Hydrocephalus Clinical Trial
Official title:
A Flow Monitor for Pediatric Hydrocephalic Shunts - Bench Study of Sensor Alignment Accuracy and Repeatability
The study hypothesis is that nurses and doctors can use a transcutaneously powered
ultrasonic flow sensor to make repeatable and accurate hydrocephalic shunt flow
measurements.
The study participants will align the flowmeter probe with a flow sensor hidden under a
thick saline pad that simulates skin. A hidden pump will provide a known flow through the
flow sensor as the participants make their measurements. Each participant will repeat these
measurements over a period of weeks, and the data will tell whether operator skill
influences flow measurement accuracy.
The study hypothesis is that a transcutaneously powered ultrasonic flow sensor for
hydrocephalic shunts can be repeatedly and accurately read by medical personnel ranging from
nurses to senior neurosurgeons.
The test will involve a prototype implantable hydrocephalic shunt flow sensor and readout
electronics. The readout electronics use a pickup coil to transcutaneously operate the
sensor once the coil is aligned with sensor's inductive coupling coils.
During the test, the shunt flow sensor will be hidden under a thick saline gel pad whose
thickness approximates the skin thickness of an older pediatric patient. A syringe pump will
provide a known flow rate through tubing connected to the sensor, but the study participants
will not know the pump flow setting.
Three neurosurgeons and six floor nurses will perform this study. Each will use the
Transonic flowmeter readout to align the pickup coils of the flowmeter with the hidden
sensor to make flow measurements.
Each participant will repeat this measurement multiple times over non-consecutive days.
The recorded data will quantify measurement accuracy and repeatability between different
operators, and will determine whether operator skill influences flow measurement accuracy.
;
Observational Model: Cohort, Time Perspective: Cross-Sectional
Status | Clinical Trial | Phase | |
---|---|---|---|
Recruiting |
NCT06040697 -
Formative Usability Assessment of Wireless Thermal Anisotropy Devices
|
||
Terminated |
NCT02408757 -
Sonographic Monitoring of Weaning of Cerebrospinal Fluid Drainages
|
||
Terminated |
NCT01863381 -
Comparison of Continuous Non-Invasive and Invasive Intracranial Pressure Measurement
|
N/A | |
Completed |
NCT00652470 -
A Study Comparing Two Treatments for Infants With Hydrocephalus
|
Phase 2 | |
Completed |
NCT00196196 -
A Precision and Accuracy Study of the Codman Valve Position Verification (VPV) System.
|
Phase 3 | |
Terminated |
NCT05501002 -
Pilot Study to Evaluate the CereVasc® eShunt® System in the Treatment of Communicating Hydrocephalus
|
N/A | |
Active, not recruiting |
NCT05068128 -
Combined Flow and Pressure Study of Craniospinal Dynamic
|
||
Completed |
NCT04496414 -
Bactiseal Catheter Safety Registry in China
|
||
Terminated |
NCT02900222 -
Study of Choroid Plexus Cauterization in Patients With Hydrocephalus
|
N/A | |
Recruiting |
NCT04099823 -
MR Evaluation of Cerebrospinal Fluid (CSF) Dynamics
|
N/A | |
Unknown status |
NCT02775136 -
An Evaluation of a Non-invasive Brain Monitor
|
N/A | |
Completed |
NCT00692744 -
Quality of Life in Elderly After Aneurysmal Subarachnoid Hemorrhage (SAH)
|
N/A | |
Recruiting |
NCT04758611 -
The ETCHES I Study (Endovascular Treatment of Communicating Hydrocephalus With an Endovascular Shunt)
|
N/A | |
Completed |
NCT03595033 -
Hydrocephalus iPad-App Based Intervention Study
|
N/A | |
Completed |
NCT04207229 -
CERTAS Programmable Valve Registry
|
||
Terminated |
NCT01973764 -
Intraventricular Drain Insertion: Comparison of Ultrasound-guided and Landmark-based Puncture of the Ventricular System
|
N/A | |
Completed |
NCT01976559 -
Comparison of Continuous Noninvasive and Invasive Intracranial Pressure Measurement--Celda Infusion Subprotocol
|
N/A | |
Completed |
NCT02381977 -
Prevalence of Acute Critical Neurological Disease in Children: a Global Epidemiological Assessment
|
N/A | |
Recruiting |
NCT06086561 -
Longitudinal Measurements of Flow in Cerebrospinal Fluid Shunts With a Wireless Thermal Anisotropy Measurement Device
|
N/A | |
Recruiting |
NCT05910944 -
European Study of Prodromal iNPH
|