Clinical Trials Logo

Clinical Trial Details — Status: Terminated

Administrative data

NCT number NCT05936294
Other study ID # E17154
Secondary ID
Status Terminated
Phase N/A
First received
Last updated
Start date September 15, 2017
Est. completion date March 31, 2021

Study information

Verified date June 2023
Source Henry Ford Health System
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

This study will address whether acute adaptive RV pacing in sinus rhythm patients with severe LV systolic dysfunction and RBBB is superior to no ventricular pacing, and is as effective as echocardiographically optimized biventricular (BIV) pacing based on Medtronic ECG belt and cardiac performance as assessed by the echocardiographic parameters of RV and LV function.


Description:

Current guidelines state that CRT implantation is less beneficial for patients with RBBB compared to those with LBBB (2). Current evidence suggests that frequent or persistent RV-only pacing in patients with narrow QRS and LV dysfunction can be harmful due to ventricular desynchronization attributable to RV apex pacing (3-5). Studies suggest that LV-only pacing is not inferior to BIV pacing in sinus rhythm patients with LV dysfunction and LBBB and ICD back up (6,7). A novel algorithm of CRT by delivering synchronized LV pacing with the intrinsic conduction in patients with severe LV systolic dysfunction and LBBB (adaptive CRT) has been demonstrated to show that it is at least as effective as protocol-driven echocardiographic optimization. It also has shown a significant reduction in the probability of 30-day readmission for both HF and all-cause hospitalizations, 46% reduced incidence of AF compared to conventional CRT, and prolongation of CRT device battery life (8-10). Patients who had an adaptive CRT algorithm that provided > 50% synchronized LV pacing or had normal AV conduction with the adaptive CRT algorithm had decreased risk of death or heart failure hospitalization when compared to those with <50% synchronized LV pacing or echocardiography-optimized BIV pacing respectively (11). In another acute study the LV dP/dtmax was higher with LV than BIV pacing when LV pacing was associated with ventricular fusion caused by intrinsic activation (12). Several studies have suggested that optimization of the programmed atrioventricular delay (AVD) and interventricular delay (VVD) delays may incrementally improve the long-term outcome of BIV pacing (13-15). The lateral LV wall contracts early in patients with RBBB when compared to LBBB so LV pacing in the CRT devices is less likely to be beneficial. . In an experimental study of 12 dogs with tachycardia-induced cardiomyopathy and RBBB (6 dogs) or LBBB (6 dogs) (16), RV-only pacing enhanced LV function and synchrony as seen by dP/dtmax measured by catheter tip placed in the LV chamber and synchrony evaluation by cardiac MRI in the RBBB group as well or better than BIV pacing. LV-only pacing worsened function in the RBBB group. RV-only pacing was also evaluated in a prospective study of 7 patients with RBBB and RV dysfunction with most patients having congenital heart disease. LV global function was intact at baseline (17). Sequential atrioventricular RV pacing with an atrioventricular delay of 90% of the PR interval was superior to atrial-only pacing for both improvement in RV dP/dtmax and LV cardiac index as seen at cardiac catheterization. The RV dP/dtmax increased by 22% in RBBB patients with RV pacing and QRS decreased from 163 +/- 39 to 126 +/- 31 ms. The two studies together suggest that most of the benefit from BIV pacing in patients with RBBB and HF is due to the RV pacing component and the benefit might be reduced due to the LV pacing component. Better timing of RV pacing in patients with RBBB might significantly decrease desynchronization from RV pacing and maximize the benefit of RV pacing as seen in adaptive LV-pacing CRT in patients with LBBB and LV systolic dysfunction. These studies have led to the current hypothesis that adaptive RV pacing using RV-only pacing synchronized to LV activation when intrinsic AV conduction is normal is more physiological and will improve RV and LV function by Echocardiography parameters and on Medtronic ECG belt for activation in patients with RBBB and LV dysfunction because of improved synchronization and narrowing the QRS duration. As a secondary goal, battery life with BIV pacing will be compared to adaptive RV pacing assuming at least 50% of RV-only pacing will be achieved with adaptive CRT for RBBB.


Recruitment information / eligibility

Status Terminated
Enrollment 5
Est. completion date March 31, 2021
Est. primary completion date March 31, 2021
Accepts healthy volunteers No
Gender All
Age group 18 Years to 89 Years
Eligibility Inclusion Criteria: Included patients will have: - sinus rhythm or isolated atrial pacing - LV systolic dysfunction (EF <35% at time of device implant) - AV conduction <250 ms, RBBB, and - have an implanted CRT device based on the ACC/AHA/HRS guidelines (2, 18) or will have a CRT device implanted during the study enrollment per ACC/AHA/HRS guidelines (2). Exclusion Criteria: - atrial fibrillation - atrial flutter - atrial tachycardia - AV delay > 250 ms - sinus tachycardia with resting heart rate at time of the study 100 bpm - frequent APCs (> 25% of the total beats/min) or PVCs (>20% of the total beats/min), or - patients with EF > 40% at time of enrollment (if LV systolic function was found to be improved from time of implant).

Study Design


Related Conditions & MeSH terms


Intervention

Diagnostic Test:
Optimization of CRT Device
Each patient will have atrial pacing 10% higher than the sinus rate or atrial pacing at a rate of 60 bpm if significant sinus bradycardia is present. Each patient will have six ECGs and echocardiographic sequences performed in the same session. In addition six electrical activation evaluations by the Medtronic ECG belt will be done as well. The first study will always be with no ventricular pacing. One study will be echocardiography-optimized BIV pacing and four studies for echocardiography-optimized (?) adaptive RV-only pacing with different AV intervals. A random-sequence will be performed to determine the order for the rest of the studies.

Locations

Country Name City State
United States Henry Ford Hospital Detroit Michigan
United States Central Arkansas Veterans Healthare System Little Rock Arkansas

Sponsors (2)

Lead Sponsor Collaborator
Henry Ford Health System Central Arkansas Veterans Healthcare System

Country where clinical trial is conducted

United States, 

References & Publications (21)

Birnie D, Lemke B, Aonuma K, Krum H, Lee KL, Gasparini M, Starling RC, Milasinovic G, Gorcsan J 3rd, Houmsse M, Abeyratne A, Sambelashvili A, Martin DO. Clinical outcomes with synchronized left ventricular pacing: analysis of the adaptive CRT trial. Heart Rhythm. 2013 Sep;10(9):1368-74. doi: 10.1016/j.hrthm.2013.07.007. Epub 2013 Jul 11. — View Citation

Boriani G, Gardini B, Diemberger I, Bacchi Reggiani ML, Biffi M, Martignani C, Ziacchi M, Valzania C, Gasparini M, Padeletti L, Branzi A. Meta-analysis of randomized controlled trials evaluating left ventricular vs. biventricular pacing in heart failure: effect on all-cause mortality and hospitalizations. Eur J Heart Fail. 2012 Jun;14(6):652-60. doi: 10.1093/eurjhf/hfs040. Epub 2012 Apr 17. — View Citation

Boriani G, Kranig W, Donal E, Calo L, Casella M, Delarche N, Lozano IF, Ansalone G, Biffi M, Boulogne E, Leclercq C; B-LEFT HF study group. A randomized double-blind comparison of biventricular versus left ventricular stimulation for cardiac resynchronization therapy: the Biventricular versus Left Univentricular Pacing with ICD Back-up in Heart Failure Patients (B-LEFT HF) trial. Am Heart J. 2010 Jun;159(6):1052-1058.e1. doi: 10.1016/j.ahj.2010.03.008. — View Citation

Byrne MJ, Helm RH, Daya S, Osman NF, Halperin HR, Berger RD, Kass DA, Lardo AC. Diminished left ventricular dyssynchrony and impact of resynchronization in failing hearts with right versus left bundle branch block. J Am Coll Cardiol. 2007 Oct 9;50(15):1484-90. doi: 10.1016/j.jacc.2007.07.011. Epub 2007 Sep 24. — View Citation

Dubin AM, Feinstein JA, Reddy VM, Hanley FL, Van Hare GF, Rosenthal DN. Electrical resynchronization: a novel therapy for the failing right ventricle. Circulation. 2003 May 13;107(18):2287-9. doi: 10.1161/01.CIR.0000070930.33499.9F. Epub 2003 May 5. — View Citation

Epstein AE, DiMarco JP, Ellenbogen KA, Estes NA 3rd, Freedman RA, Gettes LS, Gillinov AM, Gregoratos G, Hammill SC, Hayes DL, Hlatky MA, Newby LK, Page RL, Schoenfeld MH, Silka MJ, Stevenson LW, Sweeney MO, Smith SC Jr, Jacobs AK, Adams CD, Anderson JL, Buller CE, Creager MA, Ettinger SM, Faxon DP, Halperin JL, Hiratzka LF, Hunt SA, Krumholz HM, Kushner FG, Lytle BW, Nishimura RA, Ornato JP, Page RL, Riegel B, Tarkington LG, Yancy CW; American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices); American Association for Thoracic Surgery; Society of Thoracic Surgeons. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices): developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Circulation. 2008 May 27;117(21):e350-408. doi: 10.1161/CIRCUALTIONAHA.108.189742. Epub 2008 May 15. No abstract available. Erratum In: Circulation.2009 Aug 4; 120(5):e34-5. — View Citation

Gasparini M, Birnie D, Lemke B, Aonuma K, Lee KL, Gorcsan J 3rd, Landolina M, Klepfer R, Meloni S, Cicconelli M, Grammatico A, Martin DO. Adaptive Cardiac Resynchronization Therapy Reduces Atrial Fibrillation Incidence in Heart Failure Patients With Prolonged AV Conduction: The Adaptive CRT Randomized Trial. Circ Arrhythm Electrophysiol. 2019 May;12(5):e007260. doi: 10.1161/CIRCEP.119.007260. No abstract available. — View Citation

Krum H, Lemke B, Birnie D, Lee KL, Aonuma K, Starling RC, Gasparini M, Gorcsan J, Rogers T, Sambelashvili A, Kalmes A, Martin D. A novel algorithm for individualized cardiac resynchronization therapy: rationale and design of the adaptive cardiac resynchronization therapy trial. Am Heart J. 2012 May;163(5):747-752.e1. doi: 10.1016/j.ahj.2012.02.007. — View Citation

Lane RE, Chow AW, Chin D, Mayet J. Selection and optimisation of biventricular pacing: the role of echocardiography. Heart. 2004 Dec;90 Suppl 6(Suppl 6):vi10-6. doi: 10.1136/hrt.2004.043000. — View Citation

Martin DO, Lemke B, Birnie D, Krum H, Lee KL, Aonuma K, Gasparini M, Starling RC, Milasinovic G, Rogers T, Sambelashvili A, Gorcsan J 3rd, Houmsse M; Adaptive CRT Study Investigators. Investigation of a novel algorithm for synchronized left-ventricular pacing and ambulatory optimization of cardiac resynchronization therapy: results of the adaptive CRT trial. Heart Rhythm. 2012 Nov;9(11):1807-14. doi: 10.1016/j.hrthm.2012.07.009. Epub 2012 Jul 14. — View Citation

Morales MA, Startari U, Panchetti L, Rossi A, Piacenti M. Atrioventricular delay optimization by doppler-derived left ventricular dP/dt improves 6-month outcome of resynchronized patients. Pacing Clin Electrophysiol. 2006 Jun;29(6):564-8. doi: 10.1111/j.1540-8159.2006.00402.x. — View Citation

Sawhney NS, Waggoner AD, Garhwal S, Chawla MK, Osborn J, Faddis MN. Randomized prospective trial of atrioventricular delay programming for cardiac resynchronization therapy. Heart Rhythm. 2004 Nov;1(5):562-7. doi: 10.1016/j.hrthm.2004.07.006. — View Citation

Sharma AD, Rizo-Patron C, Hallstrom AP, O'Neill GP, Rothbart S, Martins JB, Roelke M, Steinberg JS, Greene HL; DAVID Investigators. Percent right ventricular pacing predicts outcomes in the DAVID trial. Heart Rhythm. 2005 Aug;2(8):830-4. doi: 10.1016/j.hrthm.2005.05.015. — View Citation

Starling RC, Krum H, Bril S, Tsintzos SI, Rogers T, Hudnall JH, Martin DO. Impact of a Novel Adaptive Optimization Algorithm on 30-Day Readmissions: Evidence From the Adaptive CRT Trial. JACC Heart Fail. 2015 Jul;3(7):565-572. doi: 10.1016/j.jchf.2015.03.001. Epub 2015 Jun 10. — View Citation

Steinberg JS, Fischer A, Wang P, Schuger C, Daubert J, McNitt S, Andrews M, Brown M, Hall WJ, Zareba W, Moss AJ; MADIT II Investigators. The clinical implications of cumulative right ventricular pacing in the multicenter automatic defibrillator trial II. J Cardiovasc Electrophysiol. 2005 Apr;16(4):359-65. doi: 10.1046/j.1540-8167.2005.50038.x. — View Citation

Surawicz B, Childers R, Deal BJ, Gettes LS, Bailey JJ, Gorgels A, Hancock EW, Josephson M, Kligfield P, Kors JA, Macfarlane P, Mason JW, Mirvis DM, Okin P, Pahlm O, Rautaharju PM, van Herpen G, Wagner GS, Wellens H; American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; American College of Cardiology Foundation; Heart Rhythm Society. AHA/ACCF/HRS recommendations for the standardization and interpretation of the electrocardiogram: part III: intraventricular conduction disturbances: a scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American College of Cardiology Foundation; and the Heart Rhythm Society. Endorsed by the International Society for Computerized Electrocardiology. J Am Coll Cardiol. 2009 Mar 17;53(11):976-81. doi: 10.1016/j.jacc.2008.12.013. No abstract available. — View Citation

Tracy CM, Epstein AE, Darbar D, Dimarco JP, Dunbar SB, Estes NA 3rd, Ferguson TB Jr, Hammill SC, Karasik PE, Link MS, Marine JE, Schoenfeld MH, Shanker AJ, Silka MJ, Stevenson LW, Stevenson WG, Varosy PD. 2012 ACCF/AHA/HRS focused update of the 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol. 2012 Oct 2;60(14):1297-313. doi: 10.1016/j.jacc.2012.07.009. Epub 2012 Sep 10. No abstract available. — View Citation

van Gelder BM, Bracke FA, Meijer A, Pijls NH. The hemodynamic effect of intrinsic conduction during left ventricular pacing as compared to biventricular pacing. J Am Coll Cardiol. 2005 Dec 20;46(12):2305-10. doi: 10.1016/j.jacc.2005.02.098. — View Citation

Vidal B, Sitges M, Marigliano A, Delgado V, Diaz-Infante E, Azqueta M, Tamborero D, Tolosana JM, Berruezo A, Perez-Villa F, Pare C, Mont L, Brugada J. Optimizing the programation of cardiac resynchronization therapy devices in patients with heart failure and left bundle branch block. Am J Cardiol. 2007 Sep 15;100(6):1002-6. doi: 10.1016/j.amjcard.2007.04.046. Epub 2007 Jul 5. — View Citation

Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H, Kutalek SP, Sharma A; Dual Chamber and VVI Implantable Defibrillator Trial Investigators. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA. 2002 Dec 25;288(24):3115-23. doi: 10.1001/jama.288.24.3115. — View Citation

Ypenburg C, Van De Veire N, Westenberg JJ, Bleeker GB, Marsan NA, Henneman MM, Van Der Wall EE, Schalij MJ, Abraham TP, Barold SS, Bax JJ. Noninvasive imaging in cardiac resynchronization therapy--Part 2: Follow-up and optimization of settings. Pacing Clin Electrophysiol. 2008 Dec;31(12):1628-39. doi: 10.1111/j.1540-8159.2008.01237.x. — View Citation

* Note: There are 21 references in allClick here to view all references

Outcome

Type Measure Description Time frame Safety issue
Primary Effectiveness of adaptive Right Ventricular (RV) pacing Echocardiographic parameters to measure cardiac function, including:
Left Ventricular Outflow Tract velocity time interval(LVOT VTI),
dP/dt of Left Ventricle (LV) and RV,
Myocardial Performance Index (MPI),
Right Ventricular Outflow Tractvelocity time interval (RVOT VTI); and
3D derived Systolic Dyssynchrony Index (SDI).
1 hour
Secondary Battery/device longevity Calculations of current drain and pacing percentages for RV-only pacing versus BIV pacing. Measured by microjoule current consumption. 1 hour
See also
  Status Clinical Trial Phase
Enrolling by invitation NCT05903313 - A Study to Evaluate Accuracy and Validity of the Chang Gung ECG Abnormality Detection Software
Completed NCT04221763 - Mechanisms and Innovations in Cardiac Resynchronisation Therapy N/A
Completed NCT04537455 - Non-invasive Mapping Using Ultra-high Frequency Electrocardiography N/A
Completed NCT01169493 - Pacing Affects Cardiovascular Endpoints in Patients With Right Bundle-Branch Block (The PACE-RBBB Trial) N/A
Terminated NCT03524001 - Bifocal Right Ventricular PAcing in Right Bundle Branch blocK and Heart Failure With Reduced Ejection Fraction. The Study Tests the Superiority of Right Ventricular Bifocal Stimulation Over VVI Implantable Defibrillator in Right Bundle Branch Block and Heart Failure. N/A
Terminated NCT02441101 - Right Ventricular Septal Pacing in Patients With Right Bundle Branch Block and Heart Failure (The SPARK Trial) N/A
Recruiting NCT05265520 - His-Bundle Corrective Pacing in Heart Failure N/A
Not yet recruiting NCT05889208 - Prophylactic Permanent Pacemaker in Patients With Right Bundle Branch Block Undergoing Transcatheter Aortic Valve Implantation N/A