Clinical Trial Summary

The objective of this study is to compare the relative capability of Stethee® (new wireless digital stethoscope) and 3M™ Littmann® Classic III™ (conventional stethoscope) in the identification of manikin-simulated heart sounds by medical doctors. A randomised, non-blinded, two-period crossover, non-inferiority design will be used to compare the acoustic performance of the two stethoscopes. We aim to determine if Stethee® is as effective as 3M™ Littmann® Classic III™ for auscultation of simulated heart sounds, and whether Stethee® could be recommended for use in clinical practice based on a non-inferior comparison with 3M™ Littmann® Classic III™.


Description:

This is a randomised, non-blinded, two-period crossover, non-inferiority study. Stethee®, a new wireless digital stethoscope, is designed to minimise ambient noise associated with electronic stethoscopes. To date, its acoustic capability in comparison to the conventional stethoscope in the auscultation of heart sounds is not known. Therefore, it is important that Stethee®'s capability as a stethoscope with sound amplification be evaluated alongside the conventional stethoscope prior to its use in the clinical setting. As acoustic performance is the primary endpoint, a non-inferiority design is proposed to enable the comparison of the mutual functionalities of the two stethoscopes. The rationale for a non-inferiority design is that: (a) electronic stethoscopes have shown a marginal benefit over conventional stethoscopes in terms of acoustic performance; (b) electronic stethoscopes have several functional advantages over conventional stethoscopes: the ability to analyse heart sounds and record heart sounds, the availability of a database for storage of heart sounds for remote analysis; and (c) electronic stethoscopes were rated better than conventional stethoscopes in terms of the overall quality of auscultation.

The relative capability of Stethee® is compared to that of 3M™ Littmann® Classic III™ by the identification of heart sounds simulated by a manikin, which functions as a surrogate patient. A random sample of non-specialist doctors stratified by health centres in Perak, Malaysia, who meet eligibility criteria and provide written consent to participate, will be randomised to initial testing with Stethee® or 3M™ Littmann® Classic III™ to perform cardiac auscultation on a manikin and identify a set of 10 heart sounds. After a 10-minute break, they will repeat the auscultation of the same 10 simulated heart sounds but played in a different sequence with the alternate stethoscope. Following the completion of both tests, the doctors will be asked to complete a questionnaire that asks about their preference for either Stethee® or 3M™ Littmann® Classic III™ with respect to ease of use, audio clarity, and diagnostic accuracy. At the end of the study, answer sheets and preference questionnaires will be collected for grading and analysis.


Recruitment information / eligibility

Status Completed
Enrollment 104
Est. completion date February 1, 2019
Est. primary completion date February 1, 2019
Accepts healthy volunteers No
Gender All
Age group N/A and older
Eligibility Inclusion Criteria:

1. Non-specialist medical doctors with up to 6 years of working experience post-house officer training

2. Non-specialist medical doctors currently working in the Medical, Paediatric, Anaesthesiology, and Emergency departments of specialist hospitals

Exclusion Criteria:

1. Non-specialist medical doctors who have worked in the Cardiology department for more than 6 months

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Stethee®
Stethee® is a new wireless digital stethoscope.
3MTM Littmann® Classic IIITM
3MTM Littmann® Classic IIITM is the conventional stethoscope.

Locations

Country Name City State
Malaysia Raja Permaisuri Bainun Hospital Ipoh Perak

Sponsors (2)

Lead Sponsor Collaborator
Clinical Research Centre, Malaysia M3DICINE Inc.

Country where clinical trial is conducted

Malaysia, 

Outcome

Type Measure Description Time frame Safety issue
Primary Correctly identified manikin-simulated heart sounds The proportion of correctly identified heart sounds between groups will be compared. The test consist of 10 cardiac conditions (represented by 10 manikin-simulated heart sounds). Subjects (medical doctors) are required to auscultate each heart sound and write the complete answer on the answer sheet. The 'all-or-nothing' marking scheme is used, i.e. answers must exactly match the marking scheme. Measured immediately after the test
Secondary User preference for the two stethoscopes User preference for the two stethoscopes with respect to ease of use, audio clarity, and diagnostic accuracy, will be compared between groups. This will be estimated as a proportion preferring Stethee® for each domain. Measured immediately after the test
Secondary Difference in the performance between subgroup of doctors from different clinical settings The performance of subgroup of doctors in specialist hospitals and doctors in non-specialist hospitals or primary care clinics will be compared. This will be estimated as a proportion difference of correctly identified heart sounds for each doctor subgroup. Measured immediately after the test
Secondary Comparison of the performance of the two stethoscopes in identifying heart sounds of different levels of difficulty The performance of the two stethoscopes in identifying heart sounds of different levels of difficulty (normal, easy, moderately difficult, difficult) will be compared. This is estimated as a proportion difference of correctly identified heart sounds for each type of heart sound. Measured immediately after the test
See also
  Status Clinical Trial Phase
Completed NCT04624412 - Different Intensities of Continuous Aerobic Exercises in Cardiac Rehab Phase 2 N/A
Completed NCT05563701 - Evaluation of the LVivo Image Quality Scoring (IQS)
Recruiting NCT04561908 - Transcatheter Microguidewire Drilling for Transseptal Left Atrial Access N/A
Active, not recruiting NCT06190743 - Perception of Cardiovascular Risk
Completed NCT04562636 - Evaluating a Messaging Campaign in the United States N/A
Completed NCT04580095 - Artificial Intelligence for Improved Echocardiography N/A
Recruiting NCT03277365 - MyGeneRank: A Digital Platform for Next-Generation Genetic Studies N/A
Active, not recruiting NCT05553106 - Evaluation of Cognitive Status, Kinesiophobia, Physical Activity Level, and Functional Performance in Coronary Intensive Care
Completed NCT03429920 - Effect of Fermented Soy Based Product on Cardiometabolic Risk Factors N/A
Recruiting NCT04390672 - Multivessel TALENT N/A
Enrolling by invitation NCT03314818 - Natural History of Carotid Plaque as Determined by 3D Ultrasound N/A
Withdrawn NCT03289104 - Improving Sternal Healing After Cardiac Surgery: Sternal Wire vs ZIPFIX N/A
Completed NCT02917213 - Imaging Silent Brain Infarct And Thrombosis in Acute Myocardial Infarction
Completed NCT02046902 - Developing and Testing a Personalized Evidence-based Shared Decision-making Tool for Stent Selection
Completed NCT01909349 - Web-based Aftercare Intervention for Cardiac Patients N/A
Completed NCT01944254 - The Precision of Pulmonary Artery Cardiac Output-measurements in Spontaneously Breathing Patients N/A
Recruiting NCT01457586 - Hemoderivative Imputable Complications in Initial Uncomplicated Heart Surgery Phase 4
Recruiting NCT01207167 - Mediators of Atherosclerosis in South Asians Living in America
Recruiting NCT01541163 - Heart and Ischemic STrOke Relationship studY N/A
Terminated NCT00968383 - Cardiovascular Magnetic Resonance for the Occluded Infarct-Related Artery Treatment N/A