Clinical Trials Logo

Clinical Trial Summary

Hemodynamic trends will be assessed using the device, in 100 dialysis sessions in 30 patients, who are prone to develop hypotensive episode during dialysis. Sitting blood pressures will be measured immediately prior to each hemodynamic measurement: before initiation of dialysis, every each hour and in the beginning of hypotension episode, just before the end and 10 min after the end of the treatment.

Gender, age, height, weight, electrode location and blood pressure data will be entered into the device. The device will measure and calculate hemodynamic parameters on each heart beat during 60 s and provides the averaged parameters.

Technology for hemodynamic measurements:

The device (NICaS, NI Medical) is a noninvasive regional bioimpedance cardiac measurement and analysis system (FDA 510k clearance no. K080941, 12 June 2009). The US Food and Drug Administration indication for use of the device states 'NICaS is intended to monitor and display hemodynamic parameters in males and females with known or suspected cardiac disorders needing cardiac assessment'.

SV will be measured by applying an alternating electrical current of 1.4mA at 30 kHz frequency through the patient's body via two pairs of tetrapolar sensors, one pair placed on the wrist of the nonaccess arm above the radial pulse and the other pair on the contralateral ankle above the posterior tibial pulse (Figure 1).

Figure 1 : Sensor location

SV is calculated by Frinerman's formula:

SV¼(dR/R) - q - (L2/Ri) - (ab)/b - KW - HF [2-4], where dR is the impedance change in the arterial system as a result of intraarterial expansion during systole, R is basal resistance, q is blood electrical resistance, L is the patient's height, Ri is basal resistance corrected for gender and age, KWis the correction of weight according to ideal values, HF is a hydration factor that takes into account the ratio between R and body mass index (BMI), which is correlated to body water volume, ab is the electrocardiogram (ECG) R-R wave interval and b is the diastolic time interval. SV is automatically calculated every 20 s and is the average of three measurements obtained consecutively during 60 s of monitoring. The SV index is calculated as SV/body surface area using the Du Bois formula [11]. Heart rate is calculated from a one channel ECG and cardiac (output) index¼SV index - heart rate/1000. Using an oscillometric method, sitting systolic and diastolic blood pressure measurements were made automatically by the dialysis machine. Mean arterial pressure [2 - (diastolicsystolic)/3], cardiac power index [CPI; mean arterial pressure (MAP) -cardiac index - 0.0022 w/m2; normal range 0.45-0.85w/m2] [12, 13] and total peripheral resistance (MAP/ cardiac index - 80 dyn - s/cm5 - m2; normal range 1600-3000 dyn - s/cm5- m2) [13] will be calculated.

As the device measures pulsatile flow and is blinded to constant flow, fluid removal during dialysis has no impact on measurement accuracy. This was recently validated by correlating SV to ECG measurements during hemodialysis treatments. Good correlation was maintained during treatment. Further, NICaS performance immunity to fluid reduction was demonstrated by the maintenance of correlation to ECG results throughout dialysis treatments [9]. The results are drawn on hemodynamic graphs showing the MAP (y-axis) as a function of cardiac index (x-axis); curves of total peripheral resistance index (TPRI) and CPI are displayed. Ranges for the normal population are depicted by a dotted octagon.


Clinical Trial Description

Hemodynamic trends will be assessed using the device, in 100 dialysis sessions in 30 patients, who are prone to develop hypotensive episode during dialysis. Sitting blood pressures will be measured immediately prior to each hemodynamic measurement: before initiation of dialysis, every each hour and in the beginning of hypotension episode, just before the end and 10 min after the end of the treatment.

Gender, age, height, weight, electrode location and blood pressure data will be entered into the device. The device will measure and calculate hemodynamic parameters on each heart beat during 60 s and provides the averaged parameters.

Technology for hemodynamic measurements:

The device (NICaS, NI Medical) is a noninvasive regional bioimpedance cardiac measurement and analysis system (FDA 510k clearance no. K080941, 12 June 2009). The US Food and Drug Administration indication for use of the device states 'NICaS is intended to monitor and display hemodynamic parameters in males and females with known or suspected cardiac disorders needing cardiac assessment'.

SV will be measured by applying an alternating electrical current of 1.4mA at 30 kHz frequency through the patient's body via two pairs of tetrapolar sensors, one pair placed on the wrist of the nonaccess arm above the radial pulse and the other pair on the contralateral ankle above the posterior tibial pulse (Figure 1).

Figure 1 : Sensor location

SV is calculated by Frinerman's formula:

SV¼(dR/R) - q - (L2/Ri) - (ab)/b - KW - HF [2-4], where dR is the impedance change in the arterial system as a result of intraarterial expansion during systole, R is basal resistance, q is blood electrical resistance, L is the patient's height, Ri is basal resistance corrected for gender and age, KWis the correction of weight according to ideal values, HF is a hydration factor that takes into account the ratio between R and body mass index (BMI), which is correlated to body water volume, ab is the electrocardiogram (ECG) R-R wave interval and b is the diastolic time interval. SV is automatically calculated every 20 s and is the average of three measurements obtained consecutively during 60 s of monitoring. The SV index is calculated as SV/body surface area using the Du Bois formula [11]. Heart rate is calculated from a one channel ECG and cardiac (output) index¼SV index - heart rate/1000. Using an oscillometric method, sitting systolic and diastolic blood pressure measurements were made automatically by the dialysis machine. Mean arterial pressure [2 - (diastolicsystolic)/3], cardiac power index [CPI; mean arterial pressure (MAP) -cardiac index - 0.0022 w/m2; normal range 0.45-0.85w/m2] [12, 13] and total peripheral resistance (MAP/ cardiac index - 80 dyn - s/cm5 - m2; normal range 1600-3000 dyn - s/cm5- m2) [13] will be calculated.

As the device measures pulsatile flow and is blinded to constant flow, fluid removal during dialysis has no impact on measurement accuracy. This was recently validated by correlating SV to ECG measurements during hemodialysis treatments. Good correlation was maintained during treatment. Further, NICaS performance immunity to fluid reduction was demonstrated by the maintenance of correlation to ECG results throughout dialysis treatments [9]. The results are drawn on hemodynamic graphs showing the MAP (y-axis) as a function of cardiac index (x-axis); curves of total peripheral resistance index (TPRI) and CPI are displayed. Ranges for the normal population are depicted by a dotted octagon. ;


Study Design


Related Conditions & MeSH terms


NCT number NCT04127877
Study type Interventional
Source Assuta Ashdod Hospital
Contact Ze'ev Katzir, MD
Phone 972052 6300563
Email zeevk@assuta.co.il
Status Recruiting
Phase N/A
Start date November 18, 2018
Completion date December 31, 2020

See also
  Status Clinical Trial Phase
Recruiting NCT06386172 - Electronic Decision-support System to Improve Detection and Care of Patients With Chronic Kidney Disease in Stockholm N/A
Recruiting NCT04910867 - APOL1 Genetic Testing Program for Living Donors N/A
Completed NCT03434145 - Changes of Ocular Structures After Hemodialysis in Patients With Chronic Kidney Diseases N/A
Recruiting NCT04984226 - Sodium Bicarbonate and Mitochondrial Energetics in Persons With CKD Phase 2
Active, not recruiting NCT05887817 - Effects of Finerenone on Vascular Stiffness and Cardiorenal Biomarkers in T2D and CKD (FIVE-STAR) Phase 4
Recruiting NCT05318196 - Molecular Prediction of Development, Progression or Complications of Kidney, Immune or Transplantation-related Diseases
Terminated NCT05022329 - COVID-19 Vaccine Boosters in Patients With CKD Phase 2/Phase 3
Not yet recruiting NCT04925661 - HEC53856 Phase Ib Study in Patients With Non-dialysis Renal Anemia Phase 1
Recruiting NCT04961164 - Resistant Starch Prebiotic Effects in Chronic Kidney Disease N/A
Completed NCT05015647 - Low Protein Diet in CKD Patients at Risk of Malnutrition N/A
Completed NCT03426787 - Helping Empower Liver and Kidney Patients N/A
Recruiting NCT06094231 - Treating Patients With Renal Impairment and Altered Glucose MetAbolism With TherapeutIc Carbohydrate Restriction and Sglt2-Inhibiton - a Pilot Study N/A
Completed NCT04363554 - The Kidneys Ability to Concentrate and Dilute Urine in Patients With Autosomal Dominant Polycystic Kidney Disease N/A
Recruiting NCT04831021 - Pre- or Per-dialytic Physical Exercise : a Cardioprotective Role? N/A
Terminated NCT04877847 - Multi-Center Trial Utilizing Low Frequency Ultrasound in the Prevention of Post-Contrast Acute Kidney Injury N/A
Recruiting NCT04422652 - Combination of Novel Therapies for CKD Comorbid Depression Phase 2
Completed NCT05055362 - Effect a Honey, Spice-blended Baked Good Has on Salivary Inflammation Markers in Adults: a Pilot Study N/A
Not yet recruiting NCT06330480 - Check@Home: General Population Screening for Early Detection of Atrial Fibrillation and Chronic Kidney Disease N/A
Recruiting NCT03176862 - Left Ventricular Fibrosis in Chronic Kidney Disease N/A
Terminated NCT02539680 - Intestinal Phosphate Transporter Expression in CKD Patients N/A