Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT04712240
Other study ID # REB20-2237
Secondary ID
Status Completed
Phase N/A
First received
Last updated
Start date November 17, 2021
Est. completion date January 13, 2022

Study information

Verified date May 2022
Source University of Calgary
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

While respiratory rate is considered a critical vital sign, it often goes unmeasured or is ignored primarily due to shortcomings of the currently used measurement methods. Respiratory rate provides important information on a person's health condition and physiological stability, and an abnormal respiratory rate is a strong indicator that a health crisis is imminent. In fact, a sudden change in respiratory rate is one of the strongest predictors of mortality. Current techniques of monitoring respiratory rate have drawbacks that limit the frequency and convenience of the respiratory monitoring. Recognizing that closer respiration monitoring can save lives and improve quality of life, reduce hospital stays, and lower medical costs, the health care industry is seeking improved respiration monitoring products. The allocation of high-risk patients to intensive care for more careful monitoring or after surgery is often arbitrary, and such care might not be available routinely. For those patients who are cared for in 'general' wards where staffing levels are limited, a practical continuous monitor of respiratory rate would be of great value. This study will act as a pilot to determine the feasibility of using respiratory sensor device to monitor respiratory rate in hospitalized patients. Deriving a measurement of respiratory rate from a respiratory rate monitoring (RRM) device is a technological approach that may overcome these limitations. The device principle is based on a piezoelectric sensor where chest expansions and contractions generate very small amounts of current by the piezoelectric sensor. The expansion and contractions are measured very accurately by over-sampling, filtering and digital signal processing to remove noise and any bias generated by the piezoelectric sensor itself or the sampling circuitry. The study will be conducted from March 2021 to September 2021. During the first phase a convenience sample of 30 patients undergoing general anesthesia with muscle paralysis and mechanical ventilation will be recruited. This phase will help to validate the RRM against capnography in mechanically ventilated patients with a set respiratory rate. Following the first phase, 120 patients undergoing a procedure using sedation or spinal anesthesia will be recruited. Patients will be breathing spontaneously, and the respiratory rate will be monitored by capnography connected to the face mask or nasal prongs. Respiratory rate detected by capnography is recorded in the electronic medical records on a minute to minute interval.


Recruitment information / eligibility

Status Completed
Enrollment 169
Est. completion date January 13, 2022
Est. primary completion date January 13, 2022
Accepts healthy volunteers No
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria: - Adult participants >= 18 years of age scheduled for elective surgery - American Society of Anesthesiologists' Status Score <=3 Exclusion Criteria: - Patients undergoing emergency surgery - Patients undergoing abdominal procedures requiring skin prep of the lower chest area - Patients undergoing breast surgery - Patients undergoing shoulder surgery - Patients with skin irritation in the area of monitor application - Pregnant or lactating women

Study Design


Related Conditions & MeSH terms

  • Validation of Respiratory Rate Monitor Versus Capnography

Intervention

Device:
Respiration rate monitor
A portable continuously monitoring respiration rate monitor will be fitted to each participant. The respiration rate monitor is housed within a comfortable chest strap that be secured to the participant for the duration of the surgical procedure

Locations

Country Name City State
Canada South Health Campus Calgary Alberta

Sponsors (1)

Lead Sponsor Collaborator
University of Calgary

Country where clinical trial is conducted

Canada, 

Outcome

Type Measure Description Time frame Safety issue
Primary Percentage of paired respiration rates within one breath per minute Percentage of respiration rates measured by respiratory rate monitor that are within one breath per minute as measured by capnography Respiration will be measured during the duration of participants' surgical procedure
Secondary Mean bias of errors Mean bias of errors between respiration rates measured with respiratory rate monitor versus capnography Respiration will be measured during the duration of participants' surgical procedure
Secondary Standard deviation of errors Standard deviation of errors between respiration rates measured with respiratory rate monitor versus capnography Respiration will be measured during the duration of participants' surgical procedure
Secondary Root mean square deviation of respiratory rate measurements Respiration will be measured during the duration of participants' surgical procedure