Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT02691741
Other study ID # ILH297-P003
Secondary ID
Status Completed
Phase N/A
First received
Last updated
Start date April 29, 2016
Est. completion date July 17, 2017

Study information

Verified date September 2019
Source Alcon Research
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The purpose of this study is to clinically evaluate the visual performance of two commercially available presbyopia-correcting Trifocal IOLs.


Description:

Subjects will attend a total of 9 visits (6 postoperative) over a 7 month period.


Recruitment information / eligibility

Status Completed
Enrollment 215
Est. completion date July 17, 2017
Est. primary completion date July 17, 2017
Accepts healthy volunteers No
Gender All
Age group 22 Years and older
Eligibility Inclusion Criteria:

- Diagnosis of bilateral cataracts with planned clear cornea cataract removal;

- Able to comprehend and willing to sign informed consent and complete all required postoperative follow-up procedures;

- Calculated lens power between 13.0 and 30.0 Diopters (D);

- Preoperative Best-corrected distance visual acuity (BCDVA) worse than 0.20 logMAR (ie, 0.22 logMAR or worse) in at least one eye;

- Potential postoperative BCDVA of 0.20 logMAR or better in both eyes. Note: Subjects with any pathology that could reduce visual potential should not be enrolled in this trial;

- Preoperative regular corneal astigmatism of < 1.00 D, in both eyes;

- Clear intraocular media other than cataract in both eyes;

- Other protocol-specified inclusion criteria may apply.

Exclusion Criteria:

- Reasonably expected to require an ocular surgical treatment at any time during the study (other than YAG capsulotomy);

- Previous refractive surgery or planned refractive surgery procedures throughout the entire duration of participation in the clinical study (including, but not limited to LASIK, astigmatic keratotomy and limbal relaxing incisions);

- Clinically significant corneal abnormalities including corneal dystrophy, inflammation or edema;

- Amblyopia;

- Previous corneal transplant;

- Any recurrent severe anterior or posterior segment inflammation of any etiology, and/or history of any disease producing an intraocular inflammatory reaction;

- Rubella, congenital, traumatic, or complicated cataracts;

- Glaucoma (uncontrolled or controlled with medication);

- Degenerative eye disorders;

- History of or current retinal conditions or predisposition to retinal conditions, previous history of, or a predisposition to, retinal detachment or presence of diabetic retinopathy;

- Optic nerve atrophy;

- Expected to require retinal laser treatment;

- Color vision deficiencies;

- Pregnant or lactating (current or planned during the course of the study);

- Other protocol-specified exclusion criteria may apply.

Study Design


Related Conditions & MeSH terms


Intervention

Device:
AcrySof® IQ PanOptix™ Presbyopia-Correcting IOL
Trifocal IOL(near, intermediate, distance) implanted for long-term use over the lifetime of the cataract patient
AT LISA® tri IOL
Trifocal IOL(near, intermediate, distance) implanted for long-term use over the lifetime of the cataract patient

Locations

Country Name City State
n/a

Sponsors (1)

Lead Sponsor Collaborator
Alcon Research

Outcome

Type Measure Description Time frame Safety issue
Primary Least Squares Mean Binocular Uncorrected Intermediate Visual Acuity (UCIVA) (60cm) at Day 120-180 VA was tested binocularly (both eyes together) with no refractive correction in place using an early treatment diabetic retinopathy study (ETDRS) chart set at 60 cm. VA was measured in "logarithm of the minimum angle of resolution" (logMAR), with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart. A lower numeric value represents better visual acuity. Day 120-180 from second eye implantation
Secondary Least Squares Mean Binocular UCIVA (60cm) at Day 120-180 VA was tested binocularly with no refractive correction in place using an ETDRS chart set at 60 cm. VA was measured in logMAR, with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart. A lower numeric value represents better visual acuity. Day 120-180 from second eye implantation
Secondary Least Squares Mean Binocular Uncorrected Distance VA (4m) at Day 120-180 VA was tested binocularly with no refractive correction in place using an ETDRS chart set at 4 meters. VA was measured in logMAR, with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart. A lower numeric value represents better visual acuity. Day 120-180 from second eye implantation
Secondary Least Squares Mean Binocular Uncorrected Near VA (40cm) at Day 120-180 VA was tested binocularly with no refractive correction in place using an ETDRS chart set at 40 cm. VA was measured in logMAR, with 0.1 logMAR increment corresponding to 5 letters, or 1 line, on an ETDRS chart. A lower numeric value represents better visual acuity. Day 120-180 from second eye implantation
Secondary Mean Photopic Binocular Defocus Curve at Day 120-180 A defocus curve is created by multiple measurements of one's visual acuity (VA) at different spherical powers. VA was measured in logMAR. A lower logMAR value indicates better visual acuity. No statistical test was performed. Day 120-180 from second eye implantation
Secondary Mean Photopic Without Glare Binocular Distance Contrast Sensitivity at Day 120-180 Contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly in lighted conditions with distance manifest correction in place and uncorrected, with no glare source illumination. Contrast sensitivity was assessed at spatial frequencies of 3, 6, 12, and 18 cycles per degree (CPD) and reported in log units. A higher numeric value represents better contrast sensitivity. No statistical test was performed. Day 120-180 from second eye implantation
Secondary Mean Photopic With Glare Binocular Distance Contrast Sensitivity at Day 120-180 Contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly in lighted conditions with distance manifest correction in place and uncorrected, with glare source illumination. Contrast sensitivity was assessed at spatial frequencies of 3, 6, 12, and 18 CPDs and reported in log units. A higher numeric value represents better contrast sensitivity. No statistical test was performed. Day 120-180 from second eye implantation
Secondary Mean Mesopic Without Glare Binocular Distance Contrast Sensitivity at Day 120-180 Contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly in dim to dark conditions with distance manifest correction in place and uncorrected, with no glare source illumination. Contrast sensitivity was assessed at spatial frequencies of 1.5, 3, 6, and 12 CPDs and reported in log units. A higher numeric value represents better contrast sensitivity. No statistical test was performed. Day 120-180 from second eye implantation
Secondary Mean Mesopic With Glare Binocular Distance Contrast Sensitivity at Day 120-180 Contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly with distance manifest correction in place and uncorrected, with glare source illumination. Contrast sensitivity was assessed at spatial frequencies of 1.5, 3, 6, and 12 CPDs and reported in log units. A higher numeric value represents better contrast sensitivity. No statistical test was performed. Day 120-180 from second eye implantation
Secondary Subject Satisfaction Recorded at Day 120-180 At the Month 6 Visit, subjects responded to the question: "Given your current postoperative vision, if you had to do it all over would you have the same lens implanted again?". Responses were reported as a percentage of subjects. No statistical test was performed. Day 120-180 from second eye implantation
See also
  Status Clinical Trial Phase
Active, not recruiting NCT06005675 - User Acceptability Evaluation of Pseudophakic Patients Previously Implanted With the TECNIS Odyssey IOL
Active, not recruiting NCT01382641 - Show Improved Vision With the HOYA AF-1 Aspheric Intraocular Lens After Cataract Surgery Phase 4
Completed NCT01382823 - Laser Cataract Surgery With the Femtosecond Laser Technology Phase 4
Completed NCT05991960 - Evaluation of Real-World Data on the Performance of the TECNIS Odyssey™ IOL
Active, not recruiting NCT05574959 - Clinical Investigation of the TECNIS® Next-Generation Intraocular Lens, Model DEN00V N/A
Completed NCT02492659 - Clinical Research of Femtosecond Laser-Assisted Cataract Surgery N/A
Completed NCT01279031 - Randomized Comparison of the Abbott WHITESTAR Signature System With Ellips Tranversal Ultrasound vs. the Alcon Infiniti With the Ozil Torsional Handpiece in Phacoemulsification: A Contralaterally-Controlled Trial Phase 4
Completed NCT01061463 - Occupational Cataracts and Lens Opacities in Interventional Cardiology : the O'CLOC Study N/A
Completed NCT01021761 - A Comparison of Prostaglandin E2 (PGE2) Inhibition of Acuvail(Ketorolac 0.45%), Xibrom (Bromfenac 0.09%)and Nevanac (Nepafenac)in Patients Undergoing Phacoemulsification Phase 4
Completed NCT00999492 - Selection of Aspheric Diffractive Multifocal Intraocular Lens (IOLS) Based on Corneal Wavefront Phase 4
Completed NCT01001806 - A Comparison of Peak Aqueous Penetration of Acuvail (Ketorolac 0.45%), Xibrom (Bromfenac 0.09%), and Nevanac (Nepafenac 0.1%)in Patients Undergoing Phacoemulsification Phase 4
Completed NCT00542581 - Acrysof Toric SN60T3corneal Astigmatism Between 0.75 and 1.00 D. N/A
Completed NCT03708367 - A Post-Market Evaluation of LipiFlow Treatment in Cataract Surgery Practice N/A
Completed NCT05575063 - Clinical Investigation of Healon EndoCoat Ophthalmic Viscosurgical Device (OVD) N/A
Recruiting NCT05531292 - PROOF-OF-CONCEPT STUDY FOR NEW INTRAOCULAR LENS, MODEL C0002 N/A
Completed NCT05058274 - Clinical Results of a Monofocal Aspheric Bitoric Intraocular Lens With Plate Haptics in Hyperopic Eyes (HIPER-AT-TORBI)
Completed NCT05053399 - Evaluation of the TECNIS Symfony® Toric Intraocular Lens N/A
Recruiting NCT02639845 - Eye Drop Application Monitor, Pilot Study N/A
Completed NCT01684007 - A Clinical Outcome Study of Two Multifocal Intraocular Lenses (IOL) in Cataract Patients N/A
Completed NCT00827073 - Comparison of Tetracaine 0.5% and Lidocaine 2% Jelly for Topical Phacoemulsification Cataract Surgery N/A