Clinical Trials Logo

Clinical Trial Details — Status: Completed

Administrative data

NCT number NCT03114124
Other study ID # ir-iis02
Secondary ID
Status Completed
Phase N/A
First received
Last updated
Start date May 23, 2017
Est. completion date November 9, 2019

Study information

Verified date August 2021
Source University of Florida
Contact n/a
Is FDA regulated No
Health authority
Study type Interventional

Clinical Trial Summary

The aim of our study is to investigate the comparative efficacy of high fidelity multi electrode ablation catheters vs the standard bipolar configuration in success of AV nodal ablation This study will provide insights on the use of new technology where application may increase efficacy, promote patient and physician safety and decrease costs.


Description:

Introduction In some elderly patients with atrial fibrillation (AF), especially in combination with heart failure, a rate control strategy may be preferred. When pharmacological therapy is ineffective or not tolerated, it is reasonable to perform atrioventricular (AV) node ablation with ventricular pacing as a class IIA indication per current guidelines. Usually, the procedure is simple and straightforward and complete heart block can be achieved without any difficulty. However, this "simple" procedure can sometimes prove to be a most difficult case. The most common reason for failure to achieve complete heart block is the inability to localize the compact AV node using the His signal with standard intracardiac electrograms. As these patients come to the laboratory in AF, the His signal may be obscured by AF waves. In some patients with a deeper intramyocardial location of the His bundle and compact AV node it becomes necessary to produce deeper ablation lesions using an irrigated catheter to achieve block. In patients with AF, the target of ablation for the "ablate and pace" approach is the compact AV node, located at the apex of the triangle of Koch. Ideally, ablation is performed at the most proximal penetrating part of the His bundle in order to maintain a proximal automatic junctional rhythm and avoid pacemaker dependence. Para Hisian pacing is most commonly used to reveal the presence of a septal accessory pathway. The His bundle is a deep insulated structure and it is difficult to capture it at usual energy outputs. Using a high- output pacing (usually 20 mA at 2 msec) it is possible to directly capture the deeply situated His bundle, which is confirmed by a narrower QRS complex on the paced electrograms. Thus, high-output pacing can be utilized to map the His bundle area in difficult situations. By applying this electrophysiologic principle of differential tissue capture to help identify the location of the compact AV node, which is in close proximity to the His bundle. Ventricular pacing was performed initially at high output to capture both the basal right ventricular myocardium and the His bundle and the output was gradually lowered to lose His bundle capture. The QRS duration is relatively narrow with high output pacing and increases when the pacing output is lowered, representing ventricular myocardial capture alone. Finally, loss of ventricular capture is seen with further reduction of pacing output. This maneuver has been shown to aid in determining the proximity of the ablation catheter to compact AV node as was validated by successful ablation at this site. Parahisian pacing in conjunction with av nodal ablation has recently been described in the literature. A novel catheter with three mini electrodes within the ablation tip (IntellaTip MiFi, Boston Scientific, Boston, MA) may enhance the available data for such a signal dependent technique. In this catheter, bipolar signals can be recorded between the three 0.8-mm-wide electrodes that are arranged radially 1.3 mm from the end of the catheter alongside the standard distal and proximal bipolar recordings. Animal studies have already demonstrated that the mini electrodes in this novel catheter are more accurate in identifying conducting gaps in linear ablations than conventional electrode recordings. Study Aims: The aim of our study is to investigate the comparative efficacy of high fidelity multi electrode ablation catheters vs the standard bipolar configuration in success of AV nodal ablation This study will provide insights on the use of new technology where application may increase efficacy, promote patient and physician safety and decrease costs. Primary and secondary objectives Primary endpoints 1. Acute success of ablation identified by a junctional rhythm or complete heart block 2. Time from application of radiofrequency energy to acute success Secondary endpoints include 1. Procedure time 2. Radiation time 3. Frequency of ablation application 4. Duration of ablation application


Recruitment information / eligibility

Status Completed
Enrollment 30
Est. completion date November 9, 2019
Est. primary completion date September 26, 2019
Accepts healthy volunteers Accepts Healthy Volunteers
Gender All
Age group 18 Years to 95 Years
Eligibility Inclusion Criteria: 1. Patients with a diagnosis of persistent or permanent atrial fibrillation documented on electrocardiography 2. Patients must meet American College of Cardiology and Heart Rhythm Society "ACC/HRS" guidelines for atrioventricular nodal ablation procedure 3. Patients must be available for at least 1 month post procedure 4. Patients must be greater than or equal to 18 years old. Exclusion Criteria: 1. Patients who do not meet ACC/HRS indications for av nodal ablation

Study Design


Related Conditions & MeSH terms


Intervention

Device:
MIFI Catheter
Multielectrode Catheter: standard ablation catheter that has the ability to record low and high frequency signals with high fidelity
Standard Ablation Catheter
Standard of care for patients with certain cardiac arrhythmias.

Locations

Country Name City State
United States University of Florida Jacksonville Florida

Sponsors (1)

Lead Sponsor Collaborator
University of Florida

Country where clinical trial is conducted

United States, 

Outcome

Type Measure Description Time frame Safety issue
Primary The Time Differences Between the 2 Arms for Acute Success of Ablation Identified by a Junctional Rhythm or Complete Heart Block Identified by a junctional rhythm or complete heart block. Time zero is when the ablation procedure begins and the successful outcomes are either complete heart block or junctional rhythm. 60 seconds
Secondary Time From Application of Radio-frequency Energy to Acute Success The target will be where the HIS electrogram amplitude is greatest and where the QRS complex from pacing is narrowest. up to 3000 seconds
See also
  Status Clinical Trial Phase
Recruiting NCT05654272 - Development of CIRC Technologies
Terminated NCT04115735 - His Bundle Recording From Subclavian Vein
Completed NCT04571385 - A Study Evaluating the Efficacy and Safety of AP30663 for Cardioversion in Participants With Atrial Fibrillation (AF) Phase 2
Completed NCT05366803 - Women's Health Initiative Silent Atrial Fibrillation Recording Study N/A
Completed NCT02864758 - Benefit-Risk Of Arterial THrombotic prEvention With Rivaroxaban for Atrial Fibrillation in France
Recruiting NCT05442203 - Electrocardiogram-based Artificial Intelligence-assisted Detection of Heart Disease N/A
Completed NCT05599308 - Evaluation of Blood Pressure Monitor With AFib Screening Feature N/A
Completed NCT03790917 - Assessment of Adherence to New Oral anTicoagulants in Atrial Fibrillation patiEnts Within the Outpatient registrY
Enrolling by invitation NCT05890274 - Atrial Fibrillation (AF) and Electrocardiogram (EKG) Interpretation Project ECHO N/A
Recruiting NCT05266144 - Atrial Fibrillation Patients Treated With Catheter Ablation
Recruiting NCT05316870 - Construction and Effect Evaluation of Anticoagulation Management Model in Atrial Fibrillation N/A
Not yet recruiting NCT06023784 - The Impact of LBBAP vs RVP on the Incidence of New-onset Atrial Fibrillation in Patients With Atrioventricular Block N/A
Recruiting NCT05572814 - Transform: Teaching, Technology, and Teams N/A
Recruiting NCT04092985 - Smart Watch iECG for the Detection of Cardiac Arrhythmias
Completed NCT04087122 - Evaluate the Efficiency Impact of Conducting Active Temperature Management During Cardiac Cryoablation Procedures N/A
Completed NCT06283654 - Relieving the Emergency Department by Using a 1-lead ECG Device for Atrial Fibrillation Patients After Pulmonary Vein Isolation
Recruiting NCT05416086 - iCLAS™ Cryoablation System Post-Market Clinical Follow-up (PMCF) Study N/A
Completed NCT05067114 - Solutions for Atrial Fibrillation Edvocacy (SAFE)
Completed NCT04546763 - Study Watch AF Detection At Home
Completed NCT03761394 - Pulsewatch: Smartwatch Monitoring for Atrial Fibrillation After Stroke N/A