Background:

Impedance cardiography (ICG) is a noninvasive method for continuous left cardiac output monitoring based on measurement of thoracic electrical bioimpedance. The objective of this study is to validate an ICG device by investigating the agreement in cardiac output (CO; left ventricular output x heart rate) measurements performed by ICG and echocardiography. The ICG device (PhysioFlow) has a new filter technology reducing signal noise and enabling measurements in patients who are moving freely.

Material and Methods

- Design: Prospective observational study.

- Inclusion: 30 healthy term infants between 24 and 48 hours of age, after parental consent

- Methods: Continuous measurement with ICG over a 30 min period and point of care echocardiography

- Primary outcome: Agreement (bis and precision) between measurement of CO with ICG and echocardiography.

- Secondary outcomes: Agreement between other echocardiographic indices of myocardial function and ICG parameters (early diastolic filling ratio, cardiac index, ventricular ejection time and left cardiac work index).

- Statistics: Agreement between ICG and echocardiography data will be evaluated using Bland-Altman statistics, including the calculation of mean bias (average difference between measurements) with corresponding standard deviation and lower and upper limits of agreements (95% confidence interval).


Clinical Trial Description

n/a


Study Design


Related Conditions & MeSH terms


NCT number NCT03867825
Study type Observational
Source University Hospital of North Norway
Contact
Status Withdrawn
Phase
Start date January 2020
Completion date December 2020

See also
  Status Clinical Trial Phase
Completed NCT04585568 - Validation of Ballistocardiographic Biosensors and Other Hemodynamic Measures for Healthy Subjects
Not yet recruiting NCT02202239 - Effect of Induction and Maintenance of Anesthesia With Etomidate on Hemodynamics and Oxidative Stress in Diabetic Patients Phase 4
Terminated NCT01116973 - Reliability of Central Venous Pressure Measurements N/A
Recruiting NCT05752409 - Comparison of Esketamine-Propofol and Fentanyl-Propofol N/A
Completed NCT04595591 - Observation of Propofol Titration at Different Speeds N/A
Not yet recruiting NCT03729817 - REstoring Flow by REvascularization With Submaximal Angioplasty in Hemodynamic IntraCranial Atherosclerotic Stenosis N/A
Completed NCT01974557 - The Effect of Age on the Hemodynamic Response During Rest and Exercise in Healthy Humans N/A
Completed NCT00620386 - Endotracheal Intubation Using Bonfils Fibrescope Without Neuromuscular Blockade Phase 2/Phase 3
Recruiting NCT06082856 - Evaluation of the Hemodynamic Effect of Dexmedetomidine in Scheduled Outpatient Surgery N/A
Recruiting NCT04665817 - Diagnostic Accuracy of CCTA-derived Versus AngiogRaphy-dErived QuantitativE Flow Ratio (CAREER) Study
Recruiting NCT03684291 - Hemodynamic Effects of Ventilation Modes
Recruiting NCT04291196 - Virtual Reality to Reduce Pre-procedure Anxiety Prior ECT N/A
Terminated NCT02495662 - The LIPMAT Study: Liposomal Prednisolone to Improve Hemodialysis Fistula Maturation Phase 2
Completed NCT01559285 - Effects of Epidurally Administered Ropivacaine Concentration on the Hemodynamic Parameters Phase 4
Completed NCT04131699 - Electrical Velocimetry (ICON Cardiometry ) Assessment of Hemodynamic Changes During Pediatric Thoracoscopic Surgery
Completed NCT05145114 - Hemodynamic Repercussions in Different Therapeutic Positions in Premature Newborn N/A
Completed NCT04419662 - Evaluation of Patients After Cardiac Surgery: Novel Ultrasound Parameters for Quantification of Renal Perfusion & Analysis of Phenylephrines' Effect on Invasive Haemodynamics and Echocardiographic Measures Phase 4
Completed NCT02902068 - Hemodynamic Monitoring in Women Throughout Cesarean Sections
Recruiting NCT00730899 - Association Study of Gene Polymorphisms With Cardiac Performance N/A
Recruiting NCT03855579 - Levosimendan Versus Milrinone in Off Pump CABG Surgery Phase 4