Clinical Trials Logo

Clinical Trial Summary

Prospective, randomized, controlled, 30-78 days trial comparing MD-Logic pump algorithm to the standard of care of patients with type 1 diabetes.

The objective of this feasibility study is to evaluate the efficacy and safety of automated determined Insulin pump settings (i.e., basal plan, correction factor, carbohydrate ration and insulin activity time) using the MD-Logic Pump Advisor in individuals with type 1 diabetes.The study will be consisted from two segments:(I)pilot study, 30-78 days trial evaluating the MD-Logic pump advisor and (II)randomized controlled 30-78 days trial comparing MD-Logic pump algorithm to the standard of care.In segment 1, the pilot study, the aim is to enroll 15-30 subjects without control group. In segment 2, the randomized controlled segment, the aim is to enroll 92 subjects, but enrolling up to 105 eligible subjects to allow for dropouts.

The randomized controled study segment will be initiated after the pilot segment.Each segment of the study will consist of 6 clinic visits taking place at intervals of 1-3 weeks, sum of 30-78 days study duration for each segment. Before each clinic visit, subjects will wear continuous glucose sensors for 6 days; the intervention group will have up to 4 iterations (e.g the pump setting will be reviewed and adjusted up to 4 times during the study period according to the MD-Logic Pump Advisor).For the control group (at segment 2 only), insulin pump settings will not be changed during the study period (patient's usual standard of care).


Clinical Trial Description

the MD-Logic Pump Advisor was design by the Diabetes Technology Center, the Institute of Endocrinology and Diabetes, National Center for Childhood Diabetes, Schneider Children's Medical Center of Israel. The MD-Logic Pump Advisor learns and adapts the patient's insulin pump settings in order to optimize glucose control. The MD-Logic Pump Advisor needs, as input, sensor readings, insulin delivery and recording of the meals during home care. During collection of the data, the patients are asked to continue their daily routine (no need for special consideration).

The present study was designed to test the automatic algorithm that will evaluate and change, when necessary, the patient's insulin pump settings based on prior sensor reading, insulin delivery, and meals data that was collected while the patient is under his regular treatment at home.

The objective of this feasibility study is to evaluate the efficacy and safety of automated determined Insulin pump settings (i.e., basal plan, correction factor, carbohydrate ration and insulin activity time) using the MD-Logic Pump Advisor in individuals with type 1 diabetes.

The study will be consisted from two segments:(I)pilot study, 30-78 days trial evaluating the MD-Logic pump advisor and (II)randomized controlled 30-78 days trial comparing MD-Logic pump algorithm to the standard of care.In segment 1, the pilot study, the aim is to enroll 15-30 subjects without control group. In segment 2, the randomized controlled segment, the aim is to enroll 92 subjects, but enrolling up to 105 eligible subjects to allow for dropouts.

The randomized controled study segment will be initiated after the pilot segment.Each segment of the study will consist of 6 clinic visits taking place at intervals of 1-3 weeks, sum of 30-78 days study duration for each segment. Before each clinic visit, subjects will wear continuous glucose sensors for 6 days; the intervention group will have up to 4 iterations (e.g the pump setting will be reviewed and adjusted up to 4 times during the study period according to the MD-Logic Pump Advisor).For the control group (at segment 2 only), insulin pump settings will not be changed during the study period (patient's usual standard of care).

Endpoints:

- Primary endpoint Increase in time spent in the normal range, defined as sensor glucose level within 70 to 180 mg/dl

- Secondary endpoints Reduced time spent above 180 mg/dl, reduced time spent below 70 mg/dl, reduced number of hypoglycemic events below 63 mg/dl, reduced glucose variability, quality of life measurements and subject treatment satisfaction.

Inclusion Criteria:

- Subject with Type 1 diabetes (>1yr since diagnosis)

- Medtronic ( MiniMed Paradigm 522, 512 , 722 or 712) insulin infusion pump CSII (continuous subcutaneous insulin infusion) therapy for at least 3 months

- Age ≥ 18 years

- HbA1c at inclusion ≥ 6.0%

- No concomitant diseases that influence metabolic control

- No current use of CGM

- Subjects do not participate in any other interventional study ;


Study Design

Allocation: Randomized, Endpoint Classification: Safety/Efficacy Study, Intervention Model: Parallel Assignment, Masking: Open Label, Primary Purpose: Treatment


Related Conditions & MeSH terms


NCT number NCT01157923
Study type Interventional
Source Rabin Medical Center
Contact
Status Completed
Phase N/A
Start date July 2010
Completion date July 2016

See also
  Status Clinical Trial Phase
Recruiting NCT05653518 - Artificial Pancreas Technology to Reduce Glycemic Variability and Improve Cardiovascular Health in Type 1 Diabetes N/A
Enrolling by invitation NCT05515939 - Evaluating the InPen in Pediatric Type 1 Diabetes
Completed NCT05109520 - Evaluation of Glycemic Control and Quality of Life in Adults With Type 1 Diabetes During Continuous Glucose Monitoring When Switching to Insulin Glargine 300 U/mL
Recruiting NCT04016987 - Automated Structured Education Based on an App and AI in Chinese Patients With Type 1 Diabetes N/A
Active, not recruiting NCT04190368 - Team Clinic: Virtual Expansion of an Innovative Multi-Disciplinary Care Model for Adolescents and Young Adults With Type 1 Diabetes N/A
Recruiting NCT05413005 - Efficacy of Extracorporeal Photopheresis (ECP) in the Treatment of Type 1 Diabetes Mellitus Early Phase 1
Active, not recruiting NCT04668612 - Dual-wave Boluses in Children With Type 1 Diabetes Insulin Boluses in Children With Type 1 Diabetes N/A
Completed NCT02837094 - Enhanced Epidermal Antigen Specific Immunotherapy Trial -1 Phase 1
Recruiting NCT05414409 - The Gut Microbiome in Type 1 Diabetes and Mechanism of Metformin Action Phase 2
Recruiting NCT05670366 - The Integration of Physical Activity Into the Clinical Decision Process of People With Type 1 Diabetes N/A
Active, not recruiting NCT05418699 - Real-life Data From Diabetic Patients on Closed-loop Pumps
Completed NCT04084171 - Safety of Artificial Pancreas Therapy in Preschoolers, Age 2-6 N/A
Recruiting NCT06144554 - Post Market Registry for the Omnipod 5 System in Children and Adults With Type 1 Diabetes
Recruiting NCT05379686 - Low-Dose Glucagon and Advanced Hybrid Closed-Loop System for Prevention of Exercise-Induced Hypoglycaemia in People With Type 1 Diabetes N/A
Recruiting NCT05153070 - Ciclosporin Followed by Low-dose IL-2 in Patients With Recently Diagnosed Type 1 Diabetes Phase 2
Completed NCT05281614 - Immune Effects of Vedolizumab With or Without Anti-TNF Pre-treatment in T1D Early Phase 1
Withdrawn NCT04259775 - Guided User-initiated Insulin Dose Enhancements (GUIDE) to Improve Outcomes for Youth With Type 1 Diabetes N/A
Active, not recruiting NCT01600924 - Study on the Assessment of Determinants of Muscle and Bone Strength Abnormalities in Diabetes
Completed NCT02855307 - Closed-loop Control of Glucose Levels (Artificial Pancreas) During Postprandial Exercise in Adults With Type 1 Diabetes Phase 2
Completed NCT02897557 - Insulet Artificial Pancreas Early Feasibility Study N/A