Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT04649177
Other study ID # BFS-OCT-01
Secondary ID
Status Completed
Phase N/A
First received
Last updated
Start date December 15, 2020
Est. completion date March 23, 2022

Study information

Verified date October 2023
Source Boston Sight
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

EYEdeal Scanning Technology enables rapid measurement of corneal and scleral topography. By accurately measuring the ocular surface with this imaging technology, the current iterative fitting method required to fit PROSE (prosthetic replacement of the ocular surface ecosystem) devices could be replaced and/or strengthened with a more rapid, automated fitting sequence. This could considerably reduce the time needed per visit, the number of visits, and the number of devices needed to be manufactured to reach the endpoint. Additionally, the scanning technology may afford the opportunity to successfully fit some pathology that were previously treatment failures. The automated technology may as well reduce the intensive clinician training time needed to fit PROSE devices, therefore increasing the availability and access to patients. The goal of this research is to evaluate the reliability and efficacy of this automated technology for fitting PROSE devices. Data from real-time measurement of the human eye ocular surface topography will be used to fabricate a prosthetic lens. The fit of the PROSE device will be evaluated, as well as the subjective comfort of the fit.


Description:

Inclusion and Exclusion Criteria The participant will be eligible to participate if the following criteria apply: 1. Written Informed Consent has been obtained prior to any study-related procedures 2. Male or female, 18 years of age and older prior to the initial visit 3. Established wearer of PROSE in the study eye with an optimized fit in the opinion of the clinician 4. No medical need for a PROSE retreatment or replacement lens at the time of enrollment in the study eye, in the opinion of the clinician 5. Initial PROSE fitting was initiated and completed at BostonSight, Needham 6. Current PROSE device does not have channels or fenestrations 7. Currently wears a PROSE device in the study eye primarily for an irregular ocular surface /irregular cornea (including but not limited to Keratoconus, Pellucid Marginal Degeneration, Ectasia, Post-Penetrating Keratoplasty, Corneal Scar, Trauma) AND/OR ocular surface disease 8. Has the ability to NOT wear the current PROSE device in the study eye for 3 days prior to scan and 3 days prior to first fitting appointment in the opinion of the investigator and subject 9. In the opinion of the investigator, the subject has the ability to follow study instructions 10. In the opinion of the investigator, the subject has the ability to complete all study procedures and visits The participant would NOT be eligible to participate if at least one of the following criteria is met: 1. Is currently participating in any other type of eye-related clinical or research study 2. Is pregnant or nursing as reported by the subject 3. Has a condition or is in a situation which, in the investigator's opinion, may put the subject at significant risk, may confound study outcomes, or may significantly interfere with the subject's participation in the study 4. Has had previous ocular surgery within the past 12 weeks 5. Intolerance to PROSE wear 6. Inability to maintain stable fixation and exposure for ocular surface imaging 7. Corneal touch by the posterior surface of the device in current PROSE device 8. Allergy to sodium fluorescein 9. Patient is an employee of BostonSight 10. Subject is currently incarcerated.


Recruitment information / eligibility

Status Completed
Enrollment 11
Est. completion date March 23, 2022
Est. primary completion date March 23, 2022
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 21 Years and older
Eligibility Inclusion Criteria: 1. Written Informed Consent has been obtained prior to any study-related procedures 2. Male or female, 18 years of age and older prior to the initial visit 3. Established wearer of PROSE in the study eye with an optimized fit in the opinion of the clinician 4. No medical need for a PROSE retreatment or replacement lens at the time of enrollment in the study eye, in the opinion of the clinician 5. Initial PROSE fitting was initiated and completed at BostonSight, Needham 6. Current PROSE device does not have channels or fenestrations 7. Currently wears a PROSE device in the study eye primarily for an irregular ocular surface /irregular cornea (including but not limited to Keratoconus, Pellucid Marginal Degeneration, Ectasia, Post-Penetrating Keratoplasty, Corneal Scar, Trauma) AND/OR ocular surface disease 8. Has the ability to NOT wear the current PROSE device in the study eye for 3 days prior to scan and 3 days prior to first fitting appointment in the opinion of the investigator and subject 9. In the opinion of the investigator, the subject has the ability to follow study instructions 10. In the opinion of the investigator, the subject has the ability to complete all study procedures and visits Exclusion Criteria: 1. Is currently participating in any other type of eye-related clinical or research study 2. Is pregnant or nursing as reported by the subject 3. Has a condition or is in a situation which, in the investigator's opinion, may put the subject at significant risk, may confound study outcomes, or may significantly interfere with the subject's participation in the study 45. Has had previous ocular surgery within the past 12 weeks 5. Intolerance to PROSE wear 6. Inability to maintain stable fixation and exposure for ocular surface imaging 7. Corneal touch by the posterior surface of the device in current PROSE device 8. Allergy to sodium fluorescein 9. Patient is an employee of BostonSight 10. Subject is currently incarcerated. -

Study Design


Intervention

Device:
Ocular surface optical coherence tomography
EYEdeal Scanning technology enables rapid measurement of the corneal and scleral topography which provides data for an automated sequence to fit PROSE devices

Locations

Country Name City State
United States BostonSight Needham Massachusetts

Sponsors (2)

Lead Sponsor Collaborator
Boston Sight EYEdeal scanning, LTC

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Central Corneal Clearance Central corneal clearance as measured by OCT after 45 minutes of lens wear Outcome measure reports data only from the study eye. 45 minutes post lens application
Primary Lens Decentration Lens decentration after 45 minutes of lens wear (x axis) Outcome measure reports data only from the study eye. 45 minutes post lens application
Primary Lens Comfort After 45 Minutes of Lens Wear 0 to 10 grading (0 = very comfortable to 10 = very uncomfortable) Outcome measure reports data only from the study eye. 45 minutes post lens application
Primary Number of Participants With Lens Centration Each study eye (one eye from each participant) was evaluated to determine if the lens was "Centered" (yes) or "Decentered" (no) on the eye after 45 minutes of wear.
Outcome measure reports data only from the study eye. All participants were reported in this outcome measure.
45 minutes post lens application
Primary Number of Participants With Aligned Scleral Landing Zone Each study eye (one eye from each participant) was evaluated to determine if the scleral landing zone was properly aligned on the eye (yes) or not properly aligned on the eye (no) after 45 minutes of wear.
Outcome measure reports data only from the study eye. All participants were reported in this outcome measure.
45 minutes post lens application
Secondary Lens Rotation Lens rotation after 45 minutes of wear. Lens rotation was measured using an embedded program in the iTRACE system. The rotation data with image guided fitting with the OCT system was collected to compare with known experience with non-image guided fitting sets (the traditional fitting process), which typically rotate approximately 20-30 degrees on the eye. A lower number may indicate the manufactured lens more closely approximates the contours of the eye.
Outcome measure reports data only from the study eye.
45 minutes post lens application
Secondary Number of Participants Achieving an Acceptable Fitting Lens Number of participants achieving an acceptable fitting lens after 45 minutes of lens wear. All participants are reported. Outcome measure reports data only from the study eye.
Is the fit acceptable for dispensing to the patient (yes) or is the fit NOT acceptable for dispensing to patient (no) as evaluated 45 minutes post lens application.
45 minutes post lens application
Secondary Number of Participants With Lens Movement Number of patients with lens movement after 45 minutes of wear are reported. Lens movement (YES) means the lens is NOT appropriately fitting/settling on the eye. Lens movement (NO) means the lens is appropriately fitting/settling on eye. Outcome measure reports data only from the study eye. 45 minutes
Secondary Number of Participants With Adequate Lens Limbal Clearance Each study eye (one eye from each participant) was evaluated to determine if there was adequate (yes) lens limbal clearance or inadequate (no) lens limbal clearance after 45 minutes of wear.
Outcome measure reports data only from the study eye. All participants were reported in this outcome measure.
45 minutes post lens application
See also
  Status Clinical Trial Phase
Active, not recruiting NCT04425551 - Effect of Micropulse Laser on Dry Eye Disease Due to Meibomian Gland Dysfunction N/A
Not yet recruiting NCT06379685 - Study to Evaluate the Safety and Tolerability of PRO-190 Ophthalmic Solution Compared to Systane Ultra® on the Ocular Surface. Phase 1
Recruiting NCT04701086 - 3 Month Study of Cationorm Pro Versus Vismed in Adults With Dry Eye Disease Related to Keratitis or Keratoconjunctivitis N/A
Active, not recruiting NCT03697876 - Safety and Tolerability of the Ophthalmic Gel PRO-165 Versus Artelac® Nightime Gel Phase 1
Terminated NCT02815293 - Topical Ophthalmic AGN-195263 for the Treatment of Evaporative Dry Eye Phase 3
Completed NCT02910713 - Evaluation of Dry Eye Symptoms in CAE With Application of Intranasal Neurostimulation N/A
Completed NCT04104997 - A Clinical Trial to Evaluate the Safety, Tolerability and Pharmacokinetic Characteristics of GLH8NDE After Single and Mutiple Ocular Administrations in Healthy Korean and Caucasian Volunteers Phase 1
Recruiting NCT02595606 - 0.3% Sodium Hyaluronate in the Treatment of Dry Eye of Diabetic Patients Phase 4
Completed NCT01711424 - An Observational Study of OPTIVE PLUS® for the Treatment of Dry Eye Disease N/A
Completed NCT01015209 - Safety and Tolerability of Chitosan-N-acetylcysteine Eye Drops in Healthy Young Volunteers Phase 1
Completed NCT01202747 - Evaluation of Screening Methods for Treatment of Meibomian Gland Dysfunction Phase 2/Phase 3
Completed NCT00969280 - Acupuncture for Dry Eye Syndrome Phase 3
Completed NCT01496482 - Comparison of Evaporimetry With the Established Methods of Tear Film Measurement N/A
Completed NCT00739713 - Effects of Sea Buckthorn Oil on Dry Eye N/A
Completed NCT00756678 - Efficacy and Acceptability of Two Lubricant Eye Drops Phase 4
Completed NCT00349440 - Efficacy of Cyclosporine for the Prevention and Treatment of Dry Eye Symptoms Following LASIK or Photorefractive Keratectomy Phase 4
Completed NCT00370747 - Efficacy and Safety Study for Ecabet Ophthalmic Solution for Treating Dry Eye Syndrome Phase 2
Completed NCT05162261 - to Evaluate the Effectiveness and Safety of the Tixel® , VS LipiFlow® in the Treatment of Meibomian Gland Dysfunction N/A
Completed NCT02871440 - A Two Comparator, Controlled Phase 3 Study in Patients With and Without Evaporative Dry Eye Phase 3
Completed NCT05042960 - Computer Screen Properties Study N/A