Clinical Trial Details — Status: Recruiting

Administrative data

NCT number NCT03811691
Other study ID # IBSY06 - CAPRI
Secondary ID
Status Recruiting
Phase Phase 4
First received
Last updated
Start date July 18, 2019
Est. completion date August 2023

Study information

Verified date June 2021
Source MicroPort CRM
Contact Anne RousseauPlasse
Phone 01 46 01 30 82
Email Anne.RousseauPlasse@crm.microport.com
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The interest of the CAPRI study is to confirm the safety and the efficacy of the ENO/TEO/OTO pacing system when used under 1.5 or 3 Tesla MRI environment and in accordance with the MRI solutions guideline.


Description:

The purpose of CAPRI study is to confirm the safety and the efficacy of the CE marked MR conditional pacing system composed of the ENO, TEO or OTO pacemaker with VEGA pacing lead(s). Conditions to undergo an MRI scan are provided in the MRI solutions guideline. The primary objective is to confirm the clinical safety of the ENO/TEO/OTO pacing system when used under 1.5 and 3 Tesla specific MRI conditions without scan exclusion zone. The study secondary main objectives are aiming to assess the performance of the MR conditional pacing system in the right atrium and ventricle at 1 month following the MRI scan: - Stability of the Pacing Capture Threshold - Stability of the lead sensed amplitude


Recruitment information / eligibility

Status Recruiting
Enrollment 270
Est. completion date August 2023
Est. primary completion date December 2021
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria: Subjects who meet all the following criteria at the time of inclusion visit may be included: - Already implanted in the left or right pectoral region for at least 6 weeks with: - ENO, TEO or OTO single chamber rate response (SR) pacemaker with a VEGA pacing lead, or - ENO, TEO or OTO dual chamber rate response (DR) pacemaker with two VEGA pacing leads. - Implanted pacing system must fulfill the following parameters: - Battery impedance is < 5 Kilo Ohm (k?) - Pacing capture threshold value is = 2 Volts (V) at 0.35 millisecond (ms) - Lead impedance value between 200? and 3000? - No diaphragmatic or pectoral stimulation at 5V/1ms - P-wave minimum sensed amplitude = 1mV for patients with sinus rhythm - R-wave minimum sensed amplitude = 4mV for patients with spontaneous conduction rhythm - Must agree to undergo a non clinically-indicated MRI scan without intravenous injection and without sedation; - Have reviewed, signed and dated informed consent. Subjects who meet any of the following criteria are not eligible to be included in the study: - Included in another clinical study that could confound the results of this study such as studies involving intra-cardiac device; - Have other active or abandoned cardiac implants already implanted; - Have other active or passive non MR conditional devices implanted such as metallic foreign body; - Have a history of brain aneurysm with ferromagnetic clipping; - Have a planned cardiac surgery within the 3 months of inclusion; - Have a medical MRI examination prescription planned within the 3 months of inclusion; - Age less than 18 years old or under guardianship or kept in detention; - Known pregnancy, women breastfeeding or in childbearing age without an adequate contraceptive method ; - Be unavailable for the scheduled follow-up associated with this clinical study or refusal to cooperate.

Study Design


Related Conditions & MeSH terms


Intervention

Diagnostic Test:
MRI exam
Pacing system safety and stability under MRI scan environment

Locations

Country Name City State
France MicroPort CRM Clamart Haut De Seine

Sponsors (1)

Lead Sponsor Collaborator
MicroPort CRM

Country where clinical trial is conducted

France, 

Outcome

Type Measure Description Time frame Safety issue
Primary Freedom from MRI-related complications following the MRI scan of the ENO/TEO/OTO pacing system An MRI-related complication is defined as a Serious Adverse Event (SAE) that led to death or to an invasive intervention or a Device Deficiency that causes the termination of any significant device function per Clinical Event Committee (CEC) adjudication.
The primary endpoint will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan).
1 month post MRI
Secondary Ventricular Pacing Capture Threshold (PCT) stability in the right atrium and ventricle following the MRI scan Subject's ventricular pacing capture threshold will be measured in Volt (V) at 0.5ms at baseline (at the MRI visit - just before the MRI scan) and at 1 month Post-MRI visit. A success is defined as a PCT increase less than or equal to 0.5V at 0.5ms between the two visits. The secondary endpoints will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan). 1 month post MRI
Secondary Atrial Pacing Capture Threshold stability in the right atrium and ventricle following the MRI scan Subject's atrial pacing capture threshold will be measured in Volt (V) at 0.5ms at Baseline (at the MRI visit - just before the MRI scan) and at 1 month Post-MRI visit. A success is defined as a PCT increase less than or equal to 0.5V at 0.5ms between the two visits. The secondary endpoints will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan). 1 month post MRI
Secondary Ventricular sensed amplitude stability in the right atrium and ventricle following the MRI scan Subject's ventricular minimum sensed amplitude will be measured in mV at Baseline (at the MRI visit - just before the MRI scan) and at 1 month Post-MRI visit. A success is defined as 50% or less decrease between the two visits. If subject's ventricular minimum sensed amplitude is less or equal to 4mV at time of baseline, then this subject will not count for this endpoint. The secondary endpoints will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan). 1 month post MRI
Secondary Atrial sensed amplitude stability in the right atrium and ventricle following the MRI scan Subject's atrial minimum sensed amplitude will be measured in mV at baseline. A success is defined as 50% or less decrease between the two visits. If subject's atrial minimum sensed amplitude is less or equal to 1mV at time of Baseline (at the MRI visit - just before the MRI scan) and at 1 month Post-MRI visit, then this subject will not count for this endpoint.The secondary endpoints will be assessed independently for each arm of the study (1.5 and 3 Tesla MRI scan). 1 month post MRI
See also
  Status Clinical Trial Phase
Completed NCT03922386 - Safety and Electrical Performances of XFINE Leads N/A
Completed NCT03294018 - Heart Rate Changes Following the Administration of Sugammadex
Not yet recruiting NCT05649891 - Checklists Resuscitation Emergency Department N/A
Not yet recruiting NCT05440071 - PoSt-market Assessment on Safety and Performance of J-sHAPed xFINE Lead
Completed NCT02153242 - The Effects of Selective Site Right Ventricular Pacing
Recruiting NCT01825473 - Study of Erythromycin in GER-Associated Apnea of the Newborn N/A
Completed NCT00950287 - Detection of Neonatal Bradycardia N/A
Completed NCT00721136 - Randomized Study of the Use of Warfarin During Pacemaker or ICD Implantation in Patients Requiring Long Term Anticoagulation N/A
Completed NCT00286858 - Beluga - Clinical Observations of Automatic Algorithms for Cardiac Pacing
Completed NCT00158925 - The EASYTRAK EPI Clinical Investigation Phase 3
Completed NCT00180557 - Austria Study - Analysis of Difference Between Active and Passive Fixation Leads Phase 4
Completed NCT01076348 - Model 4965 Post-Approval Study
Completed NCT04198220 - BIO|STREAM.ICM Obesity
Recruiting NCT05298748 - The Effect of Womb Recordings on Maturation of Respiratory Control in Preterm Infants N/A
Recruiting NCT04353960 - The Alaska Oculocardiac Reflex Study
Terminated NCT04093414 - Left Bundle Area Versus Selective His Bundle Pacing N/A
Completed NCT03786640 - Abbott Brady 3T MRI PMCF
Recruiting NCT05932602 - AVEIR DR Coverage With Evidence Development (CED) Study
Recruiting NCT05935007 - Aveir DR Real-World Evidence Post-Approval Study
Recruiting NCT04075084 - Observation of Clinical Routine Care for Patients With BIOTRONIK Implantable Cardiac Monitors (ICMs)