Clinical Trial Summary

The aim of this prospective study is to evaluate the clinical performance of habitual wearers of methafilcon A toric lenses after a refit with fanfilcon A toric lenses for 1-month of daily wear.


Recruitment information / eligibility

Status Completed
Enrollment 43
Est. completion date July 6, 2019
Est. primary completion date March 27, 2019
Accepts healthy volunteers No
Gender All
Age group 18 Years to 40 Years
Eligibility Inclusion Criteria:

- Is between 18 and 40 years of age (inclusive)

- Has had a self-reported visual exam in the last two years

- Is an adapted soft contact lens wearer

- Has a contact lens spherical prescription between +6.00 to - 8.00 (inclusive)-Have no less than -0.75D of astigmatism and no more than -2.25 D in both eyes

- Can achieve best corrected spectacle distance visual acuity of 20/25 (0.10 logMAR) or better in each eye.

- Can achieve a distance visual acuity of 20/30 (0.18 logMAR) or better in each eye with the study contact lenses.

- Has clear corneas and no active ocular disease

- Has read, understood and signed the information consent letter

- Patient contact lens refraction should fit within the available parameters of the study lenses

- Is willing to comply with the wear schedule (at least 5 days per week, >8 hours/day assuming there are no contraindications for doing so)

- Is willing to comply with the visit schedule

Exclusion Criteria:

- Has a Contact Lens (CL) prescription outside the range of the available parameters of the study lenses

- Has a spectacle cylinder less than -0.75D or more than -2.50 D of cylinder in either eye

- Has a history of not achieving comfortable Contact Lens (CL) wear (5 days per week; > 8 hours/day)

- Has contact lens best corrected distance vision worse than 20/25 (0.10 logMAR) in either eye

- Presence of clinically significant (grade 2-4) anterior segment abnormalities

- Presence of ocular or systemic disease or need of medications which might interfere with contact lens wear

- Slit lamp findings that would contraindicate contact lens wear such as:

- Pathological dry eye or associated findings

- Pterygium, pinguecula, or corneal scars within the visual axis

- Neovascularization > 0.75 mm in from of the limbus

- Giant papillary conjunctivitis (GCP) worse than grade 1

- Anterior uveitis or iritis (past or present)

- Seborrheic eczema, Seborrheic conjunctivitis

- History of corneal ulcers or fungal infections

- Poor personal hygiene

- Has a known history of corneal hypoesthesia (reduced corneal sensitivity)

- Has aphakia, keratoconus or a highly irregular cornea

- Has Presbyopia or has dependence on spectacles for near work over the contact lenses

- Has undergone corneal refractive surgery

- Is participating in any other type of eye related clinical or research study.

Study Design


Related Conditions & MeSH terms


Intervention

Device:
methafilcon A toric contact lenses
Bilateral daily wear of methafilcon A toric contact lenses
fanfilcon A toric contact lenses
Bilateral daily wear of fanfilcon A toric contact lenses

Locations

Country Name City State
Mexico Optometry Clinic, National Autonomous University Mexico City

Sponsors (1)

Lead Sponsor Collaborator
Coopervision, Inc.

Country where clinical trial is conducted

Mexico, 

Outcome

Type Measure Description Time frame Safety issue
Primary Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable). Baseline
Primary Number of Participants With Lens Centration Assessment by the Investigator - Methafilcon A Toric Contact Lenses Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable). 4 weeks
Primary Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable). Baseline
Primary Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable). 2 weeks
Primary Number of Participants With Lens Centration Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Lens centration was assessed on a 3-point scale (1=optimal centration, 2=decentration acceptable, 3=decentration unacceptable). 4 weeks
Primary Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea Baseline
Primary Percentage of Participants With Corneal Coverage Assessment by the Investigator - Methafilcon A Toric Contact Lenses Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea 4 weeks
Primary Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea Baseline
Primary Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea 2 weeks
Primary Percentage of Participants With Corneal Coverage Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Corneal coverage was collected on a YES/NO response. Yes - lens covers the cornea, No - lens does not cover the cornea 4 weeks
Primary Post-blink Movement Assessment by the Investigator - Methafilcon A Toric Contact Lenses Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement). Baseline
Primary Post-blink Movement Assessment by the Investigator - Methafilcon A Toric Contact Lenses Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement). 4 weeks
Primary Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement). Baseline
Primary Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement). 2 weeks
Primary Post-blink Movement Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Post-blink movement was assessed on a 0-4 scale (where: 0=Insufficient, unacceptable movement,1=Minimal, but acceptable movement, 2=Optimal movement, 3=Moderate, but acceptable movement, 4=Excessive, unacceptable movement). 4 weeks
Primary Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Methalfilcon A Toric Contact Lenses Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position. Baseline
Primary Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Methalfilcon A Toric Contact Lenses Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position. 4 weeks
Primary Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position. Baseline
Primary Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position. 2 weeks
Primary Lens Orientation in Primary Position of Gaze Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Lens orientation was assessed in 5-degree steps with the slit lamp by measuring the amount of mislocation of the axis mark on the lens relative to the 6 o'clock position. 4 weeks
Primary Rotational Recovery Assessment by the Investigator - Methafilcon A Toric Contact Lenses Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position Baseline
Primary Rotational Recovery Assessment by the Investigator - Methafilcon A Toric Contact Lenses Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position 4 weeks
Primary Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position Baseline
Primary Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position 2 weeks
Primary Rotational Recovery Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Lens rotational recovery was assessed in 5-degree steps 1 minute after the lens was manually rotated 45degrees temporally from the primary gaze position 4 weeks
Primary Overall Stability Assessment by the Investigator - Methafilcon A Toric Contact Lenses Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability). Baseline
Primary Overall Stability Assessment by the Investigator - Methafilcon A Toric Contact Lenses Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability). 4 weeks
Primary Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability). Baseline
Primary Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability). 2 weeks
Primary Overall Stability Assessment by the Investigator - Fanfilcon A Toric Contact Lenses Overall lens stability was assessed using a 0 - 4 scale (where 0 = very poor stability, and 4 = excellent stability). 4 weeks
Primary Overall Lens Fit Acceptance by the Investigator - Methafilcon A Toric Contact Lenses Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect. Baseline
Primary Overall Lens Fit Acceptance by the Investigator - Methafilcon A Toric Contact Lenses Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect. 4 weeks
Primary Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect. Baseline
Primary Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect. 2 weeks
Primary Overall Lens Fit Acceptance by the Investigator - Fanfilcon A Toric Contact Lenses Overall lens fit acceptance assessed using grade (0-4) 0=Should not be worn, 1=Borderline but unacceptable, 2=Minimal acceptable, early review, 3=Not perfect but okay to dispense, 4=perfect. 4 weeks
Secondary Average Daily Wearing Time - Methafilcon A Toric Contact Lenses Average daily wearing time response by the subject 4 weeks
Secondary Average Daily Wearing Time - Fanfilcon A Toric Contact Lenses Average daily wearing time response by the subject 2 weeks
Secondary Average Daily Wearing Time - Fanfilcon A Toric Contact Lenses Average daily wearing time response by the subject 4 weeks
Secondary Average Comfortable Wearing Time - Methafilcon A Toric Contact Lenses Average comfortable wearing time - response by the subject 4 weeks
Secondary Average Comfortable Wearing Time - Fanfilcon A Toric Contact Lenses Average comfortable wearing time - response by the subject 2 weeks
Secondary Average Comfortable Wearing Time - Fanfilcon A Toric Contact Lenses Average comfortable wearing time - response by the subject 4 weeks
See also
  Status Clinical Trial Phase
Completed NCT04532099 - Clinical Comparison of Two Daily Wear Frequent Replacement Silicone Hydrogel Soft Contact Lenses N/A
Completed NCT03319212 - Clinical Characterization of Symptomatic Populations N/A
Withdrawn NCT04525170 - Clinical Evaluation of HPT Treated Rigid Contact Lenses Made From Hexafocon A N/A
Completed NCT05483127 - Clinical Comparison of Two Daily Disposable Toric Soft Contact Lenses N/A
Completed NCT01741987 - Osmoprotective Containing Lubricants in Dysfunctional Tear Syndrome and Post Refractive Surgery Patients N/A
Completed NCT03688672 - Apioc Contact Lens Feasibility N/A
Completed NCT02312323 - One-month Clinical Evaluation of Silicone Hydrogel Definitive 65 HPT Contact Lenses N/A
Recruiting NCT04714424 - Comparing Vision Tests in a Virtual Reality Headset to Existing Analogues
Recruiting NCT04632784 - Clinical Trial With Artiflex Presbyopic N/A
Completed NCT04185701 - Comparison of Subjective Refraction Measurement With SiVIEW Software and by an Expert N/A
Completed NCT01706770 - Evaluation of Enfilcon A Soft Contact Lens When Compared to Galyfilcon A Soft Contact Lens N/A
Completed NCT03556579 - The Effects of Contact Lenses With Experimental Dye on Visual Function N/A
Completed NCT04654455 - Prospective Study Evaluating Dry Eye in Patients Operated on for Ametropia by LASIK Surgery and Treated With REPADROP
Not yet recruiting NCT06295536 - Optimization of a Photoretinoscopy Method for Determining the Objective Refraction of Children N/A
Completed NCT00691197 - Safety and Acceptability of Using a Rewetting Drop With Contact Lens Wear Phase 2/Phase 3
Completed NCT00597467 - Study of Soft Contact Lens Use With 7 Day Extended Wear N/A
Completed NCT04067141 - The Clinical Comparison of Somofilcon A 1 Day and Nelfilcon A Daily Disposable Contact Lenses N/A
Completed NCT04067050 - A Clinical Study of the Comfilcon A Asphere Soft Contact Lens in Users of Digital Devices N/A
Completed NCT02553681 - Clinical Evaluation of HPT Treated Rigid Contact Lenses Made From Roflufocon D N/A
Completed NCT00722891 - Ocular Responses to Short and Long-term Lens Wear N/A