Clinical Trials Logo

Clinical Trial Summary

Objective: to gain experience with in-home use of a modified algorithm that will dose insulin to minimize projected hyperglycemia overnight in addition to suspending the pump if hypoglycemia is projected overnight and to obtain feasibility, safety, and initial efficacy data.

Study Design: randomized controlled trial, with randomization on a night level within subject.

Major Eligibility Criteria: clinical diagnosis of type 1 diabetes, daily insulin therapy for at least one year and an insulin infusion pump for at least 6 months; 15.0 to <46.0 years of age; HbA1c < 10.0%; no DKA in last 6 months; no hypoglycemic seizure or loss of consciousness in last 6 months; Living with a significant other or family member ("companion") committed to participating in all study activities, and being present and available to provide assistance when the system is being used at night.

Sample Size: 30 subjects.

Study Duration and Visit Schedule: duration approximately 3 months, with preliminary run-in activities followed by up to 90 days spent in clinical trial phase of study; clinic visits at enrollment, following CGM and system assessment run-in phases, at start of clinical trial phase, at 21-day point of clinical trial phase, and after 42 nights of successful system use.

Major Efficacy Outcomes:

- Primary: time in range (70-180 mg/dl, 3.9-10.0 mmol/L) overnight.

- Secondary: time spent in hypoglycemia (≤70 mg/dl, 3.9 mmol/L) and time spent in hyperglycemia (>180 mg/dl, 10.0 mmol/L) overnight.

Major Safety Outcomes: CGM measures of hypo- and hyperglycemia, including morning blood glucose and mean overnight sensor glucose; adverse events including severe hypoglycemia and diabetic ketoacidosis.


Clinical Trial Description

Subjects who are eligible for the clinical trial initially will use a Veo insulin pump and Enlite 2 CGM sensor at home for a minimum of 6 days/week over a 2-week period to verify that the subject is able to use the CGM and insert sensors.

The first 10 subjects enrolled will participate in a 2-day overnight hotel-based pilot, collecting a total of 20 nights of data and experience with the system in a transitional hotel setting. The DSMB will review safety data from the 20 night hotel-based pilot study and make recommendations regarding proceeding to the at-home portion of the study.

Following DSMB review and approval of the safety data from the hotel study, the first 10 subjects will participate in an Algorithm Assessment Phase of approximately 10 nights of Predictive Low Glucose Suspend (PLGS) + Hyper Minimization system use each (for a nominal total of 100 nights of use at home) to determine if any adjustments to the algorithm parameters are needed and if it is safe to advance to the randomized clinical trial phase. If adjustments are needed, the Algorithm Assessment Phase will be repeated, using the same 10 subjects if possible. Once the randomized clinical trial phase begins, approximately 200 nights of randomized system use will be collected and assessed for safety by the DSMB before proceeding.

New subjects who enroll in the study after the completion of the Algorithm Assessment Phase will use the PLGS+Hyper Minimization closed-loop system at home for at least 5 days to demonstrate their ability to use the system and submit study data to the Coordinating Center.

Subjects who successfully demonstrate their ability to use the system at home as described above will be eligible for the randomized trial phase. This phase consists of use of the full system in the home for approximately 42 nights:

- Each night the blood glucose level will be checked with the BG meter and used to perform a calibration of the CGM. This calibration must occur no more than 90 minutes prior to activation of the system. NOTE: Subjects will be instructed to calibrate the CGM per manufacturer guidelines.

- Then the system will be activated, linking the CGM and insulin pump to the computer at the bedside.

- A randomization schedule on the laptop will be used to determine whether the system will run in PLGS+Hyper Minimization mode or PLGS-only.

- Subjects will be blinded as to whether the system is running in PLGS+Hyper Minimization mode or PLGS-only.

- There will not be an alarm if the pump shuts off or if automated insulin dosing occurs. The CGM alarm will be set to 60 mg/dL (3.3 mmol/L). When a CGM alarm occurs, the subject will be asked to measure the blood glucose with a BG meter, if he/she is aware of the alarm.

- The time period for outcome assessment each night will be from the time the system is activated until it is turned off in the morning.

- Pump shut off, when it occurs, will be for up to 120 minutes in a 150-minute period, and no more than 180 minutes for the entire night. Multiple instances of pump suspension can occur if there are recurrent predictions of hypoglycemia during the night.

- Small correction boluses of insulin will be delivered when the system predicts that hyperglycemia above a pre-set threshold will occur, with insulin-on-board constraints and cumulative delivery limits to minimize the likelihood of excessive insulin delivery.

- Subjects will be asked to check blood glucose with the study BG meter each morning prior to breakfast and enter the results using the controller software interface. The subject will be instructed to contact the study physician if the morning blood glucose value is <60 mg/dl (3.3 mmol/L) or >300 mg/dl (16.7 mmol/L). Monitoring processes detailed in section 3.11 will ensure that the subject can be contacted if these values are not reported as required or are out of range.

- Subjects will be asked to record all overnight carbohydrate intake using the controller software interface.

- Subjects will be asked to perform periodic CGM data uploads using the controller software interface. Monitoring processes will ensure that the subject can be contacted if these uploads do not occur as required, or if review of an upload reveals any extreme, prolonged episodes of hypoglycemia or hyperglycemia, or elevated morning blood glucose values.

Upon completion of the study, subjects as well as study clinicians will be asked to complete a human factors usability questionnaire regarding use of the study system. ;


Study Design

Allocation: Randomized, Endpoint Classification: Safety/Efficacy Study, Intervention Model: Parallel Assignment, Masking: Open Label, Primary Purpose: Prevention


Related Conditions & MeSH terms


NCT number NCT02438189
Study type Interventional
Source Jaeb Center for Health Research
Contact
Status Completed
Phase Phase 2
Start date June 2015
Completion date March 2016

See also
  Status Clinical Trial Phase
Recruiting NCT05653518 - Artificial Pancreas Technology to Reduce Glycemic Variability and Improve Cardiovascular Health in Type 1 Diabetes N/A
Enrolling by invitation NCT05515939 - Evaluating the InPen in Pediatric Type 1 Diabetes
Completed NCT05109520 - Evaluation of Glycemic Control and Quality of Life in Adults With Type 1 Diabetes During Continuous Glucose Monitoring When Switching to Insulin Glargine 300 U/mL
Recruiting NCT04016987 - Automated Structured Education Based on an App and AI in Chinese Patients With Type 1 Diabetes N/A
Active, not recruiting NCT04190368 - Team Clinic: Virtual Expansion of an Innovative Multi-Disciplinary Care Model for Adolescents and Young Adults With Type 1 Diabetes N/A
Recruiting NCT05413005 - Efficacy of Extracorporeal Photopheresis (ECP) in the Treatment of Type 1 Diabetes Mellitus Early Phase 1
Active, not recruiting NCT04668612 - Dual-wave Boluses in Children With Type 1 Diabetes Insulin Boluses in Children With Type 1 Diabetes N/A
Completed NCT02837094 - Enhanced Epidermal Antigen Specific Immunotherapy Trial -1 Phase 1
Recruiting NCT05414409 - The Gut Microbiome in Type 1 Diabetes and Mechanism of Metformin Action Phase 2
Recruiting NCT05670366 - The Integration of Physical Activity Into the Clinical Decision Process of People With Type 1 Diabetes N/A
Active, not recruiting NCT05418699 - Real-life Data From Diabetic Patients on Closed-loop Pumps
Completed NCT04084171 - Safety of Artificial Pancreas Therapy in Preschoolers, Age 2-6 N/A
Recruiting NCT06144554 - Post Market Registry for the Omnipod 5 System in Children and Adults With Type 1 Diabetes
Recruiting NCT05153070 - Ciclosporin Followed by Low-dose IL-2 in Patients With Recently Diagnosed Type 1 Diabetes Phase 2
Recruiting NCT05379686 - Low-Dose Glucagon and Advanced Hybrid Closed-Loop System for Prevention of Exercise-Induced Hypoglycaemia in People With Type 1 Diabetes N/A
Completed NCT05281614 - Immune Effects of Vedolizumab With or Without Anti-TNF Pre-treatment in T1D Early Phase 1
Withdrawn NCT04259775 - Guided User-initiated Insulin Dose Enhancements (GUIDE) to Improve Outcomes for Youth With Type 1 Diabetes N/A
Active, not recruiting NCT01600924 - Study on the Assessment of Determinants of Muscle and Bone Strength Abnormalities in Diabetes
Completed NCT02897557 - Insulet Artificial Pancreas Early Feasibility Study N/A
Completed NCT02914886 - Beneficial Effect of Insulin Glulisine by Lipoatrophy and Type 1 Diabetes (LAS) Phase 4