Blood Pressure Measurement Clinical Trial
Official title:
Evaluation of Sensifree's Alpha Sensor Signal Morphology in Comparison to Invasive Radial Arterial Line Pressure Waveform and PPG Signal Morphology
Verified date | September 2019 |
Source | Sensifree Ltd. |
Contact | n/a |
Is FDA regulated | No |
Health authority | |
Study type | Interventional |
The purpose of the protocol is to compare Sensifree Alpha sensor signal morphology, measured
at the radial artery, to an invasive arterial line pressure signal and the oxygen saturation
as read by a pulse oximeter.
The study will be split into two phases: the first phase will collect data for morphology
analysis and algorithm development, the second phase will compare device performance to the
reference invasive blood pressures. A secondary objective is to evaluate signal morphology
changes due to changes in subject arm position.
Status | Completed |
Enrollment | 15 |
Est. completion date | September 29, 2017 |
Est. primary completion date | September 29, 2017 |
Accepts healthy volunteers | Accepts Healthy Volunteers |
Gender | All |
Age group | 20 Years to 55 Years |
Eligibility |
Inclusion Criteria: - General: - Subject must have the ability to understand and provide written informed consent - Subject must be willing and able to comply with study procedures and duration - Screening and Phase 1: - Subject is from 20 to 55 years old. - Subject has wrist circumference size within 145 and 220mm for both arms. - Subjects with lateral blood pressure difference (auscultatory cuff based measurement) is less than: 15 mmHg for systolic blood pressure. 10 mmHg for diastolic blood pressure. - Subjects with investigational device signal quality higher than a predefined value - Subjects with similar investigational devices signals between their two arms. - Phase 2 (all requirements as Phase 1 except): - Subject is from 20 to 85 years old. - At least 60% of subjects will have hypertensive blood pressure values (goal) (systolic >140 mmHg and/or diastolic >90 mmHg) - At least 15% of subjects will have hypotensive blood pressure values (goal) (systolic <110 mmHg and/or diastolic <60 mmHg). - No more than 25% of subjects will have normotensive blood pressure values (goal) (systolic from 110 to 140 mmHg and/or diastolic from 60 to 90 mmHg). Exclusion Criteria: - Subject with a BMI over 39 - Deformities or abnormalities that may prevent proper application of the device under test - Females who are pregnant, who are trying to get pregnant, or have a urine test positive for pregnancy on the day of the study - Subjects with known respiratory conditions such as: (self-reported) - uncontrolled / severe asthma, - flu, - pneumonia / bronchitis, - shortness of breath / respiratory distress, - respiratory or lung surgery, - emphysema, COPD, lung disease - Subjects with known heart or cardiovascular conditions such as: (self-reported, except for blood pressure and ECG review) - have had cardiovascular surgery - have cardiac pacemakers and/or automatic internal cardio-defibrillator - Chest pain (angina) - Abnormal pulse pressure - previous heart attack - blocked artery - unexplained shortness of breath - congestive heart failure (CHF) - history of stroke - transient ischemic attack - carotid artery disease - myocardial ischemia - myocardial infarction - cardiomyopathy - Subject with a heart rate of less than 45 BPM - Self-reported health conditions as identified in the Health Assessment Form (self- reported) - diabetes, - uncontrolled thyroid disease, - kidney disease / chronic renal impairment, - history of seizures (except childhood febrile seizures), - epilepsy, - history of unexplained syncope, - recent history of frequent migraine headaches, - recent head injury - cancer / chemotherapy - Subjects with known clotting disorders (self-reported) - history of bleeding disorders or personal history of prolonged bleeding from injury - history of blood clots - hemophilia - current use of blood thinner: prescription or daily use of aspirin - Subjects with severe contact allergies to standard adhesives, latex or other materials found in pulse oximetry sensors, ECG electrodes, respiration monitor electrodes or other medical sensors (self-reported) - Subjects with a contact allergy to ultrasound gel. - Subjects with prior or known allergies to iodine or lidocaine (or similar pharmacological agents, e.g. Novocaine) - Failure of the Perfusion Index Ulnar/Ulnar+Radial Ratio test - Subject is intoxicated during the time of the visit, or was intoxicated within 24 hours prior to study visit, as reported by the subject, or per study staff judgment. - Other known health condition, should be considered upon disclosure in health assessment form - Subjects with heart rate below 45 BPM. |
Country | Name | City | State |
---|---|---|---|
United States | Clinimark Lab | Louisville | Colorado |
Lead Sponsor | Collaborator |
---|---|
Sensifree Ltd. |
United States,
Type | Measure | Description | Time frame | Safety issue |
---|---|---|---|---|
Primary | Comparison of signals acquired from the investigational device, pulse oximeter, and arterial line | BP waveform will be acquired form three sources (PPG, A-line and investigational device),data will be statistically analyzed to find the correlations between the different sources of BP waveform | Three hours | |
Secondary | Evaluation of waveform signal changes with different arm positions | Recordings will be made using the investigational device with different position of the subject arm to assess the effect of the arm positions (arm up, down and in chest level) over the acquired BP waveform signal | Three hours |
Status | Clinical Trial | Phase | |
---|---|---|---|
Completed |
NCT04976452 -
Identifying Learning Strategies to Improve Blood Pressure Measurement in Physical Therapy Education Programs
|
N/A | |
Completed |
NCT04510753 -
Data Collection of Arterial Line Waveform and BP Values
|
||
Completed |
NCT02447471 -
Agreement of the Nexfin™ Monitor With Non-invasive Blood Pressure Measurement
|
N/A | |
Not yet recruiting |
NCT06198855 -
Automated Measurement of Blood Pressure in Waiting Room (AMBP-waiting)
|
||
Completed |
NCT04132453 -
Continuous Blood Pressure and Waveform Measurements Comparison Between Invasive Arterial Line and PPG Sensor
|
||
Completed |
NCT04456179 -
Continuous Non-Invasive Blood Pressure System Data Collection in Comparison to Invasive Radial Arterial Pressure
|
N/A |