Clinical Trials Logo

Clinical Trial Details — Status: Terminated

Administrative data

NCT number NCT03394768
Other study ID # 2017/2820
Secondary ID CHF2017.08-S
Status Terminated
Phase
First received
Last updated
Start date August 10, 2018
Est. completion date January 5, 2021

Study information

Verified date May 2021
Source Changi General Hospital
Contact n/a
Is FDA regulated No
Health authority
Study type Observational [Patient Registry]

Clinical Trial Summary

Cardiac output (CO) monitoring is an important tool in critically ill patients with haemodynamic instability. Traditionally, this has been accomplished using the pulmonary artery catheter (PAC). However, its use has been associated with various complications such as pneumothorax, arrhythmia, infection, pulmonary valve injury and embolism. This has led to the development of various minimally invasive cardiac output monitoring devices such as the oesophageal Doppler, PiCCO®, NiCCO® and FloTrac®. These devices however have their inherent limitations and though minimally invasive, are still not without risks to the patient. More recently a completely non-invasive CO monitoring device, NICOM Cheetah® has been introduced. Its use is based on thoracic bioreactance and it involves placement of 4 electrodes on either side of the thorax and it analyses changes in the phase of electrical voltage signal to the current applied across the thorax. Initial validation studies on the NICOM Cheetah® device revealed promising results. This study intends to compare cardiac output monitoring capabilities of the NICOM Cheetah® device using the FloTrac® CO monitor as a reference.


Description:

This will be a single centre prospective observational study. 50 consecutive patients admitted to the surgical intensive care unit (SICU) who are deemed to require cardiac output monitoring by the treating intensivist will be studied. Data collection is planned to be carried out over 2 consecutive years. Inclusion criteria are: 1) Patients who are deemed to require cardiac output monitoring. 2) Patients 21 years old and above. 3) American Society of Anesthesiologist (ASA) physical status 1 to 4. Exclusion criteria are: 1) Patients who are not on mechanical ventilation. 2) Patients who are not in sinus rhythm. 3) Patients with allergy to adhesive tapes. Patients will be treated as per department protocols and no additional intervention will be performed. Each patient will have an arterial catheter inserted as per our usual practice. All patients will have the FloTrac® connected to the arterial catheter (standard of care in CGH SICU) and the NICOM Cheetah® electrodes placed on the skin across the anterior thoracic wall. The FloTrac CO monitor is the current standard of care for cardiac output monitoring in the SICU of CGH. All patients deemed to require cardiac output monitoring will receive the FloTrac (as per departmental practice). The NICOM CO monitor involves the application of non-invasive sensor strips. In this study, it will be applied to patients receiving the FloTrac (standard of care), on top of the standard care of monitoring with Flotrac. For each patient, standard demographic and SICU data (age, gender, comorbidities, ASA score, admission diagnosis, type of surgery, cause of haemodynamic instability, mean arterial pressure (MAP) and central venous pressure (CVP) at the time of instituting cardiac output (CO) monitoring, use of inotropic support, presence of renal replacement therapy, presence of mechanical ventilation and APACHE II score) will be collected. Cardiac Index (CI) values simultaneously recorded by the NICOM Cheetah® and FloTrac® on a 15-minutely basis for the first 10-hour period will also be collected. These data will be automatically recorded by the FloTrac® Vigilance monitor and the Cheetah NICOM monitor and transcribed to a data entering/analysis software by the principal and co-principal investigators for data analysis.


Recruitment information / eligibility

Status Terminated
Enrollment 26
Est. completion date January 5, 2021
Est. primary completion date January 5, 2021
Accepts healthy volunteers No
Gender All
Age group 21 Years and older
Eligibility Inclusion Criteria: - Patients who are deemed to require cardiac output monitoring. - Patients 21 years old and above - American Society of Anesthesiologist (ASA) physical status 1 to 4 Exclusion Criteria: - Patients who are not on mechanical ventilation. - Patients who are not in sinus rhythm. - Patients with allergy to adhesive tapes

Study Design


Related Conditions & MeSH terms


Intervention

Other:
No additional intervention
Patients will be treated as per department protocols and no additional intervention will be performed

Locations

Country Name City State
Singapore Surgical Intensive Care Unit, Changi General Hospital Singapore

Sponsors (1)

Lead Sponsor Collaborator
Changi General Hospital

Country where clinical trial is conducted

Singapore, 

References & Publications (4)

Marqué S, Cariou A, Chiche JD, Squara P. Comparison between Flotrac-Vigileo and Bioreactance, a totally noninvasive method for cardiac output monitoring. Crit Care. 2009;13(3):R73. doi: 10.1186/cc7884. Epub 2009 May 19. — View Citation

Mehta Y, Arora D. Newer methods of cardiac output monitoring. World J Cardiol. 2014 Sep 26;6(9):1022-9. doi: 10.4330/wjc.v6.i9.1022. Review. — View Citation

Raval NY, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D. Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput. 2008 Apr;22(2):113-9. doi: 10.1007/s10877-008-9112-5. Epub 2008 Mar 14. — View Citation

Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C. Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med. 2007 Jul;33(7):1191-1194. doi: 10.1007/s00134-007-0640-0. Epub 2007 Apr 26. — View Citation

Outcome

Type Measure Description Time frame Safety issue
Primary Degree of Disagreement, % Comparison of cardiac output monitoring capabilities of the NICOM Cheetah® versus the FloTrac®. Cardiac output defined as the amount of blood the heart pumps through the circulatory system in a minute. Bland-Altman analysis used to compare the level of disagreement between the 2 devices, %. 10 hours of data collected for each participant
See also
  Status Clinical Trial Phase
Completed NCT05070819 - Atrial Natriuretic Peptide in Assessing Fluid Status N/A
Recruiting NCT05492968 - The Effect of Epinephrine, Norepinephrine and Phenylephrine on Intraoperative Hemodynamic Performance Phase 4
Completed NCT03444545 - Continous Cardiac Output - Non-Invasive Evaluation (CONNIE)
Recruiting NCT04064177 - Non-invasive Continuous Cardiac Output Monitoring in Preterm Infants Study
Withdrawn NCT03058653 - Use of a Mini-fluid Bolus to Identify Fluid Responsiveness N/A
Completed NCT02880683 - Cardiac Output Autonomic Stimulation Therapy for Heart Failure (COAST-HF) - Human NeuroCatheter Study N/A
Recruiting NCT00989508 - Myocardial Protection With Perhexiline in Left Ventricular Hypertrophy Phase 2/Phase 3
Completed NCT04955184 - Comparing Cardiac Output Measurements Using a Wearable, Wireless, Non-invasive PPG-Based Device to a Swan Ganz Catheter
Completed NCT02964663 - Comparison of the Ability of the esCCO and the Volume View to Measure Trends in Cardiac Output During Cardiac Surgery N/A
Completed NCT04641949 - Methoxyflurane and Fentanyl in LBNP Phase 4
Recruiting NCT04191408 - Predictive Ability of PEEP Induced Changes in CVP to Predict Volume Responsiveness in Mechanically Ventilated Patients After Major Abdominal Surgery N/A
Completed NCT01892696 - Effects of Inspiratory Flow Waveforms on Preload N/A
Recruiting NCT02728791 - Comparison of Allows Measurement of Cardiac Output Between Clearsight System and Transpulmonary Thermodilution N/A
Completed NCT03930979 - The Effect of Hyperoxia on Cardiac Output
Recruiting NCT04755764 - Effect of Labetalol, Atenolol, and Nifedipine on Maternal Hemodynamics Measured by ICG in Early Pregnancy
Enrolling by invitation NCT02702544 - Effect of Passive Leg Raising for Fluid Responsiveness N/A
Completed NCT01880684 - Validation of Capnography as a Predictor of Cardiac Output Change as Measured by FloTrac Phase 2
Recruiting NCT05035719 - Comparison Studies Comparing Hemodynamic Parameters Provided by the Caretaker Against Respective Gold Standard References
Not yet recruiting NCT04774978 - Real-time Continuous Cardiac Output Measurements N/A
Completed NCT02485782 - Changes in Volumetric Hemodynamic Parameters Induced by Fluid Removal on Hemodialysis N/A