Heart Failure Clinical Trial
Official title:
Randomized Controlled Trial Comparing Internal vs External Cardioversion in ICD Patients
Implantation of internal defibrillators, capable of monitoring the heart and shocking life
threatening arrhythmias back to normal rhythm, for patients with severe heart failure
increases the probability of survival. Arrhythmias of the atria of the heart are common in
these patients. Administering a direct current electrical shock under anesthesia
(cardioversion) is the method of choice to reestablish normal sinus rhythm in this instance.
Safety and efficacy of external electrical cardioversion (CV) in patients with ICDs was
demonstrated in several studies. Safety of internal cardioversion (shocking the heart back
into normal rhythm via the implanted defibrillator) was described in several smaller trials.
Performing external instead of internal cardioversion in patients with implanted ICDs is more
feasible for most hospitals, as CV can be performed without a programming computer and an
additional specialist present, e.g. on the intensive care ward, and device interrogation can
be done after CV at the remote ICD/pacemaker clinic.
No scientific data on safety and efficacy endpoints comparing internal vs external CV is
currently available. The aim of the study is to compare external vs internal electrical
cardioversion for atrial arrhythmias and establish a safety and efficacy profile for external
and internal cardioversion in large cohort of ICD patients.
Introduction A mortality benefit of implanted cardiac defibrillators in patients (pts) with
severely impaired LV function has been established in multiple large trials. Atrial
fibrillation, -flutter and -tachycardia are a common comorbidity in this patient population.
Cardioversion is the method of choice to reestablish normal sinus rhythm in this instance.
Safety and efficacy of external electrical cardioversion (CV) in patients with pacemakers and
ICDs could be shown in several smaller single center trials. Safety of internal cardioversion
was described in several smaller trials. Reports on efficacy of internal CV vary between 33%
and 83% .
Performing external instead of internal cardioversion in pts with implanted ICDs is more
feasible for most hospitals, as CV can be performed without a programmer present, e.g. on the
ICU ward, and device interrogation can be done after CV at the remote ICD/pacemaker clinic. A
working group of the German cardiac society recommends internal CV for patients carrying an
ICD, suggesting a superior safety profile for internal CV, despite conflicting data and a
lack of evidence and on this issue. Thus, posing a problem in logistics for many hospitals.
No scientific data on safety and efficacy endpoints comparing internal vs external CV is
currently available.
Aim of the study To compare external vs internal electrical cardioversion for atrial
arrhythmias and establish a safety and efficacy profile for external and internal
cardioversion in large cohort of ICD patients. This study will provide vital evidence for
future recommendations regarding cardioversion in the growing population of patients carrying
implanted ICD and pacemakers. The main purpose will be to establish non-inferiority of
external electrical cardioversion to internal electrical cardioversion. This would impact day
to day practice in many arrhythmia centers and smaller hospitals around the world.
Trial populations and choice of primary analysis set The full-analysis set (FAS) is derived
from the intention-to-treat (ITT) principle. This dataset includes all trial subjects
randomized, with any valid data on the primary target variables "occurrence of severe lead
and device impairment" and "restoration of normal sinus rhythm".
The per-protocol set includes all trial subjects who were essentially treated according to
protocol (i.e. who have been cardioverted as assigned) and for whom the primary target
variables "occurrence of severe lead and device impairment" and "restoration of normal sinus
rhythm" could be determined based on complete and meaningful documentation.
In a non-inferiority trial there are special issues in the choice of primary analysis set.
Conservatism or anticonversatism of the ITT or PP analysis depends on various factors as type
of protocol deviation and missingness, and the method of handling missing data in ITT
population (see Sanchez MM, Chen X. Choosing the analysis population in non-inferiority
studies: per protocol or intent-to-treat. Stat Med. 25, 1169-81;2006). Due to our setting
(randomization immediately before intervention, allowance for cross-over from internal to
external CV, intervention in hospital setting) we do not expect a relevant amount of protocol
violations. As the follow-up period of 2 weeks is short, we expect <1% losses to follow-up.
So we assume only minor discrepancies between ITT and PP population.
We do not expect any missing data for the primary efficacy endpoint, as it is measured
immediately after CV. The primary safety endpoint is measured within two weeks after CV, and
we do not expect non-trivial missingness such as dropout due to lack of efficacy. As the
expected number of events for the primary safety endpoint is low, especially for internal CV
patients, any method of handling of missing data in ITT analysis may cause relevant bias.
Thus, the per-protocol set is considered the primary analysis population; the ITT analysis is
treated secondary/supportive and should lead to similar conclusions. Patients without any
valid data on the primary target variables will be excluded from the FAS (modified ITT), in
case of missing data on safety endpoint the valid value will be used. In a sensitivity
analysis a conservative scenario will be performed treating missing values as event in
external CV group and as no event in internal CV group. Any relevant discrepancies in
drop-out rates between CV groups will be discussed.
Outcome data and data analysis In case of cross-over from internal to external CV the primary
endpoints will be assessed after internal shock. Patient's death during intervention or
within the follow-up period is unlikely and not to be expected. In the case of patient death
during intervention, after the shock, the primary efficacy endpoint will be treated as
failure. In the case of a patient death during intervention, after the shock or during follow
up, the primary safety value will be counted as failure/severe device or lead impairment.
This may be revised after the cause of death was proven not to be lead or device associated
by interrogation and analysis by the manufacturer.
Stratification by study center is not feasible due to the low number of expected events, and
subgroup analyses by study center will be done for descriptive reasons only.
Duration and study size This study is a multicenter trial. Additional local ethics
application is required at each center.
For the primary safety endpoint "severe lead and device impairment" we expect an incidence of
10% in the external CV group and of 5% in the internal CV group. Thus, for one-sided alpha
5%, non-inferiority margin of 15%, and power 80%, a sample size of 103 patients per arm, i.e.
206 in total, is required. To account for loss to follow-up (<1% within 2 weeks after CV) and
heterogeneity between centers, 230 patients (≈206/(1-0.05)^2) (see Donner, A. Approaches to
sample size estimation in the design of clinical trials--a review. Stat Med 3, 199-214;1984)
will be randomized. A non-inferiority margin of 15% is large, however, we chose several
endpoints acting as sensitive surrogates of possible or temporary lead affectation, rather
than relying solely on hard endpoints such as permanent lead or device failure with necessity
of surgical revision.
As no data on placebo-controlled trials exists in the setting of ICD trials to determine the
non-inferiority margin on statistical grounds, the margin was chosen on clinically as well as
practicality grounds. A relevantly smaller alpha or tighter margin would lead to an
unachievable sample size and render the trial unfeasible (see D'Agostino, R. B., Massaro, J.
M. & Sullivan, L. M. Non-inferiority trials: design concepts and issues - the encounters of
academic consultants in statistics. Stat Med 22, 169-186;2003).
Of note, to detect an absolute risk reduction of 20% in the primary efficacy endpoint
"restoration of normal sinus rhythm", assuming an incidence of 90% in the external CV group
and 70% in the internal CV group, a sample size of 72 patients per arm, i.e. 144 in total, is
required (two-sided alpha 5%, power 80%, continuity corrected chi-squared test). Sample size
determination was performed with ADDPLAN™ 6.0.
Description of trial subject groups At baseline (i.e. before CV) treatment groups will be
described and compared using mean, standard deviation and percentiles (0, 25, 50, 75, 100)
for continuous variables, count and percentage for categorical variables. Variables to be
summarized include among other sex, age, BMI, comorbidities, type of device, site of
implantation, LV ejection fraction, type of anticoagulation, duration of arrhythmia, left
atrial diameter, pacing threshold, lead impedance, shock impedance, ventricular and
biventricular pacing percentage, battery voltage.
Primary safety endpoint The primary hypothesis "non-inferiority of external CV to internal
CV" regarding the primary safety endpoint "severe lead and device impairment" will be
evaluated using the confidence interval (CI) approach. A one-sided 95% CI for the difference
"observed proportion in external group - observed proportion in internal group" will be
calculated by Newcombe's method 10 and non-inferiority is declared to be shown at one-sided
level 5% if its upper bound lies below the non-inferiority bound of 15%.
Primary efficacy endpoint If non-inferiority regarding the primary safety endpoint "severe
lead and device impairment" could be demonstrated, the hypothesis "superiority of external CV
to internal CV" regarding the primary efficacy endpoint "restoration of normal sinus rhythm"
will be tested in a confirmatory way at two-sided level 5%. Due to the fixed sequence of
these two hypotheses, the type I error is not inflated. The proportion of the primary
efficacy endpoint will be compared between treatment groups by Fisher's exact test with
two-sided 95% CI calculated by Newcombe's method 10.
Secondary endpoints The components of the primary safety endpoint (i.e. a rise in threshold
(at constant duration) of >0.5V, exit block, loss of programming, rise in shock impedance by
50%, rise in charge time by 50%, drop in battery voltage of ≥0.2V within 2 weeks) will be
analyzed along the same line as the composite.
Moreover, recurrence of atrial arrhythmia at FU, lead impedance and sensing changes,
inadvertent induction of VF during CV, high-sensitive Troponin T level 3h after CV will be
summarized by treatment group and compared by statistical hypothesis tests as applicable;
confidence intervals for treatment differences/effect sizes will be calculated to aid
interpretation.
Subgroup analysis
A subanalysis of predefined subgroups will be performed for:
- CRT-D patients to assess safety of external CV in pts with LV leads.
- CRT-D patients with quadripolar leads (including thresholds for all 16 pacing vectors)
- CRT-D patients with AV-VV optimization algorithms
Safety and suspension/termination criteria Based on published data and own previous work 10
an interim safety analysis will be performed after 50% of target sample size, i.e. 115 pts,
have been included. A statistically significant difference in the primary safety endpoint
between the groups at interim analysis will lead to a subsequent early termination of the
trial (using Fisher's exact test, two-sided level 5%).
Specific termination criteria:
- Significant difference in primary safety endpoint at interim analysis
- Induction of VF in internal CV group n > 3
;
Status | Clinical Trial | Phase | |
---|---|---|---|
Recruiting |
NCT05196659 -
Collaborative Quality Improvement (C-QIP) Study
|
N/A | |
Recruiting |
NCT05650307 -
CV Imaging of Metabolic Interventions
|
||
Recruiting |
NCT05654272 -
Development of CIRC Technologies
|
||
Active, not recruiting |
NCT05896904 -
Clinical Comparison of Patients With Transthyretin Cardiac Amyloidosis and Patients With Heart Failure With Reduced Ejection Fraction
|
N/A | |
Completed |
NCT05077293 -
Building Electronic Tools To Enhance and Reinforce Cardiovascular Recommendations - Heart Failure
|
||
Recruiting |
NCT05631275 -
The Role of Bioimpedance Analysis in Patients With Chronic Heart Failure and Systolic Ventricular Dysfunction
|
||
Enrolling by invitation |
NCT05564572 -
Randomized Implementation of Routine Patient-Reported Health Status Assessment Among Heart Failure Patients in Stanford Cardiology
|
N/A | |
Enrolling by invitation |
NCT05009706 -
Self-care in Older Frail Persons With Heart Failure Intervention
|
N/A | |
Recruiting |
NCT04177199 -
What is the Workload Burden Associated With Using the Triage HF+ Care Pathway?
|
||
Terminated |
NCT03615469 -
Building Strength Through Rehabilitation for Heart Failure Patients (BISTRO-STUDY)
|
N/A | |
Recruiting |
NCT06340048 -
Epicardial Injection of hiPSC-CMs to Treat Severe Chronic Ischemic Heart Failure
|
Phase 1/Phase 2 | |
Recruiting |
NCT05679713 -
Next-generation, Integrative, and Personalized Risk Assessment to Prevent Recurrent Heart Failure Events: the ORACLE Study
|
||
Completed |
NCT04254328 -
The Effectiveness of Nintendo Wii Fit and Inspiratory Muscle Training in Older Patients With Heart Failure
|
N/A | |
Completed |
NCT03549169 -
Decision Making for the Management the Symptoms in Adults of Heart Failure
|
N/A | |
Recruiting |
NCT05572814 -
Transform: Teaching, Technology, and Teams
|
N/A | |
Enrolling by invitation |
NCT05538611 -
Effect Evaluation of Chain Quality Control Management on Patients With Heart Failure
|
||
Recruiting |
NCT04262830 -
Cancer Therapy Effects on the Heart
|
||
Completed |
NCT06026683 -
Conduction System Stimulation to Avoid Left Ventricle Dysfunction
|
N/A | |
Withdrawn |
NCT03091998 -
Subcu Administration of CD-NP in Heart Failure Patients With Left Ventricular Assist Device Support
|
Phase 1 | |
Recruiting |
NCT05564689 -
Absolute Coronary Flow in Patients With Heart Failure With Reduced Ejection Fraction and Left Bundle Branch Block With Cardiac Resynchronization Therapy
|