Clinical Trial Summary
The purpose of this study is to compare the accuracy and comparability and secondarily to
assess the values achieved by measurement of the forearm BMD by DXA and SOS by BeamMed,
relative to standard DXA evaluations at the FN and LS.
Description:
Dual-energy X-ray absorptiometry (DXA) is the current gold standard for the clinical
diagnosis of osteoporosis based on measurement of bone mineral density (BMD) [Baim et al,
2006]. As DXA technology continues to evolve, new instruments and technologies are
introduced [Shepherd et al, 2005], making it necessary to document how these advances
compare to prior densitometers.
The purpose of this study is to assess bone mineral density (BMD) comparability and
precision using a standard Hologic DXA unit and the quantitative ultrasound-driven
speed-of-sound (SOS) BeamMed technology. The physical and mechanical properties of bone that
are measured by SOS are different than bone density measured by DXA and reflect bone
strength and elasticity [Goossens et al, 2008]. The cohorts will be patients with Gaucher
disease [Zimran & Elstein, 2010] who currently are requested to undergo BMD evaluation only
bi-annually [Mistry et al, 2011], partly because of the radiation of DXA scans (admittedly
low relative to standard X-rays and CT). Patients with Gaucher disease are at added risk for
osteonecrosis and pathological fractures but to date no biomarker or technology has been
able to predict which patients are imminently at risk. Because the option of
disease-specific but very costly enzyme replacement therapy (ERT) [Barton et al, 1991] is
predicating on the estimated risk of severe disease, predicting skeletal damage (especially
osteonecrosis at the femoral neck, FN) because of osteopenia or osteoporosis has clinically
relevant ramifications [Mistry et al, 2011]. Additionally, it has been shown that there is a
correlation between low BMD at the lumbar spine (LS) and anemia in patients with Gaucher
disease [Khan et al, 2012], so that with correction of anemia with ERT it is hoped that
there will be amelioration of BMD values, and this too would be worth performing at closer
intervals.
The least inconvenient and shortest evaluation for BMD is at the forearm. For approximately
a year, all patients at the Gaucher disease have had DXA forearm evaluations when undergoing
routine testing. It is the intention of this study to prospectively compare forearm SOS
results [Olszynski et al, 2013] with those from the standard DXA equipment with the
possibility of using SOS as a quantitative measure of bone integrity on a more regular
basis.
The strategy will be to measure forearm BMD by DXA and SOS by BeamMed for comparison of
accuracy and comparability and secondarily to assess the values achieved at the forearm
relative to standard DXA evaluations at the FN and LS.
Patients who arrive at the Gaucher Clinic for routine monitoring that includes DXA
evaluation of BMD at FN, LS, and forearm will be requested to consent to undergo SunBeam
evaluation at the forearm as well. The same DXA hardware and software (Hologic, Bedford, MA)
will be used for all examinations as have been employed by the Gaucher Clinic since August
2011.
The SOS technology as developed by the BeamMed Company (originally, the Sunlight Ultrasound
Technologies, Israel) is based on the physical property that ultrasound waves travel faster
through bone than soft tissue. The bone sonometer measures the time that elapses between the
axially transmitted sound generated as an inaudible high-frequency pulsed acoustic signal at
a center frequency of 1.25 MHz and the first reception of a signal after it has traveled
through the selected bone. This is recorded as the SOS in m/sec. The ultrasonic waves are
transmitted and received by transducers embedded in the ultrasound probe. As the ultrasound
wave travels, the speed, dispersion, and attenuation of the signal is influenced by the
density, elasticity, and cohesiveness of the medium, so that the greater the density and
elasticity of the medium, the faster the speed of propagation [Ashman et al, 1984].
The device consists of a desktop main unit and 3 probes of different sizes. The small probe
(1.4cm x 2.7cm x 11cm), 900-1000 KHz will be used for this study. Calibration is performed
each day using a standardized calibrating plastic provided by the company. The SOS is
determined as the mean of 3 scans that are found to be statistically similar (coefficient of
variation below 1.2%), as calculated by the software provided with the device. Based on
these 3 readings, the 95th percentile of the measured SOS is determined and used for
statistical analysis.