Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT02053584
Other study ID # CR-0002
Secondary ID
Status Completed
Phase N/A
First received
Last updated
Start date June 2013
Est. completion date June 2013

Study information

Verified date July 2016
Source LabStyle Innovations Ltd.
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The study is performed to evaluate the accuracy of blood glucose level results obtained from fingertip using Labstyle Dario Blood Glucose Monitoring System (BGMS) compared to reference equipment (YSI 2300 STATPLUS) and to evaluate the ease of use of the Dario Device by the lay-person. The participants will review the Dario user manual and guide and will be requested to operate the Dario BGMS and perform their own glucose blood test, after which they will be asked to complete a questionnaire. As reference the participants will be tested by professional caregivers both using the Dario BGMS and by YSI 2003.


Description:

Device accuracy will be assessed based on the acceptance criteria of ISO 15197:2003 which states: "The minimum acceptable accuracy for results produced by a glucose monitoring system shall be as follows: 95% of the individual glucose results shall fall within 15 mg/dL of the results of Reference method at glucose concentrations ≥ 4,2 mmol/L (≥ 75 mg/dL)".


Recruitment information / eligibility

Status Completed
Enrollment 230
Est. completion date June 2013
Est. primary completion date June 2013
Accepts healthy volunteers No
Gender All
Age group N/A and older
Eligibility Inclusion Criteria: - Male or female between the age 18 and 80 - Type 1 and Type 2 diabetes Exclusion Criteria: - Drug and alcohol abuse - Investigator discretion

Study Design


Related Conditions & MeSH terms


Intervention

Device:
Dario™ Blood Glucose Meter
Each subject will undergo a finger prick and also provide a venous blood sample
YSI2003
Blood samples will be tested using YSI (gold standard) for comparison to evaluate accuracy.

Locations

Country Name City State
n/a

Sponsors (1)

Lead Sponsor Collaborator
LabStyle Innovations Ltd.

Outcome

Type Measure Description Time frame Safety issue
Primary BGMS Accuracy Assessment When Used by Layperson The minimum acceptable accuracy for results produced by a glucose monitoring system shall be as follows: 95% of the individual glucose results shall fall within 15 mg/dL of the results of reference method (YSI) at glucose concentrations = 4,2 mmol/L (= 75 mg/dL). (ISO 15197:2003) Within 6 seconds of applying blood drop to strip
Secondary Evaluation of the Ease of Use of the Dario BGMS by the Lay-person Each participant receives a survey to fill in describing their usage of the device and their opinions on the ease of use. Within a couple of hours from the time labelling material was provided to the subject
See also
  Status Clinical Trial Phase
Completed NCT05594446 - Morphometric Study of the Legs and Feet of Diabetic Patients in Order to Collect Data Intended to be Used to Measure by Dynamometry the Pressures Exerted by Several Medical Compression Socks at the Level of the Forefoot
Completed NCT03975309 - DHS MIND Metabolomics
Completed NCT01855399 - Technologically Enhanced Coaching: A Program to Improve Diabetes Outcomes N/A
Completed NCT01819129 - Efficacy and Safety of FIAsp Compared to Insulin Aspart in Combination With Insulin Glargine and Metformin in Adults With Type 2 Diabetes Phase 3
Recruiting NCT04984226 - Sodium Bicarbonate and Mitochondrial Energetics in Persons With CKD Phase 2
Recruiting NCT05007990 - Caregiving Networks Across Disease Context and the Life Course
Active, not recruiting NCT04420936 - Pragmatic Research in Healthcare Settings to Improve Diabetes and Obesity Prevention and Care for Our Program N/A
Recruiting NCT03549559 - Imaging Histone Deacetylase in the Heart N/A
Completed NCT04903496 - Clinical Characteristics and Disease Burden of Diabetic Patients Based on Tianjin Regional Database
Completed NCT01437592 - Investigating the Pharmacokinetic Properties of NN1250 in Healthy Chinese Subjects Phase 1
Completed NCT01696266 - An International Survey on Hypoglycaemia Among Insulin-treated Patients With Diabetes
Completed NCT04082585 - Total Health Improvement Program Research Project
Completed NCT03390179 - Hyperglycemic Response and Steroid Administration After Surgery (DexGlySurgery)
Not yet recruiting NCT05029804 - Effect of Walking Exercise Training on Adherence to Disease Management and Metabolic Control in Diabetes N/A
Recruiting NCT05294822 - Autologous Regenerative Islet Transplantation for Insulin-dependent Diabetes N/A
Completed NCT04427982 - Dance and Diabetes/Prediabetes Self-Management N/A
Completed NCT02356848 - STEP UP to Avert Amputation in Diabetes N/A
Completed NCT03292185 - A Trial to Investigate the Single Dose Pharmacokinetics of Insulin Degludec/Liraglutide Compared With Insulin Degludec and Liraglutide in Healthy Chinese Subjects Phase 1
Active, not recruiting NCT05477368 - Examining the Feasibility of Prolonged Ketone Supplement Drink Consumption in Adults With Type 2 Diabetes N/A
Completed NCT04496401 - PK Study in Diabetic Transplant récipients : From Twice-daily Tacrolimus to Once-daily Extended-release Tacrolimus Phase 4