Clinical Trials Logo

Clinical Trial Summary

The purpose of this study is to evaluate the safety and effectiveness of the G4 System when worn for up to 7-days by subjects >18 years-old with diabetes mellitus requiring insulin.

Device performance will be primarily evaluated in terms of the proportion of G4 System values that are within ±20% of YSI reference value for glucose levels >80 mg/dL and ±20 mg/dL at meter glucose levels <80 mg/dL. These G4-YSI matched pairs will be collected during in-clinic sessions on Days 1, 4, and 7 of Sensor wear. The proportion of matched pairs from this study (PTL-900360) meeting the above criteria will be compared to the proportion of such points observed in the DexCom clinical study that supported approval of the SEVEN PLUS System (PTL 300012).

Safety data of the G4 System will also be collected and safety will be characterized by the incidence of Adverse Device Effects, Serious Adverse Device Events, and Unanticipated Adverse Device Effects experienced by study participants.


Clinical Trial Description

This will be a non-randomized, single-armed study conducted in the United States. Following screening, a minimum of 60 to a maximum of 80 subjects at up to 4 clinical sites will be enrolled. All participants will wear the G4 Sensor for one 7-day wear period (~168 hours). A subset of 20-30 subjects will wear two G4 Systems (one of which will be blinded) during the 7-day wear period for the purposes of assessing precision. During home use, subjects will be asked to use the meter provided to them to take fingersticks for calibration and diabetes management. Subjects will also be asked to confirm high and low G4 System glucose alerts/alarms by taking a comparative fingerstick immediately after receiving an alert/alarm.

All subjects will participate in one, 8-hour in-clinic session on Day 1, 4, or 7 of the study to collect accuracy information against a laboratory standard (YSI) and against SMBG. During the in-clinic session, subjects will have venous blood draws for evaluation of the YSI blood glucose measurements. Carbohydrate consumption and insulin dosing may be manipulated during the in-clinic session to ensure a wide range of glucose values. For the duration of this study, subjects will be instructed to use G4 System information as an adjunct to (and not as a replacement for) standard SMBG guidance of diabetes self-management.

Subjects will perform G4 Sensor insertions at the clinic under supervision of study staff. Sensor insertion sites will be examined after removal of the Sensor(s). Study staff will document any irritation and/or Adverse Device Effects. ;


Study Design

Endpoint Classification: Safety/Efficacy Study, Intervention Model: Single Group Assignment, Masking: Open Label


Related Conditions & MeSH terms


NCT number NCT01111370
Study type Interventional
Source DexCom, Inc.
Contact
Status Completed
Phase N/A
Start date February 2010
Completion date April 2010

See also
  Status Clinical Trial Phase
Completed NCT03743779 - Mastering Diabetes Pilot Study
Completed NCT03786978 - Pharmaceutical Care in the Reduction of Readmission Rates in Diabetes Melitus N/A
Completed NCT01804803 - DIgital Assisted MONitoring for DiabeteS - I N/A
Completed NCT05039970 - A Real-World Study of a Mobile Device-based Serious Health Game on Session Attendance in the National Diabetes Prevention Program N/A
Completed NCT04507867 - Effect of a NSS to Reduce Complications in Patients With Covid-19 and Comorbidities in Stage III N/A
Completed NCT04068272 - Safety of Bosentan in Type II Diabetic Patients Phase 1
Completed NCT03243383 - Readmission Prevention Pilot Trial in Diabetes Patients N/A
Completed NCT03730480 - User Performance of the CONTOUR NEXT and CONTOUR TV3 Blood Glucose Monitoring System (BGMS) N/A
Recruiting NCT02690467 - Efficacy, Safety and Acceptability of the New Pen Needle 34gx3,5mm. N/A
Completed NCT02229383 - Phase III Study to Evaluate Safety and Efficacy of Added Exenatide Versus Placebo to Titrated Basal Insulin Glargine in Inadequately Controlled Patients With Type II Diabetes Mellitus Phase 3
Completed NCT05799976 - Text Message-Based Nudges Prior to Primary Care Visits to Increase Care Gap Closure N/A
Completed NCT06181721 - Evaluating Glucose Control Using a Next Generation Automated Insulin Delivery Algorithm in Patients With Type 1 and Type 2 Diabetes N/A
Recruiting NCT04489043 - Exercise, Prediabetes and Diabetes After Renal Transplantation. N/A
Withdrawn NCT03319784 - Analysis for NSAID VS Corticosteroid Shoulder Injection in Diabetic Patients Phase 4
Completed NCT03542084 - Endocrinology Auto-Triggered e-Consults N/A
Completed NCT02229396 - Phase 3 28-Week Study With 24-Week and 52-week Extension Phases to Evaluate Efficacy and Safety of Exenatide Once Weekly and Dapagliflozin Versus Exenatide and Dapagliflozin Matching Placebo Phase 3
Recruiting NCT05544266 - Rare and Atypical Diabetes Network
Completed NCT01892319 - An International Non-interventional Cohort Study to Evaluate the Safety of Treatment With Insulin Detemir in Pregnant Women With Diabetes Mellitus. Diabetes Pregnancy Registry
Completed NCT05031000 - Blood Glucose Monitoring Systems: Discounter Versus Brand N/A
Recruiting NCT04039763 - RT-CGM in Young Adults at Risk of DKA N/A