Clinical Trial Summary
Clinical validation of the wrist blood pressure measuring device Combei BP 880 W for home use
according to "the Universal Standard for the validation of blood pressure measuring devices
by the Association for the Advancement of Medical Instrumentation/European Society of
Hypertension/International Organization for Standardization (AAMI/ESH/ISO)" in the general
population.
Description:
Members of the AAMI, ESH and ISO committees achieved a consensus on an optimal validation
standard and agreed on a single protocol for the validation of BP measuring devices that will
replace all previous ones.
Therefore, the Association for the Advancement of Medical Instrumentation / European Society
of Hypertension / International Organization for Standardization Universal Standard
(AAMI/ESH/ISO 81060-2:2018) is now considered as the standard protocol for the validation of
non-invasive blood pressure measuring device.
This study aims to demonstrate that the Combei BP 880 W wrist device fulfills the
requirements of the new protocol of clinical validation and its Amendment 1, 2020-01 for limb
circumference distribution criteria.
this is a monocentric, prospective, non-randomized, non-interventional, cross-study.
Combei BP 880 W wrist device (Sphygmomanometer Under Test, SUT). The reference device is the
mercury sphygmomanometer The primary objective of the study is to assess the accuracy of the
automatic oscillometric BP measuring device at the wrist level, the Combei BP 880 W, in the
general population, following AAMI/ESH/ISO 81060-2:2018 protocol and its Amendment 1, 2020-01
for limb circumference distribution criteria.
The Combei BP 880 W device will pass the test if:
1. The differences of at least 255 individual paired determinations of the SUT and of the
observers' readings with the reference sphygmomanometer for all subjects, calculated
separately for systolic blood pressure and diastolic blood pressure, shall have the mean
value of the differences, be within or equal to +/- 5.0mmHg and a standard deviation no
greater than 8.0 mmHg.
2. For the systolic blood pressure and diastolic blood pressure of at least 85 subjects,
the standard deviation of the averaged paired determinations per subject of the SUT and
of the observers' readings with the reference sphygmomanometer shall meet the criteria
listed in Table 1 of AAMI/ESH/ISO 81060-2:2018.
The secondary objective is to confirm the safety of use of this device, intended for home
use.
Inclusion criteria
- Patient older than 12 years;
- Patient who signed the informed consent form;
- Patient followed-up at site (in-patient or out-patient);
- Patient with wrist circumference between 12.5 and 21.5 cm*. *in accordance with the
Indications For Use of the Combei BP 880 W device.
Non-inclusion criteria
- Patient unable to give a consent or understand properly protocol information;
- Patient suffering from arrhythmia;
- Patient with poor quality of Korotkov sounds;
- Patient for whom K5 sounds are absent. The validation team will consist of three
persons: two observers trained in accurate BP measurement and a supervisor.
1. The reference sphygmomanometer is calibrated just before the beginning of the
study.
2. The observers are trained to the use of the SUT.
3. In-patients or out-patients are included in the study after reading and signing the
ICF.
4. After inclusion, the patient is screened to verify that he/she satisfies the
inclusion/non-inclusion criteria of the protocol. If not, the patient will be
excluded from the study and no blood pressure measure will be performed on this
patient.
5. The patient is placed comfortably, and the reference sphygmomanometer is placed on
his/her left upper arm, except if left upper arm is wounded. In this case, the
reference sphygmomanometer is placed on his/her right upper arm.
6. Measurements start with alternance between measurements with the reference
sphygmomanometer and the SUT according to the same arm sequential BP measurements
method.
7. For each measurement with the reference sphygmomanometer, the two observers using
mercury sphygmomanometers and a double-headed stethoscope, blinded from each
other's result, and then by the supervisor using the tested device according to the
manufacturer recommendations. The diastolic blood pressure should be measured at
Korotkoff K5 sound. The agreement between the 2 observers will be checked all over
the evaluation period by the supervisor to make sure that the difference between
the two observers is no more than 4 mmHg for systolic and diastolic BP
values.Otherwise, the measurement should be repeated. The intra individual BP
variability for the reference BP measurements must be < 12 mmHg for SBP and < 8
mmHg for DBP.
8. The results are written down on a source document and then reported in the CRF. Any
AE, ADE or device malfunction occurring during the protocol will be also documented
and reported in the CRF. For this study, all source documents should be written in
English language.
9. At the end of the measurements, the patient participation to the study is over.
Study procedures:
1. Validation team and pre-study training
The validation team will consist of three persons: two observers trained in accurate BP
measurement and a supervisor.
The reference sphygmomanometer is calibrated just before the beginning of the study.
The observers are trained to the use of the SUT.
2. Inclusion of the patients
In-patients or out-patients of the investigational site are included in the study after
reading and signing the ICF. After inclusion, the patient is screened to verify that
he/she satisfies the inclusion/non-inclusion criteria of the protocol. If not, the
patient will be excluded from the study and no blood pressure measure will be performed
on this patient.
3. Cuffs
The cuffs used for reference auscultatory BP measurement will have an inflatable bladder
length that covers 75-100% of the arm circumference of each patient and width that
covers 37-50% of the arm circumference.
Detailed description of cuffs used for reference BP measurement will be reported for
each patient.
4. Blood pressure measurements
The validation will be performed according to the (AAMI/ESH/ISO) Universal Standard
validation protocol described in detail elsewhere10. The patients will have their reference
BP measured by the two observers using a mercury sphygmomanometer and a double-headed
stethoscope, blinded from each other's result, and then by the supervisor using the tested
devices. The agreement between the 2 observers will be checked all over the evaluation period
by the supervisor to make sure that the difference between the two observers is no more than
4 mmHg for systolic and diastolic BP values. Otherwise, the reference measurement should be
repeated. The intra individual variability of the reference BP measurements must be < 12 mm
Hg for SBP and < 8 mm Hg for DBP.
The circumference of the patient's upper arm will be measured to ensure that the cuff-size
being used is adequate for the subject. The upper arm midpoint is first determined by marking
the arm posteriorly at a point halfway between the acromion and olecranon, measured while the
arm is flexed 90 degrees at the elbow with the palm facing up. The subject's upper arm
circumference shall be determined by measuring at the midpoint of the upper arm while the
elbow is relaxed, and the arm is dangling freely to the side.
The circumference of the patient's wrist will be measured to ensure that the circumference is
within the acceptable range 12.5 - 21.5 cm for the Combei device, otherwise the subject will
be excluded.
Measurements made by the mercury sphygmomanometer, will be performed according to the "same
arm, consecutive measurements" (Figure 1), on the left arm supported at heart level.
Measurements by the Combei BP 880 W device will be performed at the wrist level of the same
arm supported at the heart level, as recommended by the manufacturer with the following
procedure.
1. Ask the subject to relax for 5-10 min.
2. make sure that the subject :
- is comfortably seated with legs uncrossed and feet flat on the floor,
- has the back, elbow and forearm supported,
- has the measurement site at the level of the left ventricle of the heart,
- and other conditions as recommended by the validation protocol and the device
manual user.
In total, nine consecutive BP measurements will be performed in each patient using the
mercury sphygmomanometers (5 times) and the tested devices (4 times).
All nine-sequential same-arm measurements will be recorded as shown in Table 2, starting with
the standard mercury sphygmomanometer, followed by the device. A resting period of a minimum
of 60 seconds shall be observed between the end of one measurement and the beginning of the
following measurement.
Table 2: Sequential measurements of BP recorded by the standard mercury sphygmomanometer and
the tested device:
R0 Entry BP, observers 1 and 2 each with the mercury standard T0 Device detection BP,
supervisor R1 Observers 1 and 2 with mercury standard T1 Supervisor with the test instrument
R2 Observers 1 and 2 with mercury standard T2 Supervisor with the test instrument R3
Observers 1 and 2 with mercury standard T3 Supervisor with the test instrument R4 Observers 1
and 2 with mercury standard
The first auscultatory and the first device measurement represent the recruitment pressures
(R0 and T0) are not used in the evaluation of the test device accuracy. Each of the reference
BP measurements (R) is the average of the simultaneous readings of the two observers. For
each subject, the device measurements T1, T2 and T3 will be compared to observer
measurements. Each of the test device measurements is compared to the average of the previous
and succeeding reference BP readings (T1 vs average of R1 and R2, T2 vs average of R2 and R3,
T3 vs average of R3 and R4).
At the end of the measurements, the patient participation to the study is over.
The Statistical Analysis: Differences are calculated by subtracting the reference BP
measurement from the test device measurement. The mean BP difference (SUT versus reference
sphygmomanometer) and its standard deviation (SD) will be calculated for the analysis. The
same criteria will be used for systolic blood pressure and diastolic blood pressure
measurements. The statistical analysis will be performed according to the requirements of the
AAMI/ESH/ISO standard 2018 and its Amendment 1, 2020-01 for limb circumference distribution
criteria.