NCT number NCT05351658
Other study ID # TEMPACE
Secondary ID
Status Recruiting
Phase N/A
First received
Last updated
Start date January 5, 2023
Est. completion date October 1, 2024

Study information

Verified date March 2024
Source ICU Medical Technologies S.L.
Contact Raimundo Vicente, MD
Phone +34 965933000
Email tempace@icumedic.com
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

Temporary cardiac pacing with active-fixation leads (TPAFL) using a reusable permanent pacemaker generator has been shown to be safer than lead systems without fixation. However, TPAFL requires the off-label (non-certificated) use of Pacemaker leads and generators. It is designed a fastening system to ensure the safety and efficacy of the procedure: the KronoSafe System®. The aim of this study is to demonstrate the safety and effectiveness of the KronoSafe System® for temporary pacing in a series of patients receiving TPAFL.


Recruitment information / eligibility

Status Recruiting
Enrollment 122
Est. completion date October 1, 2024
Est. primary completion date October 1, 2024
Accepts healthy volunteers No
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria: - Patient older than 18 years. - Need for temporary cardiac stimulation - Temporary active fixation pacemaker implantation through left or right jugular access. - Signature of informed consent - Correct functioning of the electrode after the implantation: sensing of the R wave above 5 millivolts. Ventricular pacing threshold below 1 volt. Electrode impedance between 400-1200 Ohm. - Pacemaker programming with stimulation output at 5 volts intensity and 1 millisecond pulse duration. - Placement of the KronoSafe® fastening system after electrode implantation. Exclusion Criteria: - Complication during the procedure: pneumothorax, hematoma at the puncture site, cardiac perforation, accidental arterial puncture or other that, in the investigator's opinion, poses an added risk of pacemaker dysfunction. - Need to transfer the patient to another center with loss of follow-up. - Absence of surveillance through continuous electrocardiographic monitoring with detection of arrhythmic events.

Study Design


Related Conditions & MeSH terms


Intervention

Device:
KronoSafe® System
All patients included in the study will have a temporary pacemaker with active fixation leads, held by the KronoSafe® fixation system.

Locations

Country Name City State
Spain Hospital General Universitario de Alicante Dr.Balmis Alicante
Spain Hospital Universitari Vall D'Hebron Barcelona

Sponsors (2)

Lead Sponsor Collaborator
ICU Medical Technologies S.L. Universidad Miguel Hernandez de Elche

Country where clinical trial is conducted

Spain, 

References & Publications (4)

Betts TR. Regional survey of temporary transvenous pacing procedures and complications. Postgrad Med J. 2003 Aug;79(934):463-5. doi: 10.1136/pmj.79.934.463. — View Citation

Kawata H, Pretorius V, Phan H, Mulpuru S, Gadiyaram V, Patel J, Steltzner D, Krummen D, Feld G, Birgersdotter-Green U. Utility and safety of temporary pacing using active fixation leads and externalized re-usable permanent pacemakers after lead extraction — View Citation

Vicente-Miralles R, Martin-Langerwerf DA, Nunez-Martinez JM, Marco-Juan A, Sangio-Ferreyros AD, Davila-Arias A, Montes-Jacobo L, Barrios V, Palazon-Bru A, Gil-Guillen VF, Bertomeu-Gonzalez V. A new fastening system for temporary pacing with active-fixatio — View Citation

Zei PC, Eckart RE, Epstein LM. Modified temporary cardiac pacing using transvenous active fixation leads and external re-sterilized pulse generators. J Am Coll Cardiol. 2006 Apr 4;47(7):1487-9. doi: 10.1016/j.jacc.2006.01.006. Epub 2006 Mar 15. No abstrac — View Citation

Outcome

Type Measure Description Time frame Safety issue
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer Immediately after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer Immediately after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer Immediately after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer 24 hours after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 3 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 5 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 7 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 9 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 11 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 13 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 15 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 17 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 19 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 21 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 23 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 25 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 27 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer day 29 after the procedure of lead implantation
Primary Ventricular R wave detection Ventricular R wave, measured in millivolts with pacemaker generator programmer Immediately before the procedure of lead removal
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer 24 hours after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 3 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 5 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 7 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 9 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 11 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 13 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 15 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 17 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 19 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 21 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 23 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 25 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 27 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer day 29 after the procedure of lead implantation
Primary Lead impedance detection Lead impedance, measured in ohms with pacemaker generator programmer Immediately before the procedure of lead removal
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer 24 hours after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 3 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 5 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 7 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 9 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 11 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 13 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 15 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 17 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 19 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 21 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 23 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 25 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 27 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer day 29 after the procedure of lead implantation
Primary Right ventricle threshold detection Right ventricle threshold, measured in volts with pacemaker generator programmer Immediately before the procedure of lead removal
Primary Lead displacement detection Visual observation of 2 cm electrode displacement through vein acces by clinical examination Immediately before the procedure of lead removal
Primary Capture failure detection Right ventricle threshold above 5 volts during 1 millisecond, measured with pacemaker generator programmer. Immediately before the procedure of lead removal
Primary Sensing failure detection Ventricular R wave below 1 volt, measured with pacemaker generator programmer Immediately before the procedure of lead removal
Primary Lead insulation breakdown detection Visual observation of lead insulation breakdown by clinical examination Immediately before the procedure of lead removal
Primary Generator detachment detection Visual observation of generator detachment by clinical examination. removal of the generator from the casing. Immediately before the procedure of lead removal
Primary Cardiac peroration detection New pericardial effusion detected by Echocardiography with signs of cardiac tamponade: respiratory variation of transmitral filling flow greater than 25% Immediately before the procedure of lead removal
Secondary Detection of Bacteremia Isolation of the same bacteria in the pacemaker lead and 2 blood cultures from the patient Immediately after the procedure of lead removal
Secondary Detection of Agitation Record of antipsychotic and anxiolytic treatment of the patient Immediately after the procedure of lead removal
See also
  Status Clinical Trial Phase
Completed NCT03922386 - Safety and Electrical Performances of XFINE Leads N/A
Completed NCT03294018 - Heart Rate Changes Following the Administration of Sugammadex
Not yet recruiting NCT05649891 - Checklists Resuscitation Emergency Department N/A
Not yet recruiting NCT05440071 - PoSt-market Assessment on Safety and Performance of J-sHAPed xFINE Lead
Completed NCT02153242 - The Effects of Selective Site Right Ventricular Pacing
Recruiting NCT01825473 - Study of Erythromycin in GER-Associated Apnea of the Newborn N/A
Completed NCT00950287 - Detection of Neonatal Bradycardia N/A
Completed NCT00721136 - Randomized Study of the Use of Warfarin During Pacemaker or ICD Implantation in Patients Requiring Long Term Anticoagulation N/A
Completed NCT00286858 - Beluga - Clinical Observations of Automatic Algorithms for Cardiac Pacing
Completed NCT00158925 - The EASYTRAK EPI Clinical Investigation Phase 3
Completed NCT00180557 - Austria Study - Analysis of Difference Between Active and Passive Fixation Leads Phase 4
Completed NCT01076348 - Model 4965 Post-Approval Study
Completed NCT04198220 - BIO|STREAM.ICM Obesity
Recruiting NCT05298748 - The Effect of Womb Recordings on Maturation of Respiratory Control in Preterm Infants N/A
Recruiting NCT04353960 - The Alaska Oculocardiac Reflex Study
Terminated NCT04093414 - Left Bundle Area Versus Selective His Bundle Pacing N/A
Completed NCT03786640 - Abbott Brady 3T MRI PMCF
Recruiting NCT05932602 - AVEIR DR Coverage With Evidence Development (CED) Study
Recruiting NCT05935007 - Aveir DR Real-World Evidence Post-Approval Study
Recruiting NCT04075084 - Observation of Clinical Routine Care for Patients With BIOTRONIK Implantable Cardiac Monitors (ICMs)