Clinical Trials Logo

Clinical Trial Summary

Premature ventricular contractions (PVC) are common phenomenon and occasionally retrograde conduction from ventricle to atrium occurs and may cause pacemaker mediated tachycardia. In response certain Abbott (former St Jude Medical) pacemakers have a specific PVC response algorithm ''Atrial Pace''. In this algorithm the PVARP (Post Ventricular Atrial Refractory Period) is extended to 480ms but the initial 150ms of the PVARP is deemed absolute. If atrial sensed event occurs after absolute PVARP the PVARP terminates and atrial pace follows after 330ms alert period. Two case reports exists where patients atrial tachycardia were suspected to be induced by this algorithm. In Helsinki University Hospital an index patient with multiple atrial tachycardia episodes was recognized in late 2020 where the cause of the tachycardias was suspected to be the PVC response algorithm ''Atrial Pace''. After programming the algorithm from Atrial Pace to off mode the patient had none atrial tachycardias during 2021. After investigating all the patients with physiological ICD (Implantable Cardioverter Defibrillator) or CRT-D (Cardiac Resynchronization Therapy) device and analyzing remote monitoring transmissions from the year 2020 we found 25 patients with similar atrial tachycardia episodes possibly induced by the PVC response ''Atrial Pace''.


Clinical Trial Description

The device of the subjects participating to the study is interrogated and the PVC response algorithm is switched from ''Atrial Pace'' to ''Off'' mode. Subjects are also requested to fulfill the EQ-5D-3L and ICD8 questionaries before the change in the device programming and after the 12 months surveillance. ;


Study Design


Related Conditions & MeSH terms


NCT number NCT05344456
Study type Interventional
Source Helsinki University Central Hospital
Contact
Status Completed
Phase N/A
Start date May 11, 2022
Completion date August 20, 2023

See also
  Status Clinical Trial Phase
Completed NCT00247351 - Prevention of Atrial Tachycardia After a Right Atriotomy N/A
Completed NCT04162249 - High Radiofrequency Power for Faster and Safer Pulmonary Veins Isolation - a Pilot Observational Study.
Recruiting NCT03209583 - ATrial Tachycardia PAcing Therapy in Congenital Heart
Recruiting NCT03750435 - Non-fluoroscopy Ablation of AF/AT N/A
Recruiting NCT05883631 - RESOLVE-AF: Clinical Evaluation of the Ablacathâ„¢ Mapping Catheter and Ablamap® System Utilizing Electrographic Flow (EGF) Mapping to Resolve Extra-PV Sources of Atrial Fibrillation and Guide Ablation Therapy. N/A
Recruiting NCT05434650 - Abbott Atrial Fibrillation Post Approval Study
Completed NCT06335446 - Identification, Electro-mechanical Characterisation and Ablation of Driver Regions in Persistent Atrial Fibrillation
Recruiting NCT05014802 - Baseline Atrial Fibrosis Predicts Risk for Post-operative Atrial Fibrillation in Patients Undergoing Cardiac Surgery
Completed NCT05362656 - AutoMatic disPERsion Tagging Function Preliminary Evaluation N/A
Not yet recruiting NCT05765565 - Comparison Between the Octaray and Pentaray Catheters in Patients Undergoing Ablation for Atrial Tachycardia. N/A
Recruiting NCT05521451 - Clinical Cohort Study - TRUST
Not yet recruiting NCT06327425 - MCG for Localization of Tachyarrhythmia's Origin N/A
Completed NCT03549806 - Stanford Cardiac Invasive Electrophysiology Novel Computer Experience
Completed NCT02451995 - Effectiveness of Ripple Mapping in Atrial Tachycardia Ablation N/A
Recruiting NCT03816033 - Cryotherapy Versus Radiofrequency Catheter Ablation Research Program N/A
Completed NCT05802576 - Study of Interference Between Oral Anticoagulants and Heparin During Ablation of Atrial Fibrillation (AF) or Left Atrial Tachycardia (GAD) by Catheter.
Completed NCT01108692 - Strategy of Early Detection and Active Management of Supraventricular Arrhythmia With Telecardiology (SETAM) N/A
Active, not recruiting NCT02374359 - Atrial Tachycardia as a Predictor of Atrial Fibrillation, Stroke and Mortality in Cryptogenic Stroke N/A
Completed NCT03151278 - Zero Fluoroscopic Ablation Versus Conventional Fluoroscopic Ablation for Right Atrial Arrhythmias N/A
Completed NCT03413150 - Modeling of Amiodarone Effect on Heart Rate Control in Critically Ill Patients With Atrial Tachyarrhythmias