Source Catholic University of the Sacred Heart
Contact n/a
Is FDA regulated No
Health authority
Study type Observational

Clinical Trial Summary

Growing evidences are showing the usefulness of lung ultrasound in patients with COVID-19. Sars-CoV-2 has now spread in almost every country in the world. In this study, the investigators share their experience and propose a standardized approach in order to optimize the use of lung ultrasound in covid-19 patients. The investigators focus on equipment, procedure, classification and data-sharing.


Description:

COVID-19 global emergency need a global unified approach, speaking all researchers the same language. For this reason, the investigators propose a standardization for the international use of lung ultrasound (LUS) for the management of COVID-19 patients. The LUS COVID-team is made by Italian experts in lung ultrasound currently involved in the clinical management of COVID-19 patients in different Italian areas, including the heavily involved cities of Northern Italy. Moreover, experts in ultrasound physics and image analysis are part of the team. The team developed a standardized approach regarding equipment and acquisition protocol. Moreover, the team proposed a scoring system for severity classification. To this aim, clinicians shared 20 cases of confirmed COVID-19 on an anonymized virtual database, for a total of about 44000 frames up to date. All team members discussed their clinical cases through online meetings. Images were reviewed by all team members, blinded to the clinical background, and listed in classes of severity of lung involvement based on LUS images. At the end of this process, a biomedical engineer expert in lung ultrasound collected the data and suggested a lung ultrasound grading system for COVID-19 pneumonia. Again, the biomedical engineer re-submitted the images grouped in different classes of severity to the study members, blinded of clinical data, to review again the images and evaluated agreement regarding the LUS scores. The score was defined only when all team members agreed. Methods In the setting of COVID-19, wireless probe and tablets represent the most appropriate ultrasound equipment. These devices can easily be wrapped in single use plastic covers reducing the risk of contamination and making easy the sterilization procedures. Such devices are much less expensive than usual ultrasound machines including the portable ones. In case of unavailability of these devices, portable machines dedicated to the exclusive use of COVID-19 patients can be used, although maximum care for sterilization is necessary. In these cases, probe and keyboard covers are anyway suggested, and sterilization procedures necessary following last recommendations. Sharing their real world experience in performing LUS in COVID-19 patients, the investigators propose two different ways of performing lung ultrasound with pocket devices aiming to reduce the exposition of health workers to cases. One operator uses the probe performing the ultrasound; the other one keeps the tablet and freezes images/videos. The second operator can be either in the room being at safe distance from the patient (about 2 meters), or even remain outside the door communicating by phone-call with the operator one in order to optimize the quality of images. Potentially, this last approach can reduce the operator-dependence of the ultrasound since the second operator blindly selects the images, being unaware of the clinical conditions of the patient. The two operators will follow an agreed, tested and standardized images acquisition protocol. Acquisition protocol Fourteen areas (three posterior, two lateral and two anterior) should be scanned per patient along the lines here indicated. Scans need to be intercostal, as to cover the widest surface possible with one scan. Standard sequence of evaluations is proposed using landmarks on chest anatomic lines. Echographic scans can be identified with a progressive numbering starting from right posterior basal regions. For patient able to maintain the sitting position: 1. Right basal on paravertebral line above the curtain sign 2. Right middle on paravertebral line at the inferior angle of shoulder blade 3. Right upper on paravertebral line at spine of shoulder blade 4. Left basal on paravertebral line above the curtain sign 5. Left middle on paravertebral line at the inferior angle of shoulder blade 6. Left upper on paravertebral line at spine of shoulder blade 7. Right basal on mid-axillary line below the internipple line 8. Right upper on mid axillary line above the internipple line 9. Left basal on mid-axillary line below the internipple line 10. Left upper on mid axillary line above the internipple line 11. Right basal on mid-clavicular line below the internipple line 12. Right upper on mid-clavicular line above the internipple line 13. Left basal on mid-clavicular line below the internipple line 14. Left upper on mid-clavicular line above the internipple line In case of performance of LUS in critical care settings (such as patients on invasive ventilation) and for patients that are not able to maintain sitting position, the posterior areas might be difficult to be evaluated. In these cases, the operator should try to have a partial view of the posterior basal areas, currently considered a "hot-area" for COVID-19, and however, start echographic assessment from landmark number 7. - Use Convex or Linear probes, according to the patient's body size - Use single focal point modality (no multifocusing), setting the focal point on the pleura line. Employing a single focal point, and setting it at the right location, has the benefit to optimize the beam shape for sensing the lung surface. At focus, the beam has the smallest width as is therefore set to best respond to the smallest details. - Keep the mechanical index (MI) low (start from 0.7 and reduce it further if allowed by the visual findings). High MIs, employed for a long observation time, may result in damaging the lung. - Avoid as much as possible saturation phenomena, control gain and diminish MI if needed (see example of lung ultrasound images in the figures). Saturation phenomena occurs, e.g., when the signal streght of the echo signals is too high for the receiving electronics to be converted into electrical signals conserving a linear relation with the pressure amplitude. This has the effect of distorting the signals, and produces images where the dynamics of the actual signal is lost. The visual appearance of this phenomenon is the presence of areas which are completely white. In this case it is therefore not possible to appreciate local variations in the response to insonifications. - Avoid the use of cosmetic filters and specific imaging modalities such as Harmonic Imaging, Contrast, Doppler, Compounding. - Achieve the highest frame rate possible (e.g. no persistence, no multifocusing) - Save the data in DICOM format. In case this is not possible, save the data directly as a video format. Visual findings, especially when related to very small alterations, do not appear on every frame. It is thus advantageous to acquire movies, where the lung surface below the landmark can be monitored for few seconds during breathing. Scoring procedures - Score 0: The pleura line is continuous, regular. Horizontal artifacts (A-line) are present. These artifacts are generally referred as A-lines. They are due to the high-reflectivity of the normally aerated lung surface, and characterize the visual representation of the multiple reflections happening between the ultrasound probe and the lung surface itself. - Score 1: The pleura line is indented. Below the indent, vertical areas of white are visible. These are due to local alterations in the acoustical properties of the lung, as for example the replacement of volumes previously occupied by air in favor of media which are acoustically much more similar to the intercostal tissue (water, blood, tissue). This phenomenon opens channels accessible to ultrasound, which can explain the appearance of the vertical artifacts. - Score 2: The pleura line is broken. Below the breaking point, small to large consolidated areas (darker areas) appear with associated areas of white below the consolidated area (white lung). The darkening of the consolidated areas signals the loss of aeration and the transition of these areas towards acoustic properties similar to soft tissue over the entire area represented by the consolidation itself. Beyond the consolidations, the appearance of areas of white lung signals the presence of areas not yet fully deaerated, where air inclusions are still present but embedded in tissue like material. This highly scattering environment can explain this peculiar pattern. - Score 3: The scanned area shows dense and largely extended white lung with or without larger consolidations. At the end of the procedure, the clinician will write for each area the highest score obtained (e.g quadrant 1, score 2; quadrant 10, score 1; and so on).


Recruitment information / eligibility

Status Completed
Enrollment 220
Est. completion date December 31, 2021
Est. primary completion date December 31, 2021
Accepts healthy volunteers No
Gender All
Age group N/A and older
Eligibility Inclusion Criteria: - suspected COVID-19 patients, - known COVID-19 patients. Exclusion Criteria: - confirmed disease different from COVID-19.

Study Design


Intervention

Diagnostic Test:
Lung ultrasound
A proposal of a standardized approach in order to optimize the use of lung ultrasound in COVID-19 patients. We focus on equipment, procedure, classification and data-sharing.

Locations

Country Name City State
Italy Bresciamed Brescia
Italy Pulmonary Medicine Unit, Lodi General Hospital Lodi
Italy 118 USL Nordovest Toscana Lucca
Italy Diagnostic and Interventional Ultrasound Unit, Valle del Serchio General Hospital Lucca
Italy Emergency Department, Fondazione IRCCS Policlinico San Matteo, and Department of Internal Medicine and Therapeutics, University of Pavia Pavia
Italy Department of woman and child health and public health, Fondazione Policlinico Universitario A. Gemelli IRCCS Roma
Italy Pulmonary Medicine Unit, Dept. Medical and Surgical Sciences, Fondazione Policlinico Universitario A. Gemelli IRCCS Roma
Italy Department of Information Engineering and Computer Science, Ultrasound Laboratory Trento, University of Trento Trento
Italy Emergency Medicine Unit, General Hospital Voghera

Sponsors (8)

Lead Sponsor Collaborator
Catholic University of the Sacred Heart 118 USL Nordovest Toscana, Lucca, Italy, Bresciamed, Brescia, Italy, Fondazione IRCCS Policlinico San Matteo, University of Pavia, Pavia, Italy, General Hospital, Voghera, Italy, Lodi General Hospital, Lodi, Italy, Università degli Studi di Trento, Valle del Serchio General Hospital, Lucca, Italy

Country where clinical trial is conducted

Italy, 

Outcome

Type Measure Description Time frame Safety issue
Primary Lung ultrasound grading system for COVID-19 pneumonia Scoring procedures
Score 0: The pleura line is continuous, regular. Horizontal artifacts (A-line) are present. These artifacts are generally referred as A-lines.
Score 1: The pleura line is indented. Below the indent, vertical areas of white are visible.
Score 2: The pleura line is broken. Below the breaking point, small to large consolidated areas (darker areas) appear with associated areas of white below the consolidated area (white lung).
Score 3: The scanned area shows dense and largely extended white lung with or without larger consolidations At the end of the procedure, the clinician will write for each area the highest score obtained.
At enrollment.
See also
  Status Clinical Trial Phase
Completed NCT05065827 - Lung Ultrasound Findings in Patients With COVID-19 in a UK ED
Recruiting NCT05359770 - Association of Inspiratory Muscle Training With HD-tDCS for Assistance to Patients With Long Covid-19 N/A
Completed NCT04515147 - A Dose-Confirmation Study to Evaluate the Safety, Reactogenicity and Immunogenicity of Vaccine CVnCoV in Healthy Adults for COVID-19 Phase 2
Not yet recruiting NCT06025812 - Clinical Study of Omicron BA.4/5-Delta Strain Recombinant Novel Coronavirus Protein Vaccine (CHO Cells) N/A
Recruiting NCT05289115 - Protocol for Assistance to Patients With Long Covid-19 Undergoing Treatment With HD-tDCS N/A
Completed NCT04953078 - A Study to Evaluate Safety, Tolerability, and Reactogenicity of an RBD-Fc-based Vaccine to Prevent COVID-19 Phase 1
Completed NCT04779138 - Increasing Vaccine Uptake in Underresourced Public Housing Areas N/A
Not yet recruiting NCT05868239 - Impact of Aerosol Box Use During Cardiopulmonary Arrest: A Multicenter Study N/A
Completed NCT04690413 - NOWDx Test for the Detection of Antibodies to COVID-19 N/A
Completed NCT04818164 - Prone Position Improves End-Expiratory Lung Volumes in COVID-19 Acute Respiratory Distress Syndrome
Terminated NCT04530448 - Coronavirus Induced Acute Kidney Injury: Prevention Using Urine Alkalinization Phase 4
Completed NCT04572399 - UVA Light Device to Treat COVID-19 N/A
Recruiting NCT04610567 - Treatment of Patients With Mild Coronavirus-19 (COVID-19) Disease With Methotrexate Associated to LDL Like Nanoparticles (Nano-COVID19) Phase 1/Phase 2
Recruiting NCT04772170 - Observational Digital Biomarker Discovery in Respiratory Virus Challenge Studies
Recruiting NCT04581954 - Inflammatory Signal Inhibitors for COVID-19 (MATIS) Phase 1/Phase 2
Completed NCT04405934 - COG-UK Project Hospital-Onset COVID-19 Infections Study N/A
Enrolling by invitation NCT04484025 - SPI-1005 Treatment in Moderate COVID-19 Patients Phase 2
Completed NCT05572840 - Wear Your Mask, Wash Your Hands, Don't Get COVID-19 N/A
Withdrawn NCT04838847 - A Study to Evaluate the Immunogenicity and Safety of the SARS-CoV-2 mRNA Vaccine CVnCoV in Elderly Adults Compared to Younger Adults for COVID-19 Phase 3
Completed NCT04556149 - imPulseā„¢ Una Full-spectrum, Over Clothing E-stethoscope