Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT02283411
Other study ID # ADC-US-VAL-14135
Secondary ID
Status Completed
Phase N/A
First received October 31, 2014
Last updated January 22, 2018
Start date November 2014
Est. completion date March 2015

Study information

Verified date January 2018
Source Abbott Diabetes Care
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The purpose of the study is to evaluate the safety and effectiveness of the Abbott Sensor Based Glucose Monitoring System Personal (System-P) and Abbott Sensor Based Glucose Monitoring System Professional (System-Pro) when used as an adjunct to blood glucose testing over a 14-day wear period in adult subjects. The Abbott Sensor Based Glucose Monitoring System - Professional Sensors will be worn by adult subjects over a 14-day wear period.

The primary objective is to characterize the Systems performance with respect to Yellow Spring Instrument (YSI) reference venous sample measurements. The device performance will be primarily evaluated in terms of point and rate accuracy of the Abbott Sensor Based Glucose Monitoring Systems in reference to YSI.

Safety of the Abbott Sensor Based Glucose Monitoring Systems will be characterized by Adverse Device Effects and Serious Adverse Device Effects experienced by study participants.

240 subjects will be enrolled and additional healthy subjects may also be enrolled in the study as training subjects.


Description:

Sensors are inserted into the back of the subject's upper arm for up to fourteen (14) days. The subjects are expected to perform capillary BG tests during the 14-day sensor wear. Three in clinic visits are scheduled during the 14-day sensor wear period. During each in-clinic visit, study staff performed IV blood draws to obtain venous blood for YSI reference glucose measurements.


Recruitment information / eligibility

Status Completed
Enrollment 156
Est. completion date March 2015
Est. primary completion date March 2015
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria:

1. Subject must be at least 18 years of age.

2. Subject must have a diagnosis of type 1 or type 2 diabetes mellitus.

3. Use one of the following for their diabetes management:

- intensive insulin therapy (insulin therapy through an insulin pump and/or multiple daily insulin injections) or

- non-intensive insulin therapy (Basal insulin, injectable non-insulin anti-diabetic agents such as a pramlintide or GLP-1 agonists, NPH insulin either alone or in combination with regular insulin or fast acting insulin analogue, or oral medication)

4. Intensive-insulin using (IIT) subjects only - During one of the in-clinic sessions, subject is willing to have their blood glucose levels manipulated into high and low glucose levels. Insulin-to-carbohydrate ratio and insulin sensitivity formula must be documented prior to having blood glucose levels manipulated. NIT subjects will only be observed during each in-clinic session.

5. For subjects that exercise routinely (at least 3 times per week), willing to exercise during at least one of the in-clinic sessions, if asked.

6. Willing to perform a minimum of 8 finger sticks per day during the study.

7. Subject must be able to read and understand English.

8. In the investigator's opinion, the subject must be able to follow the instructions provided to him/her by the study site and perform all study tasks as specified by the protocol.

9. Subject must be available to participate in all study visits.

10. Subject must be willing and able to provide written signed and dated informed consent.

Exclusion Criteria:

1. Subject has known allergy to medical grade adhesive or isopropyl alcohol used to disinfect skin.

2. Subject is pregnant, is attempting to conceive or is not willing and able to practice birth control during the study duration (applicable to female subjects only).

3. Subject has extensive skin changes/diseases at the proposed application sites that could interfere with device placement or the accuracy of interstitial glucose measurements. Such conditions include, but are not limited to extensive psoriasis, recent burns or severe sunburn, extensive eczema, extensive scarring, dermatitis herpetiformis, skin lesions, redness, infection or edema.

4. Subject is on dialysis.

5. Subject has a history of untreated hypothyroidism, adrenal gland failure or insufficiency, celiac disease, renal failure, cystic fibrosis, unstable coronary heart disease, or serious psychiatric disorders.

6. Subjects with a history of hypoglycemia unawareness, hypoglycemic seizures or unconsciousness.

7. Subjects who have experienced diabetes-related complications requiring assistance from another person in the last six months.

8. Subject currently is participating in another clinical trial.

9. Subject has donated blood within 112 days (3.7 months) prior to the beginning of the study activities.

10. Subject is anemic.

11. Subject does not have a known insulin to carbohydrate ratio at the time of enrollment (IIT subjects only).

12. Subject has concomitant medical condition which, in the opinion of the investigator, could interfere with the study or present a risk to the safety or welfare of the subject or study staff. Such conditions include but are not limited to:

- History of HIV, Hepatitis B or C

13. Subject has X-ray, MRI or CT appointment scheduled during the period of study participation, and the appointment cannot be rescheduled for a time before study participation starts or after study participation ends.

14. Subject has an abnormal EKG, unless cleared for study participation by a cardiologist.

15. Subject is unsuitable for participation due to any other cause as determined by the Investigator.

Additional Criteria

16. The following subjects may participate in the study but will be excluded from all glycemic and exercise challenges:

- Subject's taking medications known to mask symptoms or hypoglycemia, such as beta blockers;

- Subjects with any of the following: clinically significant history of cardiovascular or cerebrovascular disease, cardiac arrhythmia, neurological disorders such as seizures, CVA, or syncope, or hypokalemia.

Study Design


Intervention

Device:
Abbott Sensor Based Glucose Monitoring System
Subjects will wear the Abbott Sensor Based Glucose Monitoring Systems and will receive no treatment except for safety purposes.

Locations

Country Name City State
United States Atlanta Diabetes Associates Atlanta Georgia
United States Rocky Mountain Diabetes & Osteoporosis Center Idaho Falls Idaho
United States Rainier Clinical Research Center Renton Washington
United States Clinical Trials of Texas San Antonio Texas
United States William Sansum Diabetes Center Santa Barbara California
United States Diablo Clinical Research Walnut Creek California

Sponsors (1)

Lead Sponsor Collaborator
Abbott Diabetes Care

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary The Device Performance Was Evaluated in Terms of Point Accuracy of the Abbott Sensor Based Glucose Monitoring Systems in Reference to Yellow Spring Instrument (YSI). Point accuracy of the system was evaluated as the proportion of System readings that are within ±20% of the YSI reference value for glucose levels = 80 mg/dL and within ±20 mg/dL for YSI glucose levels <80 mg/dL. All 125 subjects wore both System-P and System-Pro Sensors and were included in the outcome measure. System P, the Personal System which is intended for single patient use and System Pro is intended for use by healthcare professionals (HCP). Each subject had up to 136 YSI samples collected during the study which were paired with sensor glucose readings measured at the same time. Point accuracy of the system was evaluated as the percentage of sensor glucose readings that are within ±20% of the YSI reference value for glucose levels = 80 mg/dL and within ±20 mg/dL for YSI glucose levels <80 mg/dL. 14 days
Primary Safety of the Abbott Sensor Based Glucose Monitoring Systems Was Characterized by Adverse Device Effects and Serious Adverse Device Effects Experienced by Study Participants. Safety of the Abbott Sensor Based Glucose Monitoring Systems, including adverse device effects and serious adverse device effects were assessed for all participants enrolled in the study. Safety was evaluated throughout the subject's study participation.
Secondary The Trend Accuracy of the System P Device Performance Was Evaluated. Trend accuracy of the device performance was assessed by comparing of the glucose rate of change results between the sensor glucose readings and YSI reference results for System P only. 14 days
Secondary The Temporal System Accuracy at Different Glucose Rates of Change and Different Glucose Ranges (Hypoglycemic, Euglycemic, and Hyperglycemic Ranges) Was Evaluated. The Continuous Glucose Error Grid Analysis (CG-EGA) in the Hypoglycemic range (YSI = 70 mg/dL), in the Euglycemic range (70 < YSI = 180 mg/dL) and in the Hyperglycemic range (YSI > 180 mg/dL) were evaluated. 14 days
Secondary The Point Accuracy of the Device Performance Was Evaluated in Reference to Capillary Blood Glucose (BG). Point accuracy of the system was evaluated as the proportion of System readings that are within ±20% of the capillary blood glucose (BG) value for glucose levels = 80 mg/dL and within ±20 mg/dL for capillary blood glucose (BG) levels <80 mg/dL. All 125 subjects wore both System-P and System-Pro Sensors and were included in the outcome measure. System P, the Personal System which is intended for single patient use and System Pro is intended for use by healthcare professionals (HCP). Each subject had approximately112 capillary blood glucose (BG) samples collected during the study which were paired with sensor glucose readings measured at the same time. Point accuracy of the system was evaluated as the percentage of sensor glucose readings that are within ±20% of the BG value for glucose levels = 80 mg/dL and within ±20 mg/dL for BG levels <80 mg/dL. 14 days
See also
  Status Clinical Trial Phase
Active, not recruiting NCT05666479 - CGM Monitoring in T2DM Patients Undergoing Orthopaedic Replacement Surgery
Completed NCT05647083 - The Effect of Massage on Diabetic Parameters N/A
Active, not recruiting NCT05661799 - Persistence of Physical Activity in People With Type 2 Diabetes Over Time. N/A
Completed NCT03686722 - Effect of Co-administration of Metformin and Daclatasvir on the Pharmacokinetis and Pharmacodynamics of Metformin Phase 1
Completed NCT02836704 - Comparison of Standard vs Higher Starting Dose of Insulin Glargine in Chinese Patients With Type 2 Diabetes (Glargine Starting Dose) Phase 4
Completed NCT01819129 - Efficacy and Safety of FIAsp Compared to Insulin Aspart in Combination With Insulin Glargine and Metformin in Adults With Type 2 Diabetes Phase 3
Completed NCT04562714 - Impact of Flash Glucose Monitoring in People With Type 2 Diabetes Using Non-Insulin Antihyperglycemic Therapy N/A
Completed NCT02009488 - Treatment Differences Between Canagliflozin and Placebo in Insulin Secretion in Subjects With Type 2 Diabetes Mellitus (T2DM) Phase 1
Completed NCT05896319 - Hyaluronic Acid Treatment of the Post-extraction Tooth Socket Healing in Subjects With Diabetes Mellitus Type 2 N/A
Recruiting NCT05598203 - Effect of Nutrition Education Groups in the Treatment of Patients With Type 2 Diabetes N/A
Completed NCT05046873 - A Research Study Looking Into Blood Levels of Semaglutide and NNC0480-0389 When Given in the Same Injection or in Two Separate Injections in Healthy People Phase 1
Terminated NCT04090242 - Impact of App Based Diabetes Training Program in Conjunction With the BD Nano Pen Needle in People With T2 Diabetes N/A
Completed NCT04030091 - Pulsatile Insulin Infusion Therapy in Patients With Type 1 and Type 2 Diabetes Mellitus Phase 4
Completed NCT03604224 - A Study to Observe Clinical Effectiveness of Canagliflozin 300 mg Containing Treatment Regimens in Indian Type 2 Diabetes Participants With BMI>25 kg/m^2, in Real World Clinical Setting
Completed NCT03620357 - Continuous Glucose Monitoring & Management In Type 2 Diabetes (T2D) N/A
Completed NCT01696266 - An International Survey on Hypoglycaemia Among Insulin-treated Patients With Diabetes
Completed NCT03620890 - Detemir Versus NPH for Type 2 Diabetes Mellitus in Pregnancy Phase 4
Withdrawn NCT05473286 - A Research Study Looking at How Oral Semaglutide Works in People With Type 2 Diabetes in Germany, as Part of Local Clinical Practice
Not yet recruiting NCT05029804 - Effect of Walking Exercise Training on Adherence to Disease Management and Metabolic Control in Diabetes N/A
Completed NCT04531631 - Effects of Dorzagliatin on 1st Phase Insulin and Beta-cell Glucose Sensitivity in T2D and Monogenic Diabetes Phase 2