Study information

Verified date May 2023
Source Johnson & Johnson Vision Care, Inc.
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

This is a single-masked, bilateral, controlled, randomized, 2x2 crossover, non-dispensing clinical trial to characterize the effects of an HEV-Blocker on several measures of visual function in a presbyopic population.


Recruitment information / eligibility

Status Completed
Enrollment 37
Est. completion date April 14, 2022
Est. primary completion date April 14, 2022
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 40 Years to 70 Years
Eligibility Inclusion Criteria: - Potential subjects must satisfy all of the following criteria to be enrolled in the study: 1. Read, understand, and sign the STATEMENT OF INFORMED CONSENT and receive a fully executed copy of the form 2. Appear able and willing to adhere to the instructions set forth in this clinical protocol. 3. Be between at least 40 and not more than 70 years of age at the time of screening. 4. Possess a wearable pair of spectacles that provide correction for distance vision. 5. Be an adapted soft contact lens wearer in both eyes (i.e. worn lenses a minimum of 2 days per week for at least 8 hours per wear day, for 1 month of more duration). 6. Be wearing a presbyopic contact lens correction (e.g., reading spectacles over contact lenses, multifocal or monovision contact lenses, etc.) or if not respond positively to at least one symptom on the "Presbyopic Symptoms Questionnaire". 7. The subject must have normal color perception as evidenced by reading 17 out of the first 21 Ishihara plates correctly. 8. The subject's distance spherical equivalent of their refraction must be in the range of -1.00 D to -4.00 D. 9. The subject's refractive cylinder must be =0.75 D in each eye. 10. The subject's ADD power must be in the range of +0.75 D to +2.50 D in each eye. 11. The subject must have best corrected distance visual acuity of 20/20-3 or better in each eye. Exclusion Criteria: - Potential subjects who meet any of the following criteria will be excluded from participating in the study: 1. Be currently pregnant or lactating. 2. Any systemic disease (e.g., Sjögren's Syndrome), allergies, infectious disease (e.g., hepatitis, tuberculosis), contagious immunosuppressive diseases (e.g., HIV), autoimmune disease (e.g. rheumatoid arthritis), or other diseases, by self-report, which are known to interfere with contact lens wear and/or participation in the study. 3. Use of any of the following oral medications within 1 week prior to enrollment: oral retinoids, oral tetracyclines, oral phenothiazines, anticholinergics, oral/inhaled steroids, oral phosphodiesterase type 5 inhibitors, interferon alfa, antimycobacterial antibiotics, and nitroimidazole antibiotics. See Section 9.1 for additional details regarding excluded systemic medications. 4. Any current use of ocular medication with the exception of rewetting drops. 5. Any known hypersensitivity or allergic reaction to single use preservative free rewetting drops or sodium fluorescein. 6. Any previous or planned (during the course of the study) ocular surgery (e.g., radial keratotomy, PRK, LASIK dacryocystorhinostomy, cataract surgery, retinal surgery, peripheral iridotomy/iridectomy, etc.). 7. Participation in any contact lens or lens care product clinical trial within 7 days prior to study enrollment. 8. Employee or immediate family member of an employee of clinical site (e.g., Investigator, Coordinator, Technician). 9. Any active or ongoing ocular or systemic allergies that may interfere with contact lens wear. 10. A history of amblyopia, strabismus or binocular vision abnormality. 11. History of optic nerve or retinal disease or trauma 12. Abnormal color vison. 13. Any other ocular abnormalities that are known to interfere with contact lens wear and/or participation in the study. This may include, but not be limited to entropion, ectropion, chalazia, recurrent styes, glaucoma, history of recurrent corneal erosions, or corneal distortion 14. Any grade 3 or greater slit lamp findings (e.g., edema, corneal neovascularization, corneal staining, tarsal abnormalities, conjunctival injection) on the FDA scale. 15. Any current ocular infection or inflammation.

Study Design


Related Conditions & MeSH terms


Intervention

Device:
JJVC Investigational Multifocal Contact Lens TEST Lens
TEST Lens
ACUVUE® OASYS Multifocal
CONTROL Lens

Locations

Country Name City State
United States University of Georgia Athens Georgia

Sponsors (1)

Lead Sponsor Collaborator
Johnson & Johnson Vision Care, Inc.

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Starburst Diameter (mm) The starburst diameter (mm) was measured each subject binocularly. Starburst diameter was quantified by the horizontal spread of the spokes emerging from the center of a small, point-like stimulus. A calibrated custom-made micrometer (two sides with reverse threading) was used to spread two posts out from a central mid-point. The posts attach to a custom-designed ruler that indicates the lateral spread of the starburst. Measurements for this metric are positive. Smaller diameter indicates better lens performance. The average diameter (mm) for each lens type was reported by period. 30-minutes post lens insertion
Primary Halos Diameter (mm) Halo diameter (mm) was quantified and measured using the same micrometer as for starbursts diameter (mm). The Halo diameter (mm) was measured each subject binocularly. Measurements for halo diameter are positive, where smaller values indicate better performance. The average diameter (mm) for each lens type. 30-minutes post lens insertion
Primary Glare Disability Threshold Glare disability threshold was collected binocularly for each subject. Subjects were exposed to a target stimulus for 2 seconds; before the measurement was taken the annulus was set to a level below that which would cause the target stimulus to be veiled. The experimenter would then adjust the intensity of the annulus until subjects could no longer see the target stimulus. Subjects would indicate this by pressing a buzzer. Glare disability thresholds take on positive values, where higher values indicate better lens performance. The average glare disability level (change in log relative energy) for each lens type was reported. 30-minutes post lens insertion
Secondary Glare Discomfort (mm) The Glare discomfort was evaluated by calculating the change in height of the palpebral fissure measured while the participant views the mid-wave central grating target, compared to the maximal squint during a 5-second Photostress exposure induced by a solid, 10-degree broadband field, which completely obscures the target (~10-deg diameter). Both eyes were recorded and averaged, to produce an average palpebral fissure height before and during the Photostress exposure. Glare discomfort can take on negative and positive values, where smaller values indicate better lens performance. The average glare discomfort (change in palpebral fissure height) for each lens type was reported. 30-minutes post lens insertion
Secondary Heterochromatic Contrast Threshold (HCT) Using the same optical bench set-up described Glare Disability Threshold, to measure HCT, participants viewed the mid-wave (580 nm) grating target presented on presented on a short wave (460 nm) "sky-light background." The intensity of the background was adjusted, until the participant was no longer able to resolve the grating target. The log relative energy of the background needed to obscure the central target was recorded as HCT. Heterochromatic contrast thresholds are positive values, where higher values indicate better lens performance. The average threshold each lens type was reported. 30-minutes post lens insertion
Secondary Two-Point Threshold (mm) With 403nm Filter To create the two-point threshold stimulus, a single point- source like stimulus was spread into two distinct points of light, using a custom-built, collapsible light baffle. The two-point threshold was quantified as the minimum distance (mm) at which a participant could detect two distinct points of light, separated by a small black space. Measurements for this metric are positive. Smaller distances indicate better lens performance. The average distance for each lens type and period. 30-minutes post lens insertion
Secondary Two-Point Threshold (mm) Without 403nm Filter To create the two-point threshold stimulus, a single point- source like stimulus was spread into two distinct points of light, using a custom-built, collapsible light baffle. The two-point threshold was quantified as the minimum distance (mm) at which a participant could detect two distinct points of light, separated by a small black space. Measurements for this metric are positive. Smaller distances indicate better lens performance. The average distance for each lens type. 30-minutes post lens insertion
See also
  Status Clinical Trial Phase
Recruiting NCT06019806 - Effects of Forward Head Posture Correction on Visual Acuity in Low Level Visually Impaired University Students N/A
Completed NCT04567186 - Evaluation of a Daily Disposable Novel Multifocal Contact Lens in a Hyperopic Population N/A
Completed NCT02568254 - Evaluation of Three Daily Disposable Contact Lenses N/A
Completed NCT01951703 - Senofilcon A Investigational Manufacturing Process N/A
Terminated NCT00637468 - EAGLE - Multicenter Study of the European Assessment Group for Lysis in the Eye Phase 3
Recruiting NCT04632901 - A New Parameter for the Assessment of Distance Visual Capacity: the Critical Visual Acuity (CVA)
Active, not recruiting NCT03747601 - Temporal Interference Brain Stimulation N/A
Completed NCT06131476 - Clinical Investigation of Visual Acuity in Contact Lens Wearers After Instillation of a Lipid- Based Lubricating Eye Drop Phase 1
Recruiting NCT05414565 - Post-Market Study of Alcon Intraocular Lenses
Completed NCT05056870 - Clinical Evaluation of Spherical Soft Contact Lenses, Toric Soft Contact Lenses and Spectacles in Low Astigmats N/A
Completed NCT04615507 - Evaluation of a Daily Disposable Novel Multifocal Contact Lens in a Myopic Population; Part 2 N/A
Completed NCT05099380 - Validation of Senofilcon A With New UV / HEV Filter N/A
Completed NCT03713281 - Evaluation of a Toric Multifocal Contact Lens Manufactured in Etafilcon Material in a Low ADD Hyperopic Population
Completed NCT05582304 - Role of Glare and Spectral Filtering on Contrast Sensitivity: A Pilot Study N/A
Completed NCT03742271 - Evaluation of Senofilcon A With New UV-blocker on a Neophyte Population N/A
Completed NCT05554640 - Clinical Performance of Two Daily Disposable Silicone Hydrogel Contact Lenses N/A
Completed NCT01926704 - MTF Image Modifications N/A
Completed NCT04649606 - Dynamic Characterisation of Meibomian Gland Structure
Completed NCT05101252 - Comparison of a Daily Disposable Multifocal Contact Lens to a Marketed Product N/A
Not yet recruiting NCT06054763 - Buyang Huanwu Decoction and Normal Tension Glaucoma N/A