The SOMNOtouch NIBP is a cuffless device for continuous blood pressure measurement. The device is widely used, but validation for 24-hour ambulatory blood pressure measurement (ABPM) is limited. This study seeks to validate the SOMNOtouch device for use in 24-hour ABMP as compared to a validated oscillometric device for 24-hour ABPM. A total of 60 patients with indication for 24-hour BP measurement will be recruited from our tertiary hypertension Clinic. All included patients will conduct dual 24-hour ABPM with both the SOMNOtouch NIBP and a the oscillometric device. Furthermore, patients will undergo measurement of arterial stiffness with a Sphygmocor device as well as complete a questionaire on device related discomfort.


Clinical Trial Description

Recruitment of study participants: A total of 60 patients with indication for 24-hour BP measurement will be recruited from our tertiary hypertension clinic at Aarhus University Hospital, Aarhus, Denmark. Patients will be eligible for inclusion if they are 18 years or older and have indication for ABPM with an automated oscillometric device. Exclusion criteria will be as follows: Known significant arterial abnormalities in the upper extremities, known cardiac arrythmia, upper arm circumference >16,5" or <6", or extreme uncontrolled hypertension (i.e. resting office BP > 230/140 mmHg). To enable testing of the device at both normal and elevated BP-levels, we will aim at including at least 15 patients in each of the following categories: Normal BP, stage 1 hypertension and stage ≥2, based on NICE guidelines for ABPM17. Informed consent will obtained from all patients prior to inclusion in the study. At inclusion, a simultaneous bilateral BP-measurement will be performed with three consecutive measurements using an automated oscillometric device (Microlife WatchBP Office®) after 5 minutes of rest. If the difference in Systolic Blood Pressure (SBP) between the arms is found to be >5 mmHg, the patient will not be included in the study. 24-hour BP measurements: All staff related to the project are familiar with the SOMNOtouch NIBP™ (SomBP) device, which has been used in our clinic for selected patients with high cuff-based BP and at the same time symptoms of hypotension, since 2015. Mounting and calibration of both devices will be conducted according to manufacturers' instructions. With the patient sitting in an upright position with back-support, a conventional oscillometric BP-device (Spacelabs 90217 / 90207, OscBP) will be fitted on the right arm, while the SomBP will be fitted on the left arm. The SomBP will be calibrated with a single BP measurement with a cuff-based sphygmomanometer. The OscBP will record BP three times every hour during daytime and two times every hour during nighttime, while the SomBP will measure BP continuously. The patients will be instructed to engage in normal activities but refrain from strenuous exercise and, at the time of cuff-inflation, to stop moving and talking, while keeping the arm still with the cuff at heart level. Pulse wave velocity (PWV) as well as pulse wave analysis (PWA) will be performed in duplicate with the SphygmoCor system (version 8.2, Atcor Medical, Sydney, Australia). A small questionnaire on device-related pain and/or discomfort as well as sleeping habits will be provided to all participants. As per ESH guidelines, a minimum of 70% useful measurements on the oscillometric device during the ABPM will be required for the data to be considered useful. Assessment the acceptability of data-quality of SomBP will be made on the basis of discussion between the primary investigator and KLC, who, at the time of evaluation will both be blinded to the corresponding OscBP measurement. No SomBP measurements will be accepted with more than 50% failed measurement in the report. Data from both devices will be excluded from the study if data-collection from either device is deemed to have insufficient quality to be usable. For these instances, patients will be asked to repeat examination with both devices. SomBP1 data will extracted ad hoc using Domino Light 1.4.0 software. Statistical analysis: All statistical analysis will be performed using STATA software. A p-value of 0.05 will be prespecified as statistically significant. Variables will be checked for distribution through histograms and QQ-plots. Means will be compared using paired t-test. Standard regression analysis will be performed to test for association in continuous variables. Categorical variables will be compared using chi-squared tests. ;


Study Design


Related Conditions & MeSH terms


NCT number NCT04278001
Study type Interventional
Source Central Jutland Regional Hospital
Contact
Status Completed
Phase N/A
Start date September 1, 2018
Completion date February 1, 2020

See also
  Status Clinical Trial Phase
Recruiting NCT05684055 - "Community-based, eHealth Supported Management of Cardiovascular Risk Factors by Lay Village Health Workers (ComBaCaL aHT TwiC 1 & ComBaCaL aHT TwiC 2) N/A
Active, not recruiting NCT05436730 - "Escape" Phenomenon of the Antihypertensive Therapy Efficacy
Not yet recruiting NCT03294070 - Fimasartan Plus Amlodipine on Hemodynamic Parameters and Arterial Stiffness in Patients With Hypertension Phase 4
Recruiting NCT01959997 - Comparison of Redo PVI With vs. Without Renal Denervation for Recurrent AF After Initial PVI Phase 2
Completed NCT00983632 - Selective Vagus Nerve Stimulation in Human N/A
Recruiting NCT06098300 - Effects of Change in Blood Pressure on Retinal Capillary Rarefaction in Patients With Arterial Hypertension
Terminated NCT04677322 - TO ASSESS THE EFFECTIVENESS OF THE INTERVENTION OF THE LOW-SODIUM DIET IN PATIENTS WITH HTA
Completed NCT04564118 - A Retrospective Analysis of the Adherence to Clinical Practice Guidelines Using the "MedicBK" Digital Platform in Patients With Hypertension and Atrial Fibrillation
Withdrawn NCT03047538 - Fixed Combination for Lipid and Blood Pressure Control Phase 4
Completed NCT03046264 - Invasive Validation of Non-invasive Central Blood Pressure Measurements Using Oscillometric Pulse Wave Analysis N/A
Completed NCT02620995 - Effects of Sildenafil on Penile Vascular Function in Hypertensive Men With Erectile Dysfunction Phase 4
Completed NCT02041832 - Detection of Subclinical Atrial Fibrillation in High Risk Patients Using Implantable Loop Recorder N/A
Completed NCT01459120 - Comparison of Door-to-door Versus Community Gathering to Provide HIV Counseling and Testing Services in Rural Lesotho N/A
Completed NCT01546181 - Retinal Imaging by Adaptive Optics in Healthy Eyes and During Retinal and General Diseases
Recruiting NCT01132001 - Ambulatory Versus Home Blood Pressure Measurement N/A
Completed NCT01454583 - Registry for Ambulant Therapy With RAS-Inhibitors in Hypertension-patients in Germany N/A
Recruiting NCT03917758 - Antidiuretic Function Before and During Treatment With SGLT2 Inhibitors N/A
Completed NCT03722524 - The Use of TrIple Fixed-dose Combination in the Treatment of Arterial Hypertension
Completed NCT03539627 - Azilsartan Medoxomil in Hypertensive pAtients With Stable Ischemic Heart Disease and DiabEtes MEllitus.
Recruiting NCT02002585 - Renal Protection Using Sympathetic Denervation in Patients With Chronic Kidney Disease Phase 2