Continuous distending airway pressure (CPAP) has gained popularity as a means to provide non-invasive respiratory support in neonates to reduce ventilator induced lung injury (VILI). However, maintaining CPAP in preterm infants has been challenging, often related to issues with nasal interface such as nasal septal injury, problems with keeping the prongs in the nose and leak around the prongs with are important factors in proving effective CPAP. RAM cannula was recently approved for providing supplemental oxygen and soon adapted by many centers to provide CPAP in preterm infants. Concerns have been raised that RAM cannula interface results in sub-optimal pressure delivery compared to standard nasal prongs.


Clinical Trial Description

Objective: To compare efficacy of delivering continuous distending pressure in preterm neonates using RAM cannula versus the standard prongs (Hudson prongs) by measuring intra-oral pressure. It is hypothesize that CPAP delivered with RAM cannula is comparable to pressure delivered via standard nasal interface in routine clinical practice; and the desired CPAP level is often not achieved in presence of variable leak from mouth irrespective of type of nasal interface. Aim 1: Compare the level of CPAP in oral cavity between RAM cannula and Hudson prong nasal interface in preterm Hypothesis: The intraoral pressure with CPAP will not be different with use of RAM cannula or Hudson prong nasal interface Aim 2: Evaluate the effect of nasal interface (RAM cannula and Hudson prong) on respiratory variables Hypothesis: There will be no difference in respiratory variables with CPAP delivered by RAM cannula compared to Hudson prongs. Method: This was a single center prospective study. Inclusion criteria: Any preterm with respiratory distress requiring CPAP but not mechanical ventilation or NIPPV was eligible for the study. 12 preterm infant more than 48 hours old and stable on bubble CPAP were recruited for the study Exclusion criteria: Infants with major congenital anomalies, neuromuscular disorders, upper airway anomalies, critically sick infants CPHS approval and parental consent was obtained prior to enrollment in the study. Each infant was studied for ~1 hour during which the oral pressure was measured while receiving CPAP with RAM nasal cannula interface and then with Hudson prongs. Blinding of the intervention (nasal interface) was not feasible because of the visible differences in the two nasal interfaces. Procedure: Intra-oral pressure measured via a differential pressure transducer while infant in receiving CPAP, and digitized data recorded with each nasal interface ( goal to record at least 20 sec of stable waveform on each nasal interface). Standard care provided during the study period and no changes made with CPAP level or flow. The difference between set pressure and measured pressure with RAM cannula and Hudson prong was compared with standard paired student t-test with 95% confidence limits. Difference in other variables will be similarly compared with t-test. STAT 12 will be used for statistical analysis. ;


Study Design


Related Conditions & MeSH terms


NCT number NCT03212508
Study type Observational
Source Dartmouth-Hitchcock Medical Center
Contact
Status Completed
Phase
Start date August 1, 2016
Completion date March 22, 2017

See also
  Status Clinical Trial Phase
Recruiting NCT04016246 - Respiratory Effect of the LISA Method With Sedation by Propofol Versus Absence of Sedation. Phase 3
Enrolling by invitation NCT04118400 - Ventilator Weaning Outcome Between NIV-NAVA and Nasal CPAP (or IMV ) Modes in Premature Neonates
Completed NCT01941745 - Efficacy of Recombinant Human Clara Cell 10 Protein (rhCC10) Administered to Premature Neonates With Respiratory Distress Syndrome Phase 2
Not yet recruiting NCT04019106 - Budesonide With Intratracheal Surfactants in Extremely Preterm Infants Phase 1/Phase 2
Not yet recruiting NCT04862377 - Intratracheal Budesonide With Surfactant to Prevent Bronchopulmmonary Dysplasia. Phase 3
Not yet recruiting NCT05791331 - REspiratory MEchanics for Delivering Individualised Exogenous Surfactant N/A
Completed NCT04086095 - Feasibility Study - Neofact N/A
Not yet recruiting NCT05609877 - The NONA-LISA Trial N/A
Recruiting NCT04326270 - Crossover Comparison of Tidal Volume Delivery During Nasal Intermittent Positive Pressure Ventilation in Preterm Infants: Infant Cannula vs. Nasal Continuous Positive Airway Pressure Prongs N/A
Active, not recruiting NCT03969992 - A Dose-Ranging Study to Determine the Efficacy, Safety and Tolerability of AeroFact Phase 2
Completed NCT03700606 - Physiological Changes With High-Flow Nasal Cannula N/A
Recruiting NCT04359134 - Combined Lung Ultrasounds and Transthoracic Electrical Bioimpedance in Preterm Infants With Respiratory Distress.
Not yet recruiting NCT06229821 - Safety and Efficacy of Intrapulmonary Percussive Ventilation in Preterm Infants N/A
Recruiting NCT05446389 - PAL to Improve Oral Feeding for Infants With Chronic Lung Disease N/A
Withdrawn NCT04914715 - nHFOV Versus Invasive Conventional Ventilation for Preterm Neonates With Respiratory Distress Syndrome N/A
Recruiting NCT03825835 - 30% or 60% Oxygen at Birth to Improve Neurodevelopmental Outcomes in Very Low Birthweight Infants N/A
Recruiting NCT03562182 - Effect of Steroid Administration on Maternal Blood Levels of hLPCAT1 mRNA
Completed NCT05031650 - Open Lung Strategy During Non-Invasive Respiratory Support of Very Preterm Infants in the Delivery Room N/A
Completed NCT03306524 - The Role of Circuit Flow During Mechanical Ventilation of Neonates N/A
Completed NCT01473264 - Safety, Pk and Anti-inflammatory Effects of CC10 Protein in Premature Infants With Respiratory Distress Syndrome (RDS) Phase 1/Phase 2