Clinical Trial Summary

Pulse oximetry is a standard non-invasive method of measuring blood oxygen saturation (SpO2). In developing countries, pulse oximeters are rare because of expense and electricity requirements. Our ECEM group has developed the Phone Oximeter, which uses a cell phone (widely available in developing countries) to compute and analyze information from a pulse oximeter sensor. To further reduce costs, we use the integrated flashlight of a cell phone's camera to extract SpO2 (Camera Oximeter), eliminating the need for external hardware. This study aims to calibrate the Camera Oximeter by exposing 30 healthy adult volunteers to various altitudes in UBC's hypoxia chamber.


Clinical Trial Description

Purpose:

We at the Electrical and Computer Engineering in Medicine Group (ECEM) at the University of British Columbia, Vancouver, Canada, plan to make pulse oximetry available to resource poor countries by designing a low-cost, battery-powered pulse oximeter device. We have developed the Phone Oximeter, which uses a mobile phone to compute and analyze the information received from a pulse oximeter sensor. The use of mobile phones as patient monitors is appealing as they are widely available in many developing countries. Utilizing battery power, mobile phones do not rely on a continuous source of electricity. This is essential, as most low-resource settings lack adequate infrastructure and thus cannot provide the uninterrupted power supply required for conventional patient monitoring. Furthermore, a mobile phone has the efficiency, integrated display, and processing power required to analyze and store the data derived from the pulse oximeter sensors. Data from the pulse oximeter can be transmitted to referral centers for diagnostic and advisory purposes where cellular and networking services permit.

At present, the ECEM Phone Oximeter software runs on smartphones and receives input from an external proprietary oximeter sensor which is relatively expensive ($300 US). To further reduce cost and increase ubiquity of oximeters, we are developing two novel pulse oximeter systems for mobile phones that will be more integrated: the Audio Oximeter (AudioOx) and the Camera Oximeter. This study aims at calibrating the Camera Oximeter.

The Camera Oximeter uses the camera of a mobile phone as a photo detector. Illumination of the finger is provided through the integrated flash light. This allows completely eliminating external hardware, which allows the camera phone to become a standalone vital sign recording device at minimal cost (no additional cost other than the phone and the software). This principle has been applied in many mobile phone software applications available in various software repositories to measure the HR of the user. Preliminary research has shown that the extraction of SpO2 from a video recorded with a mobile phone is also possible. We have further extended this approach and want to calibrate and validate a software implementation of this approach.

Hypothesis:

We hypothesize that this study will allow us to successfully calibrate the Camera Oximeter.

Justification:

For the safety of the patient, the development of any new pulse oximeter equipment requires calibration and evaluation for accuracy. This is required to ensure that the pulse oximeter device is functional. Results of accuracy evaluations have to be reported so the clinician is informed and he/she can understand the limits of the device that he/she is using. The International Organization for Standardization (ISO) and a technical committee have created standards that regulate the calibration and accuracy evaluation procedures of pulse oximeters. These standards are enforced by regulatory bodies, such as the federal drug administration (FDA), to approve the medical device for its use.

Pulse Oximeter Theory and Calibration

The underlying principle of pulse oximeters is the Beer-Lambert law which stipulates that light intensity diminishes exponentially when traveling in an absorbing medium. Oxygenated (O2Hb) and de-oxygenated (Hb) hemoglobin have different extinction coefficients for different wavelengths and are the most important light absorbers in blood. In some wavelength ranges, Hb is more absorbent than O2Hb, and in others the absorption property is inverted. These regions are separated by isosbestic points. With each pulse, the volume of blood in the vessels and the optical path length increase, and therefore, the total absorbance also increases. As a result, the variation of volume can be registered as a variation of transmitted light intensity and recorded with a photodetector. This produces the PPG waveform. The difference in total absorbance at two distinct wavelengths allows the calculation of a modulation ratio R. The theoretical relationship between R and SaO2 can be described as:

SaO2 = (εHb(λ1) - εHb(λ2) x R) / (εHb(λ1) - εO2Hb(λ1) + (εO2Hb(λ2) - εHb(λ2)) x R)

The Beer-Lambert law ignores scattering of light. Various other design aspects of individual oximeters (e.g. photodetector quantum efficiency, probe type) are not taken into account in the theoretical equation. To account for these effects, the relationship between R and SaO2 has to be determined empirically using calibration.

The calibration of pulse oximeters for use on human subjects is regulated by ISO 80601-2-61. This standards document defines minimal requirements for devices, provides classification methods for systems, instructions how to measure accuracy of SpO2 and HR, and lists possible fault conditions. The standards document also provides informative recommendations on how to measure accuracy (Annex CC), standards for calibration (Annex DD), and guidelines for evaluating and documenting SpO2 accuracy in human subjects.

There is no standard tool for calibrating pulse oximeters. According to the ISO pulse oximetry standard document, "there is today no accepted method of verifying the correct calibration of a pulse oximeter probe/pulse oximeter monitor combination other than testing on human beings. This is due to the complexity of the optical intricacies of the interaction of light and human tissue upon which pulse oximetry depends". In-vitro calibration device prototypes using whole blood have been suggested in various research projects, but these systems are, to-date, not commercially available and very expensive to build and operate. They have not been accepted as standard tools for calibration because of their lack of simulating tissue interaction with pulse oximeter light. It is therefore necessary to calibrate pulse oximeters on human subjects.

For calibration and accuracy evaluation of SpO2 measurements, the ISO accepts two methods of testing:

Invasive testing: The SpO2 readings of the test pulse oximeter are compared to values of arterial blood oxygen saturations (SaO2) measured with a "gold standard" blood gas analysis using a CO-oximeter.

Non-invasive testing: The SpO2 readings of the test pulse oximeter are compared to values from a secondary standard pulse oximeter, which can be used as a transfer standard because its calibration is directly traceable to a blood gas analysis.

These tests can be conducted on healthy volunteers who consent to induced hypoxemia or patients from whom arterial blood samples/reference pulse oximeter readings are available.

Objectives:

Our main objective is to calibrate the Camera Oximeter. We will also evaluate the accuracy and reliability of the calibration to measure SpO2, respiratory rate, and heart rate measurements. Measurements produced by two secondary standard pulse oximeters and a capnometer (CO2 breath analyzer) will be used as reference monitors.

Research Method:

This will be a non-invasive testing on healthy subjects using two secondary standard, clinically approved pulse oximeters. The approved oximeters will provide the reference measurements for both SpO2 and HR. This is the least invasive approach that is recommended by the ISO standards.

Study subjects will be put into a hypoxic state by exposing them to normobaric (sea-level atmospheric pressure) hypoxia (low oxygen) by administrating an air mix containing a reduced O2 concentration. This is achieved in a hypoxia chamber where O2 concentration is gradually reduced to simulate high altitude (around 4500 m).

The goals of the current study are very similar to another study conducted in the hypoxia chamber (REB ID#H12-02365, Calibration and evaluation of an Audio Pulse Oximeter Sensor (AudioOx) at ascent and descent from simulated altitude), the same methodology is applied. This will allow recruiting subjects for both studies and will reduce the total number of subjects necessary for achieving our goal.

Statistical Analysis:

Calibration of SpO2 Data from the initial set of subjects (at least 10) in the study will be used to calibrate the Camera Oximeter oximetry data. Firstly, the red and infra-red (IR) light signals from oximeter sensors are used to calculate ratio R, where

R = ( ACRED / DCRED ) / ( ACIR / DCIR )

ACRED and ACIR are pulsatile components of the red and infra-red light detected by the oximeter photo detector. DCRED and DCIR are constant components of the red and infra-red light detected by the oximeter photo detector.

R values are paired to SpO2 readings from the reference oximeter taken at the same time. Linear regression on a second order polynomial equation is then used to estimate the calibration parameters.

Evaluation of SpO2 Readings from later subjects (at least 10) will be used to evaluate the accuracy and validity of the SpO2 measurements from both novel oximeters. Paired readings from the test oximeters and the reference oximeter are grouped into six ranges (70-75%, 76-80%, 81-85%, 86-90%, 91-95% and 96-100%). Mean bias, precision and accuracy will be calculated as per ISO definitions for each group of SpO2 ranges and the overall range (70-100%) [5].

Local bias, b, is the difference between the expectation of the test results (SpO2) and an accepted reference value (SR). For pulse oximetry, this is, at a given value of the reference SpO2, the difference between the y-value of the regression line at that coordinate and the y-value of the line of identity, in a plot of SpO2 versus SR, or given by:

Local bias, bi = SpO2 fit,i − SRi

where SpO2 fit,i is the value of the curve fitted to the test data at the ith reference SpO2 value, SRi.

Mean bias, B, is mean difference between the test and reference values, preserving sign (average of all local bias values):

Mean bias, B = (∑i=1 to n (bi))/n

Precision is the closeness of agreement between independent test results obtained under stipulated conditions, defined as the standard deviation of the residuals (Sres), given by:

Precision, Sres = √(∑i=1 to n (SpO2i - SpO2fit,i)^2 )/(n-2))

where n is the number of data pairs in the sample within the range of interest; (SpO2,i − SpO2 fit,i) is the difference between the ith SpO2 datum and the value of the fitted curve corresponding to the ith reference SpO2 value, SRi.

Accuracy of the pulse oximeter shall be stated in terms of the root-mean-square (rms) difference between measured values (SpO2,i) and reference values (SRi), as given by:

Accuracy = √((∑i=1 to n(SpO2i- SRi)^2 )/n)

To express ACCURACY relative to the "gold-standard" blood gas analysis, the error of the secondary standard pulse oximeter (errorref) has to be included, thereby:

Accuracycorrected = √Arms^2 + error(ref)^2

Evaluation of HR Readings from later subjects (at least 10) will be used to evaluate the accuracy and validity of HR measurements from the novel oximeter. Paired HR readings from the test oximeter and the reference oximeters are grouped into four ranges (40-65, 66-90, 91-115, and 116-140 beats per minute (bpm)). Mean bias, precision and accuracy are calculated as per ISO definitions for each group of HR ranges and the overall range (40-140 bpm), as detailed previously for SpO2.

Evaluation of Respiratory Rate Readings from later subjects (at least 10) will be used to evaluate the accuracy and validity of respiratory rate measurements extracted from the plethysmogram that was recorded with the Camera Oximeter. Paired respiratory rate readings from the test oximeter and the reference capnometry are grouped into three ranges (0-8, 9-16, and >17 breaths per minute). Mean bias, precision and accuracy are calculated for each group of respiratory range and the overall range. ;


Study Design


Related Conditions & MeSH terms


NCT number NCT01732016
Study type Interventional
Source University of British Columbia
Contact
Status Completed
Phase N/A
Start date January 2013
Completion date August 2013

See also
  Status Clinical Trial Phase
Completed NCT03832205 - Validation of Respiratory Rate and Heart Rate Measurements by Capaciflectors Placed in Four Locations on the Chest
Completed NCT03723954 - Heart Rate Variations of Golf Fans During the 2018 the Ryder Cup
Withdrawn NCT01026961 - Safety of Phenylephrine Hydrochloride, Acetaminophen, Dimethindene Maleate Compared to Phenylephrine Hydrochloride Alone in Healthy Volunteers Phase 1/Phase 2
Completed NCT00760253 - Compare the Side Effects and Difference Awake Level of Three TCI Propofol Formula in TVOR Patients N/A
Completed NCT00380341 - Effects of Instrument-Applied Spinal Manipulative Therapy on Postureal Control and Autonomic Balance N/A
Completed NCT04179279 - Comparative Analysis of a Non-contact Respiratory and Heart Rate Monitor Vs. a Conventional Clinically Validated Reference Monitor
Completed NCT02469844 - Heart Rate Variability (HRV) Analysis in Patients With Nocturnal Epileptic Seizures
Not yet recruiting NCT00503282 - Urinary Catheter Exchange on Admission of Septic Elderly With Permanent Catheter Before Initiation of Antibiotics N/A
Completed NCT05160207 - Intubation-induced Decrease in Heart Rate as an Indicator for Intraoperative Bradycardia
Withdrawn NCT01701960 - Effect of Local Anesthetic on Hemodynamic Measures During Nasal Surgery N/A
Completed NCT02997800 - The Effect of Intraoperative Labetalol on Time to Discharge Phase 2
Completed NCT03359928 - Acute Physiological and Perceptual Responses to Novel Forms of High-intensity Interval Exercise N/A
Completed NCT05600725 - Novel Cardiac Pacing to Initiate Cardiac Remodeling in Heart Failure N/A
Completed NCT05366049 - The Effect of Methods Used in Oral Antipyretic Administration on the Vital Findings of the Child With Fever N/A
Recruiting NCT05401435 - Reliability of Measuring Blood Pressure With a Smartwatch N/A
Completed NCT01638169 - The Contribution of Heels for Walking Quality Among Children With Muscular Dystrophy N/A
Completed NCT00143988 - Heart Rate and Blood Pressure Response During Exercise and Sexual Activity in Normotensive and Hypertensive Volunteers N/A
Completed NCT03131375 - Dexmedetomidine Reduces Emergence Delirium in Children Undergoing Tonsillectomy With Propofol Anesthesia Phase 2/Phase 3
Completed NCT01612130 - Valerian for Conscious Sedation in Patients Submitted to Impacted Lower Third Molars Surgery N/A
Completed NCT06095635 - Acute Influence of Resveratrol Supplementation and Beet Extract in Recovery After Exercise in Coronary Paths N/A