Clinical Trials Logo

Clinical Trial Details — Status: Active, not recruiting

Administrative data

NCT number NCT03214601
Other study ID # IP-0019-16
Secondary ID
Status Active, not recruiting
Phase N/A
First received
Last updated
Start date December 4, 2017
Est. completion date August 31, 2028

Study information

Verified date August 2023
Source Bolton Medical
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The purpose of this study is to conduct an early clinical evaluation of the Relay Branch System, which will provide initial insight into the clinical safety and function of the device. This Early Feasibility Study (EFS) will assess the safety and effectiveness of the device at the index procedure and at 30-day follow-up. The study will evaluate the delivery and deployment of the device, patency of branches and branch vessels, and exclusion of the aortic pathology. The data will help determine if modifications need to be made to the device, the procedural steps, operator technique, or the indications for use.


Description:

This Early Feasibility Study (EFS) protocol describes the clinical study requirements for the Relay Branch System; a device designed to provide an option for patients with arch and proximal descending chronic thoracic aortic pathologies. As an EFS, this investigation is intended to provide proof of principle and initial clinical safety data on the Relay Branch System. The study is planned as an initial investigation of the device for aortic arch and proximal descending thoracic aortic aneurysmal disease, PAU (including IMH) and uncomplicated chronic Type B aortic dissection (including IMH). The study will yield information on procedural techniques; assessing the safety and effectiveness of the device at the index procedure and at 30 days, focusing on device delivery and deployment, and circulatory exclusion of the pathologic process. As a branched device, patency of the endograft branches will also be assessed. The study will evaluate three-dimensional (3D) imaging data, both at baseline and through follow-up. Baseline 3D anatomy will augment information on the precise anatomic configuration of patients presenting aortic arch pathology treatable with the device. Follow-up imaging will provide information on the effectiveness of the device with respect to endoleaks in patients with aneurysms, sealing of dissections, PAU and IMH, and stability of the device at the deployed position, response, endograft patency, and short-term device integrity. The data from this EFS will yield insights into the following aspects of the device, preceding a traditional feasibility or pivotal study: - The clinical safety of the device-specific aspects of the procedure, - Determination of delivery and deployment of the device, - Operator-dependent aspects of device use, - Human factors associated with the design and use of the device, - Safety of the device as assessed by device-related adverse events, - Effectiveness of the device in performing its intended purpose over short-term follow-up. Observations from the study will guide the instructions for use (IFU) for the device. Finally, the study will collect imaging data to augment the current use conditions data set. It is anticipated that information collected will be used to make applicable design changes, or be combined with a prospective, investigational device exemption (IDE) study for submission of an original premarket approval application (PMA) to the U.S. Food and Drug Administration (FDA) for approval to commercially distribute the system.


Recruitment information / eligibility

Status Active, not recruiting
Enrollment 30
Est. completion date August 31, 2028
Est. primary completion date September 18, 2023
Accepts healthy volunteers No
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria: - Age =18 years - Anatomy that would require coverage of the brachiocephalic trunk (BCT) and/or left carotid arteries if a non-branch endograft were implanted - At least one of the following: a. Aneurysm of the ascending aorta, aortic arch, or proximal descending aorta that meets at least one of the following: i. = 5.5 cm in diameter ii. = 4.0 cm in diameter that has increased in size by 0.5 cm in the last 6 months iii. Measures twice the size of the normal aorta diameter iv. Is saccular in configuration b. PAU within the ascending aorta, aortic arch, or proximal descending thoracic aorta (DTA) with or without intramural hematoma (IMH) c. Chronic, uncomplicated aortic dissection distal to the BCT (type B), with either aortic diameter =5.5 cm or =4.0 cm with an increase in size by 0.5 cm in the last 6 months d. IMH distal to the BCT with recurrent symptoms, i.e. hypertension or intractable pain, despite best medical therapy. - A non-aneurysmal proximal aortic neck diameter ranging between 28 mm and 43 mm and a non-aneurysmal distal aortic neck diameter ranging between 19 mm and 43 mm. - A proximal attachment zone of the arch graft, meeting the oversizing requirement, must be a minimum of 25mm in length when measured on the inner curve of the ascending aorta, proximal to the BCT. - Total length from the coronaries to the proximal edge of the BCT must be a minimum of 70mm. - The length of the distal landing zone should be 20mm minimum. - Coverage of the left subclavian artery is permitted. Revascularization of the left subclavian artery may be considered in all cases by the treating physician and, especially, in anatomies where revascularization is determined to be clinically necessary - The distal landing zone must contain a straight segment (non-tapered, non-reverse tapered; defined by < 10% diameter change) with length equal to or greater than the required attachment length of the intended device - Non-aneurysmal BCT and left common carotid arteries with diameters ranging > 6.0mm across the entire length of the treatment zone. - Distal branch landing zone must be <20.0 mm in diameter and a minimum of 25.0 mm in length. - Adequate arterial access for introduction and delivery of the Relay Branch System. Note: Alternative methods to gain proper access can be utilized (e.g., iliac conduit) - Considered high risk or prohibitive risk for conventional surgery by treating physician or aortic team - Must be willing to comply with the follow-up evaluation schedule - Subject or legally authorized representative must sign the informed consent form prior to implant. Exclusion Criteria: - Significant stenosis, calcification, thrombus, or tortuosity of intended fixation sites that would compromise fixation or seal of the device - Aortic angulations (radius) less than 15mm at intended proximal landing zone. - Ascending aortas that would require the arch graft to be deployed less than 15mm distal to the coronaries - Pre-procedure occlusion or planned coverage of both subclavian arteries - Anatomic variants which would compromise circulation to both vertebral arteries after placement of the stent-graft - Prior endovascular repair in the ascending/descending thoracic aorta or aortic arch. The device may not be placed within any prior endovascular graft - Concomitant aneurysm/disease of the abdominal aorta requiring repair - Prior abdominal aortic aneurysm repair (endovascular or surgical) that was performed less than 6 months prior to the planned stent implant procedure - Prior mechanical aortic valve replacement except for hybrid valves with biological leaflets. - Major surgical or medical procedure within 45 days prior to the planned procedure or is scheduled for a major surgical or medical procedure within 60 days post implantation. Except for any planned procedures for the prospective stent-graft placement, e.g., left subclavian artery bypass or transposition - Untreatable allergy or sensitivity to contrast media or device components - Blood coagulation disorder or bleeding diathesis in which the treatment cannot be suspended for one week pre and post repair - Acute Coronary Syndrome (ACS) including unstable angina - Severe Congestive Heart Failure (New York Heart Association functional class IV) - Stroke within 12 months of the planned treatment date - Myocardial infarction within 3 months of the planned treatment date. - Chronic atrial fibrillation or other hypercoagulable condition that requires treatment with anticoagulants or presence of left atrial appendage thrombus - Severe pulmonary disease (documented FEVI <30% or as assessed by the study physician) at screening - Acute renal failure (defined as serum creatinine > 2.5mg/dL Note: Patients that have chronic renal failure and are managed medically or through hemodialysis, and can tolerate the follow-up imaging schedule can be included. - Significant carotid bifurcation disease (>70% diameter reduction by duplex ultrasound or angiography) - Hemodynamic instability - Active systemic infection at the time of treatment - Morbid obesity or other condition that may compromise or prevent the necessary imaging requirements - Connective tissue disorders, mycotic aneurysms, or infected aorta - Less than two-year life expectancy - Current or planned participation in an investigational drug or device study that has not completed primary endpoint evaluation - Currently pregnant or planning to become pregnant during the course of the study - Medical, social, or psychological issues that the Investigator believes may interfere with treatment or follow-up

Study Design


Intervention

Device:
penetrating atherosclerotic ulcer, aorta branch cardiovascular implant
The Relay Branch System is intended to provide an option for patients with arch and proximal descending chronic thoracic aortic pathologies.

Locations

Country Name City State
United States Emory University Atlanta Georgia
United States University of Maryland Baltimore Maryland
United States Massachusetts General Hospital Boston Massachusetts
United States Cleveland Clinic Cleveland Ohio
United States Duke University Durham North Carolina
United States University of Florida Gainesville Florida
United States Baylor College of Medicine Houston Texas
United States University of Southern California Los Angeles California
United States Columbia University Medical Center New York New York
United States Hospital of the University of Pennsylvania/Penn Presbyterian Philadelphia Pennsylvania
United States Baylor Scott and White Research Institute Plano Texas
United States Barnes Jewish Hospital at Washington University Saint Louis Missouri

Sponsors (1)

Lead Sponsor Collaborator
Bolton Medical

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary Composite of Major Adverse Events (MAEs) The primary safety outcome determined by the rate of MAEs, including
All-cause mortality
Disabling stroke: a Modified Rankin Score (mRS) of 2 or more at 90 days and an increase in at least one mRS category from an individual's pre-stroke baseline
Permanent Paralysis/Paraparesis
Spontaneous Myocardial Infarction according to the SCAI definition
Renal Failure
Procedural Blood Loss >1,000mL
30 days after the index procedure
Secondary Composite of the following Criteria Acute technical success evaluation, beginning with the insertion of the introducer sheath and as a composite of the following:
Success is defined as a successful delivery of the device through (i.e. ability to deliver the implant to the intended implantation site, without the need for unanticipated corrective intervention related to delivery):
Successful deployment of the device as defined as
i. Deployment of the endovascular stent-graft at the intended implantation site; ii. Patency (<50%) of all components of the implant with absence of device deformations requiring unplanned placement of an additional device within the endovascular stent-graft; and iii. absence of inadvertent covering of aortic branch vessels; and
c. Successful withdrawal of the delivery system
Patency of all endograft components at 30 days
Technical success will be evaluated in 2 stages; the first evaluation occurs during the implant procedure. Technical success is then re-evaluated again 30 days post operation to ensure the graft placement was successful as outlined above
See also
  Status Clinical Trial Phase
Recruiting NCT04149600 - Identification of Genetic Causes of Calcific Aortic Valve Disease
Enrolling by invitation NCT04035356 - HAART Annuloplasty Device Valve Repair Registry
Recruiting NCT02729064 - Intraoperative Nasal Insulin Effect on Plasma and CSF Insulin Concentration and Blood Glucose Phase 1
Completed NCT02467062 - Implementation of Non-size Markers Derived From 4D Flow MRI of Patients With Aortic Disease. N/A
Completed NCT02283307 - Dual Energy CT Urography With Reduced Iodinated Contrast N/A
Completed NCT01678261 - X-chromosome Inactivation, Epigenetics and the Transcriptome N/A
Terminated NCT01055275 - Cook Iliac Branch Graft Post-market Registry N/A
Completed NCT00615888 - Fast Track Management in Elective Open Infrarenal Aortic Aneurysm Repair N/A
Completed NCT03510793 - Microcirculation and Anesthesia in Vascular Surgery
Completed NCT03207568 - RE-GENERATION: The Safety and Performance of the Relay Pro and Relay NBS Pro Stent-graft Devices in the European Union (EU) N/A
Recruiting NCT05073991 - Incidence of Mortality and Complications After Lung Surgery, Open Thoracic Aortic Repair, TEVAR, EVAR.
Not yet recruiting NCT01918969 - Reference Values of Circulating Endothelial Progenitor Cells N/A
Completed NCT02000544 - Clinical Evaluation of a Modular Extracorporeal Circulation Circuit N/A
Not yet recruiting NCT01918982 - Circulating Endothelial Progenitor Cells and Aortic Aneurysm N/A
Completed NCT01256372 - An Trial of Two Dosing Regimens of AP214 for the Prevention of Kidney Injury in Patients Undergoing Cardiac Surgery Phase 2
Recruiting NCT00661518 - PET/CT Imaging of Aneurysm Wall Inflammation N/A
Completed NCT00094575 - Standard Open Surgery Versus Endovascular Repair of Abdominal Aortic Aneurysm (AAA) Phase 4
Recruiting NCT04471909 - NEXUS Aortic Arch Clinical Study to Evaluate Safety and Effectiveness N/A
Recruiting NCT05603520 - Phenotyping Heterogeneity and Regionality of the Aorta
Completed NCT03727542 - Influence of Short AV Delay Permanent Pacing on Matrix Metalloproteinase Levels N/A