Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT01999751
Other study ID # EH12274
Secondary ID
Status Completed
Phase
First received November 26, 2013
Last updated April 11, 2018
Start date October 2012
Est. completion date April 11, 2018

Study information

Verified date April 2018
Source Essentia Health
Contact n/a
Is FDA regulated No
Health authority
Study type Observational [Patient Registry]

Clinical Trial Summary

This study is a prospective, non-randomized, unblinded case series of patients with permanent pacemakers and implantable cardioverter-defibrillators undergoing medically-required magnetic resonance imaging (MRI) scanning, at St. Mary's Medical Center, Essentia Health. Patients will be enrolled over a 60-month period and followed for 12 months, with data collected to evaluate the study's primary endpoint, change in pacing thresholds over time, as well as a series of secondary endpoints, including: adverse events, symptoms, need to make pacemaker programming changes, and possible artifacts created by the pacemaker systems on the MRI scans. This study will allow for carefully monitoring of the safety MRI scanning in this population and improve practice models and protocols in the future.


Description:

General Approach: This study is a prospective, non-randomized, unblinded case series of patients with permanent pacemakers or implantable cardioverter-defibrillators (ICDs) undergoing medically-required MRI scanning, followed for a 12 month period. MRI scanning will be performed using accepted, standard protocols using FDA-approved equipment.

Methods and Materials: Patients will be enrolled over a 60-month period by the Principal Investigator (PI) and/or Sub-Investigator (Sub-I), at the Regional Heart Center, located at St. Mary's Medical Center, Essentia Health, Duluth, MN, during evaluations for medically-required MRI scanning. Eligible patients includes any person with a permanent pacemaker or ICD implanted for at least 6 weeks, referred for a medically-required MRI scan by a physician not participating in the trial. ICD generators must have been market released after 2001. Both pacemaker dependent and non-dependent patients are eligible. The primary endpoint is any change greater than 1 V at a pulse width of 0.5 ms in a pacemaker or ICD lead at any time within 12 months of the MRI scan.

On the day of the MRI scan, the pacemaker or ICD will be interrogated and programmed per usual practice at Essentia Health for patients undergoing an MRI scan with device in place. Three separate baseline capture threshold measurements (V) will be obtained at a fixed pulse width of 0.5 msec. Three separate baseline measurements will also be made of lead pacing impedances (Ohms) and sensed P- or R-waves (mV) using programmer default methods. Histograms will be cleared. Leads will be programmed to bipolar modality. If the patient is not device dependent, the device will be programmed to ODO/OVO/OAO mode (monitor only) if available, otherwise it will be reprogrammed to DDI or VVI mode with a lower rate of 40 bpm with pacing outputs programmed to subthreshold levels in all leads. If the participant is device dependent, the device will be programmed to an asynchronous pacing modality (DOO/VOO) with maximum pacing output. If the patient has an ICD, therapies will be disabled. The attending investigator may alter the pacing modality during the course of the scan if he or she deems it necessary to maximize patient safety or comfort. Rate responsiveness, any rate enhancements including Ventricular Rate Regulation mode, and magnet mode will be turned off.

During MRI scanning, resuscitation equipment will be available. If the participant is pacemaker dependent, the study PI or sub-I must be in the MRI suite during the scan. If the participant is not pacemaker dependent, then a pacemaker nurse will be in the MRI suite to perform pre- and post-checks, and an electrophysiologist will be available in the building for questions and troubleshooting. Patients will be asked to report all symptoms including: palpitations; pain; heat or warmth sensation; device movement; or light-headedness.

Scans will be performed using usual protocols with standard peak Specific Absorption Rate (SAR) limitations for the scan. The MRI scans will be evaluated for any artifact created by the pacemaker system. During the scan, continuous electrocardiographic and pulse oxymetry with waveform monitoring will be performed using usual protocols. Following the scan, repeat thresholds will be performed for each lead and histograms will be obtained. Three separate capture threshold measurements (V) will be obtained immediately after the patient leaves the scanner room at a fixed pulse width of 0.5 msec. Three separate measurements will also be made immediately following the scan of the lead pacing impedances (Ohms) and sensed P- or R-waves (mV) using programmer default methods. Participants who are pacemaker dependent will undergo a one-hour post-scan check using the same threshold protocol. These one-hour post scan measurements will be performed either in the pacemaker clinic if the patient is an outpatient or in the hospital if the patient is an inpatient. Any stored events detected on the device as electromagnetic interference (EMI) artifact during the scan may be retrieved and stored to diskette for later analysis. The pacemaker or ICD will be reprogrammed to original settings prior to discharge from the MRI suite unless the attending investigator deems it necessary to alter device setting to maximize patient safety or comfort. If the patient has an ICD, the capacitor will be reformed and charge time noted prior to dismissal from the MRI suite.

Patients will be followed up in the pacemaker clinic at 3 and 12 months from the MRI scan using usual clinic protocols for device follow-up. If a patient undergoes a second MRI scan, the follow-up schedule will be set based upon the most recent MRI scan.


Recruitment information / eligibility

Status Completed
Enrollment 124
Est. completion date April 11, 2018
Est. primary completion date April 2, 2018
Accepts healthy volunteers No
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria:

1. Permanent Pacemaker or ICD implanted for at least 6 weeks

2. Referred for medically-required MRI scan

3. Age greater than or equal to 18 years

Exclusion Criteria:

1. ICD market released before 2001 including GDT Ventak-AV, GDT MINI-II, GDT MINI-IV, GDT MINI-III, GDT Prizm, GDT Prizm-2 and MDT GEM-I series

2. Epicardial lead or subcutaneous array

3. Device has reached elective replacement indicator (ERI) or end of life (EOL)

4. Other usual contraindications to MRI scanning

5. Known or suspected lead fracture or abandoned lead

6. Pacemaker dependent patients with an ICD that cannot be programmed to asynchronous pacing (DOO/VOO)

7. Clinical factors that are deemed by the PI or Sub-I to increase the participant's risk above an acceptable clinical threshold

Study Design


Related Conditions & MeSH terms


Intervention

Device:
MRI Scan
Clinically indicated MRI scan of patients with pacemaker or implantable cardioverter-defibrillator in situ.

Locations

Country Name City State
United States St. Mary's Medical Center Duluth Minnesota

Sponsors (1)

Lead Sponsor Collaborator
Essentia Health

Country where clinical trial is conducted

United States, 

References & Publications (12)

Erlebacher JA, Cahill PT, Pannizzo F, Knowles RJ. Effect of magnetic resonance imaging on DDD pacemakers. Am J Cardiol. 1986 Feb 15;57(6):437-40. — View Citation

Fetter J, Aram G, Holmes DR Jr, Gray JE, Hayes DL. The effects of nuclear magnetic resonance imagers on external and implantable pulse generators. Pacing Clin Electrophysiol. 1984 Jul;7(4):720-7. — View Citation

Gimbel JR, Kanal E, Schwartz KM, Wilkoff BL. Outcome of magnetic resonance imaging (MRI) in selected patients with implantable cardioverter defibrillators (ICDs). Pacing Clin Electrophysiol. 2005 Apr;28(4):270-3. — View Citation

Goldschlager N, Epstein A, Friedman P, Gang E, Krol R, Olshansky B; North American Society of Pacing and Electrophysiology (NASPE) Practice Guideline Committee. Environmental and drug effects on patients with pacemakers and implantable cardioverter/defibrillators: a practical guide to patient treatment. Arch Intern Med. 2001 Mar 12;161(5):649-55. Review. — View Citation

Hayes DL, Holmes DR Jr, Gray JE. Effect of 1.5 tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers. J Am Coll Cardiol. 1987 Oct;10(4):782-6. — View Citation

Martin ET, Coman JA, Shellock FG, Pulling CC, Fair R, Jenkins K. Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla. J Am Coll Cardiol. 2004 Apr 7;43(7):1315-24. — View Citation

Mollerus M, Albin G, Lipinski M, Lucca J. Magnetic resonance imaging of pacemakers and implantable cardioverter-defibrillators without specific absorption rate restrictions. Europace. 2010 Jul;12(7):947-51. doi: 10.1093/europace/euq092. Epub 2010 Mar 30. — View Citation

Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, Caffo BS, Frick KD, Kraut MA, Kamel IR, Calkins H, Berger RD, Bluemke DA, Halperin HR. A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011 Oct 4;155(7):415-24. doi: 10.7326/0003-4819-155-7-201110040-00004. — View Citation

Nazarian S, Roguin A, Zviman MM, Lardo AC, Dickfeld TL, Calkins H, Weiss RG, Berger RD, Bluemke DA, Halperin HR. Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla. Circulation. 2006 Sep 19;114(12):1277-84. Epub 2006 Sep 11. — View Citation

Roguin A, Zviman MM, Meininger GR, Rodrigues ER, Dickfeld TM, Bluemke DA, Lardo A, Berger RD, Calkins H, Halperin HR. Modern pacemaker and implantable cardioverter/defibrillator systems can be magnetic resonance imaging safe: in vitro and in vivo assessment of safety and function at 1.5 T. Circulation. 2004 Aug 3;110(5):475-82. Epub 2004 Jul 26. — View Citation

Shellock FG, O'Neil M, Ivans V, Kelly D, O'Connor M, Toay L, Crues JV. Cardiac pacemakers and implantable cardioverter defibrillators are unaffected by operation of an extremity MR imaging system. AJR Am J Roentgenol. 1999 Jan;172(1):165-70. — View Citation

Sommer T, Vahlhaus C, Lauck G, von Smekal A, Reinke M, Hofer U, Block W, Träber F, Schneider C, Gieseke J, Jung W, Schild H. MR imaging and cardiac pacemakers: in-vitro evaluation and in-vivo studies in 51 patients at 0.5 T. Radiology. 2000 Jun;215(3):869-79. — View Citation

* Note: There are 12 references in allClick here to view all references

Outcome

Type Measure Description Time frame Safety issue
Primary Pacing threshold increase Any change greater than 1 V at a pulse width of 0.5 msec in a pacemaker or implantable cardioverter-defibrillator lead at any time within 12 months of the MRI scan 12 months
Secondary Ectopy Increased cardiac ectopy at the time of the MRI scan 1 hour
Secondary Pacemaker reprogramming Unanticipated pacemaker or implantable cardioverter-defibrillator reprogramming because of alterations in device behavior possibly related to the MRI scan 12 months
See also
  Status Clinical Trial Phase
Completed NCT03079726 - Use of Device Data to Predict Frailty in Individuals
Recruiting NCT06014996 - Comparison of PFA vs. RFA in Patients With Symptomatic Paroxysmal Atrial Fibrillation. N/A
Recruiting NCT04092985 - Smart Watch iECG for the Detection of Cardiac Arrhythmias
Completed NCT05404789 - Tolerance of Centrifuge-induced Acceleration in Subjects With Diabetes Mellitus and Cardiac Arrhythmia N/A
Recruiting NCT04554160 - Arrhythmias in Pulmonary Hypertension Assessed by Continuous Long-term Cardiac Monitoring
Recruiting NCT04471584 - Comparative Analysis in Detection of Atrial Arrhythmia and ECG Quality in Three Different Insertable Cardiac Monitors N/A
Recruiting NCT02906189 - Safety of Magnetic Resonance Imaging in Non-MRI Conditional Pacemakers and ICD Devices N/A
Completed NCT02550340 - Munich Study for Beer Related ECG Change Workup N/A
Completed NCT02466633 - Effect of Monitoring System Design on Response Time to Cardiac Arrhythmias N/A
Completed NCT01871090 - Remote Device Interrogation In The Emergency Department N/A
Completed NCT04884100 - enHEART - Exploring Full Content of Optical Signals to Enhance Cardiac Arrhythmia Screening N/A
Withdrawn NCT04556240 - RECORD-VP: Real-time Evaluation of Cardiac Outpatient Recording Device With VitalPatch RTM
Active, not recruiting NCT05481359 - AF-FLOW Registry: This Study is to Evaluate Ablamap Software in Patients Undergoing Ablation for Atrial Fibrillation.
Recruiting NCT05047835 - Comparison of Prone Position and Standard Electrocardiogram in COVID-19 Patients
Not yet recruiting NCT05974306 - Evaluation of a Proactive Clinical Management and Early Diagnosis of Arrhythmias in Patients With Heart Failure and Non Severely Reduced Left Ventricular Function Through a Telemonitoring System N/A
Not yet recruiting NCT05957315 - Mobile Cardiac Outpatient Telemetry for Unexplained Syncope N/A
Completed NCT06260670 - FLOW EVAL-AF: FLOW Mapping Electrogram VALidation in Patients With Persistent Atrial Fibrillation N/A
Active, not recruiting NCT04876963 - HOLT-ED: Holter-monitoring in End-stage Renal Disease
Completed NCT04943354 - Associations of Combinations of Single-nucleotide Polymorphisms in Women With Premature Ovarian Failure N/A
Recruiting NCT03857711 - Randomized Clinical Trial PULVAB (Prophylactic Pulmonary Veins Ablation) N/A