Clinical Trials Logo

Clinical Trial Details — Status: Recruiting

Administrative data

NCT number NCT06054503
Other study ID # H03
Secondary ID
Status Recruiting
Phase N/A
First received
Last updated
Start date September 26, 2023
Est. completion date October 2023

Study information

Verified date September 2023
Source Archeon
Contact Stéphanie François
Phone +33 (0)3 81 66 23 80
Email s.francois@archeon-medical.com
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

Recently, the threat of viral pandemics (Covid-19, severe acute respiratory syndrome, avian flu H5N1, and H1N1), which might result in thousands of patients requiring mechanical ventilation, has accelerated the need for mechanical ventilation equipment. Disaster may create thousands of critically ill patients requiring mechanical ventilation and may force difficult allocation decisions when demand greatly exceeds supply. Creating a new monitor will only add one more product to the one already used and create confusion for the user. Therefore, the aim is now to develop an independent ventilation module, compatible with the already existing cardiac monitors, which integrates advanced ventilation monitoring functions (mechanical ventilation and RCP). This module could be used by the teams already equipped with multiparametric monitors and be a real added value as the monitoring of the ventilation is critical, especially in emergency situations. Then, it could answer to the clinical need and massively equip every hospital care center in the event of mass casualty incident or viral pandemic. Moreover, this device could be used by emergency teams during daily operations. The aim of this study is to validate a prototype of sensor intended to monitor ventilator parameters of ventilated patients and guide healthcare professionals to provide safe ventilation.


Recruitment information / eligibility

Status Recruiting
Enrollment 56
Est. completion date October 2023
Est. primary completion date October 2023
Accepts healthy volunteers No
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria: 1. Male or female patient aged 18 years or above 2. Patient ventilated with the reference ventilators: Servo U - Maquet Critical Care AB, Röntgenvägen 2, 171 54 Solna, Sweden for intensive care unit or Perseus - Drägerwerk AG & Co. KGaA, Moislinger Allee 53-55, 23542 Lübeck, Germany for operating room 3. Sedated patient with or without NBA (Neuromuscular Blocking Agent) 4. Patient with a positive end-expiratory pressure (PEEP) = 10 cm H2O 5. Patient stable without recent (less than one hour) or planned intervention (such a bronchoscopy, prone positioning, physiotherapy) 6. The primary attending physician approves use of the protocol on the patient 7. Anticipated length of intubation > 2 hours per patient 8. The patient or legally authorized representative has signed and dated a written Informed Consent Form (ICF) prior to the initiation of any clinical investigation procedures 9. Patient affiliated to the French social security regimen 10. Procedure will be performed at least 30 minutes after inhaled treatment (eg salbutamol) Exclusion Criteria: 1. Patient with a serious disease, for which the ventilation unplugging during a few seconds could have a serious impact (particularly patients with PaO2/Fi02 < 150 with pH < 7,30, with PEEP > 10 cm H20, or with Fi02 > 50%) 2. Patient with disease judged by the investigator to be incompatible with the conduct of the clinical investigation procedures or the interpretation of the clinical investigation results (example: IG-bronchopleural fistula) 3. Patient with a significant lower airway obstruction (asthma…), confirmed by ventilator spirometry. 4. Inability to maintain the airway intubation 5. Patient with a recent history of cardiac and/or respiratory arrest 6. Patient with multiple-organ system failure 7. Patient with hemodynamic instability (tachycardia, rapid change of vasopressor) 8. Pregnancy 9. Participation in another clinical investigation at the same time as the present clinical investigation 10. Nitric oxide use

Study Design


Related Conditions & MeSH terms


Intervention

Device:
prototype of sensor
The investigational device is composed of a sensor placed on the respiratory line of the ventilated patients and connected to a ventilation measure unit which records the ventilation parameters

Locations

Country Name City State
France CHU de Besançon Besançon

Sponsors (1)

Lead Sponsor Collaborator
Archeon

Country where clinical trial is conducted

France, 

Outcome

Type Measure Description Time frame Safety issue
Primary Number and proportion of patients for whom the insufflated volume measured by the sensor prototype is equivalent to the one measured by the ventilatory monito 2 hours
Secondary Number and proportion of patients for whom the exhaled volume measured by the sensor prototype are equivalent to the one measured by the ventilatory monitor 2 hours
Secondary Number and proportion of patients for whom the pressure measured by the sensor prototype are equivalent to the one measured by the ventilatory monitor 2 hours
Secondary Number and proportion of patients for whom the flow rate measured by the sensor prototype are equivalent to the one measured by the ventilatory monitor 2 hours
Secondary Precision range of the insufflated volume between the sensor prototype and the one measured by the reference ventilator 2 hours
Secondary Number and proportion of patients for whom the leak percentage calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and proportion of patients for whom the insiratory time calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and proportion of patients for whom the expiratory time calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and proportion of patients for whom the pause time calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and proportion of patients for whom the plateau time calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and proportion of patients for whom the respiratory rate calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and proportion of patients for whom the mechanical power calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and proportion of patients for whom the lung resistance calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and proportion of patients for whom the lung compliance calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and proportion of patients for whom the lung elastance calculated by the sensor prototype are equivalent to the one calculated by the ventilatory monitor ventilator 2 hours
Secondary Number and frequency of device deficiencies 2 hours 30 minutes
Secondary Number and frequency of adverse events 2 hours 30 minutes
See also
  Status Clinical Trial Phase
Completed NCT01997931 - The Impact of Bispectral Index Monitoring on Sedation Administration in Mechanically Ventilated Patients N/A
Completed NCT04134481 - Effect Of Clustered Nursing Interventions On Physiological Response In Critically Ill Patients
Completed NCT05588180 - The Ratio Of Femoral Vein Diameter To Femoral Artery Diameter With Pulse Pressure Variation As A Diagnostic Tool N/A
Recruiting NCT06153498 - PK/PD Properties and Safety of Remazolam Besylate for Injection in ICU Patients With Impaired Renal Function Phase 2
Recruiting NCT01256866 - Sedation of Mechanically Ventilated Critically Ill Patients: Midazolam Versus Dexmedetomidine Phase 4
Completed NCT01237886 - Weaning And Variability Evaluation N/A
Completed NCT00470821 - Oral Melatonin in Critically Ill High-risk Patients Phase 4
Completed NCT00322010 - Early Directed Physical Therapy in the Management of Mechanically Ventilated Patients in a Medical Intensive Care Unit Phase 2
Completed NCT01090258 - Pilot Evaluation of Closed Loop Ventilation and Oxygen Controller N/A
Completed NCT00199641 - Evaluation of Two Enteral Nutrition Strategies in Mechanically Ventilated Patients N/A
Not yet recruiting NCT06182553 - Effect of ERCC and/or PEEP-ZEEP Maneuver on Oxygenation, Ventilation, and Airway Secretions Removal in MV Patients N/A
Completed NCT03524937 - Prevention of Delirium in Intensive Care by Melatonin Phase 2/Phase 3
Completed NCT03333395 - Nebulized Heparin and Salbutamol in Mechanically Ventilated Patients With AECOPD Phase 4
Completed NCT05387720 - Concurrent Trigger Sensitivity Adjustment And Diaphragmatic Facilitation On Weaning From Mechanical Ventilation N/A
Recruiting NCT03800849 - Mechanical Ventilation Practices in Uganda's Intensive Care Units
Not yet recruiting NCT05480371 - Cough Assist Device in Mechanically Ventilated Patients N/A
Recruiting NCT04983615 - Opioid vs. Benzodiazepine-based Sedation for Mechanically Ventilated Patients in the Internal Medicine Ward N/A
Completed NCT06124404 - Step 1 of A Two-step Trial to Evaluate the Effectiveness and Safety of Remimazolam Besylate for Sedation in ICU Patients Phase 2
Completed NCT00935896 - High Tidal Volume Induces Inflammation In Normal Lungs N/A