Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT00626093
Other study ID # CRD 388
Secondary ID
Status Completed
Phase Phase 4
First received
Last updated
Start date January 2008
Est. completion date April 2010

Study information

Verified date February 2019
Source Abbott Medical Devices
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The purpose of this prospective study is to evaluate the effect of cardiac resynchronization therapy (CRT) on the defibrillation threshold (DFT) estimates in cardiac resynchronization therapy defibrillators (CRT-D) patients. The hypothesis of the study is that defibrillation threshold (DFT) will decrease with 6 months of cardiac resynchronization therapy (CRT).


Description:

Cardiac Resynchronization Therapy (CRT) has emerged as a promising therapeutic addition in patients with drug refractory heart failure (HF). Along with providing relief of symptoms of HF, cardiac resynchronization therapy defibrillators (CRT-D) are used for the prevention of sudden cardiac death (SCD). Although there are concerns that the defibrillation threshold (DFT) estimates are elevated in the heart failure (HF) patient population due to lower LV ejection fraction (EF) and left ventricular (LV) dysfunction, there is paucity of data available to evaluate this theory.

Recently, two different studies in a retrospective manner evaluated the energy requirements in patients receiving cardiac resynchronization therapy defibrillators (CRT-D). Burke et al analyzed DFTs in 50 patients each implanted with a cardiac resynchronization therapy defibrillators (CRT-D) device and an implantable cardioverter defibrillator (ICD). Although the ejection fraction (EF) in cardiac resynchronization therapy defibrillators (CRT- D) group was lower than the implantable cardioverter defibrillator (ICD) group, the mean defibrillation thresholds (DFTs) between the two groups were not significantly different (10.2 ± 6.1 J for the cardiac resynchronization therapy (CRT) group vs. 9.5 ± 5.0 J for the control group)14. In the ASSURE study, Doshi et al. showed that patients receiving cardiac resynchronization therapy defibrillators (CRT-D) devices do not have higher energy requirements when compared to patients receiving modern single or dual chamber implantable cardioverter defibrillators (ICDs). Although, there was a trend toward higher energy requirements found among patients with higher degrees of heart failure.

Major cardiac resynchronization therapy (CRT) trials have showed that cardiac resynchronization therapy (CRT) therapy has positive effects on ejection fraction (EF) and the heart failure (HF) condition of the patient over time. But, there is no study that has evaluated the defibrillation thresholds (DFTs) in the heart failure (HF) patients over time of receiving cardiac resynchronization therapy (CRT). This information will help the clinicians decide if they need to perform more invasive procedures during device implant to lower DFTs in patients who do not meet the defibrillation safety margin or they should just wait over time for the CRT to reduce the defibrillation threshold (DFT). Also, there is no published data about the stability of defibrillation thresholds (DFTs) in heart failure (HF) patients over time. The results from this study will also help to clarify whether it is safe to never test the defibrillation thresholds (DFTs) post-implant in this patient population.


Recruitment information / eligibility

Status Completed
Enrollment 77
Est. completion date April 2010
Est. primary completion date April 2010
Accepts healthy volunteers No
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria:

- Patient meets standard indications for a Cardiac Resynchronization Therapy - Defibrillator (CRT-D).

- Patient will be implanted with an FDA approved St Jude Medical (SJM) Cardiac Resynchronization Therapy - Defibrillator (CRT-D) and compatible defibrillation lead system.

- Patient is able to tolerate defibrillation threshold (DFT) testing.

- Patient is geographically stable and willing to comply with the required follow-up schedule.

- Patient has a life expectancy of greater than 6 months from the time of implant.

- Patient has stable heart failure (HF) medications at least one month prior to enrollment.

Exclusion Criteria:

- Inability to successfully implant an intravascular lead Cardiac Resynchronization Therapy - Defibrillator (CRT-D) device. (i.e. exclude epicardial leads).

- Patient is getting his Cardiac Resynchronization Therapy - Defibrillator (CRT-D) device replaced.

- Inability to successfully obtain the defibrillation threshold (DFT) at implant.

- Currently participating in a clinical trial that includes an active treatment arm or another data collection registry.

- Recent (within 24 hours) administration of Nesiritide™.

- Patient is on amiodarone (other antiarrhythmic agents known to affect defibrillation thresholds - DFTs) at the time of enrollment.

- Patient is pregnant.

- Patient is less than 18 years old.

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Cardiac Resynchronization Therapy - Defibrillator (CRT-D)
Patients in the study who receive Cardiac Resynchronization Therapy - Defibrillator (CRT-D) are indicated for Cardiac Resynchronization Therapy (CRT). Patient undergoes Defibrillation threshold (DFT) testing at implant and at 6 months. Defibrillation threshold (DFT) testing will include 3 ventricular fibrillation (VF) inductions.

Locations

Country Name City State
United States Penrose Hospital Colorado Springs Colorado
United States Pacific Heart Institute Santa Monica California

Sponsors (1)

Lead Sponsor Collaborator
Abbott Medical Devices

Country where clinical trial is conducted

United States, 

References & Publications (1)

Gold MR, Hedayati A, Alaeddini J, Payne JP, Bailin S, Sturdivant JJ, Pradhan S, Oza AL. Temporal stability of defibrillation thresholds with cardiac resynchronization therapy. Heart Rhythm. 2011 Jul;8(7):1008-13. doi: 10.1016/j.hrthm.2011.02.006. Epub 201 — View Citation

Outcome

Type Measure Description Time frame Safety issue
Primary Defibrillation Threshold Difference Obtained in Volts (V) Between Implant and 6 Months All patients underwent defibrillation threshold testing at cardiac resynchronization therapy-defibrillator (CRT-D) implant and then at 6 months. The outcome measure is the difference in DFT (defibrillation threshold) in volts between implant and 6 months. Baseline and 6 months
Secondary Defibrillation Threshold Difference Obtained in Joules (J) Baseline and 6 months
See also
  Status Clinical Trial Phase
Completed NCT02923726 - Assessment of Primary Prevention Patients Receiving An ICD - Systematic Evaluation of ATP N/A
Recruiting NCT05175937 - REal World Assessment for Patients Implanted With Implantable CardioverTer DefibrIllatOr Using Bluetooth Technology
Completed NCT02026102 - A Pilot Trial of Patient Decision Aids for Implantable Cardioverter-Defibrillators (ICDs) N/A
Completed NCT01262508 - Development of Algorithms to Predict Hemodynamic Instability N/A
Completed NCT00998218 - Effect of Ranolazine on Arrhythmias and Microvolt T- Wave Alternans (MVTWA) Patients With LV Dysfunction Phase 3
Completed NCT00845286 - Marathon, Genetics, Inflammation and the Cardiovascular System: MAGIC-Trial N/A
Active, not recruiting NCT00562757 - Prospective Study to Identify Patients at Risk of Dangerous Ventricular Arrhythmias N/A
Completed NCT00622453 - Arrhythmias in Myotonic Muscular Dystrophy N/A
Completed NCT03833843 - Sudden Cardiac Death in Systemic Right Ventricle
Completed NCT04024865 - Domperidone and Risk of Serious Cardiac Events in Postpartum Women
Recruiting NCT05799833 - Low QRS Voltages in Young Healthy Individuals and Athletes
Recruiting NCT05694572 - Post-market Surveillance of Micorport CRM Cardiac Implantable Electronic Devices
Not yet recruiting NCT03622307 - Subcutaneous ICD Therapy Combined With VT Ablation for the Secondary Prevention of Sudden Cardiac Death N/A
Recruiting NCT02058771 - Utilising Lifemap to Investigate Malignant Arrhythmia Therapy
Recruiting NCT02056509 - The Effect of Chest Compression and Ventilation Coordination During Cardiopulmonary Resuscitation. N/A
Terminated NCT01948206 - Study of Paced QRS Duration as a Marker of Sudden Cardiac Death N/A
Completed NCT01227785 - Next Generation INCEPTA Implantable Cardioverter Defibrillator (ICD) and Cardiac Resynchronization Therapy (CRT-D) Field Following Study N/A
Completed NCT00534300 - Intravenous n-3 Fatty Acids and Ventricular Tachycardia in Patients With Implantable Cardioverter Defibrillator (ICD)-Pacemaker Phase 1/Phase 2
Terminated NCT00524862 - Standard Drug Therapy vs. Implanted Defibrillator for Primary Prevention of Sudden Cardiac Death N/A
Recruiting NCT00181233 - Imaging Techniques for Identifying Factors of Sudden Cardiac Death Risk