Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT03444545
Other study ID # CONNIE 1
Secondary ID
Status Completed
Phase
First received
Last updated
Start date October 29, 2015
Est. completion date December 31, 2018

Study information

Verified date January 2019
Source Karolinska Institutet
Contact n/a
Is FDA regulated No
Health authority
Study type Observational

Clinical Trial Summary

The capnodynamic method non-invasively calculates effective pulmonary blood flow (EPBF) continuously during surgery. In this study EPBF is compared to cardiac output (CO) measured with Transpulmonary Thermodilution (TPTD) att baseline and during hemodynamic changes in patients scheduled for open abdominal surgery at the Karolinska University Hospital, Solna, Sweden.


Description:

The capnodynamic method continuously calculates effective pulmonary blood flow (EPBF) with the help of a capnodynamic equation:

ELV x (FACO2(n) - FACO2(n-1)) = deltat(n) x EPBF (CvCO2 - CvCO2(n)) - VTCO2

ELV Effective lung volume [L] EPBF Effective pulmonary blood flow [L/min] n current breath n-1 previous breath FACO2 mean alveolar carbon dioxide fraction CvCO2 mixed venous carbon dioxide content [Lgas/Lblood] CcCO2n pulmonary end-capillary carbon dioxide content [Lgas/Lblood] VTCO2n volume [L] of carbon dioxide eliminated by the current, nth, breath delta t n current breath cycle time [min]

The equation above describes the mole balance between the CO2 delivered to lungs (EPBF), the volume taking part in the gas exchange (ELV) and CO2 excreted from the lungs (VTCO2). Normally there is no difference in CO2 between the actual and the preceding breath as the same amount of CO2 as delivered to the lungs as is excreted. When small changes in CO2 concentration are inserted into the equation obtained with short inspiratory or expiratory pauses in three out of nine breaths, nine different equations are obtained. The three unknown variables; ELV, EPBF and CvCO2can be solved with a linear least square optimization, a well-known numerical mathematical principle. The breathing pattern is automatically controlled by the ventilator which provides continuous calculations of EPBF where each value represents the average of the preceding nine breaths and renews with each breath as the newest replaces the oldest in the equation system.

At the day of surgery, included patients arrive 45 minutes earlier than otherwise at the surgical unit. After safe surgical checklist, vital signs are measured and epidural catheter inserted during a light sedation. Patients are anesthetized and muscle relaxed. An endotracheal tube is inserted in the trachea. Central arterial and vein catheters are inserted in sterile conditions with ultrasound guidance. Anesthesia is maintained with propofol in target controlled infusion and a short acting opioid is added as needed.

The protocol in the first 25 patients comprises simultaneous measurements of EPBF and CO at 1) Baseline (three measurments), 2) PEEP 15cmH2O, 3) PEEP 5cmH2O 4) before epidural activation, 5) 10-15 minutes after epidural activation 6) before volume infusion (when the patient is considered hypovolemic by the attending anaesthesiologist) and 7) shortly after volume infusion.

The attending anesthesiologist has the final responsibility on deciding the order of the protocol, for example if volume infusion should precede PEEP 15 cmH20 if hypovolemia is suspected.

The succeeding 10 patients (patients 26 to 35) are PEEP optimized before start of protocol using a standardiazed open lung tool recruitment. During the PEEP elevation (step 2 in the protocol above) 10cmH2O is added to the optimal PEEP and then released back to the baseline optimal PEEP (step 3).

Each comparative measurement of CO and EPBF in the protocol includes as an average of three TPTD measured with PiCCO2 and and the average of EPBF registered in the beginning of the first TPTD and in the end of the last TPTD.


Recruitment information / eligibility

Status Completed
Enrollment 35
Est. completion date December 31, 2018
Est. primary completion date September 28, 2018
Accepts healthy volunteers No
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria:

- Elective major abdominal surgery requiring advanced hemodynamic monitoring

Exclusion Criteria:

- Symptomatic coronary artery disease, chronic obstructive pulmonary disease with emphysema

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Capnodynamic method
Expired carbon dioxide is measured with a mainstream infrared sensor (Capnostat-3, Respironics Inc, Wallingford, CT, USA) and gas flow analysed with the flow sensor incorporated in the ventilator (Servo-i, Maquet Critical Care, Solna, Sweden) which was connected to a computer where all the mathematical analysis is carried out with a software written in Matlabâ„¢ (The Mathworks Inc, Natick, MA, USA). Instead of a fixed ratio between inspiration and expiration, three breaths out of every nine are prolonged with a 3-4 seconds longer expiration creating small changes in end- expiratory CO2
Transpulmonary thermodilution
Each cardiac output measurement comprises an average of three thermodilutions performed successively one after another. TPTD is performed with a PiCCO2 monitor (Pulsion Medical Systems SE, Feldkirchen, Germany)

Locations

Country Name City State
Sweden Karolinska University Hospital Stockholm

Sponsors (2)

Lead Sponsor Collaborator
Karolinska Institutet Stockholm County Council, Sweden

Country where clinical trial is conducted

Sweden, 

References & Publications (2)

Sander CH, Sigmundsson T, Hallbäck M, Sipmann FS, Wallin M, Oldner A, Björne H. A modified breathing pattern improves the performance of a continuous capnodynamic method for estimation of effective pulmonary blood flow. J Clin Monit Comput. 2017 Aug;31(4):717-725. doi: 10.1007/s10877-016-9891-z. Epub 2016 Jun 1. — View Citation

Sigmundsson TS, Öhman T, Hallbäck M, Redondo E, Sipmann FS, Wallin M, Oldner A, Hällsjö Sander C, Björne H. Performance of a capnodynamic method estimating effective pulmonary blood flow during transient and sustained hypercapnia. J Clin Monit Comput. 2018 Apr;32(2):311-319. doi: 10.1007/s10877-017-0021-3. Epub 2017 May 11. — View Citation

Outcome

Type Measure Description Time frame Safety issue
Primary Absolute values of EPBF compared to Cardiac output Mean cardiac output as measured with transpulmonary thermodilution compared to mean effective pulmonary blood flow measured with the capnodynamic method at the start of first thermodilution and during the time interval zero to 4 hours. Mean CO vs EPBF from zero to 4 hours perioperatively
Secondary Changes in EPBF compared to changes in Cardiac output Measurement of cardiac output and EPBF as described in Outcome 1 before and after hemodynamic changes. Mean CO vs mean EPBF from zero to 4 hours perioperatively
See also
  Status Clinical Trial Phase
Completed NCT05070819 - Atrial Natriuretic Peptide in Assessing Fluid Status N/A
Recruiting NCT05492968 - The Effect of Epinephrine, Norepinephrine and Phenylephrine on Intraoperative Hemodynamic Performance Phase 4
Recruiting NCT04064177 - Non-invasive Continuous Cardiac Output Monitoring in Preterm Infants Study
Withdrawn NCT03058653 - Use of a Mini-fluid Bolus to Identify Fluid Responsiveness N/A
Completed NCT02880683 - Cardiac Output Autonomic Stimulation Therapy for Heart Failure (COAST-HF) - Human NeuroCatheter Study N/A
Recruiting NCT00989508 - Myocardial Protection With Perhexiline in Left Ventricular Hypertrophy Phase 2/Phase 3
Completed NCT04955184 - Comparing Cardiac Output Measurements Using a Wearable, Wireless, Non-invasive PPG-Based Device to a Swan Ganz Catheter
Completed NCT02964663 - Comparison of the Ability of the esCCO and the Volume View to Measure Trends in Cardiac Output During Cardiac Surgery N/A
Completed NCT04641949 - Methoxyflurane and Fentanyl in LBNP Phase 4
Recruiting NCT04191408 - Predictive Ability of PEEP Induced Changes in CVP to Predict Volume Responsiveness in Mechanically Ventilated Patients After Major Abdominal Surgery N/A
Completed NCT01892696 - Effects of Inspiratory Flow Waveforms on Preload N/A
Recruiting NCT02728791 - Comparison of Allows Measurement of Cardiac Output Between Clearsight System and Transpulmonary Thermodilution N/A
Completed NCT03930979 - The Effect of Hyperoxia on Cardiac Output
Recruiting NCT04755764 - Effect of Labetalol, Atenolol, and Nifedipine on Maternal Hemodynamics Measured by ICG in Early Pregnancy
Enrolling by invitation NCT02702544 - Effect of Passive Leg Raising for Fluid Responsiveness N/A
Completed NCT01880684 - Validation of Capnography as a Predictor of Cardiac Output Change as Measured by FloTrac Phase 2
Recruiting NCT05035719 - Comparison Studies Comparing Hemodynamic Parameters Provided by the Caretaker Against Respective Gold Standard References
Not yet recruiting NCT04774978 - Real-time Continuous Cardiac Output Measurements N/A
Completed NCT02485782 - Changes in Volumetric Hemodynamic Parameters Induced by Fluid Removal on Hemodialysis N/A
Completed NCT00446017 - The Effects of Adrenaline and Milrinone in Patients With Myocardial Dysfunction After CABG Phase 3

External Links