Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT04974697
Other study ID # CR-6435
Secondary ID
Status Completed
Phase N/A
First received
Last updated
Start date July 19, 2021
Est. completion date October 29, 2021

Study information

Verified date October 2022
Source Johnson & Johnson Vision Care, Inc.
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The study is a bilateral, single-masked, single-arm, 4-visit dispensing study. There will be one study treatment, with the subject being dispensed lenses for 6 to 8 days, then lens power will be optimized (if necessary) and dispensed for a further 2 weeks to assess the clinical performance of investigational multifocal toric contact lenses.


Recruitment information / eligibility

Status Completed
Enrollment 95
Est. completion date October 29, 2021
Est. primary completion date October 29, 2021
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 40 Years to 70 Years
Eligibility Inclusion Criteria: - Potential subjects must satisfy all of the following criteria to be enrolled in the study: 1. Read, understand, and sign the STATEMENT OF INFORMED CONSENT and receive a fully executed copy of the form. 2. Appear able and willing to adhere to the instructions set forth in this clinical protocol. 3. Be at least 40 and not more than 70 years of age at the time of screening. 4. Own a wearable pair of spectacles if required for their distance vision. 5. Be an adapted soft contact lens wearer in both eyes (i.e. worn lenses for at least 8 hours per day at least two days per week for the past 4 weeks). 6. Be already wearing a presbyopic contact lens correction (e.g., reading spectacles over contact lenses, multifocal or monovision contact lenses, etc.) or if not respond positively to at least one symptom on the "Presbyopic Symptoms Questionnaire". 7. The subject's distance spherical component of their refraction must be in the range of either -1.25 D to -3.75 D, or +1.25 D to +3.75 D. 8. The subject's refractive cylinder must be -1.00 D to -1.50 D in each eye, with the cylinder axes in the range of 90°±30°. 9. The subject's ADD power must be in the range of +0.75 D to +2.50 D in each eye. 10. The subject must have best corrected distance visual acuity of 20/20-3 or better in each eye. Exclusion Criteria: - Potential subjects who meet any of the following criteria will be excluded from participating in the study: 1. Be currently pregnant or lactating. 2. By self-report, have any systemic disease (e.g. Sjögren's Syndrome), allergies, infectious disease (e.g., hepatitis, tuberculosis), contagious immunosuppressive diseases (e.g., HIV), autoimmune disease (e.g. rheumatoid arthritis), or other diseases, by self-report, which are known to interfere with contact lens wear and/or participation in the study. 3. Use systemic medications that may interfere with contact lens wear or cause blurred vision. 4. Currently use ocular medication (with the exception of rewetting drops). 5. Have any known hypersensitivity or allergic reaction to single use preservative free rewetting drops or sodium fluorescein. 6. Have had any previous, or have any planned, ocular or intraocular surgery (e.g. radial keratotomy, PRK, LASIK, cataract surgery, retinal surgery, etc.). 7. Have had previous eyelid injuries, surgeries or procedures which are known to have caused abnormal eyelid position or movement, by self-report. 8. Have participated in any contact lens or lens care product clinical trial within 7 days prior to study enrollment. 9. Be an employee or immediate family member of an employee of clinical site (e.g., Investigator, Coordinator, Technician). 10. Have a history of amblyopia or strabismus, by self-report. 11. Have a history of herpetic keratitis, by self-report. 12. Have ocular allergies, infections or other ocular abnormalities that are known to interfere with contact lens wear and/or participation in the study. This may include, but not be limited to entropion, ectropion, chalazia, recurrent styes, glaucoma, history of recurrent corneal erosions, aphakia, or corneal distortion. 13. Have any Grade 3 or greater slit lamp findings (e.g., edema, corneal neovascularization, corneal staining, tarsal abnormalities, conjunctival injection) on the FDA scale.

Study Design


Related Conditions & MeSH terms


Intervention

Device:
JJVC Investigational Multifocal Toric Contact Lens
TEST

Locations

Country Name City State
United States Birmingham Vision Care Bloomfield Township Michigan
United States Stam & Associates Eye Care Jacksonville Florida
United States VRC-West Jacksonville Florida
United States Maitland Vision Center Maitland Florida
United States Eye Associates of New York New York New York
United States Kannarr Eye Care Pittsburg Kansas
United States Tyler Eye Associates Tyler Texas
United States Sacco Eye Group Vestal New York

Sponsors (1)

Lead Sponsor Collaborator
Johnson & Johnson Vision Care, Inc.

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Overall Vision Overall vision was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20),where higher scores indicate a more favorable/positive response with a range of 0-120. 2-Week Follow-up
Primary LogMAR Visual Acuity Binocular visual performance on logMAR (Logarithm of Minimal Angle of Resolution) scale was evaluated at distance (4 m),intermediate (64 cm) and near (40 cm) under high luminance high contrast lighting condition using ETDRS (Early Treatment Diabetic Retinopathy Study) charts. The room illuminance was required to be between 7.3 and 7.9 EV (394-597 lux). For Distance (4 m), the acceptable range for the chart luminance was 10.5-10.7 EV (181-208 cd/m2). For Intermediate (64cm) and Near (40cm), the acceptable range for the chart luminance 10.8-11.1 EV (223-274 cd/m2). Letter by-letter results calculated the visual performance score for each chart read. A logMAR score closer to zero, or below zero, indicates a better visual acuity. A logMAR visual performance score of 0.0 is equivalent to Snellen visual acuity of 20/20. Hyperopes and myopes were combined based on historical data for subjects in this study with regard to visual performance logMAR. 2-Week Follow-up
Secondary Change From Baseline CLUE Vision Score Overall vision was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The change from baseline was calculated as the 2 week follow-up scores minus baseline scores. Change from baseline scores range from -120 to 120, where higher change from baseline scores indicate better performance. The average change from baseline was reported. Up to 2-Week Follow-up
Secondary Change From Baseline CLUE Comfort Score Overall comfort was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The change from baseline was calculated as the 2- week follow-up scores minus baseline scores. Change from baseline scores range from -120 to 120, where higher change from baseline scores indicate better performance. The average change from baseline was reported. Up to 2-Week Follow-up
Secondary Change From Baseline CLUE Handling Score Overall handling was assessed using the Contact Lens User Experience™ (CLUE) questionnaire. CLUE is a validated patient reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18- 65. Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120. The change from baseline was calculated as the 2- week follow-up scores minus baseline scores. Change from baseline scores range from -120 to 120, where higher change from baseline scores indicate better performance. The average change from baseline was reported. Up to 2-Week Follow-up
See also
  Status Clinical Trial Phase
Completed NCT06019806 - Effects of Forward Head Posture Correction on Visual Acuity in Low Level Visually Impaired University Students N/A
Completed NCT04567186 - Evaluation of a Daily Disposable Novel Multifocal Contact Lens in a Hyperopic Population N/A
Completed NCT02568254 - Evaluation of Three Daily Disposable Contact Lenses N/A
Completed NCT01951703 - Senofilcon A Investigational Manufacturing Process N/A
Terminated NCT00637468 - EAGLE - Multicenter Study of the European Assessment Group for Lysis in the Eye Phase 3
Recruiting NCT04632901 - A New Parameter for the Assessment of Distance Visual Capacity: the Critical Visual Acuity (CVA)
Active, not recruiting NCT03747601 - Temporal Interference Brain Stimulation N/A
Completed NCT06131476 - Clinical Investigation of Visual Acuity in Contact Lens Wearers After Instillation of a Lipid- Based Lubricating Eye Drop Phase 1
Recruiting NCT05414565 - Post-Market Study of Alcon Intraocular Lenses
Completed NCT05056870 - Clinical Evaluation of Spherical Soft Contact Lenses, Toric Soft Contact Lenses and Spectacles in Low Astigmats N/A
Completed NCT04615507 - Evaluation of a Daily Disposable Novel Multifocal Contact Lens in a Myopic Population; Part 2 N/A
Completed NCT05099380 - Validation of Senofilcon A With New UV / HEV Filter N/A
Completed NCT03713281 - Evaluation of a Toric Multifocal Contact Lens Manufactured in Etafilcon Material in a Low ADD Hyperopic Population
Completed NCT05582304 - Role of Glare and Spectral Filtering on Contrast Sensitivity: A Pilot Study N/A
Completed NCT03742271 - Evaluation of Senofilcon A With New UV-blocker on a Neophyte Population N/A
Completed NCT04995055 - Evaluation of Multifocal Contact Lens Designs With and Without an HEV Blocker on Visual Function N/A
Completed NCT05554640 - Clinical Performance of Two Daily Disposable Silicone Hydrogel Contact Lenses N/A
Completed NCT01926704 - MTF Image Modifications N/A
Completed NCT04649606 - Dynamic Characterisation of Meibomian Gland Structure
Completed NCT05101252 - Comparison of a Daily Disposable Multifocal Contact Lens to a Marketed Product N/A