Clinical Trial Summary

The purpose of the study is to test the accuracy benefit of having two glucose sensors (over one sensor alone) when they are positioned: 2mm, 10mm, 20mm, or 30mm apart. It is not yet known how close two sensors can be and still work correctly.


Description:

An artificial pancreas system will likely require multiple glucose sensing elements to function safely. It is not yet known how close two glucose sensors can be placed and still work properly. Meaning, if one is reading inaccurately,and the other is positioned adjacent to it, will it read inaccurately also, defeating the purpose of having two sensors. Twenty adult subjects with Type 1 diabetes will wear four Medtronic REAL-Time glucose sensors during two separate 10-hour studies. The inter-sensor distances of each pair to be tested will be: 2, 10, 20, and 30mm apart. The sensors will be inserted the day prior to the study day to allow for signal stabilization and calibration. During the study day, arterialized venous blood will be drawn every 15 minutes to measure blood glucose. The sensor values and interstitial values will be recorded from each sensor receiver at each time point. Subjects will be fed two standardized meals and given pre-meal insulin based on their typical insulin to carbohydrate ratio.


Recruitment information / eligibility

Status Completed
Enrollment 20
Est. completion date January 2011
Est. primary completion date January 2011
Accepts healthy volunteers No
Gender Both
Age group 18 Years to 65 Years
Eligibility Inclusion Criteria:

1. Type 1 diabetes mellitus diagnosed for at least 6 months

2. Current usage of multiple daily injections or subcutaneous insulin pump treatment

3. Age 18-65 years

4. HbA1c of 6.0 - 9.5% at screening visit

5. BMI 21-35

6. Willingness to follow all study procedures, including attending all clinic visits

7. Willingness to sign informed consent and HIPAA documents

Exclusion Criteria: Subjects presenting with any of the following will not be included in the study.

1. Pregnancy or Lactation: for women of childbearing potential there is a requirement for a negative urine pregnancy test.

2. Renal insufficiency (serum creatinine of 2.0 mg/dL or greater). Serum ALT or AST equal to or greater than 3 times the upper limit of normal; hepatic synthetic insufficiency as defined as a serum albumin of less the 3.3 g/dL; or serum bilirubin of over 2.

3. Hematocrit of less than or equal to 34%.

4. Congestive heart failure, NYHA class III or IV.

5. Visual impairment preventing reading of glucose meter values.

6. Active coronary artery disease as manifested by unstable angina, or a myocardial infarction or therapeutic coronary percutaneous procedure (e.g., PTCA, stent placement) or CABG within the past 4 months.

7. Active foot ulceration.

8. Severe peripheral arterial disease characterized by ischemic rest pain or severe claudication.

9. Cerebrovascular accident within the past 6 months.

10. Active alcohol abuse, substance abuse, or severe mental illness (as judged by the principal investigator).

11. Active malignancy, except basal cell or squamous cell skin cancers.

12. Major surgical operation within 30 days prior to screening.

13. Seizure disorder (epilepsy).

14. Any concurrent illness, other than diabetes, that is not controlled by a stable therapeutic regimen.

15. Currently administration of oral or parenteral corticosteroids.

16. Use of an investigational drug within 30 days prior to screening.

17. Bleeding disorder, treatment with warfarin, or platelet count below 50,000.

Study Design

Observational Model: Case-Only, Time Perspective: Prospective


Related Conditions & MeSH terms


Locations

Country Name City State
United States Oregon Health & Science University Portland Oregon

Sponsors (2)

Lead Sponsor Collaborator
Oregon Health and Science University The Leona M. and Harry B. Helmsley Charitable Trust

Country where clinical trial is conducted

United States, 

References & Publications (3)

Castle JR, Engle JM, El Youssef J, Massoud RG, Yuen KC, Kagan R, Ward WK. Novel use of glucagon in a closed-loop system for prevention of hypoglycemia in type 1 diabetes. Diabetes Care. 2010 Jun;33(6):1282-7. doi: 10.2337/dc09-2254. Epub 2010 Mar 23. — View Citation

Schmidtke DW, Pishko MV, Quinn CP, Heller A. Statistics for critical clinical decision making based on readings of pairs of implanted sensors. Anal Chem. 1996 Sep 1;68(17):2845-9. — View Citation

Ward WK, Casey HM, Quinn MJ, Federiuk IF, Wood MD. A fully implantable subcutaneous glucose sensor array: enhanced accuracy from multiple sensing units and a median-based algorithm. Diabetes Technol Ther. 2003;5(6):943-52. — View Citation

Outcome

Type Measure Description Time frame Safety issue
Primary Difference between very large sensor accuracy errors (=50% from reference glucose) with the simultaneous use of four sensors vs one sensor. 9-hour collection period, X 2 No
Secondary Difference between mean absolute relative difference with the simultaneous use of four sensors vs one sensor. 9-hour glucose level collection, X 2 No
See also
  Status Clinical Trial Phase
Completed NCT04476472 - Omnipod Horizon™ Automated Glucose Control System Preschool Cohort N/A
Completed NCT03635437 - Evaluation of Safety and Diabetes Status Upon Oral Treatment With GABA in Patients With Longstanding Type-1 Diabetes Phase 1/Phase 2
Completed NCT04909580 - Decision Coaching for Youth and Parents Considering Insulin Delivery Methods for Type 1 Diabetes N/A
Active, not recruiting NCT00679042 - Islet Transplantation in Type 1 Diabetic Patients Using the University of Illinois at Chicago (UIC) Protocol Phase 3
Completed NCT03293082 - Preschool CGM Use and Glucose Variability in Type 1 Diabetes N/A
Completed NCT04016662 - Automated Insulin Delivery in Elderly With Type 1 Diabetes (AIDE T1D) Phase 4
Completed NCT02527265 - Afrezza Safety and Pharmacokinetics Study in Pediatric Patients Phase 2
Completed NCT03738865 - G-Pen Compared to Glucagen Hypokit for Severe Hypoglycemia Rescue in Adults With Type 1 Diabetes Phase 3
Completed NCT03240432 - Wireless Innovation for Seniors With Diabetes Mellitus N/A
Completed NCT03168867 - Effectiveness Trial of an E-Health Intervention To Support Diabetes Care in Minority Youth (3Ms) N/A
Completed NCT03674281 - The VRIF Trial: Hypoglycemia Reduction With Automated-Insulin Delivery System N/A
Completed NCT03669770 - Ultrasound Classification and Grading of Lipohypertrophy and Its Impact on Glucose Variability in Type 1 Diabetes
Recruiting NCT03682640 - Azithromycin Insulin Diet Intervention Trial in Type 1 Diabetes Phase 2
Recruiting NCT04096794 - Chinese Alliance for Type 1 Diabetes Multi-center Collaborative Research
Completed NCT02882737 - The Impact of Subcutaneous Glucagon Before, During and After Exercise a Study in Patients With Type 1 Diabetes Mellitus N/A
Recruiting NCT02745808 - Injectable Collagen Scaffold™ Combined With HUC-MSCs for the Improvement of Erectile Function in Men With Diabetes Phase 1
Completed NCT02562313 - A Trial Investigating the Continuous Subcutaneous Insulin Infusion of a Liquid Formulation of BioChaperone Insulin Lispro in Comparison to Humalog® Phase 1
Withdrawn NCT02518022 - How to be Safe With Alcoholic Drinks in Diabetes N/A
Completed NCT02558491 - Feasibility of a Decision Support System to Reduce Glucose Variability in Subject With T1DM N/A
Completed NCT02596204 - Diabetes Care Transformation: Diabetes Data Registry and Intensive Remote Monitoring N/A