Clinical Trial Details — Status: Not yet recruiting

Administrative data

NCT number NCT06393582
Other study ID # REC-23-05-12-02
Secondary ID
Status Not yet recruiting
Phase N/A
First received
Last updated
Start date May 1, 2024
Est. completion date September 1, 2024

Study information

Verified date April 2024
Source University of Sharjah
Contact Ahmed Aziz, PhD
Phone +971 65057312
Email a.aziz@sharjah.ac.ae
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The aims of this clinical study are: 1) to compare the accuracy of conventional versus digital impression techniques for implant impressions; 2) to assess the passive fit of the prosthesis constructed from the most accurate impression. Thirty edentulous patients who received 4 - 6 implant and indicated for implant-supported fixed prosthesis will be invited. Full-arch conventional and digital impressions will be made for each patient. Accuracy evaluation will be made and the prosthesis will be constructed from the most accurate impression. Passive fit of the zirconia implant-supported prostheses will be verified at the try-in stage clinically and radiographically.


Description:

The aims of this clinical study are: 1) to compare the accuracy of conventional versus digital impression techniques for implant impressions; 2) to assess the passive fit of the prosthesis constructed from the most accurate impression. Study design The ethical approval will be obtained from the research ethics board. Thirty edentulous patients who received 4 - 6 implant and indicated for implant-supported fixed prosthesis will be invited. All participants will require to read and sign the informed consent form before initiation of the treatment. The impression procedures will be performed by one experienced prosthodontist. The prostheses will be evaluated by an independent prosthodontist. Sample size calculation The sample size is calculated based on the tests for paired means. With α= 0.05, β= 0.2, mean of distance deviation, IOS= 17 μm, and standard deviation= 34 μm, a sample size of 25 is necessary, which would have the power of 90% to detect the difference. Therefore, a total of 25 patients will be invited to participate in this clinical study. This sample size is more than previous studies evaluating IOS and SPG workflows (15 - 17 patients). However, it has been reported that the dropout rate in clinical studies is 20%. For this reason, a total of 30 patients will be invited to compensate for the anticipated loss of patients. Treatment procedure Conventional impression Abutment-level impression copings will hand tightened onto the abutments and splinted with autopolymerizing acrylic resin (Pattern Resin; GC), and allowed to polymerize. A custom open-tray splinted impressions will be made with a polyether material (Impregum Penta Soft; 3M ESPE). After 5 minutes, the impression copings will be unscrewed from the implants, and the impression will be removed. Type IV dental stone (dentostone 220; dentona AG) will be poured into the impression to generate a definitive cast. The models will be digitalized with a high-resolution laboratory scanner (E4; 3Shape A/S) with an accuracy of 4 mm to obtain the 3D data of the models in standard tessellation language (STL) format. Digital impressions After tightening the scan bodies (CARES® RC Mono Scan body, Straumann, Basel, Switzerland), the intraoral scanner will be used according to the manufacturer's recommendations. Three scanners will be used since they have different ways of obtaining 3D images and working principles. The technology of Trios scanner (TRIOS 4, 3Shape A/S) depends on confocal microscopy, ultrafast optical scanning and auto-correction; Cerec Primescan (Dentsply Sirona) depends on optical triangulation and confocal microscopy; Medit i700 (Medit, Seoul, Korea) depends on 3D in-motion video technology, 3D full-color streaming capture and advance algorithm to stitch images. The scanning path will begin with the occlusal surfaces, then turned to the palatal or lingual surfaces, and lastly to the buccal surfaces of the scan bodies. Then, removing the scan body with a trim tool from the occlusal view and exporting the STL file. Accuracy measurement All STL files generated from the intraoral scan (STL digital) and the STL files generated from cast digitization (STL conventional) will be imported into 3D inspection software (Geomagic, 3D Systems Inc., Rock Hill, SC) for merging and evaluation of 3D deviations. Following software best-fit alignment, the 3D comparison will be used to evaluate the 3D deviations between the digital files (STL files from IOS and digitized conventional working casts) limited to the area of scan bodies. Assessment of the root man square (RMS) deviations will be the primary outcome. The effect of the type of edentulous jaw (maxilla vs mandible) on the 3D deviations will be the secondary outcome, assessing whether the jaw has an effect on the amount of 3D deviations. Prostheses fabrication The final restorative phase included definitive with one-piece, screw-retained zirconia implant-supported prostheses. The prosthesis will be constructed from the most accurate workflow after confirmation of the passive fit at the try-in appointment. In addition, the fit of the zirconia implant-supported prostheses will be confirmed prior to delivery, clinically and radiographically. Statistical analysis Statistical analysis will be performed by SPSS version 29 (IBM Corp, Armonk, NY). Data will be presented as mean, median, and standard deviation. The normality of data will be tested by the Shapiro-Wilk test. The generalized linear mixed model will be used to compare the influence of scan bodies and impression copings on the accuracy of digital and conventional impressions. The fixed factors will be the 'IOS', 'type of scan body', 'type of impression coping' and their interaction. The precision will be calculated using the gamma distribution with the log-link function. The p values will be adjusted by the Bonferroni method for pairwise comparison (>.05). All clinical outcomes will be assessed by one examiner. Since some patients will received more than one implant, linear mixed effect models (LMM) will be fitted using restricted maximum likelihood estimations to assess the effect of the follow-up time on repeated measurements of marginal bone loss, plaque index, calculus index, gingival index, bleeding index, and probing depth while accounting for within-person variability. LMM calculates means with standard errors as it estimates variability across multiple samples of a population (i.e., the different patients, implants, and repeated measures). The multivariable models included the fixed effects of the baseline value and follow-up in months. The interaction between the baseline and follow-up value (baseline value*follow-up) will be tested for model improvement. The included random effect will be the patient (i.e., random intercepts). The random effect of the follow-up (i.e., random slopes) will be also tested. Statistical comparisons will be performed using the type III analysis of variance test with Satterthwaite's method to estimate the degrees of freedom. A significance level of p< 0.05 will be chosen for all the analyses.


Recruitment information / eligibility

Status Not yet recruiting
Enrollment 30
Est. completion date September 1, 2024
Est. primary completion date August 2, 2024
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria: Edentulous patients who received implant surgery and indicated for implant-supported fixed prosthesis. Patients rehabilitated with fixed full-arch interim prostheses with 4 - 6 implants Exclusion Criteria: Patients with severe systemic diseases, dementia, or muscle tremors will be excluded from the study.

Study Design


Intervention

Procedure:
Accuracy of full-arch implant impressions
Conventional impressions for full-arch implant impressions will be digitized and used as a reference for comparison.
Accuracy of full-arch implant impressions using intraoral scanners
Digital scanners impressions for full-arch implant impressions will be compared to the reference (conventional) impressions

Locations

Country Name City State
n/a

Sponsors (1)

Lead Sponsor Collaborator
University of Sharjah

References & Publications (8)

Akalin ZF, Ozkan YK, Ekerim A. Effects of implant angulation, impression material, and variation in arch curvature width on implant transfer model accuracy. Int J Oral Maxillofac Implants. 2013 Jan-Feb;28(1):149-57. doi: 10.11607/jomi.2070. — View Citation

Al Quran FA, Rashdan BA, Zomar AA, Weiner S. Passive fit and accuracy of three dental implant impression techniques. Quintessence Int. 2012 Feb;43(2):119-25. — View Citation

Alikhasi M, Siadat H, Nasirpour A, Hasanzade M. Three-Dimensional Accuracy of Digital Impression versus Conventional Method: Effect of Implant Angulation and Connection Type. Int J Dent. 2018 Jun 4;2018:3761750. doi: 10.1155/2018/3761750. eCollection 2018. — View Citation

Lee H, So JS, Hochstedler JL, Ercoli C. The accuracy of implant impressions: a systematic review. J Prosthet Dent. 2008 Oct;100(4):285-91. doi: 10.1016/S0022-3913(08)60208-5. — View Citation

Lee SJ, Gallucci GO. Digital vs. conventional implant impressions: efficiency outcomes. Clin Oral Implants Res. 2013 Jan;24(1):111-5. doi: 10.1111/j.1600-0501.2012.02430.x. Epub 2012 Feb 22. — View Citation

Papaspyridakos P, Hirayama H, Chen CJ, Ho CH, Chronopoulos V, Weber HP. Full-arch implant fixed prostheses: a comparative study on the effect of connection type and impression technique on accuracy of fit. Clin Oral Implants Res. 2016 Sep;27(9):1099-105. doi: 10.1111/clr.12695. Epub 2015 Sep 16. — View Citation

Sorrentino R, Gherlone EF, Calesini G, Zarone F. Effect of implant angulation, connection length, and impression material on the dimensional accuracy of implant impressions: an in vitro comparative study. Clin Implant Dent Relat Res. 2010 May;12 Suppl 1:e63-76. doi: 10.1111/j.1708-8208.2009.00167.x. Epub 2009 May 8. — View Citation

Vigolo P, Millstein PL. Evaluation of master cast techniques for multiple abutment implant prostheses. Int J Oral Maxillofac Implants. 1993;8(4):439-46. — View Citation

Outcome

Type Measure Description Time frame Safety issue
Primary Accuracy of conventional versus digital impression techniques for implant impressions Following the best-fit algorithm, the accuracy will be measured in terms of 3D comparison to evaluate the 3D deviations between the digital files (STL files from IOS and digitized conventional working casts) limited to the area of scan bodies. 6 months
Primary The passive fit of the prosthesis constructed from the most accurate impression Clinical assessment of the prosthesis at the try-in stage using one screw test. 6 months
Primary The passive fit of the prosthesis Radiographic assessment of the prosthesis by periapical radiograph 6 months
See also
  Status Clinical Trial Phase
Recruiting NCT05466006 - Peri-implant Soft Tissue Volume Changes at Small Buccal Dehiscences N/A
Withdrawn NCT00313989 - Evaluation of Osseo-integration After Placement of Dental Implants With Patients Receiving Radiotherapy for Buccopharyngeal Oncology N/A
Completed NCT05081050 - Short Implants Supporting Single Crowns in the Posterior Region
Completed NCT04707677 - Comparative Aesthetic Analysis Between Titanium and Ceramic Implants N/A
Recruiting NCT05989061 - PEEK Versus Titanium Customized Healing Abutments N/A
Recruiting NCT05722041 - Soft Tissue Volume Changes in Posterior Socket Sealing Abutment.
Completed NCT05512845 - The Double Factor Technique: a Computer-guided Implant Surgery Technique for Fully Edentulous Patients
Recruiting NCT03837158 - Titanium Zirconium (TiZr) Mini Implants N/A
Completed NCT01944267 - Procedures for Improving the Mucosa Around Implants N/A
Completed NCT00971516 - Healing Profile of Titanium Dental Implants Placed in Patients With Diabetes N/A
Completed NCT04305080 - Effect of Anodic Oxidation of Implant Abutment Collar on the Peri-implant Soft Tissue Health N/A
Completed NCT06191510 - Xenogenic Bone Intentionally Left Exposed
Completed NCT04459494 - In the Treatment of Implants With Different Methods; Early Bone Loss and Patient Satisfaction N/A
Recruiting NCT05644470 - Association of Implant Position and Crown Contour With Tissue Health
Completed NCT05101954 - Platelet Rich Fibrin and Concentrated Growth Factors on Osseointegration of Immediate Implants N/A
Recruiting NCT05748379 - Atraumatic Zirconia Abutment Versus Customized Composite Healing Abutment in Maxilla or Mandible. N/A
Recruiting NCT05627037 - Marginal Bone Level Around Titanium Implants Produced by Additive Manufacture N/A
Enrolling by invitation NCT06187805 - Retrospective Case Series - OSSIX Volumax Collagen Xenograft Scaffolding
Recruiting NCT06100120 - A Triple-blinded Non-inferiority Multicentre Randomized Controlled Trial Comparing Hybrid and Fully Etched Implants P1 N/A
Recruiting NCT06155799 - Accuracy of Dynamic Navigation System on Zygomatic Implant Placement N/A