Clinical Trial Summary

Hypotheses:

• A wearable sensor system can be used to accurately monitor three common gait parameters: gait speed, stride length, and torso motion.

Objectives:

This study will examine the feasibility of using wearable sensors to monitor common gait parameters: gait speed, stride length and torso motion. A wearable sensor system of 8 commercially available inertial measurement units (IMU) will be composed. These sensors will work in unison to monitor the gait parameters.

Technical Objectives

- Gather information on commercially available IMUs

- Use computer software to monitor and record data from IMUs

- Develop an algorithm that can monitor volunteer gait speed, stride length and torso motion

- Develop a graphical algorithm that compares healthy patient data to potential mild traumatic brain injury (mTBI ) candidates

- Determine if there is a potential for mTBI determination using the wearable sensors

- Accurately validate the wearable sensor system to the gait parameters measured using a Vicon motion analysis system


Clinical Trial Description

This study will examine the feasibility of using wearable sensors to monitor common gait parameters. A sensor system consisting of 8 wearable inertial measurement units (IMU) that include accelerometers, gyroscopes and magnetometers will be examined. These sensors will be programmed to work in unison to monitor gait parameters. The wearable sensor suite will have the capability to collect and store gait parameter data. The size and weight of each sensor should be similar to a standard metal wrist watch. The sensor suite will monitor three gait parameters: gait speed, stride length, and torso motion. These gait parameters are important for monitoring and detection of mild traumatic brain injury and as well as other neurological and vestibular disorders.

The wearable sensor suite will have the following design parameters:

- Devices must be able to both stream data real-time and record data onboard for later transfer to a computer.

- Data must be processed using a software analysis package such as MATLAB®

- Device must be no heavier than the size of a metal wrist watch

- Device must be easily put on and be easily operated by clinicians and users

- Device must have a battery life of greater than 4 hours The wearable sensor suite will then be validated using an eight camera Vicon infrared optical capture motion analysis system for comparison. Ten healthy volunteers will be asked to complete four different types of gait trials 5 times for each trial type while being monitored by the Vicon motion analysis system and the wearable sensor system.

The types of gait trials will be the following:

- Walk down a 20 ft path at self-selected normal stride length and cadence

- Walk down a 20 ft path with increased stride length while decreasing gait speed

- Walk down a 20 ft path with increased stride length while increasing gait speed

- Walk down a 20 ft path with decreased stride length while decreasing gait speed

- Walk down a 20 ft path with decreased stride length while increasing gait speed ;


Study Design

Observational Model: Case Control, Time Perspective: Prospective


Related Conditions & MeSH terms


NCT number NCT01620021
Study type Observational
Source University of South Florida
Contact
Status Completed
Phase N/A
Start date April 2011
Completion date June 2014

See also
  Status Clinical Trial Phase
Active, not recruiting NCT04867148 - The Prediction and Prevention of Disease by Using Big Data in Motion Analysis N/A
Completed NCT05538221 - Reliability of a Gait Analysis System
Recruiting NCT05530213 - Interlaboratory Reliability of 3D Gait Analysis N/A
Recruiting NCT05277181 - Wearable Technology as an Objective Tool for Measuring Running Gait
Recruiting NCT05989113 - Psychometric Qualities IPAQ-SF
Not yet recruiting NCT05305755 - Reliability of a Gait Analysis System
Not yet recruiting NCT06264999 - Robotic Assisted Knee Arthroplasty - Retain the Anterior Cruciate Ligament or Not N/A
Completed NCT03898544 - Gait Kinematics After Primary Total Knee Arthroplasty (TKA) Versus Revision TKA
Completed NCT05957783 - Changes in Upper Limb Kinematics in Children With Cerebral Palsy After Lower Limb Surgery
Recruiting NCT05751408 - Technical Evaluation of Commercial IMUs Within Clinical Gait Analysis
Completed NCT04788719 - Pre and Postoperative Gait Analysis of Randomized Patient to 2 Different Cement Fixation Methods N/A
Completed NCT04791605 - Gait Pattern Between a Cemented and Non-cemented Femur Stem N/A
Withdrawn NCT03858556 - Effect of Chuna Manipulation Therapy on Gait in Patients With Lumbar Disc Herniation
Completed NCT04529642 - Is the 6-Minute Walking Test Combined With an IMU Useful in Stroke Rehabilitation Assessment?
Recruiting NCT05452655 - Intensive Multidisciplinary Rehabilitation and Biomarkers in Parkinson's Disease N/A
Not yet recruiting NCT05855577 - Motor Function Efficacy of Pharmacological Treatments Targeting Energy Metabolism, in Parkinson's Patients Phase 4
Recruiting NCT06253624 - Gait Pattern Analysis After All Arthroscopic ATFL Repair Surgery N/A
Completed NCT04230733 - Pelvis Displacement and Gait Indicators in Patient With Lumbar Disc Herniation
Recruiting NCT06304077 - Restorative Environments for Gait Therapy With VR N/A
Completed NCT04434313 - Treatment of Hemiparetic Gait Impairments Using Telehealth With the Moterum iStride Solution™ N/A