Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT03033849
Other study ID # IDT-1608-AL
Secondary ID
Status Completed
Phase N/A
First received January 16, 2017
Last updated September 11, 2017
Start date January 2017
Est. completion date March 2017

Study information

Verified date September 2017
Source Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The user performance evaluation shows whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system. In this study, user performance evaluation will be performed for Contour® Next One (Ascensia Diabetes Care GmbH), Accu-Chek® Aviva Connect (Roche Diabetes Care GmbH), FreeStyle Freedom Lite (Abbott Diabetes Care Inc.), OneTouch® Verio (LifeScan Europe) and GlucoMen® areo A. Menarini Diagnostics S.r.l. (based on ISO 15197:2013; EN ISO 15197:2015, clause 8.

For each BGMS, measurement procedures for user performance evaluation will be performed with 1 test meter and 1 reagent system lot by the study subjects.

The same meter and an additional test meter will be used for double measurements performed by study personnel (with the same reagent system lot used by subjects).


Recruitment information / eligibility

Status Completed
Enrollment 189
Est. completion date March 2017
Est. primary completion date March 2017
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years and older
Eligibility Inclusion Criteria:

- Male or female, with type 1 diabetes, type 2 diabetes or subjects without diabetes

- Signed informed consent form

- Minimum age of 18 years

- Subjects are legally competent and capable to understand character, meaning and consequences of the study.

- If blood glucose values < 80 mg/dl or > 300 mg/dl shall be measured after short term alteration in insulin therapy:

- Male or female with type 1 diabetes and intensified insulin therapy or insulin pump therapy.

- Signature of subjects to document consent with these procedures on informed consent form.

Exclusion Criteria:

- Pregnancy or lactation period

- Severe acute disease (at the study physician's discretion)

- Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)

- Current constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)

- Being unable to give informed consent

- < 18 years

- Legally incompetent

- Being committed to an institution (e.g. psychiatric clinic)

- Language barriers potentially compromising an adequate compliance with study procedures

- Dependent on investigator or sponsor

- If blood glucose values < 80 mg/dl shall be measured after short term alteration in insulin therapy, subjects with type 1 diabetes, suffering from:

- Coronary heart disease

- Condition after myocardial infarction

- Condition after cerebral events

- Peripheral arterial occlusive disease

- Hypoglycemia unawareness

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Blood glucose monitoring system for self-testing Contour® Next One
This is a user performance study, that is intended to show whether people with diabetes are able to obtain accurate measurement results with a blood glucose monitoring system (BGMS). The subjects' measurement technique (e.g. applying blood onto the system reagent, reading the result) will be observed and documented by study personnel. Immediately after the measurement by the subject, the study personnel will perform a measurement with the same test meter. Measurement results obtained by the subjects with the BGMS will be compared to measurement results obtained with the comparison method. Sample collection for the comparison measurement will be performed by study personnel within 5 min after the subject's measurement with the test meter. The samples will be collected from the subjects' skin puncture (if possible), an additional skin puncture (if required) will be documented. An additional skin puncture will be performed by study personnel with single-use lancing devices.
Blood glucose monitoring system for self-testing Accu-Chek® Aviva Connect

Blood glucose monitoring system for self-testing FreeStyle Freedom Lite

Blood glucose monitoring system for self-testing OneTouch® Verio

Blood glucose monitoring system for self-testing GlucoMen® areo


Locations

Country Name City State
Germany Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm Ulm Baden-Württemberg

Sponsors (2)

Lead Sponsor Collaborator
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm Ascensia Diabetes Care

Country where clinical trial is conducted

Germany, 

Outcome

Type Measure Description Time frame Safety issue
Primary Acceptance criteria defined by ISO 15197:2013 (E) will be applied 95 % of the individual glucose measured values shall fall within ± 15 mg/dl (0.83 mmol/l) of the measured values of the comparison measurement procedure at glucose concentrations < 100 mg/dl (5.55 mmol/l) and within ± 15 % at glucose concentrations = 100 mg/dl (5.55 mmol/l). For each subject, the experimental phase has an expected duration of up to 3 hours
See also
  Status Clinical Trial Phase
Completed NCT03743779 - Mastering Diabetes Pilot Study
Completed NCT03786978 - Pharmaceutical Care in the Reduction of Readmission Rates in Diabetes Melitus N/A
Completed NCT01804803 - DIgital Assisted MONitoring for DiabeteS - I N/A
Completed NCT05039970 - A Real-World Study of a Mobile Device-based Serious Health Game on Session Attendance in the National Diabetes Prevention Program N/A
Completed NCT04507867 - Effect of a NSS to Reduce Complications in Patients With Covid-19 and Comorbidities in Stage III N/A
Completed NCT04068272 - Safety of Bosentan in Type II Diabetic Patients Phase 1
Completed NCT03243383 - Readmission Prevention Pilot Trial in Diabetes Patients N/A
Completed NCT03730480 - User Performance of the CONTOUR NEXT and CONTOUR TV3 Blood Glucose Monitoring System (BGMS) N/A
Recruiting NCT02690467 - Efficacy, Safety and Acceptability of the New Pen Needle 34gx3,5mm. N/A
Completed NCT02229383 - Phase III Study to Evaluate Safety and Efficacy of Added Exenatide Versus Placebo to Titrated Basal Insulin Glargine in Inadequately Controlled Patients With Type II Diabetes Mellitus Phase 3
Completed NCT05799976 - Text Message-Based Nudges Prior to Primary Care Visits to Increase Care Gap Closure N/A
Completed NCT06181721 - Evaluating Glucose Control Using a Next Generation Automated Insulin Delivery Algorithm in Patients With Type 1 and Type 2 Diabetes N/A
Recruiting NCT04489043 - Exercise, Prediabetes and Diabetes After Renal Transplantation. N/A
Withdrawn NCT03319784 - Analysis for NSAID VS Corticosteroid Shoulder Injection in Diabetic Patients Phase 4
Completed NCT03542084 - Endocrinology Auto-Triggered e-Consults N/A
Completed NCT02229396 - Phase 3 28-Week Study With 24-Week and 52-week Extension Phases to Evaluate Efficacy and Safety of Exenatide Once Weekly and Dapagliflozin Versus Exenatide and Dapagliflozin Matching Placebo Phase 3
Recruiting NCT05544266 - Rare and Atypical Diabetes Network
Completed NCT01892319 - An International Non-interventional Cohort Study to Evaluate the Safety of Treatment With Insulin Detemir in Pregnant Women With Diabetes Mellitus. Diabetes Pregnancy Registry
Completed NCT05031000 - Blood Glucose Monitoring Systems: Discounter Versus Brand N/A
Recruiting NCT04039763 - RT-CGM in Young Adults at Risk of DKA N/A