Chronic Kidney Diseases Clinical Trial
Official title:
Regional Electric Bio Impedance-assisted Monitoring of Cardiac, Vascular Resistance and Fluid Balance, in Chronic Hemodialysis Patients
Hemodynamic trends will be assessed using the device, in 100 dialysis sessions in 30
patients, who are prone to develop hypotensive episode during dialysis. Sitting blood
pressures will be measured immediately prior to each hemodynamic measurement: before
initiation of dialysis, every each hour and in the beginning of hypotension episode, just
before the end and 10 min after the end of the treatment.
Gender, age, height, weight, electrode location and blood pressure data will be entered into
the device. The device will measure and calculate hemodynamic parameters on each heart beat
during 60 s and provides the averaged parameters.
Technology for hemodynamic measurements:
The device (NICaS, NI Medical) is a noninvasive regional bioimpedance cardiac measurement and
analysis system (FDA 510k clearance no. K080941, 12 June 2009). The US Food and Drug
Administration indication for use of the device states 'NICaS is intended to monitor and
display hemodynamic parameters in males and females with known or suspected cardiac disorders
needing cardiac assessment'.
SV will be measured by applying an alternating electrical current of 1.4mA at 30 kHz
frequency through the patient's body via two pairs of tetrapolar sensors, one pair placed on
the wrist of the nonaccess arm above the radial pulse and the other pair on the contralateral
ankle above the posterior tibial pulse (Figure 1).
Figure 1 : Sensor location
SV is calculated by Frinerman's formula:
SV¼(dR/R) - q - (L2/Ri) - (ab)/b - KW - HF [2-4], where dR is the impedance change in the
arterial system as a result of intraarterial expansion during systole, R is basal resistance,
q is blood electrical resistance, L is the patient's height, Ri is basal resistance corrected
for gender and age, KWis the correction of weight according to ideal values, HF is a
hydration factor that takes into account the ratio between R and body mass index (BMI), which
is correlated to body water volume, ab is the electrocardiogram (ECG) R-R wave interval and b
is the diastolic time interval. SV is automatically calculated every 20 s and is the average
of three measurements obtained consecutively during 60 s of monitoring. The SV index is
calculated as SV/body surface area using the Du Bois formula [11]. Heart rate is calculated
from a one channel ECG and cardiac (output) index¼SV index - heart rate/1000. Using an
oscillometric method, sitting systolic and diastolic blood pressure measurements were made
automatically by the dialysis machine. Mean arterial pressure [2 - (diastolicsystolic)/3],
cardiac power index [CPI; mean arterial pressure (MAP) -cardiac index - 0.0022 w/m2; normal
range 0.45-0.85w/m2] [12, 13] and total peripheral resistance (MAP/ cardiac index - 80 dyn -
s/cm5 - m2; normal range 1600-3000 dyn - s/cm5- m2) [13] will be calculated.
As the device measures pulsatile flow and is blinded to constant flow, fluid removal during
dialysis has no impact on measurement accuracy. This was recently validated by correlating SV
to ECG measurements during hemodialysis treatments. Good correlation was maintained during
treatment. Further, NICaS performance immunity to fluid reduction was demonstrated by the
maintenance of correlation to ECG results throughout dialysis treatments [9]. The results are
drawn on hemodynamic graphs showing the MAP (y-axis) as a function of cardiac index (x-axis);
curves of total peripheral resistance index (TPRI) and CPI are displayed. Ranges for the
normal population are depicted by a dotted octagon.
Hemodynamic trends will be assessed using the device, in 100 dialysis sessions in 30
patients, who are prone to develop hypotensive episode during dialysis. Sitting blood
pressures will be measured immediately prior to each hemodynamic measurement: before
initiation of dialysis, every each hour and in the beginning of hypotension episode, just
before the end and 10 min after the end of the treatment.
Gender, age, height, weight, electrode location and blood pressure data will be entered into
the device. The device will measure and calculate hemodynamic parameters on each heart beat
during 60 s and provides the averaged parameters.
Technology for hemodynamic measurements:
The device (NICaS, NI Medical) is a noninvasive regional bioimpedance cardiac measurement and
analysis system (FDA 510k clearance no. K080941, 12 June 2009). The US Food and Drug
Administration indication for use of the device states 'NICaS is intended to monitor and
display hemodynamic parameters in males and females with known or suspected cardiac disorders
needing cardiac assessment'.
SV will be measured by applying an alternating electrical current of 1.4mA at 30 kHz
frequency through the patient's body via two pairs of tetrapolar sensors, one pair placed on
the wrist of the nonaccess arm above the radial pulse and the other pair on the contralateral
ankle above the posterior tibial pulse (Figure 1).
Figure 1 : Sensor location
SV is calculated by Frinerman's formula:
SV¼(dR/R) - q - (L2/Ri) - (ab)/b - KW - HF [2-4], where dR is the impedance change in the
arterial system as a result of intraarterial expansion during systole, R is basal resistance,
q is blood electrical resistance, L is the patient's height, Ri is basal resistance corrected
for gender and age, KWis the correction of weight according to ideal values, HF is a
hydration factor that takes into account the ratio between R and body mass index (BMI), which
is correlated to body water volume, ab is the electrocardiogram (ECG) R-R wave interval and b
is the diastolic time interval. SV is automatically calculated every 20 s and is the average
of three measurements obtained consecutively during 60 s of monitoring. The SV index is
calculated as SV/body surface area using the Du Bois formula [11]. Heart rate is calculated
from a one channel ECG and cardiac (output) index¼SV index - heart rate/1000. Using an
oscillometric method, sitting systolic and diastolic blood pressure measurements were made
automatically by the dialysis machine. Mean arterial pressure [2 - (diastolicsystolic)/3],
cardiac power index [CPI; mean arterial pressure (MAP) -cardiac index - 0.0022 w/m2; normal
range 0.45-0.85w/m2] [12, 13] and total peripheral resistance (MAP/ cardiac index - 80 dyn -
s/cm5 - m2; normal range 1600-3000 dyn - s/cm5- m2) [13] will be calculated.
As the device measures pulsatile flow and is blinded to constant flow, fluid removal during
dialysis has no impact on measurement accuracy. This was recently validated by correlating SV
to ECG measurements during hemodialysis treatments. Good correlation was maintained during
treatment. Further, NICaS performance immunity to fluid reduction was demonstrated by the
maintenance of correlation to ECG results throughout dialysis treatments [9]. The results are
drawn on hemodynamic graphs showing the MAP (y-axis) as a function of cardiac index (x-axis);
curves of total peripheral resistance index (TPRI) and CPI are displayed. Ranges for the
normal population are depicted by a dotted octagon.
;
Status | Clinical Trial | Phase | |
---|---|---|---|
Recruiting |
NCT06386172 -
Electronic Decision-support System to Improve Detection and Care of Patients With Chronic Kidney Disease in Stockholm
|
N/A | |
Recruiting |
NCT04910867 -
APOL1 Genetic Testing Program for Living Donors
|
N/A | |
Completed |
NCT03434145 -
Changes of Ocular Structures After Hemodialysis in Patients With Chronic Kidney Diseases
|
N/A | |
Recruiting |
NCT04984226 -
Sodium Bicarbonate and Mitochondrial Energetics in Persons With CKD
|
Phase 2 | |
Active, not recruiting |
NCT05887817 -
Effects of Finerenone on Vascular Stiffness and Cardiorenal Biomarkers in T2D and CKD (FIVE-STAR)
|
Phase 4 | |
Recruiting |
NCT05318196 -
Molecular Prediction of Development, Progression or Complications of Kidney, Immune or Transplantation-related Diseases
|
||
Terminated |
NCT05022329 -
COVID-19 Vaccine Boosters in Patients With CKD
|
Phase 2/Phase 3 | |
Not yet recruiting |
NCT04925661 -
HEC53856 Phase Ib Study in Patients With Non-dialysis Renal Anemia
|
Phase 1 | |
Recruiting |
NCT04961164 -
Resistant Starch Prebiotic Effects in Chronic Kidney Disease
|
N/A | |
Completed |
NCT05015647 -
Low Protein Diet in CKD Patients at Risk of Malnutrition
|
N/A | |
Completed |
NCT03426787 -
Helping Empower Liver and Kidney Patients
|
N/A | |
Recruiting |
NCT06094231 -
Treating Patients With Renal Impairment and Altered Glucose MetAbolism With TherapeutIc Carbohydrate Restriction and Sglt2-Inhibiton - a Pilot Study
|
N/A | |
Completed |
NCT04363554 -
The Kidneys Ability to Concentrate and Dilute Urine in Patients With Autosomal Dominant Polycystic Kidney Disease
|
N/A | |
Recruiting |
NCT04831021 -
Pre- or Per-dialytic Physical Exercise : a Cardioprotective Role?
|
N/A | |
Terminated |
NCT04877847 -
Multi-Center Trial Utilizing Low Frequency Ultrasound in the Prevention of Post-Contrast Acute Kidney Injury
|
N/A | |
Recruiting |
NCT04422652 -
Combination of Novel Therapies for CKD Comorbid Depression
|
Phase 2 | |
Completed |
NCT05055362 -
Effect a Honey, Spice-blended Baked Good Has on Salivary Inflammation Markers in Adults: a Pilot Study
|
N/A | |
Not yet recruiting |
NCT06330480 -
Check@Home: General Population Screening for Early Detection of Atrial Fibrillation and Chronic Kidney Disease
|
N/A | |
Recruiting |
NCT03176862 -
Left Ventricular Fibrosis in Chronic Kidney Disease
|
N/A | |
Terminated |
NCT02539680 -
Intestinal Phosphate Transporter Expression in CKD Patients
|
N/A |