Clinical Trials Logo

Clinical Trial Summary

The aim of this study is to analyze similarities and evaluate interchangeability of Visante and Cirrus OCT (optical coherent tomography)and to compare agreement between gonioscopy and the two systems for evaluation of anterior segment parameters.


Clinical Trial Description

All participants underwent routine ophthalmic examination by their physician. Three independent examiners performed gonioscopy using Spaeth Gonioscopic Grading System. Findings were recorded independently along with risk assessment of angle being closed as: high, medium, and low or no risk. Patient information was not available to doctor during gonioscopy. One eye of each patient was imaged using both Visante and Cirrus OCT. Imaging was performed under scotopic and photopic conditions and obtained from horizontal (3:00 and 9:00 o'clock) and vertical (6:00 and 12:00 o'clock) meridians. The same experienced independent examiner performed all Cirrus measurements. Visante was performed by 3 trained technicians using same protocol.

OCT images were exported and masked to be evaluated for presence of open or closed anterior chamber angle. Masked images were presented with no other patient information available to two examiners with glaucoma training working together to score on basis of iridocorneal angle as, closed (0 degree), high possibility (10 degree), medium risk (20 degree), and low or no risk (> 30 degree). ;


Study Design


Related Conditions & MeSH terms


NCT number NCT01895686
Study type Observational
Source Wills Eye
Contact
Status Completed
Phase
Start date November 2011
Completion date November 2012

See also
  Status Clinical Trial Phase
Completed NCT03284853 - Safety and Efficacy Study of PG324 (Netarsudil/Latanoprost 0.02% / 0.005%) Ophthalmic Solution Compared to GANFORT® Ophthalmic Solution in Open Angle Glaucoma or Ocular Hypertension Phase 3
Terminated NCT03611530 - CoQun Study - (COQUN): a Study to Evaluate the Effects of CoQun in Patients Affected by Open-angle Glaucoma N/A
Not yet recruiting NCT06441643 - Next Generation Rocklatan Phase 2
Not yet recruiting NCT02868502 - Different Glaucoma Treatments Effect on Intraocular Pressure Fluctuation With Postural Change in Eyes With Open-angle Glaucoma N/A
Completed NCT02272569 - STARflo European Safety and Efficacy Study N/A
Completed NCT02628223 - 180 Degree vs. 360 Degree Selective Laser Trabeculoplasty as Initial Therapy for Glaucoma N/A
Completed NCT01936389 - A Prospective Study to Assess the Hypotensive Efficacy of Rho-Kinase Inhibitor AR-12286 Ophthalmic Solution 0.5% and 0.7% in Patients With Exfoliation Syndrome and Ocular Hypertension or Glaucoma Phase 2
Completed NCT00941525 - Central Corneal Thickness and 24-hour Fluctuation of Intraocular Pressure Phase 4
Completed NCT00539526 - Evaluation of Hyperemia With the Use of Ocular Prostaglandin Analogues Phase 4
Completed NCT00121147 - Additivity Study: Additive Effect on Eye Pressure of Azopt and Alphagan P to Travatan N/A
Active, not recruiting NCT05035394 - Swedish Microinvasive Glaucoma Surgery Study (SMIGS) N/A
Completed NCT03450629 - Evaluation of Safety and Efficacy of PDP-716 Phase 3
Active, not recruiting NCT06061718 - Travoprost Intraocular Implant in Conjunction With Cataract Surgery Phase 3
Recruiting NCT05241938 - PSLT Compared to Prostaglandin Analogue Eye Drops N/A
Completed NCT01342406 - Effects of Selective Laser Trabeculoplasty on Aqueous Humor Dynamics
Completed NCT02250612 - SYL040012, Treatment for Open Angle Glaucoma Phase 2
Completed NCT01937312 - Effect of SIMBRINZA® Suspension as an Added Therapy to a Prostaglandin Analogue Phase 4
Terminated NCT01983579 - TF (SENSIMED Triggerfish) in Intraocular Anti-VEGF (Vascular Endothelial Growth Factor) Injection Phase 4
Completed NCT01937299 - Effect of SIMBRINZA® Suspension as an Added Therapy to TRAVATAN Z® Phase 4
Completed NCT01229982 - A Study Evaluating the Latanoprost Punctal Plug Delivery System (L-PPDS) in Subjects With Ocular Hypertension or Open-Angle Glaucoma Phase 2