Failed Back Surgery Syndrome Clinical Trial
Official title:
Efficacy of High Voltage VS Standard Voltage Radiofrequency on Unilateral Lower Limb Neuralgia in a Patient With Failed Back Surgery Syndrome
Postoperative lumbar spine pain syndrome (FBSS) refers to patients who have undergone surgery for lumbar spine disease one or more times and still have intractable lumbosacral pain after surgery, with or without lower limb sensory and motor dysfunction. Pain is the main reason for reducing patients' quality of life. Hussain and Erdek believe that despite up to 3 months of postoperative intervention, pain may still exist. Since persistent pain seriously affects patients' daily life and work, optimizing the FBSS treatment plan and improving the quality of life of FBSS patients cannot be ignored. At present, the treatment methods for neuropathic pain radiated to the lower extremity after lumbar spine surgery mainly include oral drug therapy, nerve block, pulsed radio frequency (PRF) and spinal cord electrical stimulation, and other minimally invasive interventional methods. Still, there is no domestic or foreign treatment method is reported in the literature to be effective.
Selective nerve root PRF is an effective method for treating neuropathic pain, but its analgesic mechanism is still unclear. The traditional PRF adopts the standard 42 ℃ mode, but due to its low voltage (40 V), short duration (120 s), and limited intensity of action, the patient cannot get the best therapeutic effect. In recent years, high-voltage long-term PRF has been gradually applied to treating neuropathic pain such as postherpetic neuralgia, trigeminal neuralgia, and glossopharyngeal neuralgia, and achieved ideal results. . Selective spinal nerve puncture under ultrasound and X-ray guidance: After the patient enters the operating room, the venous access is inserted, and the patient is placed in a prone position. ECG monitoring is given, and the patient's vital signs are continuously monitored. Routine sterilization and laying of sheets, puncture with an 20 G radiofrequency puncture needle under the guidance of ultrasound, repeated X-rays to confirm the position of the needle tip, when the needle tip reaches the target position, the X-ray front view will show that the needle tip of the puncture needle is located below the pedicle, and the lateral view shows The tip of the puncture needle is located at the posterior 1/2of the intervertebral foramen, then connect the radiofrequency treatment apparatus, perform sensory and motor tests, and perform sensory tests The pain in the corresponding innervated area is reproduced within 0.5 V and 50 Hz, and is consistent with the patient's original pain location; if >0.5 V does not induce pain in the original pain area, fine-tune the needle tip position and then perform the sensory test until 0.1 The needle tip position can be fixed when the patient's original pain can be induced within ~0.5 V. In the exercise test, muscle beats in the corresponding area were induced within 1.0 V and 2 Hz. After the test was successful, 1 ml of contrast agent will be injected, and the X-ray should show that the contrast agent spread along the nerve root and epidural space. ;
Status | Clinical Trial | Phase | |
---|---|---|---|
Completed |
NCT01701804 -
The Effectiveness of Non-surgical Integrative Package on Failed Back Surgery Syndrome
|
||
Not yet recruiting |
NCT01185665 -
Predicting Outcome of Transcutaneous Electrical Nerve Stimulator (TENS) in Failed Back Surgery Syndrome
|
N/A | |
Terminated |
NCT00974623 -
Bone Graft Materials Observational Registry
|
N/A | |
Recruiting |
NCT01181817 -
Evaluation of the Brain Activity During Spinal Cord Stimulation in Failed Back Surgery Syndrome Using Functional MRI and MRS
|
N/A | |
Recruiting |
NCT04732325 -
Sensory Testing of Multiple Forms of Spinal Cord Stimulation for Pain
|
N/A | |
Completed |
NCT05018377 -
Effectiveness and Safety of Percutaneous Adhesiolysis Using Rac'z Catheter Versus Navi Catheter in Management of Failed Back Surgery Syndrome
|
N/A | |
Not yet recruiting |
NCT04268602 -
The Effect of Intradermal Local Anesthetic Injection in FBSS (Failed Back Surgery Syndrome)
|
N/A | |
Recruiting |
NCT03957395 -
Comparison of Effectiveness of Tonic, High Frequency and Burst Spinal Cord Stimulation in Chronic Pain Syndromes
|
N/A | |
Completed |
NCT04121104 -
Electrodermal Activity and Respiration in Patients Treated With Spinal Cord Stimulation
|
N/A | |
Recruiting |
NCT05068011 -
Differential Target Multiplexed Spinal Cord Stimulation
|
||
Not yet recruiting |
NCT06034041 -
The Effect of Mediclore as an Anti-adhesion Agent and Safety in Full-endoscopic Spine Surgery: a Preliminary Study
|
Phase 4 | |
Completed |
NCT02837822 -
Effects of Spinal Cord Stimulation on Sensory Perceptions of Chronic Pain Patients
|
N/A | |
Completed |
NCT01966250 -
Electroacupuncture as a Complement to Usual Care for Patients With Non-acute Pain After Back Surgery
|
N/A | |
Completed |
NCT00018876 -
Low-Dose Radiation to Prevent Complications of Back Surgery
|
Phase 1 | |
Completed |
NCT03523000 -
Determining the Prognostic Value of Continuous Intrathecal Infusion
|
Phase 4 | |
Completed |
NCT01776749 -
Subcutaneous Stimulation as Add on Therapy to SCS toTreat Low Back Pain in FBSS
|
N/A | |
Recruiting |
NCT04244669 -
Impact of the SCS With Different Waveforms Over the Quality of Life (SCS-Quality)
|
N/A | |
Completed |
NCT02974101 -
AdaptiveStim Objectifying Subjective Pain Questionnaires
|
N/A | |
Terminated |
NCT01990287 -
SENSE (Subcutaneous and Epidural Neuromodulation System Evaluation) Study
|
N/A | |
Terminated |
NCT01711619 -
SubQStim: A Post-market Study of Subcutaneous Nerve Stimulation in Failed Back Surgery Syndrome (FBSS) Patients
|
N/A |