Stress Urinary Incontinence Clinical Trial
Official title:
Pressure-Volume Measurements on the AMS 800TM Cuff
he idea of the experiment is to calculate the volume of fluid that is required to raise cuff pressure from 20 cmH2O to 120 cmH2O in the American Medical System's AMS 800 Artificial Urinary Sphincter (AUS).
The current AMS 800 is not personalized to the patient's needs. Most men only need 20 cm H2O
of urethral compression to remain dry when they are sedentary, yet need 120 cm H2O of
urethral compression when they are active. The current device compromises with an in between
pressure - as high as it can be without exceeding safety tolerance thresholds. As such many
men still leak when they are active. In fact, this compromise is so central to the device
that a new term has evolved to describe men's suboptimal level of continence after the AMS
800. We call it "social continence", meaning that they are dry enough to only have to wear
1-2 pads per day for protection.
The market is hungry for a device that can adapt to the patient's level of activity, reducing
the pressure most of the day to protect the urethra and then briefly increasing the pressure
when he is more active.
The experimental procedure is explained below and care will be taken to make sure the
pressure sensor is sterile. The experiment involves minimal risk and will be conducted by the
PI who has 10+ years of experience in AUS implantations.
The procedure for implanting the AMS 800 is mentioned in the Operating Manual attached with
this protocol. The treatment procedure is linked with the experimental setup designed by the
PI and the research student in the following manner:
1. Follow surgical preparation steps mentioned on pages 14-15 of the AMS 800 Operating
Manual. Once the cuff sizer is used to determine the cuff length (usually 4.0cm or
4.5cm), move on to next step. For graphical representation, follow steps 1 to 9b on
pages 18- 20 of the AMS 800 Operating Manual.
2. The AMS 800 cuff is filled with deionized water such that pressure is 20cmH2O. Follow
steps on pages 13-14 of the AMS 800 Operating Manual for filling the cuff.
3. The AMS 800 cuff around the urethra has a kink-free tubing which goes into a female
connector
4. The female connector connects to the male connector of the IV tubing
5. The IV tubing is connected to a T-Connector
6. One end of the T-Connector goes to the TruWave Pressure Transducer
7. Other end goes to a syringe
8. The system is pumped with deionized water and is therefore sterile.
9. Water is pumped into the cuff with the syringe and as the volume in the cuff increases,
corresponding cuff pressure changes. This is done until we have a volume for a pressure
reading of 120cmH2O on the pressure transducer.
10. Once the reading is obtained, the PI will flush out the excess water from the cuff, and
fill it with an optimal volume as required by the treatment procedure which he has
performed for over 10 years.
11. The treatment procedure will continue to follow the steps hereafter as has been designed
for AUS placement.
For all other details including device description and post-operative care, refer to the AMS
800 operating Manual. All devices including the AMS 800 (PMA-P000053) and the Edward
LifeSciences Pressure Transducer (501K- K142749) are FDA approved.
;
Status | Clinical Trial | Phase | |
---|---|---|---|
Recruiting |
NCT04829357 -
Post Market Clinical Follow-up Study on TVT ABBREVO® Continence System
|
||
Completed |
NCT05493735 -
Lidocaine for Pessary Check Pain Reduction
|
Phase 3 | |
Completed |
NCT04512053 -
A Phase 2 Study of TAS-303 in Female Patients With Stress Urinary Incontinence
|
Phase 2 | |
Active, not recruiting |
NCT06224335 -
Measurement of Intravaginal and Intra-abdominal Pressure and Pad Test During Sports Activities (SPORTVAGPRES)
|
||
Recruiting |
NCT05304312 -
The Role of Kegel Exercises Book to Improve Treatment in Stress Urinary Incontinence Women
|
N/A | |
Not yet recruiting |
NCT05527665 -
Sexual Fonction and Discomfort in Women After Midurethral Sling Surgery, Using PPSSQ
|
||
Not yet recruiting |
NCT04558762 -
Ten Years Follow-up After Insertion of a MUS (Mid Urethral Sling) Due to Stress Urinary Incontinence
|
||
Withdrawn |
NCT02524366 -
A Study of Transcorporal Versus Standard Artificial Urinary Sphincter Placement
|
N/A | |
Completed |
NCT01924728 -
Efficacy of Magnetic Stimulation for Stress Urinary Incontinence
|
N/A | |
Completed |
NCT01676662 -
Solace European Confirmatory Trial
|
N/A | |
Unknown status |
NCT01455779 -
Lyrette: Renewing Continence Objective and Subjective Efficacy Study
|
N/A | |
Terminated |
NCT01029106 -
Gynecare TVT Secur for the Management of Stress Urinary Incontinence (SUI)
|
N/A | |
Completed |
NCT01123096 -
Is the Cough Stress Test Equivalent to the 24 Hour Pad Test in the Assessment of Stress Incontinence?
|
N/A | |
Completed |
NCT01770691 -
Preliminary Performance Study of the New TIPI Device in the Prevention of Stress Urinary Incontinence
|
N/A | |
Withdrawn |
NCT00573703 -
Laparoscopic Burch Colposuspension Versus Transobturatory Tape for the Treatment of Female Urinary Stress Incontinence
|
Phase 4 | |
Completed |
NCT00234754 -
Trans-Obturator Tape Versus Trans-Vaginal Tape for Stress Urinary Incontinence in Women
|
N/A | |
Completed |
NCT00441454 -
Retropubic vs. Transobturator Tension-free Vaginal Tape
|
N/A | |
Completed |
NCT03985345 -
Prospective Evaluation of the Connected EMY Biofeedback Probe in the Management of Stress Urinary Incontinence.
|
N/A | |
Active, not recruiting |
NCT03671694 -
Laser Vaginal Treatment for SUI
|
N/A | |
Completed |
NCT04097288 -
Effects of Single Dose Citalopram and Reboxetine on Urethral and Anal Closure Function on Healthy Female Subjects
|
Phase 1 |