Urinary Retention Clinical Trial
Official title:
A Prospective Evaluation of the Catheter Science C3 "Umbrella Catheter" for Patients With Urinary Retention Due to Bladder Dysfunction (C3 Catheter Study)
The Catheter Science C3 "Umbrella Catheter" has been developed for management of urinary retention in patients who are either using a Foley or Self Intermittent Catheter. This device is inserted into the bladder in an extended state and once the stylet is removed the device folds into collapsed state. A thread is attached to the device and traverses thru the urethra and is attached to a bobber. With gentle traction on the thread the device engages the bladder neck and the central tube extends which allows for the flow of urine. Once voiding is complete, the thread is released and the device collapses back to the folded state into the bladder. Each voiding cycle is accomplished in a similar fashion. This is a novel technique in which the bladder is catheterized with the device from within the bladder into the bladder neck and urethra. The standard self catheter technique is for placement of a catheter from the outside into the urethra and into the bladder.
Interventional (clinical trial): Participants are assigned prospectively to an intervention
or interventions according to a protocol to evaluate the effect of the intervention(s) on
biomedical or other health related outcomes.
This single arm, prospective study is designed to produce valid scientific evidence
regarding:
1. Safety and efficacy of the Catheter Science C3 "Umbrella Catheter" in establishing
urinary drainage and allowing the control of micturition when indwelling for up to 7
days.
2. The reduction in Catheter Acquired Urinary Tract Infection (CAUTI)
A single arm study design was chosen because there is no alternative treatment to serve as an
appropriate control. The C3 catheterizes the bladder with engagement into the bladder neck
under the control of the patient or care giver. A Foley catheter, a pre- amendment device,
was considered for a control treatment. However, although a Foley catheter is used to
establish urinary drainage with close to 100% efficacy, it does not restore voiding function
like the C3, and severely impairs activities of daily life.
The intrinsic design flaws of a Foley catheter traversing thru the urinary passage which
violates the anatomical protective points which help to avoid bacterial contamination. The
distal 1/3 portion of the urethra in females is commonly colonized with bacteria. Standard
Catheter technique traverses from the outside into the entire length of the urethra to drain
the bladder. The standard self catheter flow characteristics are impacted due to the
placements of the "eyes" to the catheter lumen. The C3 lumen is open in line with the flow.
;
Status | Clinical Trial | Phase | |
---|---|---|---|
Withdrawn |
NCT05538273 -
Optimal Management of Urinary Catheter in Gyn Onc
|
||
Recruiting |
NCT05065255 -
Study of the Criteria for Choosing Speedicath Catheters During the Initiation of Self-catheterization
|
||
Completed |
NCT04236596 -
Pudendal Nerve Mapping Towards Improved Neuromodulation for Urinary Retention
|
N/A | |
Completed |
NCT05526651 -
Improvement of Urinary Retention Symptoms With Electroacupuncture Therapy
|
N/A | |
Not yet recruiting |
NCT06070545 -
The Effect of Modified Supine Position on Urinary Elimination After Percutaneous Coronary Intervention
|
N/A | |
Recruiting |
NCT03314025 -
Prophylactic Tamsulosin in Prevention of Post-operative Urinary Retention in Men After Transanal Endoscopic Microsurgery
|
Phase 2 | |
Recruiting |
NCT05439902 -
Efficacy of Alpha-blockers (Tamsulosin) in the Treatment of Symptomatic Dysuria in Multiple Sclerosis in Women
|
N/A | |
Enrolling by invitation |
NCT05259800 -
Peppermint Oil for Urinary Retention
|
N/A | |
Recruiting |
NCT06417346 -
Comparison of Laparoscopic and Open Inguinal Hernia Repair in Elderly Patients
|
N/A | |
Completed |
NCT03296280 -
Evaluation of Implementation of a National Point-of-Care Ultrasound Training Program
|
||
Completed |
NCT04715971 -
Urinary Retention on an Acute Geriatric Hospitalisation Unit.
|
||
Completed |
NCT03302936 -
Assessing Pyridium for Post-Sling Urinary Retention
|
Phase 4 | |
Completed |
NCT02886143 -
Effect of Active vs. Passive Voiding Trials on Time to Discharge, Urinary Tract Infection, and Urinary Retention
|
Phase 3 | |
Completed |
NCT02749188 -
Evaluation of Bladder Stimulation as a Noninvasive Technique of Urine Collection in Infant Who Have Not Acquired Walking
|
N/A | |
Completed |
NCT01220362 -
Effects of Thoracic Epidural Analgesia and Surgery on Lower Urinary Tract Function: A Randomized, Controlled Study
|
N/A | |
Completed |
NCT01284361 -
Comparison of Two Intermittent Urinary Catheters
|
N/A | |
Terminated |
NCT00804557 -
Efficacy And Safety Of The Uro-Ease® Urinary Catheter During Clean Intermittent Catheterization
|
N/A | |
Terminated |
NCT00563654 -
Randomized Trial Comparing Alfuzosin GITS 10 mg Daily With Placebo for Trial Without Catheter in Acute Urinary Retention With Long Term Follow up
|
N/A | |
Terminated |
NCT00563485 -
Randomized Trial Comparing Terazosin 5 mg Daily and Doxazosin GITS 4 mg Daily for Trial Without Catheter in Acute Urinary Retention With Long Term Follow up
|
N/A | |
Terminated |
NCT04682366 -
The Effect of Tamsulosin on Postoperative Urinary Retention
|