Clinical Trial Summary

Aerosol delivery during mechanical ventilation has long been a long debated topic. As evidence-based knowledge about the delivery of aerosol to the lungs of mechanically ventilated patients increases, one piece of the puzzle has remained unexplored; measurement of clinically relevant outcomes. The primary aim of this research is to compare clinical outcomes (ventilator-associated events (VAEs), length of stay (LOS) in intensive care unit (ICU), and total days on mechanical ventilation) when using a traditional jet nebulizer versus a newer generation vibrating mesh nebulizer during mechanical ventilation. The secondary aim of this research is to identify source of bacteria by obtaining cultures of each nebulizer and ventilator circuit and plating them for colony growth and identification.


Description:

This pilot study was designed as a randomized, controlled trial comparing clinical outcomes and contamination rates in jet nebulizers versus vibrating mesh nebulizers. The sample size will be one of consecutive convenience over an entire year (September 2013-September 2014) to include all seasons. The study will be performed at Rush University Medical Center (RUMC) in Chicago, IL. Each mechanically ventilated patient within inclusion criteria, with a physician order for aerosol treatment, will be randomized to either the jet nebulizer or mesh nebulizer group using SPSS computer software. The respiratory care staff will place the nebulization device in the ventilator circuit per hospital protocol. Both devices, jet and mesh nebulizers, are currently standard practice; therefore hospital protocol will be followed regarding placement and administration of ordered aerosol treatment. The study subject will remain on the device to which they were randomized for the duration of their hospital stay to obtain clinical outcome data. Study staff to be notified of patient extubation from mechanical ventilation, discharge from intensive care unit, or expiration; this will conclude participant involvement in the study. Retrospective clinical outcome data will be obtained from the subject's electronic medical chart at conclusion of study. Retrospective data to include: 1. LOS in ICU (days) 2. LOS on mechanical ventilator (days) 3. Drug administration information 1. Drug and dose ordered 2. Number of treatments delivered 3. Treatment type - Continuous - Intermittent 4. Ventilator-Associated Tracheobronchitis (# VATs)16 a. Fever, increased volume and purulence of secretions, a positive culture (quantitative or semi-quantitative) of a respiratory sample (tracheal aspirates or bronchoscopic specimens), and the absence of a new or an evolving pulmonary infiltrate in the chest x-ray in a patient on mechanical ventilation >48 hours 5. Ventilator-Associated Events in Adults15 1. Ventilator-Associated Condition (# VACs) - After a period of stability or improvement on the ventilator, the patient has at least one of the following indications of worsening oxygenation: - Minimum daily FiO2 values increase ≥ 0.20 over baseline and remain at or above that increased level for ≥ 2 calendar days. - - Minimum daily PEEP values increase ≥ 3 cm H2O over baseline and remain at or above that increased level for ≥ 2 calendar days. 2. Infection-related Ventilator-Associated Complication (# IVACs) - On or after calendar day 3 of mechanical ventilation and within 2 calendar days before or after the onset of worsening oxygenation, the patient meets both of the following criteria: - Temperature > 38˚ C or < 36˚ C, or white blood cell count ≥ 12,000 cells/mm3 or ≤ 4,000 cells/mm3. AND - A new antimicrobial agent(s) is started, and is continued for ≥ 4 calendar days. 3. Possible Ventilator-Associated Pneumonia (# Possible VAPs) - On or after calendar day 3 of mechanical ventilation and within 2 calendar days before or after the onset of worsening oxygenation, ONE of the following criteria is met: - Purulent respiratory secretions - Positive culture 4. Probably Ventilator-Associated Pneumonia (# Probable VAPs) - On or after calendar day 3 of mechanical ventilation and within 2 calendar days before or after the onset of worsening oxygenation, ONE of the following criteria is met: - Purulent respiratory secretions (from one or more specimen collections- and defined as for possible VAP) and one of the following (positive culture of endotracheal aspirate, positive culture of bronchoalveolar lavage, positive culture of lung tissue, positive culture of protected specimen brush - One of the following without requirement for purulent respiratory secretions (positive pleural fluid culture, positive lung histopathology, positive diagnostic test for Legionella, positive diagnostic test on respiratory secretions for influenza virus, respiratory syncytial virus, adenovirus, parainfluenza virus) The Jet Nebulizer Protocol: - Physician order received for subject; randomization to jet nebulizer occurred. - Jet nebulizer, Misty Max 10™ (Carefusion, CA), placed into spring-loaded t-piece between the humidifier and the ventilator (approximately 15 cm from the gas outlet). - Aerosol treatment delivered per physician order at flow rates of 8-10 L/min. - Jet nebulizer to be replaced every 3 days per hospital protocol. 1. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures. 2. Cultures - Jet Nebulizer Culture (3 plates): Sample to be collected from jet nebulizer by running oxygen at a flow of 8-10 L/min through nebulizer (3 mL normal saline in the reservoir) while directing the nebulizer towards plate for 1 minute. Process to be repeated for 3 different plates. Plates to be labeled, have edges sealed with tape, be placed in biohazard bag and sent to microbiology lab for analysis. 1. Plate 1: MacConkey agar (id gram (-) organisms) 2. Plate 2: Chocolate agar (id fastidious organisms) 3. Plate 3: Blood agar (id gram(+) and gram(-) organisms) - Ventilator Circuit Culture: Material to be collected from rim of inspiratory ventilator circuit at wye by rubbing a sterile swab along circuit 3x in a circular motion. Swab to be sealed in test tube with lid and sent to lab for analysis. - Sputum samples: Data obtained per electronic medical chart. Sputum sample data to be analyzed in study when ordered by physician. - After obtaining cultures, the jet nebulizer will be replaced with new one and labeled per hospital protocol. - Further cultures are not necessary unless patient is identified by study staff to have VAE. (Appendix A) - Study staff to be notified of patient extubation from mechanical ventilation, discharge from intensive care unit or expiration. The Vibrating Mesh Nebulizer protocol: - Physician order received for subject; randomization to vibrating mesh nebulizer occurred. - Aeroneb® Solo (Aerogen, Galway, Ireland) vibrating mesh nebulizer to be placed before the heater on the "dry side" of the heater water chamber on the inspiratory ventilator circuit. - Aerosol treatment delivered per physician order. - Mesh nebulizer to be replaced every 30 days per hospital protocol. 1. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures. 2. Cultures - Vibrating Mesh Nebulizer Culture (3 plates): Sample to be collected from vibrating mesh nebulizer while directing the nebulizer(3 mL normal saline in the reservoir) towards plate for 1 minute. Process to be repeated for 3 different plates. Plates to be labeled, have edges sealed with tape, be placed in biohazard bag and sent to microbiology lab for analysis. 1. Plate 1: MacConkey agar (id gram (-) organisms) 2. Plate 2: Chocolate agar (id fastidious organisms) 3. Plate 3: Blood agar (id gram(+) and gram(-) organisms) - Ventilator Circuit Culture: Material to be collected from rim of inspiratory ventilator circuit at wye by rubbing a sterile swab along circuit 3x in a circular motion. Swab to be sealed in test tube with lid and sent to lab for analysis. - Sputum samples: Data obtained per electronic medical chart. Sputum sample data to be analyzed in study when ordered by physician. - Vibrating mesh nebulizer replaced with new one and labeled per hospital protocol. - Further cultures not necessary unless patient identified by study staff to have VAE. (appendix A) - Study staff to be notified immediately of patient extubation from mechanical ventilation, discharge from intensive care unit or expiration.


Recruitment information / eligibility

Status Completed
Enrollment 120
Est. completion date October 2014
Est. primary completion date October 2014
Accepts healthy volunteers No
Gender All
Age group 1 Year and older
Eligibility Inclusion Criteria: - Age > 1 y/o - Intubated with endotracheal tube (ETT) and on mechanical ventilation - Patient in all intensive care units at Rush University Medical Center - Physician order for aerosolized breathing treatment in medical record Exclusion Criteria: - Age < 1 y/o - Diaphragmatic paralysis - Paraplegic or high spinal cord injury - Brain death

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Vibrating Mesh Nebulizer Arm
Physician order received for subject, randomization to vibrating mesh nebulizer occurred. Aeroneb® Solo (Aerogen, Galway, Ireland) vibrating mesh nebulizer to be placed before the heater on the "dry side" of the heater water chamber on the inspiratory ventilator circuit. Aerosol treatment delivered per physician order. Mesh nebulizer to be replaced every 30 days per hospital protocol. a. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.
Jet Nebulizer Arm
Physician order received for subject, randomization to jet nebulizer occurred. Jet nebulizer, Misty Max 10™ (Carefusion, CA), placed into spring loaded t-piece between the humidifier and the ventilator (approximately 15 cm from the gas outlet). Aerosol treatment delivered per physician order at flow rates of 8-10 L/min. Jet nebulizer to be replaced every 3 days per hospital protocol. a. Prior to every nebulizer change or every 3 days, study staff to be notified in order to obtain cultures.

Locations

Country Name City State
United States Rush University Medical Center Chicago Illinois

Sponsors (1)

Lead Sponsor Collaborator
Rush University Medical Center

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Venilator Associated Events (VAE) Ventilator associated conditions (VAC) and infection related ventilator associated conditions (IVAC) were identified based on the 2013 NHSN Surveillance for VAEs (e,g., "See page 92 at https://www.cdc.gov/nhsn/PDFs/pscManual/validation/2013-PSC-Manual-validate.pdf for detailed criteria.") duration of mechanical ventilation or discharge from the intensive care unit
Secondary Participants With Bacterial Growth While on mechanical ventilator, bacterial growth will be assessed from the vent circuit and nebulizer, and participant sputum. duration of mechanical ventilation or discharge from the intensive care unit
Secondary Length of Stay in Intensive Care Unit Total length of stay in intensive care unit in days discharge from the intensive care unit
Secondary Days on Mechanical Ventilation Total days spent on Mechanical Ventilator discharge from the intensive care unit
See also
  Status Clinical Trial Phase
Recruiting NCT05904652 - High Flow Nasal Oxygen at Extubation for Adults Requiring a Breathing Tube for Treating Severe Breathing Difficulties N/A
Completed NCT04030208 - Evaluating Safety and Efficacy of Umbulizer in Patients Requiring Intermittent Positive Pressure Ventilation N/A
Recruiting NCT03697785 - Weaning Algorithm for Mechanical VEntilation N/A
Recruiting NCT02989051 - Fluid Restriction Keeps Children Dry Phase 2/Phase 3
Completed NCT02930525 - Effect of High Flow Nasal Cannula vs. Standard Care on Respiratory Stability in Pediatric Procedural Sedation N/A
Recruiting NCT02539732 - Prediction of Outcome of Weaning From Mechanical Ventilation Using the Electrical Activity of the Diaphragm
Enrolling by invitation NCT02290236 - Monitored Saturation Post-ICU N/A
Terminated NCT01583088 - Early Stage Amyotrophic Lateral Sclerosis Phrenic Stimulation Phase 3
Withdrawn NCT00990119 - High Flow Therapy (HFT) to Treat Respiratory Insufficiency in Chronic Obstructive Pulmonary Disease (COPD) N/A
Completed NCT00741949 - Broncho-alveolar Lavage Under Noninvasive Ventilation With Propofol TCI in Patient With AHRF Phase 3
Completed NCT01411722 - Electrical Activity of the Diaphragm During the Weaning Period Phase 2
Recruiting NCT00339053 - Immunonutrition and Thoracoabdominal Aorta Aneurysm Repair Phase 4
Recruiting NCT00187434 - Comparison of Two Methods of Continuous Positive Airway Pressure (CPAP) to Support Successful Extubation of Infants of Birth Weights ≤ 1500 Grams (C2CPAP) N/A
Completed NCT02687802 - Respiratory Mechanics and Patient-ventilator Asynchrony Index in Patients With Invasive Mechanical Ventilation
Not yet recruiting NCT05906030 - Diaphragm Dysfunction and Ultrasound Perioperatively
Completed NCT03850977 - Is There an Association Between Chronic Pancreatitis and Pulmonary Function
Completed NCT02845375 - Effect of Neuromuscular Blockade and Reversal on Breathing Phase 4
Recruiting NCT03309423 - Is Venous to Arterial Conversion (v-TAC) of Blood Gas Reliable in Critical Ill Patients in the ICU? N/A
Completed NCT04115969 - Outcome After Non-invasive Ventilation.
Not yet recruiting NCT03687385 - The Effect of High-flow Nasal Oxygenation to the Saturation During Analgo-sedation in Different ASA Risk Class Patients N/A