count less than 1800
cells/mm2, glaucoma or intraocular pressure greater than 21 mmHg, amblyopic eyes, retinal
abnormalities, diabetes mellitus, steroid or immunosuppressive treatment, and connective
tissue diseases.
The selected lenses used in this study were the Tecnis Z9001® (Model Z9001, Advanced Medical
Optics, Santa Ana, CA) and the ClariFlexTM (Advanced medical Optics, Santa Ana, CA) (Table
1). Eyes in Group 1 (n=25) received the Tecnis Z9001® (Model Z9001) and eyes in Group 2
(n=25) received the ClariFlexTM (Table 1).
The Holladay formula was used to calculate the IOL power when there was a short axial length
(< 22.0 mm) and the SRK/T formula was used when there was an average axial length (≥ 22.0
mm). The A-constant used was 119.1 for the Tecnis Z9001 IOL and 118.0 for the ClariFlex IOL.
Axial length was measured with the IOL Master (Carl Zeiss Meditec AG), and the targeted
postoperative refractive error was 0.0.
Preoperative and postoperative evaluation included uncorrected distance visual acuity
(UCDVA), best corrected visual acuity (BCVA), spherical equivalent (SE), slitlamp
biomicroscopy, applanation tonometry, fundus examination, B-scan biometry, specular
microscope and corneal topography. Topography was performed in all patients using the EyeSys
unit (Version 3.03; EyeSys Technologies, Houston, Texas). The corneal endothelial cell count
was also recorded for all eyes (noncontact specular microscope, NonCon Robo, Konan). The
visual acuity measurements were recorded with logMAR UCDVA and BCVA. Clinical data were
collected preoperatively and 1 and 3 months postoperatively for each eye.
All patients were operated in the same fashion by the same surgeon. All patients received
topical anesthesia by lidocaine 2% gel before surgery. A 2.75 mm self-sealing clear cornea
incision was made on the temporal side. Viscoelastic solution of sodium hyaluronate 3% and
chondroitin sulfate 4% (Viscoat®) was used to reform and stabilize the surgical planes and
protect the endothelium. A 5.00 to 5.25 mm continuous curvilinear capsulorhexis was
initially performed with a 26-gauge needle and completed with forceps. The nucleous was
removed without intraoperative complications such as posterior capsule rupture.
Phacoemulsification was performed using the Infinite (Alcon Surgical) or Sovereign machine
(Allergan Surgical). All IOLs were inserted in the capsular bag with the injector system.
The viscoelastic material was completely removed at the end of the procedure. No sutures
were used in any case. After 1 week the patient had the fellow eye operated. Postoperative
medication included moxifloxacin (Vigamox®) or gatifloxacin (Zymar®) 4 times a day for 2
weeks, 0.1% diclofenaco sodium (Voltaren) 3 times a day for 4 weeks and steroid (Predfort)
eyedrops 4 times a day for 6 weeks. Postoperative follow-up was 1 day, 1 and 3 months.
Postoperative evaluations were performed at 1 day, 1 month, and 3 months. At 1 month and 3
months, visual acuity, refraction, contrast sensitivity (CSV-1000 HGT, Vector Vision Inc.),
and wavefront analysis with the Zywave aberrometer (Bauch & Lomb, Rochester, New York) were
performed.
Contrast sensitivity was measured using the CSV-1000 HGT testing instrument (Vector Vision
Inc.), which presents a translucent chart divided into 4 cycles with spatial frequencies of
3, 6, 12, and 18 cycles per degree (cpd). The background illumination of the translucent
chart does not depend on room lighting; rather, it is provided by a fluorescent luminance
source of the instrument and is automatically calibrated to 85 candelas/m2. All measurements
were obtained under mesopic (5 cd/m2) and photopic (85 cd/m2) conditions. Each cycle
contains 17 round patches that are 1.5 inches in diameter. The first patch has a
high-contrast grating and presents the sample. The test patches are arranged in 2 rows with
8 levels of contrast. The levels decreased from left to right along the row in a logarithmic
fashion in 0.17 log units for steps 1 through 3 and 0.15 log units for steps 3 through 8.
The examinations were performed unilaterally at a distance of 2.5 m with best corrected
visual acuity (BCVA) and an undilated pupil. All measurements were conducted under the same
conditions by an examiner who was unaware of the type of IOL implanted.
Wavefront analysis was performed by the Zywave aberrometer. The Zywave aberrometer uses the
Hartmann-Shack method of analysis of the outgoing wavefront that measures up to fifth-order
Zernike aberrations, including coma, trefoil, and spherical aberrations.
For statistical analysis of visual acuity, logarithm of minimum angle of resolution (logMAR)
acuity values were used. Similarly, the recorded contrast sensitivity values were
transformed into log values as described by Vector Vision. All data analyses were performed
using SPSSX statistical programs (SPSS Inc, Chicago, IL). The 2 IOLs were compared between
eyes intraindividually. The analysis was based on a non-normal distribution of the data. The
nonparametric Mann-Whitney U test was used to compare data between the 2 IOL groups. A P
value less than 0.05 was considered statistically significant.
;
Allocation: Randomized, Endpoint Classification: Efficacy Study, Intervention Model: Parallel Assignment, Masking: Double Blind (Subject, Caregiver, Investigator), Primary Purpose: Treatment
Status | Clinical Trial | Phase | |
---|---|---|---|
Completed |
NCT04685538 -
Chloroprocaine 3% Gel Eye Drop as Topical Anestheticsin Phacoemulsification.
|
Phase 3 | |
Recruiting |
NCT06060041 -
IC-8 Apthera IOL New Enrollment Post Approval Study
|
||
Recruiting |
NCT05518539 -
Evaluation of Quality of Vision and Visual Outcomes With Bilateral Implantation of the Clareon PanOptix Intraocular Lens
|
||
Recruiting |
NCT05271942 -
Tilt and Tumble vs Divide and Conquer - a Unique Comparison of the Two Cataract Surgery Methods
|
N/A | |
Active, not recruiting |
NCT04778501 -
PMCF Study on Monofocal Toric IOL (PODEYE TORIC) in Asia
|
N/A | |
Completed |
NCT05062564 -
Efficacy of LipiFlow in Patients Affected by Meibomian Gland Dysfunction in Reducing Post-cataract Surgery Dry Eye
|
N/A | |
Completed |
NCT03751033 -
Influence of DisCoVisc Ophthalmic Viscosurgical Device (OVD) on Intraoperative Aberrometry Readings
|
N/A | |
Completed |
NCT02529488 -
Investigation of AcrySof® IQ PanOptix™ Presbyopia-Correcting Intraocular Lens (IOL) Model TFNT00
|
N/A | |
Completed |
NCT04539548 -
A Study Assessing the Safety and Efficacy of Dextenza® for the Treatment of Ocular Pain and Inflammation Following Surgery for Pediatric Cataract
|
Phase 3 | |
Completed |
NCT03740659 -
Evaluation Of Aqueous Humor Of Levofloxacin-Dexamethasone Eye Drops And Of Its Components In Patients Undergoing Cataract Surgery
|
Phase 2 | |
Completed |
NCT03494257 -
Effect of Fixed Brinzolamide-brimonidine Combination on Intraocular Pressure After Phacoemulsification
|
N/A | |
Completed |
NCT05119127 -
Rotational Stability of Acrysof IQ Vivity Extended Vision Toric IOL and Refractive Visual Outcome.
|
N/A | |
Active, not recruiting |
NCT04271709 -
Manhattan Vision Screening and Follow-Up Study (NYC-SIGHT)
|
N/A | |
Recruiting |
NCT03713268 -
Intraoperative OCT Guidance of Intraocular Surgery II
|
||
Completed |
NCT03739528 -
Levo-Dexa vs. Tobra+Dexa for Prevention and Treatment of Inflammation and Prevention of Infection in Cataract Surgery
|
Phase 3 | |
Completed |
NCT02888210 -
A Study Assessing Safety and Efficacy of MD-15 Intraocular Lens in Patients With Aphakic Eye After Cataract Surgery
|
Phase 3 | |
Completed |
NCT03356847 -
Evaluation of the Rotational Stability of the Monofocal SISA Implant Following Cataract Surgery
|
N/A | |
Completed |
NCT04332640 -
Clinical Evaluation of the Next Generation Phaco System
|
N/A | |
Recruiting |
NCT03638726 -
Subconjunctival Atropine and Intracameral Epinephrine for Pupil Dilation in Phacoemulsification
|
Phase 4 | |
Completed |
NCT03050697 -
Evaluation of the Safety and Performance of the HARMONI® Toric Lens
|
N/A |