Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT03567096
Other study ID # SJM-CIP-10231
Secondary ID
Status Completed
Phase N/A
First received
Last updated
Start date October 9, 2018
Est. completion date September 30, 2023

Study information

Verified date January 2024
Source Abbott Medical Devices
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The objective of this clinical investigation is to evaluate the clinical benefits of left ventricle (LV) only pacing combined with automatic adjustment of AV timing (SyncAV) in patients receiving cardiac resynchronization therapy (CRT) after 6 months of therapy. This clinical investigation is a prospective, two-arm, randomized 1:1, multicenter feasibility study designed to evaluate the effectiveness of LV only with multipoint pacing (MPP) and SyncAV compared to bi-ventricular pacing with MPP and SyncAV. The clinical investigation will be conducted at approximately 7 centers in Europe and Canada. Approximately 120 subjects will be enrolled in the study. No site may enroll more than 33% of the total subjects. Data will be collected at enrollment, CRT implant procedure, hospital pre-discharge, one and 6 months post implant. Enrollment data collection will include demographics, cardiovascular history, medication, echocardiography measurements and quality of life questionnaire. CRT implant procedure data collection will include implanted system information and lead location. The electrical conduction recording procedure will include surface ECG and device IEGM recordings during various pacing configurations at implant or up to 45 days post implant. In patients who consent to invasive measurements (expected target of at least 80 patients), a hemodynamic recording procedure will include invasive hemodynamic measurements during various pacing configurations which may take place during device implant or up to 45 days post implant. Hospital pre-discharge data collection will take place within 3 days after the CRT implant, electrical conduction recordings visit or hemodynamic recordings visit and will include system information, surface ECG, and device IEGMs. In a subset of patients from selected centers that have access to this technology (expected 20 patients), non-invasive electrical activation data will be collected with body surface mapping within 45 days of the implant procedure. Patients will be randomized 1:1 to receive either biventricular pacing with multipoint pacing (MPP) or LV-only pacing with MPP at the one-month (± 15 days) visit. The 6-month (± 15 days) post randomization follow up visit will include surface ECG, IEGMs, echocardiographic parameters and quality of life questionnaire. Subjects participating in this clinical investigation will follow the hospital center standard of care from implant to 6 month follow up. The expected duration of enrollment is 1.5 year. The total duration of the clinical investigation is expected to be 2 years.


Recruitment information / eligibility

Status Completed
Enrollment 101
Est. completion date September 30, 2023
Est. primary completion date February 24, 2023
Accepts healthy volunteers No
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria: - Patients with approved indication for CRT scheduled to be implanted with an MPP-enabled ABT Quadripolar CRT pacing system, de novo or upgrade from non-CRT system - Patient must be > 18 years of age, able to provide informed consent and willing to comply with study requirements - Documented permanent left bundle branch block (LBBB) - Intrinsic QRS duration = 150 ms - Sinus (or atrial paced) rhythm with intact AV conduction with PR interval = 250 ms Exclusion Criteria: - Resting heart rate > 100 bpm - AV Block (1st degree with PR> 250 ms, 2nd or 3rd degree) - Documented persistent atrial tachycardia or atrial fibrillation at the moment of enrollment or patients not likely to remain in sinus (or atrial paced) rhythm for the duration of the study - Patients scheduled for AV node ablation to treat paroxysmal atrial arrhythmias - Recent (< 3 months) myocardial infarction, ablation, electrolyte imbalance, or any condition within the last 90 days that would contraindicate for CRT programming changes in the opinion of the investigator - Women who are pregnant or plan to become pregnant during the study course

Study Design


Intervention

Device:
BiV pacing with MPP and SyncAV
This cohort will receive Bi-Ventricular Pacing (left ventricle and right ventricle pacing) with MultiPoint Pacing of the left ventricle (two stimulation sites) and the SyncAV algorithm active.
LV only pacing with MPP and SyncAV
This cohort will receive left ventricular only pacing (no right ventricle pacing) with MultiPoint Pacing of the left ventricle (two stimulation sites) and the SyncAV algorithm active.

Locations

Country Name City State
Canada Montreal Heart Institute Montreal
France CHU Rennes Rennes
Italy Policlinico Casilino Roma
United Kingdom Queen Elizabeth Hospital Birmingham
United Kingdom St. Bartholomew's Hospital London
United Kingdom John Radcliffe Hospital Oxford

Sponsors (2)

Lead Sponsor Collaborator
Abbott Medical Devices Abbott

Countries where clinical trial is conducted

Canada,  France,  Italy,  United Kingdom, 

Outcome

Type Measure Description Time frame Safety issue
Other Hemodynamic response Acute changes in LV dP/dtmax (measured with pressure wire, pressure-volume loop catheter or non-invasive system) resulting from various CRT and SyncAV pacing configurations (subset of patients consenting to invasive measurements intraoperative
Other Left Ventricle Electrical Activation Acute changes in LV activation time resulting from various CRT and SyncAV pacing configurations (subset of patients undergoing body surface mapping). intraoperative
Primary Clinical response Patient clinical response to LV only pacing with SyncAV optimization at six-month follow-up evaluated by the Packer clinical composite score. up to 6 months
Secondary Left Ventricle reverse remodeling Patient reverse remodeling response to LV only pacing at six-month follow-up defined as end-systolic volume reduction >15%. up to 6 months
Secondary Electrical resynchronization Acute changes in surface ECG QRS duration resulting from various CRT and SyncAV pacing configurations. intraoperative
See also
  Status Clinical Trial Phase
Recruiting NCT05650307 - CV Imaging of Metabolic Interventions
Recruiting NCT05654272 - Development of CIRC Technologies
Recruiting NCT05196659 - Collaborative Quality Improvement (C-QIP) Study N/A
Active, not recruiting NCT05896904 - Clinical Comparison of Patients With Transthyretin Cardiac Amyloidosis and Patients With Heart Failure With Reduced Ejection Fraction N/A
Completed NCT05077293 - Building Electronic Tools To Enhance and Reinforce Cardiovascular Recommendations - Heart Failure
Recruiting NCT05631275 - The Role of Bioimpedance Analysis in Patients With Chronic Heart Failure and Systolic Ventricular Dysfunction
Enrolling by invitation NCT05564572 - Randomized Implementation of Routine Patient-Reported Health Status Assessment Among Heart Failure Patients in Stanford Cardiology N/A
Enrolling by invitation NCT05009706 - Self-care in Older Frail Persons With Heart Failure Intervention N/A
Recruiting NCT04177199 - What is the Workload Burden Associated With Using the Triage HF+ Care Pathway?
Terminated NCT03615469 - Building Strength Through Rehabilitation for Heart Failure Patients (BISTRO-STUDY) N/A
Recruiting NCT06340048 - Epicardial Injection of hiPSC-CMs to Treat Severe Chronic Ischemic Heart Failure Phase 1/Phase 2
Recruiting NCT05679713 - Next-generation, Integrative, and Personalized Risk Assessment to Prevent Recurrent Heart Failure Events: the ORACLE Study
Completed NCT04254328 - The Effectiveness of Nintendo Wii Fit and Inspiratory Muscle Training in Older Patients With Heart Failure N/A
Completed NCT03549169 - Decision Making for the Management the Symptoms in Adults of Heart Failure N/A
Recruiting NCT05572814 - Transform: Teaching, Technology, and Teams N/A
Enrolling by invitation NCT05538611 - Effect Evaluation of Chain Quality Control Management on Patients With Heart Failure
Recruiting NCT04262830 - Cancer Therapy Effects on the Heart
Completed NCT06026683 - Conduction System Stimulation to Avoid Left Ventricle Dysfunction N/A
Withdrawn NCT03091998 - Subcu Administration of CD-NP in Heart Failure Patients With Left Ventricular Assist Device Support Phase 1
Recruiting NCT05564689 - Absolute Coronary Flow in Patients With Heart Failure With Reduced Ejection Fraction and Left Bundle Branch Block With Cardiac Resynchronization Therapy