Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT01605877
Other study ID # J-11-037
Secondary ID
Status Completed
Phase N/A
First received May 22, 2012
Last updated June 8, 2015
Start date May 2012
Est. completion date December 2013

Study information

Verified date June 2015
Source Alcon Research
Contact n/a
Is FDA regulated No
Health authority Japan: Ministry of Health, Labor and WelfareJapan: Institutional Review Board
Study type Interventional

Clinical Trial Summary

The purpose of this study is to evaluate safety and effectiveness of AcrySof® IQ ReSTOR® +2.5 D Multifocal IOL Model SN6AD2 [SV25T0] in Japanese cataract patients.


Description:

Each subject completed a preoperative examination of both eyes, implantation of IOL at the operative visit for each eye, and up to 8 postoperative visits (each eye examined at Day 1-2, Day 7-14, and Day 30-60, with binocular visits at Day 120-180 and Day 330-420 after the second implantation). The second implantation occurred within 30 days of the first.


Recruitment information / eligibility

Status Completed
Enrollment 70
Est. completion date December 2013
Est. primary completion date December 2013
Accepts healthy volunteers No
Gender Both
Age group 20 Years and older
Eligibility Inclusion Criteria:

- Sign informed consent;

- Diagnosed with bilateral cataracts;

- Planned cataract removal by phacoemulsification;

- Potential postoperative visual acuity of 0.5 decimal or better in both eyes;

- Preoperative astigmatism = 1.5 diopter;

- Clear intraocular media other than cataract in study eyes;

- Calculated lens power within the available range;

- Able to undergo second eye surgery within 30 days of the first eye surgery;

- Other protocol-defined inclusion criteria may apply.

Exclusion Criteria:

- Significant irregular corneal aberration as demonstrated by corneal topography;

- Any inflammation or edema (swelling) of the cornea;

- Diagnosed degenerative visual disorders predicted to cause future acuity losses to a level worse than 0.5 decimal;

- Diabetic retinopathy;

- Previous refractive surgery, retinal detachment, corneal transplant;

- Glaucoma;

- Pregnancy;

- Currently participating in another investigational drug or device study;

- Exclusion criteria during surgery;

- Other protocol-defined exclusion criteria may apply.

Study Design

Intervention Model: Single Group Assignment, Masking: Open Label, Primary Purpose: Treatment


Related Conditions & MeSH terms


Intervention

Device:
AcrySof® IQ ReSTOR® +2.5 D Multifocal IOL Model SN6AD2 [SV25T0]
Multifocal IOL with extended secondary focal point implanted for long-term use over the lifetime of the cataract patient

Locations

Country Name City State
Japan Tokyo Dental College Suidobashi Hospital Chiyoda-ku Tokyo
Japan Hayashi Eye Hospital Fukuoka

Sponsors (1)

Lead Sponsor Collaborator
Alcon Research

Country where clinical trial is conducted

Japan, 

Outcome

Type Measure Description Time frame Safety issue
Primary Uncorrected Decimal VA (5 m) Visual acuity (VA) was tested monocularly (each eye separately) at all visits and binocularly (both eyes together) at Day 30-60, Day 120-180, and Day 330-420 unaided at a distance of 5 meters (m) using a chart. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Baseline (preoperative), Day 1-2, Day 7-14, Day 30-60, Day 120-180, Day 330-420 No
Primary Uncorrected Decimal VA (50 cm) VA was tested monocularly at all visits and binocularly at Day 30-60, Day 120-180, and Day 330-420 unaided at a distance of 50 centimeters (cm) using a chart. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Baseline (preoperative), Day 1-2, Day 7-14, Day 30-60, Day 120-180, Day 330-420 No
Primary Best Corrected Decimal VA (5 m) VA was tested monocularly at all visits and binocularly at Day 30-60, Day 120-180, and Day 330-420 with the participant's best spectacle correction at a distance of 5 m using a chart. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Baseline (preoperative), Day 1-2, Day 7-14, Day 30-60, Day 120-180, Day 330-420 No
Primary Best Corrected Decimal VA (50 cm) VA was tested monocularly at all visits and binocularly at Day 30-60, Day 120-180, and Day 330-420 with the participant's best spectacle correction at a distance of 50 cm using a chart. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Baseline (preoperative), Day 1-2, Day 7-14, Day 30-60, Day 120-180, Day 330-420 No
Primary Distance-Corrected Decimal VA (50 cm) VA was tested monocularly at the preoperative, Day 30-60, Day 120-180, and Day 330-420 visits and binocularly at Day 30-60, Day 120-180, and Day 330-420 using a chart. Distance-corrected VA at 50 cm is the VA at 50 cm measured under a corrected condition in which best-corrected VA at 5 m was obtained. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Baseline (preoperative), Day 30-60, Day 120-180, Day 330-420 No
Secondary Uncorrected Decimal VA (1 m) VA was tested binocularly unaided at a distance of 1 m using a chart. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Day 120-180, Day 330-420 No
Secondary Distance Corrected Decimal VA (1 m) VA was tested binocularly using a chart. Distance-corrected VA at 1 m is the VA at 1 m measured under a corrected condition in which best-corrected VA at 5 m was obtained. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Day 120-180, Day 330-420 No
Secondary Uncorrected Decimal VA (40 cm) VA was tested binocularly unaided at a distance of 40 cm using a chart. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Day 120-180, Day 330-420 No
Secondary Distance Corrected Decimal VA (40 cm) VA was tested binocularly using a chart. Distance-corrected VA at 40 cm is the VA at 40 cm measured under a corrected condition in which best-corrected VA at 5 m was obtained. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Day 120-180, Day 330-420 No
Secondary Uncorrected Decimal VA at Best Distance VA was tested monocularly at all visits and binocularly at Day 30-60, Day 120-180, and Day 330-420 using a chart at the distance of best near vision (cm) as decided by the participant. VA was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Baseline (preoperative), Day 1-2, Day 7-14, Day 30-60, Day 120-180, Day 330-420 No
Secondary Mean Best Distance (cm) for Uncorrected Decimal Near VA VA was tested monocularly at all visits and binocularly at Day 30-60, Day 120-180, and Day 330-420 unaided using a chart. The participant indicated the distance (cm) at which best near vision was attained. Baseline (preoperative), Day 1-2, Day 7-14, Day 30-60, Day 120-180, Day 330-420 No
Secondary Mean Best Distance (cm) for Distance Corrected Decimal Near VA VA was tested monocularly at Day 1-2, Day 7-14, Day 30-60, Day 120-180, and Day 330-420 and binocularly at Day 30-60, Day 120-180, and Day 330-420 with the participant's best spectacle correction using a chart. The participant indicated the distance (cm) at which best near vision was attained. Day 1-2, Day 7-14, Day 30-60, Day 120-180, Day 330-420 No
Secondary Best Corrected Far (3 m) Contrast Sensitivity Far contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly with the participant's best spectacle correction at a distance of 3 m using the Vector Vision CSV 1000 illuminated box (one site) and the Vision Contrast Test System (other site). Contrast sensitivity was assessed at spatial frequencies of 1.5, 3, 6, 12, and 18 cycles per degree (cpd). For CSV 1000, contrast sensitivity was not measured at spatial frequency 1.5 cpd because there was no option for this frequency. Raw scores were transformed to logMar (logarithm of the minimum angle of resolution) units. A higher numeric value represents better contrast sensitivity. Day 120-180 from second eye implantation No
Secondary Best Corrected Near (46 cm) Contrast Sensitivity Near contrast sensitivity (ie, the ability to detect slight changes in luminance before they become indistinguishable) was assessed binocularly with the participant's best spectacle correction at a distance of 46 cm using the Functional Acuity Contrast Test (FACT). Contrast sensitivity was assessed at spatial frequencies of 1.5, 3, 6, 12, and 18 cycles per degree (cpd). Raw scores were transformed to logMar units. A higher numeric value represents better contrast sensitivity. Day 120-180 from second eye implantation No
Secondary Mean Defocus Decimal VA (5 m) Defocus VA (an indicator of the expected range of vision with a presbyopia-correcting IOL) was tested binocularly with the participant's best spectacle correction at a distance of 5 m using a chart. Lenses of different spherical powers were placed in front of the eyes to produce varying levels of defocus. The VA at each spherical power was measured in decimal, with 1.0 decimal corresponding to 20/20 Snellen. A higher numeric value represents better visual acuity. Day 120-180 from second eye implantation No
Secondary Percentage of Participants With Positive Response, Stereoscopic Vision Test Stereopsis (the ability to perceive depth and 3-dimensional structure) was assessed binocularly under well-lit conditions at best distance using the Stereo Optical Company, Inc., Original Stereo Fly Test and polarized viewers. The participant was shown a series of symbols, with a positive response defined as correct identification of the 3-dimensional symbol. Responses were recorded for 3 symbols: Fly (easiest to perceive), animal, and circle (hardest to perceive). Day 120-180 from second eye implantation No
Secondary Percentage of Participants With Positive Response, Quality of Life Questions The participant completed a questionnaire, indicating (yes/no) if he/she experienced visual difficulty in every-day activities while not wearing spectacles. A positive response was defined as, "Yes=visual difficulty." Day 120-180 from second eye implantation No
See also
  Status Clinical Trial Phase
Active, not recruiting NCT06005675 - User Acceptability Evaluation of Pseudophakic Patients Previously Implanted With the TECNIS Odyssey IOL
Active, not recruiting NCT01382641 - Show Improved Vision With the HOYA AF-1 Aspheric Intraocular Lens After Cataract Surgery Phase 4
Completed NCT01382823 - Laser Cataract Surgery With the Femtosecond Laser Technology Phase 4
Completed NCT05991960 - Evaluation of Real-World Data on the Performance of the TECNIS Odyssey™ IOL
Active, not recruiting NCT05574959 - Clinical Investigation of the TECNIS® Next-Generation Intraocular Lens, Model DEN00V N/A
Completed NCT02492659 - Clinical Research of Femtosecond Laser-Assisted Cataract Surgery N/A
Completed NCT01279031 - Randomized Comparison of the Abbott WHITESTAR Signature System With Ellips Tranversal Ultrasound vs. the Alcon Infiniti With the Ozil Torsional Handpiece in Phacoemulsification: A Contralaterally-Controlled Trial Phase 4
Completed NCT01061463 - Occupational Cataracts and Lens Opacities in Interventional Cardiology : the O'CLOC Study N/A
Completed NCT01001806 - A Comparison of Peak Aqueous Penetration of Acuvail (Ketorolac 0.45%), Xibrom (Bromfenac 0.09%), and Nevanac (Nepafenac 0.1%)in Patients Undergoing Phacoemulsification Phase 4
Completed NCT00999492 - Selection of Aspheric Diffractive Multifocal Intraocular Lens (IOLS) Based on Corneal Wavefront Phase 4
Completed NCT01021761 - A Comparison of Prostaglandin E2 (PGE2) Inhibition of Acuvail(Ketorolac 0.45%), Xibrom (Bromfenac 0.09%)and Nevanac (Nepafenac)in Patients Undergoing Phacoemulsification Phase 4
Completed NCT00542581 - Acrysof Toric SN60T3corneal Astigmatism Between 0.75 and 1.00 D. N/A
Completed NCT03708367 - A Post-Market Evaluation of LipiFlow Treatment in Cataract Surgery Practice N/A
Completed NCT05575063 - Clinical Investigation of Healon EndoCoat Ophthalmic Viscosurgical Device (OVD) N/A
Recruiting NCT05531292 - PROOF-OF-CONCEPT STUDY FOR NEW INTRAOCULAR LENS, MODEL C0002 N/A
Completed NCT05058274 - Clinical Results of a Monofocal Aspheric Bitoric Intraocular Lens With Plate Haptics in Hyperopic Eyes (HIPER-AT-TORBI)
Completed NCT05053399 - Evaluation of the TECNIS Symfony® Toric Intraocular Lens N/A
Recruiting NCT02639845 - Eye Drop Application Monitor, Pilot Study N/A
Completed NCT01684007 - A Clinical Outcome Study of Two Multifocal Intraocular Lenses (IOL) in Cataract Patients N/A
Completed NCT00827073 - Comparison of Tetracaine 0.5% and Lidocaine 2% Jelly for Topical Phacoemulsification Cataract Surgery N/A