Clinical Trial Summary
Introduction and objectives: Left bundle branch area pacing (LBBAP), is a novel modality of
physiologic pacing that requires an adequate assessment of the endocardium as well as the
body-surface electrocardiographic signals for a successful pacing-lead implantation. The
objective of the study consist of assessing if LBBAP criteria, specifically septal and left
bundle branch pacing (LBBP) criteria, can be satisfactory measured using the signals
resulting from different cardiac pacing analyzers (PSAs) by comparing them to the
'conventional' signals obtained from multichannel polygraph systems (MPSs) during the
implantation procedure.
Methods: Comparative observational, prospective, multicenter study between the signals
obtained by PSA vs MPS of consecutive patients who underwent a LBBAP strategy as first
intention.
Description:
METHODS All participants will provide written informed consent before undergoing the
procedure. The study was approved by the Institutional Review Board of Hospital del Mar
(Barcelona) and complied with the declaration of Helsinki.
Study protocol and data collection This was a non-randomized, prospective multicentric study.
From july 2023 to may 2024 consecutive patients over 18 years of age who will undergo a LBBAP
system as first intention, will be enrolled from five participating hospitals. Baseline
characteristics including demography, pacing indication, basal QRS and left ventricular
ejection fraction will be collected. For the implantation procedure analysis, total
procedural time, total fluoroscopy time, total radiation dosage, pacing electrocardiogram
parameters including the paced QRS morphology and duration, V6 R wave peak time (V6-RWPT),
and changes between selective and non-selective LBBP will be recorded, as well as the pacing
lead parameters including R-wave amplitude, threshold and impedance with unipolar and bipolar
configurations. First 24hs complications will be all documented.
In a prephase of the study, a comparison between MPS and PSA surface signals, was performed.
Medtronic and Biotronik PSAs' use a four-leads monitoring system, three of them detecting
bipolar signals. The Right Arm (RA) lead was placed at the back of the patient, at the level
of the heart, to diminish the distance between leads and generating two 'pseudomonopolar'
signals together with the two other leads, (Left Arm (LA) and Left Foot (LF)) placed at "V1"
and "V6" standard positions ("modified" leads position). Boston Scientific (BSC) and Abbott
PSAs use a five-leads monitoring system, one of them (V1 lead) capable of detecting a
monopolar signal. In this case, the V1 lead was placed at the standard "V1" position and the
LA lead was moved to the "V6" position ("modified" lead position). The similarities on the
QRS morphology compared to the MPS were confirmed in all PSAs.
Implantation procedure All patients received standard medical treatment according to their
clinical conditions and underwent the LBBAP implantation in accordance with the standardized
pacemaker implantation technique. Bipolar electrograms were filtered through a bandpass of
30-500 Hertz (Hz) and recorded on a MPS at a speed of 100milimeter/second (mm/s) and a gain
of 0.1 miliVolt/centimeter (mV/cm). Before the implantation procedure, the PSA-leads will be
placed as previously described. Signals will be optimized and displayed in each PSA screen at
a speed range between [50-100] mm/sec and at gain of 0.05 mV/mm. The basal QRS morphology and
duration corresponding to 'V1' and 'V6' PSA-leads will be compared to those of the MSP, to
assess the correct position of the electrodes. In case of detecting a QRS morphology and/or
duration on the PSA significantly different from the MSP (more than 2 deflections in opposite
directions or a difference in QRS duration > 10 msec), repositioning of the leads wil be
suggested to obtain a similar QRS morphology and duration between both recording systems. The
sheath as well as the pacing-lead characteristics and final position (LVSP vs LBBP) will be
left at operator's discretion.
Left bundle branch (LBB) capture was defined in the presence of a paced QRS with right bundle
branch block (RBBB) morphology in V1 (initial r wave in lead V1) and at least one of the
following criteria:
1. V6-RWPT < 80 msec.
2. V1-V6 inter-peak interval > 33 msec.
3. QRS transition from non-selective left bundle branch pacing (NS-LBBP) to selective left
bundle branch pacing (S-LBBP) or NS-LBBP to left ventricular septal pacing (LVSP) during
pacing threshold.
4. Sudden increase of V6-RWPT > 15 msec because of reducing pacing output.
5. Left bundle potential-V6RWPT = stim-V6RWPT (±10 msec). For patients without initial r
wave in V1, LBBP will be considered in the presence of QRS transition during pacing.
LVSP will be defined by a paced QRS showing a typical terminal R-wave in V1, anatomical
position of the lead in the left ventricular septum documented by fluoroscopy and not
fulfilling the additional criteria previously described for LBBP.
During the implant procedure, initial septal unipolar pacing will be performed looking for an
"W" morphology in V1 in the mid-septal area, and the last recommendations of international
guidelines will be followed for placing the lead. Briefly, when the selected initial position
at operator's discretion is fulfilled, the lead is deployed using rapid turns and lead depth
within the septum is assessed using pacing QRS morphology, pacing impedance, unfiltered
electrogram amplitude from lead tip, or using contrast through the sheath. The objective of
the study is to validate and compare the criteria for the LBBAP in both recording systems.
Specifically, paced QRS in V1 and V6, V6-RWPT, V1-V6 inter-peak interval, abrupt Δ V6-RWPT >
15 msec at decremental pacing output and QRS transitions between non-selective and selective
capture of the conduction system during pacing were measured and confirmed in both, the PSA
and the MSP, to assess the final lead position. Endocavitary signals from the lead tip will
be evaluated and measured to detect congruent pre-ventricular signals from the conduction
system as well.
Monitoring and discharge Patients will be in-hospital for a period of 24 hours. Before
discharge, all patients will undergo a 12-lead electrocardiogram, an X-ray in
postero-anterior and lateral views, as well as clinical monitoring to rule out acute
complications.