Clinical Trial Details — Status: Active, not recruiting

Administrative data

NCT number NCT04744467
Other study ID # PHY2101
Secondary ID
Status Active, not recruiting
Phase N/A
First received
Last updated
Start date May 5, 2021
Est. completion date December 2024

Study information

Verified date May 2024
Source Beaver-Visitec International, Inc.
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

This is a single-center, prospective, open-label PMCF study whereby patients undergoing routine cataract surgery will have mono- or bilateral implantation of hydrophobic acrylic monofocal toric intraocular lens PODEYE TORIC (study no. PHY2101, ClinicalTrials.gov ID: not yet assigned)


Description:

This is a single-center, prospective, open-label PMCF study whereby patients undergoing routine cataract surgery will have mono- or bilateral implantation of hydrophobic acrylic monofocal toric intraocular lens PODEYE TORIC. The device under investigation is a hydrophobic acrylic monofocal toric intraocular lens (IOL) manufactured by the sponsor of this study. The refractive power of the lens has a spherical and a cylindrical component. The IOLs will be implanted in the course of routine cataract surgery and will benefit to patients suffering from cataract development and showing pre-existing corneal astigmatism. If only one eye qualifies for toric lens implantation due to small amount of pre-existing corneal astigmatism, one eye will be implanted with PODEYE TORIC and the contralateral eye will be received an implantation with the monofocal parent lens PODEYE. The study purpose is to determine safety, rotational stability, refractive and visual performance, contrast sensitivity, patient satisfaction and PCO rate after mono- or bilateral implantation of PODEYE TORIC intraocular lens. In total approx. 23 patients will be recruited for this clinical study and undergo mono- or bilateral implantation of PODEYE TORIC intraocular lens. The maximum time between 1st and 2nd eye implantation is 30 days. Subjects participating in the trial will attend a total of maximum 12 study visits over a period of 630-780 days (preoperative, operative, 1-2 days, 7-14 days, 30-60 days, 120-180 days, 330-420 days and 630-780 days postoperative visits). Subjects would have the option for unscheduled visits if required medically.


Recruitment information / eligibility

Status Active, not recruiting
Enrollment 23
Est. completion date December 2024
Est. primary completion date July 2024
Accepts healthy volunteers No
Gender All
Age group 50 Years and older
Eligibility Inclusion Criteria: - Male or female adults ages 50 years or older on the day of screening who have clinically documented cataracts in both eyes; - Calculated IOL power (sphere and cylinder) is within the range of the investigational IOLs; - Regular corneal astigmatism (measured by a topographer); - Corneal astigmatism =0.75 D and = 4.25 D (measured by an automatic keratometer) in one or both eyes; - Capability to understand and sign an IRB approved informed consent form and privacy authorization; - Clear intraocular media other than cataract; - Dilated pupil size large enough to visualize IOL axis markings postoperatively; - Best corrected visual acuity projected to be better than 0.2 logMAR after toric IOL implantation; - Willing and able to conform to the study requirements. Exclusion Criteria: - Age of patient <50 years at the day of screening; - Regular corneal astigmatism <0.75 D or >4.25 D (measured by an automatic keratometer) in both eyes - Irregular astigmatism (measured by a topographer); - Subjects with diagnosed degenerative visual disorders (e.g. macular degeneration or other retinal or optic disorders); - Subjects with AMD suspicious eyes as determined by OCT examination; - Previous intraocular or corneal surgery; - Traumatic cataract; - History or presence of macular edema; - Instability of keratometry or biometry measurements; Acceptable maximum standard deviation: AL: ± 150 µm; ACD: ± 150 µm; K1 / K2: ± 0.15 D; - Clinically significant, uncontrolled glaucoma with expected negative impact on Contrast Sensitivity and/or visual acuity outcomes; - Pregnant, lactating or, if able to bear children, unwilling to use medically acceptable birth control over the course of the study; - Concurrent or previous (within 30 days) participation in another drug or device investigation; - Clinically significant dry eye as determined by the investigator' s judgement; - Ocular surface disease (clinical symptoms or keratitis); - Patients showing contraindications as listed in the current Instructions for use (IFU); - Unsuitable for study participation for any other reason, as determined by Investigator's clinical judgment (reason to be documented on eCRF). In addition to above mentioned in- and exclusion criteria, subjects shall be discontinued when certain conditions are present at the time of surgery, including: - zonular instability; - need for iris manipulation; - capsular fibrosis or other opacity; and - inability to fixate IOL in desired position. In such cases, the subject shall be followed until the condition has stabilized.

Study Design


Related Conditions & MeSH terms


Intervention

Device:
IOL implantation experimental
Mono- or bilateral implantation of toric intraocular lenses PODEYE TORIC. If only one eye qualifies for toric lens implantation due to small amount of pre-existing corneal astigmatism, one eye will be implanted with PODEYE TORIC and the contralateral eye will be receive an implantation with the monofocal parent lens PODEYE (PhysIOL sa/nv, Liège, Belgium).

Locations

Country Name City State
Czechia Gemini Eye Clinic Zlin

Sponsors (2)

Lead Sponsor Collaborator
Beaver-Visitec International, Inc. targomedGmbH

Country where clinical trial is conducted

Czechia, 

References & Publications (2)

Kretz FT, Breyer D, Klabe K, Hagen P, Kaymak H, Koss MJ, Gerl M, Mueller M, Gerl RH, Auffarth GU. Clinical Outcomes After Implantation of a Trifocal Toric Intraocular Lens. J Refract Surg. 2015 Aug;31(8):504-10. doi: 10.3928/1081597X-20150622-01. — View Citation

McAlinden C, Pesudovs K, Moore JE. The development of an instrument to measure quality of vision: the Quality of Vision (QoV) questionnaire. Invest Ophthalmol Vis Sci. 2010 Nov;51(11):5537-45. doi: 10.1167/iovs.10-5341. Epub 2010 May 26. — View Citation

Outcome

Type Measure Description Time frame Safety issue
Other Intraocular pressure (IOP) measurement The IOP will be measured with non-contact tonometer as part of the routine follow up examinations. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. Pre-Op, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Keratometry Keratometric measurements are performed to calculate the required IOL power. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. Pre-Op, 1 month postoperative, 6 months postoperative
Other Biometry Biometry measurements are performed to calculate the required IOL power. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. Pre-Op
Other IOL power and target refraction The parameter of the implanted IOL must be recorded as well as the target refraction given by the IOL calculator. This parameter is needed to calculate the accuracy of achieving the target refraction. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. during surgery
Other Fundus OCT An OCT (optical coherence tomography) image will be taken at the preoperative visit to identify AMD suspicious eyes that need to be excluded from this study. The examination is performed and documented according to the guidelines in ISO 11979-7:2018. Pre-Op, 6 months postoperative
Other Pupil Size Photopic and mesopic pupil diameters are measured with a precision of at least +/-0.5 mm. For the photopic pupil diameters, eye illumination should be identical to that used for photopic contrast sensitivity testing. For the mesopic pupil diameters, eye illumination should be identical to that used for mesopic contrast sensitivity testing. It is required that pupil measurements be made with an infrared pupillometer / CCD camera. Pupil measurements should only be made after the eye has had time to fully adapt to the testing conditions (at least 10 minutes).
The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
Pre-Op, 6 months postoperative
Other Manifest refraction The manifested refraction is measured by means of a phoropter. The data contains values for sphere, cylinder and axis of cylinder according to ISO 11979-7:2018. Pre-Op, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Uncorrected Distance Visual Acuity (UDVA) - monocular UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2014. This assessment is done monocularly under photopic conditions. Pre-Op, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Uncorrected Distance Visual Acuity (UDVA) - monocular mesopic UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2014. This assessment is done monocularly under mesopic conditions. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Uncorrected Distance Visual Acuity (UDVA) - binocular UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2014. This assessment is done binocularly under photopic conditions. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Uncorrected Distance Visual Acuity (UDVA) - binocular mesopic UDVA is measured with ETDRS charts placed in 4m distance according to ISO 11979-7:2014. This assessment is done binocularly under mesopic conditions. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Corrected Distance Visual Acuity (CDVA) - monocular CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2014. This assessment is done monocularly under photopic conditions. Pre-Op, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Corrected Distance Visual Acuity (CDVA) - monocular mesopic CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2014. This assessment is done monocularly under mesopic conditions. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Corrected Distance Visual Acuity (CDVA) - binocular CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2014. This assessment is done binocularly under photopic conditions. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Corrected Distance Visual Acuity (CDVA) - binocular mesopic CDVA is measured with ETDRS charts placed in 4m distance with best aided corrective glasses according to ISO 11979-7:2014. This assessment is done binocularly under mesopic conditions. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Binocular Contrast Sensitivity under photopic light conditions Contrast Sensitivity under photopic light conditions using the standardized contrast sensitivity device Clinical Trial Suite (M&S Technologies). This examination is performed binocularly. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Binocular Contrast Sensitivity under mesopic light conditions Contrast Sensitivity under mesopic light conditions using the standardized contrast sensitivity device Clinical Trial Suite (M&S Technologies). This examination is performed binocularly. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Binocular Contrast Sensitivity under photopic light conditions and using a glare source Contrast Sensitivity under photopic light conditions with a glare source using the standardized contrast sensitivity device Clinical Trial Suite (M&S Technologies). This examination is performed binocularly. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Binocular Contrast Sensitivity under mesopic light conditions and using a glare source Contrast Sensitivity under mesopic light conditions with a glare source using the standardized contrast sensitivity device Clinical Trial Suite (M&S Technologies). This examination is performed binocularly. 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Patient reported outcomes, Quality of Vision Questionnaire Outcomes measures of a questionnaire to address subjective study outcomes and disturbance by photic phenomena. For this study, the validated and verified questionnaire Quality of Vision (QoV) will be used.
The QoV Questionnaire consists of a Rasch-tested, linear-scaled, 30-item instrument on three scales providing a QoV score in terms of symptom frequency, severity, and bothersome. It is suitable for measuring QoV in patients with all types of refractive correction, eye surgery, and eye disease that cause QoV problems [McAlinden, 2010]. The score ranges are 0 to 100 with smaller scales indicating better results.
Pre-Op, 6 months postoperative, 24 months postoperative
Other Slitlamp examination - Corneal Status The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
• Corneal Status
Pre-Op, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Fundus examination with dilated pupil The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
• Fundus
Pre-Op, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Slitlamp examination - Signs of inflammation The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
Signs of inflammation
Anterior chamber cells,
Anterior chamber flare,
Cystoid macular oedema,
Hypopyon, and
Endophthalmitis
Pre-Op, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Slitlamp examination - Pupillary block The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
• Pupillary block
Pre-Op, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Slitlamp examination - Retinal detachment The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
• Retinal detachment
Pre-Op, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Slitlamp examination - Status of anterior and posterior capsule The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
• Status of anterior and posterior capsule
Pre-Op, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Slitlamp examination - IOL decentration The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
• IOL decentration
1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Slitlamp examination - IOL tilt The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
• IOL tilt
1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Slitlamp examination - IOL discoloration The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
• IOL discoloration
1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Slitlamp examination - IOL opacity The slitlamp examination is one examination to analyse the integrity of eye structures and the implanted IOL. The examination is performed and documented according to the guidelines in ISO 11979-7:2018.
With the slitlamp the ophthalmologist can observe the eyes stereoscopically. A focussed slit of light, which can be width-adjusted, is projected on the eye to be examined. The investigator observes this projection on the eye through a reflected light microscope. The slitlamp is used to observe the anterior and posterior part of the eye, including cornea, lens and anterior chamber. By dilating the pupil the fundus can also be examined. Following conditions shall be examined with the slitlamp:
• IOL opacity
1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Manual assessment of IOL axis position IOL axis will be subjectively measured by the examiner during the slit lamp examination with dilated eyes. Only for the PODEYE toric implanted eyes. within 10 minutes after surgery, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other PCO grading Posterior capsule grading should be performed after pupil dilation at the follow up visits mentioned in the examination matrix. The status of the posterior capsule should be noted to be either "intact" or "not intact". If the capsule is intact, the presence of PCO and/or Elschnig's pearls will be documented as "present" or "absent". If "present", the location and clinical significance of the finding will be documented. Posterior capsular opacification (PCO) will be graded using the following scale:
0: None
Minimal
Mild
Moderate
Severe
1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Halo and Glare Scores (binocular) A Halo and Glare simulator software (Eyeland Design Network GmbH, Germany) will be used to assess the subjective perception of photic phenomena. The patients will be asked to adjust the halo and glare level in a simulated image to the amount they perceive of such photic phenomena. This test is performed pre- and postoperatively. The preoperative outcomes serve as a baseline measurement.
This simulator utilizes a scale for intensity, size of the halo, and glare that ranges from zero (none) to 100 (extremely disturbing). It also allows for classifying the halos into three types, as follows: T1 (diffuse halo ring), T2 (starburst type), and T3 (distinct halo ring) [Kretz, 2015].
Pre-Op, 6 months postoperative, 24 months postoperative
Other Safety outcomes The SPE rates of the investigational product will be compared to the results of historical data of EN ISO 11979-7:2018, Annex E (Table E.2).
Rates of adverse events as determined by EN ISO 11979-7:2018 will be reported through the complete study follow-up. The proportion of all subjects with these events will be reported over time. Statistical comparisons to EN ISO 11979-7:2018 SPE (Safety and Performance Endpoints) rates (Annex E) will be based on all eyes implanted with the investigational device using a one-sided, exact test based on the binomial distribution.
Pre-Op, during surgery, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Other Medication Revision Prior to the study start, a standard medication will be defined that all patients will receive and that is standard for cataract surgery. Only changes to the standard concomitant medication will be documented in the CRFs. Pre-Op, during surgery, 1 day postoperative, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Primary Rotational stability - Photograph of IOL with dilated pupil to asses IOL rotation The primary performance endpoint is to show statistically non-inferior rotational stability outcomes on PODEYE TORIC lens compared to CER data obtained for a multifocal toric IOL (POD FT) by the same manufacturer sharing the same haptic design but based on a hydrophilic lens material. The null hypothesis is that the absolute mean IOL rotation measured on PODEYE TORIC is not larger than the absolute mean IOL rotation reported in the current CER on POD FT lens.
The alternative hypothesis is that the absolute mean IOL rotation measured on PODEYE TORIC is larger than the outcomes reported in the current CER on POD FT lens. The measure is taken at 6 months postoperative visit compared to the measurement performed within 10 minutes after surgery.
6 months postoperative
Secondary Best Corrected Distance Visual Acuity compared to historical Data The objective is to compare CDVA above defined thresholds of the investigational product to the results of historical data of EN ISO 11979-7:2018, Annex E (Table E.3 and E.4). The null hypothesis for the CDVA is that / when the true rate of cases with visual acuity 0.3 logMAR or better is less than or equal to the SPE rate. The alternative hypothesis is that the "success" rate is higher than the SPE rate. If the distance corrected visual acuity is 0.3 logMAR or better in 30 eyes or more (with 36 eyes included), the null hypothesis could be rejected, and the alternative hypothesis could be accepted. Pre-Op, 1 week postoperative, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
Secondary Change in cylindrical Power of the Eye Change in cylindrical power, defined as the difference between magnitude of preoperative keratometric cylinder and postoperative manifest cylinder is calculated at every Form. The outcomes of this calculation will be stratified per 0.25D step of preoperative keratometric cylinder and descriptive statistics (mean, standard deviation, median, minimum, maximum) will be provided.
In addition, the following outcomes will be tabulated (stratified per 0.25D step of preoperative keratometric cylinder):
The proportion that showed a "reduction in cylindrical power of the eye" <0.50 D
The proportion that showed a "reduction in cylindrical power of the eye" =0.50 D
The proportion that showed change in absolute value "reduction in cylindrical power of the eye" = 0.50 D Also, regression analysis will be performed comparing "Reduction in cylindrical power of the eye" (corneal plane) and preoperative keratometric cylinder.
Pre-Op, 1 month postoperative, 6 months postoperative, 12 months postoperative, 24 months postoperative
See also
  Status Clinical Trial Phase
Completed NCT04685538 - Chloroprocaine 3% Gel Eye Drop as Topical Anestheticsin Phacoemulsification. Phase 3
Recruiting NCT06060041 - IC-8 Apthera IOL New Enrollment Post Approval Study
Recruiting NCT05518539 - Evaluation of Quality of Vision and Visual Outcomes With Bilateral Implantation of the Clareon PanOptix Intraocular Lens
Recruiting NCT05271942 - Tilt and Tumble vs Divide and Conquer - a Unique Comparison of the Two Cataract Surgery Methods N/A
Active, not recruiting NCT04778501 - PMCF Study on Monofocal Toric IOL (PODEYE TORIC) in Asia N/A
Completed NCT05062564 - Efficacy of LipiFlow in Patients Affected by Meibomian Gland Dysfunction in Reducing Post-cataract Surgery Dry Eye N/A
Completed NCT03751033 - Influence of DisCoVisc Ophthalmic Viscosurgical Device (OVD) on Intraoperative Aberrometry Readings N/A
Completed NCT02529488 - Investigation of AcrySof® IQ PanOptix™ Presbyopia-Correcting Intraocular Lens (IOL) Model TFNT00 N/A
Completed NCT04539548 - A Study Assessing the Safety and Efficacy of Dextenza® for the Treatment of Ocular Pain and Inflammation Following Surgery for Pediatric Cataract Phase 3
Completed NCT03740659 - Evaluation Of Aqueous Humor Of Levofloxacin-Dexamethasone Eye Drops And Of Its Components In Patients Undergoing Cataract Surgery Phase 2
Completed NCT03494257 - Effect of Fixed Brinzolamide-brimonidine Combination on Intraocular Pressure After Phacoemulsification N/A
Completed NCT05119127 - Rotational Stability of Acrysof IQ Vivity Extended Vision Toric IOL and Refractive Visual Outcome. N/A
Active, not recruiting NCT04271709 - Manhattan Vision Screening and Follow-Up Study (NYC-SIGHT) N/A
Recruiting NCT03713268 - Intraoperative OCT Guidance of Intraocular Surgery II
Completed NCT03739528 - Levo-Dexa vs. Tobra+Dexa for Prevention and Treatment of Inflammation and Prevention of Infection in Cataract Surgery Phase 3
Completed NCT02888210 - A Study Assessing Safety and Efficacy of MD-15 Intraocular Lens in Patients With Aphakic Eye After Cataract Surgery Phase 3
Completed NCT03356847 - Evaluation of the Rotational Stability of the Monofocal SISA Implant Following Cataract Surgery N/A
Completed NCT04332640 - Clinical Evaluation of the Next Generation Phaco System N/A
Recruiting NCT03638726 - Subconjunctival Atropine and Intracameral Epinephrine for Pupil Dilation in Phacoemulsification Phase 4
Completed NCT03050697 - Evaluation of the Safety and Performance of the HARMONI® Toric Lens N/A